MUST2 SAMURAI_2018 S3-LEB-LPC SUPERNEMO MUGAST EXPAND SCALP GALATRON HiCARI VELO
  MUST2, Page 1 of 48  Not logged in MUST2
New entries since:Thu Jan 1 01:00:00 1970
Status ExpNbr Date Author Type Category Target-Source Subject Text Attachments
Pinnede748 Thu Mar 23 11:19:13 2017 Adrien Matta   ScheduleThis message will contain the update schedule
of the preparation of e748.
  
PinnedGeneral Fri Mar 24 09:32:11 2017 Adrien MattaMUST2Software Usefull document for MUST2 operationsDAS Configuration:
http://wiki.ganil.fr/gap/wiki/Documentation/Software/DAS
 NOTICE_UTILISATION_MUVI_HARD_et_SOFT_(ver_081013).pdfMUVI_card_USER_MANUAL_(ver_081013).pdf 
Fixede748 Wed May 24 09:07:59 2017 Adrien MattaOTHERGeneral e748 PlanningFeel free to edit this spread sheet.

  
Pinnede748 Wed May 24 09:19:57 2017 Adrien MattaOTHERSocial e748 CollaboratorsList of Collaborators (Please edit if anyone
is missing):
  
Fixede748 Wed May 24 09:32:41 2017 Adrien MattaMUST2Hardware Mounting of the telescope in G3The telescope has been mounted in there stand
in G3. 
  
Fixede748 Wed May 24 09:35:00 2017 Adrien MattaCATSHardware BTD chambers mounting in D6Sebastient started mounting the BTD chamber
in D6. 
The frame is in place, BTD chambers one and
  
Fixede748 Wed May 24 15:48:24 2017 Adrien MattaMUST2Hardware PCB number for each telescope[LIST]
[*] T1 PCB 5
[*] T2 PCB 9
  
Fixede748 Thu Jun 1 09:29:50 2017 Adrien MattaMUST2Hardware Telescope cabling debbug in G3Test performed without vaccum, in the dark,
cooling at 20.
  
Fixede748 Thu Jun 1 09:32:38 2017 Adrien MattaMUST2Hardware MUST2 cabling schemeThe MUST2 cabling scheme has been change so
Telescope 1 to 4 and 5 to 8 are on two independent
set of distribution box and HV Card.
  
Fixede748 Fri Jun 2 12:03:32 2017 Adrien, Laurent, ShumpeiMUST2Hardware M2C vacuum testWe mounted 2 dry pump on the chamber and started
pumping at 11h.
Pressure was about 6e-1 mbar around 11h45.
  
Fixede748 Tue Jun 6 09:07:34 2017 Adrien MattaBEAMGeneral Moving to D6M2C Chamber now in D6 after a smooth ride,
ready to be aligned.
  
Ongoinge748 Tue Jun 6 17:57:07 2017 Valérian et FrançoisEXLHardware Hardware coupling of EXL with MUST2
EXL seems to fit in Adrien's setup. We still
have to check if it fits in the chamber.
 EXL_MUST2_top.png 
Fixede748 Tue Jun 6 18:17:03 2017 Valérian et FrançoisEXLHardware Hardware coupling of EXL with MUST2EXL seems to fit with Adrien's setup. 
We still have to test if it fits inside the
chamber.
 EXL_MUST2_top.jpgEXL_MUST2_side.jpg 
Fixede748 Wed Jun 7 11:29:55 2017 AdrienMECHHardwareN/ASurveyor mesurementThe surveyor are carrying out the measurement
of the detectors in D6.
The measurement should take the day, starting
  
Fixede748 Wed Jun 7 18:03:02 2017 FranckOTHERHardware LPC electronicsCaen N568B: 509, 490
Ortec TAC: 1102, 1479, 1480, 1481
Lecroy GDG222: E2557, 01175C, 831
  
Fixede748 Thu Jun 8 18:57:14 2017 Valérie LapouxMUST2Hardware Currents and Voltages on MUST2 Thursday 8th June. 
Vacuum in the chamber 9e-05 mbar; MUST2 temperature
at -5°C.
  
Fixede748 Fri Jun 9 12:30:49 2017 AdrienVACUUMHardware Vacuum LeaksWe reach a stable 1e-5 mbar. We performed
a leak check and identify that the Ionisation
chamber gate valve is not totally tight,
  
Fixede748 Fri Jun 9 12:33:25 2017 AdrienMECHGeneral MUST2 positionAfter a the surveyor came we figured out the
MUST2 telescope are 1cm too high with respect
to the beam line.
  
Fixede748 Fri Jun 9 12:34:18 2017 AdrienDAQHardware HubWe are currentlky waiting for the hub to be
configured so we can start the MUST2 DAQ.
Should be done in the afternoon.
  
Fixede748 Fri Jun 9 12:35:46 2017 AdrienMUST2Hardware CsI trip on T6The detector were left on during the night.
T6 CsI trip and cannot be switch on again.
Once we established that all the MUFFEE works
  
Fixede748 Fri Jun 9 14:41:54 2017 e748CATSGeneral Runs 1 - 4Tests alpha source, dummy data   
Pinnede748 Fri Jun 9 15:49:46 2017 Valérie LapouxMUST2Hardware MUST2 configurationThe MUST2 configurations for telescopes in
the chamber and low/High voltage Caen crate
are given in the attached file.
 Confige748MUST2alimHVandLV.pdfpréparationMUST2forE748.pdf 
Pinnede748 Fri Jun 9 15:51:44 2017 Valérie LapouxMUST2Hardware General Trigger GMT for MUST2 electronicsThe GMT configuration for the trigger inputs
and the inputs from MUST2 electronics are
given in the file.
 DaqElectronicsMUST2gmtforE748.pdf 
Fixede748 Mon Jun 12 11:34:07 2017 e748MUST2Hardware MUVI cars setupWe had problem with MUVI1 (pre serie 1) not
working properly when a telescope was pluggeg
(but fine otherwise).
  
Fixede748 Mon Jun 12 11:36:34 2017 e748MUST2Hardware I2C error on TelescopeAt the Moment we got I2C error on the following
detectors:
  
Fixede748 Mon Jun 12 11:50:34 2017 e748CATSSoftware Settings QDCGates QDC CATS1X, CATS1Y, CATS2X, CATS2Y:
0.996
Pedestal QDC CATS1X, CATS1Y, CATS2X: 0.5
  
Fixede748 Mon Jun 12 11:52:16 2017 e748CATSGeneral Run 6 and 7: Calibration Runs CATSRun 6: Calib CATS1 X&Y
--> Full, x2, x5, pedestal
 Capture.png 
Fixede748 Mon Jun 12 11:55:27 2017 e748CATSGeneral Run 5 - JUNK pulser run on CATS1Dummy run with pulser.
JUNK
  
Fixede748 Mon Jun 12 14:09:24 2017 e748CATSSoftware calibration coefficients Runs 6-7columns are for 1st order calib (y= ax + b):
pedestal - b - a
 CATS_Pulser_Calib_Run6-7.pngCATS1_X.e748CATS1_Y.e748CATS2_X.e748CATS2_Y.e748 
Fixede748 Mon Jun 12 16:56:36 2017 e748OTHERGeneral e748 ProposalExperiment proposal  Matta_PAC_GANIL_June_2016_v3.pdf 
Fixede748 Mon Jun 12 22:48:45 2017 Valérie LapouxMUST2General Tests of the DAQ with the Alpha Source12Th June. 3-peak Alpha source rotated towards
the telescopes MM1 to MM4 (square wall of
4 telescopes).
 8x 
Pinnede748 Tue Jun 13 11:14:23 2017 Valérie LapouxMUST2General GMT configuration, FAGThe Gatconf parameter is now defined in the
action file.
The new configuration for GMT (see chronogramme
 GMTconfigOKMust2FAG100ns.pngGMTtrigParamActions.pngGMTparamDASgatconf.png 
Fixede748 Wed Jun 14 12:13:00 2017 e748MUST2Hardware MUST2 I2C Err debuggingIt appears several SL cables are faulty.
To avoid manipulation of all the cable we
now have T1-4 working. These required to
  
Pinnede748 Thu Jun 15 10:16:03 2017 Valérie LapouxMUST2Hardware Electronics for GMT trigger and FAGThe GMT is used in a mode where we do not
make coincidence and decision, GMT entries
are single and all accepted as triggers of
 DaqElectronicsMUST2gmtforE748.pdf 
Fixede748 Thu Jun 15 17:51:15 2017 e748RUNGeneral3-alphaRun 5-7[SIZE=3]run5[/SIZE]

alpha source 
 MUST2_EXT2.png 
Fixede748 Fri Jun 16 09:57:54 2017 e748RUNGeneral3-alphaRun 11RUN 11 : alpha calibration with suppression
of zeros ( & alignment of pedestals) for
all night
  
Fixede748 Fri Jun 16 09:59:00 2017 e748RUNGeneral60CoRun 12RUN 12 : EXL trigger in GMT
+ gamma source 60Co
 exl_co_source.pngexl_co_source_i.png 
Fixede748 Sat Jun 17 10:22:43 2017 Adrien MattaEXLHardware EXL ADC gateEXL ADC is now setup with individual gate
on the ADC. 
  
Fixede748 Sat Jun 17 10:27:26 2017 Adrien MattaONLINESoftware Starting the GRU online analysisWe setup a script to start the right environment
to run GRU, ganil2root and nptool.
  
Fixede748 Sat Jun 17 10:32:25 2017 Adrien MattaONLINESoftware Online/Offline list of taskWe need to do the following improvement to
the online / offline environment:
  
Fixede748 Sat Jun 17 10:56:09 2017 Adrien MattaBEAMGeneral JC Thomas plans for the beam productionHere are a few notes by Jean Charles about
the expected LISE setting based on 2013 and
2014 16C tuning.
 REX-E628.pdf 
Fixede748 Mon Jun 19 16:03:18 2017 e748EXLGeneral Progress On EXLWe fiddled around with EXL trying to get a
reasonnable 60Co spectrum.
  
Fixede748 Mon Jun 19 16:04:51 2017 e748RUNGeneral3-alphaRun 15Run 15 is started.
Source: We have a 3 alpha source in target
position and a 60Co on the window flange.
  
Fixede748 Mon Jun 19 16:36:07 2017 e748EXLHardware Gain changed on EXLRun 15 online analysis showed waht we believe
was 4 MeV range.
  
Fixede748 Mon Jun 19 16:36:31 2017 e748RUNGeneral60CoRun 16Run 16 is started.
Source: We have a 3 alpha source in target
position and a 60Co on the window flange.
  
Fixede748 Mon Jun 19 18:02:16 2017 e748RUNGeneral60CoRun 17Run 15 is started.
We have a 3 alpha source in target position
and a 60Co on the window flange.
  
Fixede748 Mon Jun 19 18:07:55 2017 e748EXLHardware Threshold changed on EXLRun 15 online analysis showed what we believe
was 4 MeV range.
  
Fixede748 Tue Jun 20 09:08:46 2017 e748EXLGeneral EXL Spectrum with 60Co sourceHere is a spectrum of EXL after one night
of accumulating stat on 60Co.
 exl_60co.ps 
Fixede748 Wed Jun 21 09:05:26 2017 Adrien Matta, Phil AdsleyMUST2General MUST2 Dead TimeWe cabled the Dead time in the scaler.
We have a 100Hz pulser (entry 2 of U2M) in
coincidence with the DT output of the GMT
  
Fixede748 Wed Jun 21 09:39:38 2017 e748DAQGeneral Backup of config fileganacq manip back up.  must2.tar.gz 
Fixede748 Thu Jun 22 10:24:12 2017 e748CATSSoftware Manip config file backup before mergeManip config file backup before merge  e748_cats.tar.gz 
Fixede748 Thu Jun 22 11:10:35 2017 ValérianEXLSoftware EXL MappingMapping of the EXL detector in the analysis
program versus how it is written on the cables.
 Mapping_of_EXL_in_analysis.pdf 
Fixede748 Thu Jun 22 17:46:45 2017 e748MUST2Hardware MUST2 SL cable repairThree cables came back from ASCO, two of them
were the original spare cables.
  
Fixede748 Thu Jun 22 17:49:20 2017 Adrien and Phil MUST2General MUST2 telescope test with alpha sourceAll Telescope now pass the I2C test.

We ran with an alpha source (attached pictures
 T14_X.pdfT14_Y.pdfT58_X.pdfT58_Y.pdf 
Fixede748 Thu Jun 22 17:55:52 2017 e748MUST2Hardware Surveyor mesurementThe surveyor are missing some point for the
telescope measurement.
  
Fixede748 Fri Jun 23 09:18:14 2017 e748OTHERGeneral Plan for 23rd of JuneWe have to open today, I started warming up
of the telescope at 9h18 today.
  
Fixede748 Fri Jun 23 09:20:33 2017 e748OTHERGeneral Plan for coming days23rd of June:
    Fix T5
    Surveyor measurement
  
Fixede748 Fri Jun 23 16:37:54 2017 Adrien, Laurent, VincentOTHERHardware MUST2 chamber openingWe opened the chamber and the surveyor took
there measurement.
 IMG_1936.JPG 
Fixede748 Fri Jun 23 20:49:49 2017 AdrienMUST2General RestartChamber at 2e-5, cooling -9deg outside; -5
on copper LV off.
  
Fixede748 Sat Jun 24 21:41:30 2017 e748MUST2General Pedestal alignement and calibrationMUSt2 pedestal aligned for all telescope.

Started alpha run with all telescope in trigger
  
Fixede748 Sun Jun 25 09:53:44 2017 Adrien MattaRUNGeneral3-alphaRun 18Junk   
Fixede748 Sun Jun 25 09:54:08 2017 Adrien MattaRUNGeneral3-alphaRun 19 Source: 3-alpha + 60Co
Trigger: T1-8 or EXL
  
Fixede748 Mon Jun 26 09:15:31 2017 Adrien MattaRUNGeneral3-alphaRun 19 : disk quota exceededRun 19 was still running this morning, but
file stop writting at 6 O'Clock, when the
disk quota exceeded (we wrtoe about 14Go
  
Fixede748 Mon Jun 26 10:26:16 2017 ValérianEXLSoftware60CoAdd-backWe implemented the add-back correction with
cross-talk subtraction.
 addback.pdf 
Fixede748 Mon Jun 26 17:38:32 2017 e748MUST2GeneralN/ACALIMEROLuc and Greg fixed Calimero, We have to perform
the calibration but there is some work going
on with CATS.
  
Fixede748 Tue Jun 27 10:15:38 2017 e748DAQGeneral ScalersCATS were added to the scalers:
T CATS1 : 3rd scaler channel
T CATS2 : 4th scaler channel 
  
Fixede748 Tue Jun 27 13:47:15 2017 e748RUNGeneralN/ARun 21Calib CATS1:
pedestal - x5  - x2 - full
  
Fixede748 Tue Jun 27 13:47:57 2017 e748RUNGeneralN/ARun 22, 23JUNK   
Fixede748 Tue Jun 27 13:48:22 2017 e748RUNGeneralN/ARun 24CALIB CATS2:
x5 - x2 - full - pedestal
  
Fixede748 Tue Jun 27 15:39:04 2017 e748RUNGeneralN/ARun 26Calib Cats 2

pedestal, x5, x2, full
  
Fixede748 Tue Jun 27 15:41:53 2017 e748RUNGeneralN/ARun 27Calib CATS 1
pedestal, x5, x2, full
  
Fixede748 Tue Jun 27 17:41:17 2017 e748RUNGeneral3-alpharun 28Started run #28
3-alpha
  
Fixede748 Tue Jun 27 17:51:24 2017 e748RUNGeneral3-alphaRun 29Started run #29 at 17h45
3-alpha around 90deg
  
Fixede748 Wed Jun 28 10:52:15 2017 Valérie LapouxMUST2General Memo for CALIMERO MUST2 Time CalibrationTemporary file indicating the main steps for
the time calibration of MUST2 using the Calimero
DAQ package program.
 ConfigDAQDasCalimerov2017vlxTMP.pdf 
Fixede748 Wed Jun 28 10:54:34 2017 Valérie LapouxRUNSoftware DAQ problems due to TCP/IP errorsProblems occurred during the run #29 at 2h38
Wed. 28th June, with the following error
messages written in the Run Control GUI window:
 ProblemRun29TCPerror.png 
Fixede748 Wed Jun 28 11:02:38 2017 Valérie LapouxRUNGeneral3-alphaCalibration run #31 with Alpha source - MM5 and MM8Run #31 started at 11:01, Wednesday 28th June.
The alpha source is rotated towards the telescopes
5, 8.
 Run31startNarvalAlphaSource.png 
Fixede748 Wed Jun 28 11:42:06 2017 e748CATSGeneralN/ASurveyor's positions CATS   Position_geometre_CATS.png 
Fixede748 Wed Jun 28 15:07:43 2017 ThomasCHIOSoftwareN/ATrapezoidal filter for CHIOconst int nCHIO=200; // nb samples. sera ajuste
en manip    
 
  
Fixede748 Wed Jun 28 15:17:24 2017 Valérie LapouxRUNGeneral End of run 31-statusRun #31 started at 11h01; stopped at 15h13
Problem on ganlx9 VME Chio
 TIMEOUT on the stop process.
 Run31stopProbTimeoutGanlx9Chio.png 
Fixede748 Thu Jun 29 10:44:45 2017 e748TARGETHardwareCD2Target and Mask are MountedWe mounted the target and mask in the chamber
before closing:
 IMG_1973.JPGIMG_1954.JPGIMG_1953.JPGIMG_1974.JPG 
Fixede748 Thu Jun 29 23:26:10 2017 Valérie LapouxRUNGeneralEmptyALL CALIMERO Time Calibration MUVI129 JUNE 2017 
CONDITIONS FOR CALIMERO CALIBRATIONS
Time calibrator Period 20ns ; range 512 ns
 CalimeroRunControl.pngConfigCalimeroStat.pngCalibTmuvi1Yperiod20nsrange600ns.pngTimeMust2X_MUVI1_uncalib.pngTimeMust2Y_MUVI1_uncalib.png 
Fixede748 Fri Jun 30 00:33:36 2017 Valérie LapouxRUNGeneralEmptyALL CALIMERO Time Calibration MUVI230 JUNE 2017
CONDITIONS FOR CALIMERO CALIBRATIONS
Time calibrator Period 20ns ; range 512 ns
 CalibTimeMUVI2stripXT.pngSpectresMUVI2stripXT300ns.png 
Fixede748 Fri Jun 30 01:01:08 2017 Valérie LapouxRUNGeneralEmptyALL CALIMERO Time Calibration MUVI2 GOOD RUNS 600ns[B]RANGE TAC MUST2 600ns
CALIMERO TIME CALIBRATOR PERIOD 20ns; Range
512ns[/B]
 TimeMust2X_MUVI2_uncalib.pngTimeMust2Y_MUVI2_uncalib.png 
Fixede748 Fri Jun 30 02:00:44 2017 e748MUST2HardwareN/AWarming up of detector startedAll LV and HT off, vaccumm at 1.3e-5 mbar,
starting warm up sequence.
  
Fixede748 Fri Jun 30 11:09:59 2017 FreddyOFFLINEGeneral Analysis of DSSD Energy Calibration - alpha sourceWe made a calibration with the last alpha
source runs:
-run 19 for T1-2-3-4
 run19_Summary_EX.pdfrun19_Summary_EY.pdf 
Fixede748 Fri Jun 30 12:24:17 2017 Valérie LapouxDAQGeneral DAQ Parameters: TAC, Caviar, LISE D4, D6, Caviar   DASParamVXIMust2CatsTACLise.pngDASParamVXIChioQDC.pngDASParamVXIChioADClisepar.png 
Fixede748 Fri Jun 30 16:49:49 2017 Valérie LapouxDAQGeneral GMT triggers and chronographConfigurations for the GMT inputs.
30 June 2017 pm
all Must2 telescopes,
 DASConfigTriggerGMT.pngChronogrammeGMT30juinFAG100nsCT5µs.png 
Fixede748 Fri Jun 30 16:55:04 2017 Valérie LapouxDAQGeneral Run control for global DAQ with NUMEXOThe run start/stop sequence for the global
DAQ including the NUMEXO:
 RunControlAll748ModeMonitoring.pngNarvalRunControlAllE748.png 
Fixede748 Fri Jun 30 17:53:35 2017 Valérie LapouxDAQGeneral Scalers parameters U2M   Actionse748ganlx11withU2M.pngScalerAnalyzerU2M.png 
Fixede748 Fri Jun 30 20:46:16 2017 Valérie LapouxTARGETGeneral TARGET configurationList of the reaction target in the chamber
from bottom to up:
  
Fixede748 Sat Jul 1 09:27:46 2017 Adrien MattaOTHERGeneralN/ASHIFT TABLE    
Fixede748 Sat Jul 1 10:47:18 2017 Adrien MattaBEAMGeneralN/APrimary beam energy measurementPCP delivered the primary beam energy measurement  IMG_1984.JPG 
Fixed  Sat Jul 1 12:45:25 2017 Anastasia CATSGeneral CATS Calibratedrun_0026 CATS 1 X Y
run_0027 CATS 2 X Y
 cats_calibrated.png 
Fixede748 Sat Jul 1 14:32:02 2017 Valérie LapouxDAQGeneral Scalers U2M, parametersU2M configuration +Actions ganlx11

The signals for the scalers are converted
 ScalersU2Mok.pngCompilActionsganlx11ScalerGr1to4..png 
Fixede748 Sat Jul 1 14:45:17 2017 Valérie LapouxDAQGeneral Parameters for correlation checkingsThe [B]parameters[/B] to be used for the checking
of the correlations between crates/VME/VXI
and good running of the merging
  
Fixede748 Sat Jul 1 15:12:01 2017 Valérie LapouxONLINEHardware Conditions for Bidiv and TAC 1st July 15pm(Update the file CONDTIONS748.txt during the
experiment)
  
Fixed  Sat Jul 1 17:09:04 2017 e748MUST2General TIME CAL CHECKrun_0040_CalMust2Histos.root  must2_time_cal_run40.png 
Fixede748 Sat Jul 1 22:16:46 2017 e748DAQGeneralN/AProblem of MergeThe DAQ was not merging properly.
We figured out that the centrum was goiging
out of sync when the run was stopped.
  
Fixede748 Sun Jul 2 03:46:11 2017 e748OTHERHardware UPDATE OF TAC settingsHere are several changes made during the night
to optimize timing for TACs:
  
Fixede748 Sun Jul 2 04:58:36 2017 e748RUNGeneralN/ARun 74: MASK INStarting Run 74 with trigger plastic and MASK
at target position.
  
Fixede748 Sun Jul 2 05:45:18 2017 e748CATSSoftwareN/AMask CATS 2Run 75 

shift surveyor
 Run75MaskCATS2.png 
Fixede748 Sun Jul 2 05:56:12 2017 e748CATSSoftwareN/AMask on targetRun 74 Mask on target

With surveyor offsets and the inversion X
 Run74MaskTarget.pngRun74MaskTarget_Offsetmv.png 
Fixede748 Sun Jul 2 08:06:53 2017 Valérie LapouxRUNGeneralCD2Run 77 global checkConditions of Run 77

2 July - start at 6h29 - stop at 8h38 
 6x 
Fixede748 Sun Jul 2 08:40:25 2017 Valérie LapouxRUNGeneralCD2Run 78 - 2 July - 18O beam   Run #78 CD2 2.5 mg/cm2 position 3 - 18O
beam 
Same conditions as for Run 77
 Run78ConfdevDataTrig.pngRUN78tPLcats1CORREL.pngRun78CorrelWindowtimePlastCats1.pngRun78TACplastCats.pngRun78TrigMM2impact.png 
Fixede748 Sun Jul 2 08:46:13 2017 Valérie LapouxDAQHardwareCD2Summary of the TAC conditionsJULY 2nd 8pm

Bidiv =  1 for CATS1 and 2
  
Fixede748 Sun Jul 2 09:54:05 2017 Valérie LapouxDAQGeneralCD2Test of the TAC plast CATS ADC VXI Run78   InspectLinesVXIMust2Run78LTtAiCatsFAG.pngInspectVXIchioADCsignalTplchCATS1Run78.png 
Fixede748 Sun Jul 2 10:31:48 2017 Valérie LapouxONLINEHardwareCD2Change of the electronics for TAC Plast CATS1Shift U (Béa +Vlx) entered in the room to
check the electronics of Cats1 and of the
TAC plast-Cats1
  
Fixede748 Sun Jul 2 10:44:59 2017 Valérie LapouxRUNGeneralCD2Run 79 Beam of 18Orun_0079.dat.02-07-17_10h37m20s
Triggers : all MUST2 telescopes 1 to 8.
The pulser is ON again in the DAQ and Scaler
 8x 
Fixede748 Sun Jul 2 11:38:03 2017 Valérie LapouxRUNGeneralCD2Runs 80-81 TESTS change of TAC and triggersFor runs 80- 81: 
for test purposes, we changed the coincidence
mode for CATS: we put FAG instead of the
  
Fixede748 Sun Jul 2 12:08:56 2017 Valérie LapouxRUNGeneralCD2Run 82Run82 start at 12h00 stop at 12h09
e748 Exxx    6684672  2 juil. 12:09 run_0082.dat.02-07-17_12h00m43s
Before starting run 82, we enter in the D6
  
Fixede748 Sun Jul 2 12:16:57 2017 Valérie LapouxRUNGeneralCD2Runs 83-84 only CATS1 triggerRun83 start at 12h11 stop at 12h16
 e748 Exxx   21561344  2 juil. 12:16 run_0083.dat.02-07-17_12h11m16s
 ScalersRun84Cats1trigg.pngConfigTRigCorrelMErgeRun84.png 
Fixede748 Sun Jul 2 13:22:46 2017 Valérie LapouxRUNGeneralCD2Run 85+summary TACStarted run #85 at 13h21 stop at 13h31
e748 Exxx    5439488  2 juil. 13:31 run_0085.dat.02-07-17_13h21m57s
Triggers all MUST2
  
Fixede748 Sun Jul 2 13:31:21 2017 Valérie LapouxRUNGeneralCD2Run 86 trigger CATS1Run 86 start at 13h32 stop at 13h43
 e748 Exxx   34340864  2 juil. 13:43 run_0086.dat.02-07-17_13h32m13s
 
 CorrelRun86mergeOKruncontrol.pngRun86TacPlastCATStrigCats1.png 
Fixede748 Sun Jul 2 13:49:20 2017 Valérie LapouxRUNGeneralCD2Run 88 trigger CATS1Run 88 start at 13h48 
TRIGGER CATS1 div =100 only
18O beam CD2 target
  
Fixede748 Sun Jul 2 14:49:41 2017 Valérie LapouxONLINEHardwareCD2TAC and signals for Plast-CATS1Now we return to the situation of 2 TACs for
the coding of the same time Plast-Cats1 in
the 2 DAQs systems VXI_CHIO and VXI_MUST2.
  
Fixede748 Sun Jul 2 14:54:39 2017 Valérie LapouxRUNGeneral Runs 90-92 Tuning of the cocktail beam in D4Runs 90 to 92 Tuning of LISE 
GMT Trigger IN13=Trig_LISE  (D4)
  
Fixede748 Sun Jul 2 15:13:49 2017 Valérie LapouxRUNGeneral Runs 93-94 Tuning of the cocktail beam in D6Runs 93-94 Tuning of LISE 
GMT Trigger IN13=Trig_LISE (=D6)
 e748 Exxx   25493504  2 juil. 15:11 run_0093.dat.02-07-17_15h07m01s
 ScalerAnalyzerCaviar15h29.png 
Fixede748 Sun Jul 2 17:25:22 2017 Valérie LapouxDAQHardwareCD2ADC parameters-new ones for LISE +Pulser for CorrelationsNew parameter for LISE in VXI_CHIO ADC_CHIO

parameter TED4HF address=4 channel 5 (cable
 DASVXIchioAdcLiseParam.pngConfigDASnew2ndJulyVXIMust2.png 
Fixede748 Sun Jul 2 18:19:24 2017 Valérie LapouxDAQHardwareCD2Change of DAQ parameters for LISE   DASparamLISEVXIChioNEW.png 
Fixede748 Sun Jul 2 18:27:48 2017 Valérie LapouxONLINESoftware Narval configuration, Watcher portsNew file xml is loaded with various Watchers,
the corresponding ports 
Watcher 10205
 RunControlRun101allwtch.png 
Fixede748 Sun Jul 2 19:40:02 2017 Valérie LapouxDAQGeneral Correlation parameters for ChiO in VXI-ChIO VXI_MUST2- Parameters
VXI_CHIO
ADC_CHIO_V15 channel 16, add=15
 Run104ChioCorrelPulser.png 
Fixede748 Sun Jul 2 22:03:45 2017 e748RUNGeneralN/ARun 118
 -> bidon: Beam in D4 but Trig <> LISE
  
Fixede748 Sun Jul 2 23:20:43 2017 LISE team JCT+ChrSBEAMGeneralN/AIdentification plot LISE Run 119Run 119 2nd July started at 22h20 (around)
Lise trigger
PID ED4_TED4HF without wedge 
 Capture-2.png 
Fixede748 Mon Jul 3 03:48:57 2017 e748OFFLINEGeneral Correlation Test on Run 104
We analysed run 104 to check the correlation
of events coded in different crates. It is
 run_104_shifting.pdf 
Fixede748 Mon Jul 3 06:26:24 2017 e748OTHERGeneralN/ADamned ELOG We just recovered the ELOG after 5-6 hrs out
of order
  
Fixede748 Mon Jul 3 06:30:54 2017 e748BEAMGeneralN/A12Be beam rate Final 12Be beam rate as obtained by JCT with
present wedge is about 6.10^3 per euA in
D6
  
Fixede748 Mon Jul 3 06:30:58 2017 e748DAQSoftware PROBLEM of conversion for RUN with even number[COLOR=red][B]IMPORTANT[/B][/COLOR]

We realized that starting at run 160 (included)
  
Fixede748 Mon Jul 3 07:04:30 2017 e748OTHERGeneralN/AtritonsCount rate in ZD plastic is bout 2;5 times
larger than in CATS1 and CATS2, probably
due to tritons 
  
Fixede748 Mon Jul 3 07:06:17 2017 e748RUNGeneralCD2RUN 173 - First Physics with 12BeRUN 173

START: 6H11
 run173_MUST2_Ecal.pngrun173_MUST2_dEE.pngCapture-1.png 
Fixede748 Mon Jul 3 07:52:36 2017 e748RUNGeneralCD2RUN 175[B]RUN 175
[/B]
 
  
Fixede748 Mon Jul 3 07:52:39 2017 e748RUNGeneralCD2RUN 175[B]RUN 175
[/B]
 
  
Fixede748 Mon Jul 3 14:28:09 2017 Franck et ValérianCHIOHardwareN/ACHIO's electronic schemeA quick scheme of the CHIO electronic. The
patch outputs are the one from the LISE acquisition
room.
 Schema_electronique_CHIO.pdf 
Fixede748 Mon Jul 3 17:28:05 2017 e748 -Thomas R, Georges V, Sébastien HCHIOGeneralN/A(Tentative) Solution to recover the CHIO: it works! An interview of Th, Seb and Georges by Valérie
on July 3rd.
Please indicate complements or corrections
 P2280767schemeChioElog.JPGP2280756OscilloChio.JPG 
Fixede748 Mon Jul 3 17:31:08 2017 e748OFFLINEGeneralCD2BEAM and FRAGMENT PIDWe used the Si in D6 and the TOF(CATS1-CAVIAR)
to identify the beam nuclei.
 run_203_PID_SiD6.pdfrun_203_PID_beam.pdfrun_203_PID_reactionfrag.pdf 
Fixede748 Mon Jul 3 18:45:41 2017 Valérie LDAQGeneralN/AChecking ALL RUNS with Ganil_Tape The problems with the runs which could not
be converted during the night was due
to a double header buffers of the runs with
 GanilTapeRun207.pngMessageRunControlNArval.pngRunControlOnNarvalAllWatcher.png 
Fixede748 Mon Jul 3 19:43:16 2017 FranckCHIOHardwareN/ACHIO logic signals for TAC and CAEN digitizerSTOP of CHIO CAEN digitizer = VAL AND Delayed
CHIO (Delay = 700 ns). Before run 217, the
CHIO was not delayed.
  
Fixede748 Mon Jul 3 19:59:02 2017 Valérie LRUNGeneralCD2Series of runs - afternoon/evening 3rd July
Series of runs during the afternoon with
the tuning of the ChIo and the tuning of
  
Fixede748 Mon Jul 3 20:02:46 2017 e748OFFLINEGeneralCD2run217PID run 217
trigger = plast
 Capture.png 
Fixede748 Mon Jul 3 20:29:56 2017 Valérie LRUNGeneralCD2Data runs after LISE tuningThe beam was tuned by the LISE team.
Beam conditions during the run 222 were changed
to increase the beam at 1µA.
 ScalersRun222trigPlastCATS1.png 
Fixede748 Mon Jul 3 20:42:14 2017 Valérie LRUNGeneralCD2Conditions for data runs, Run 223Triggers for the experiments :MM1...MM8,
CATS1 div, Cats2div Plast div
All divisions on triggers are set at 100.
 TriggRun223AllMust2CatsPlast.png 
Fixede748 Mon Jul 3 21:13:18 2017 e748OFFLINEGeneralN/Acatsrun222 cats  Capture.png 
Fixede748 Mon Jul 3 21:41:57 2017 Valérie LRUNGeneralCD2Runs 225-227---------------------------------------------------------------------------------
3 µA at 21h25 ; CAVIAR D6 28000/s
CATS 20000/s
 Run225l3julyScalerAnalyzer3µA.png 
Fixed  Mon Jul 3 22:10:28 2017 Valérie LCATSHardwareCD2VXI CATS crate problem solved We had problems with CATS VME/VXI crate.
It was idle and it was not possible to restart
it by breakup and init. 
 VisuCATproblemTerm.pngVXICATSreset.png 
Fixede748 Mon Jul 3 22:24:05 2017 Valérie LRUNGeneralCD2Run series of problems 3rd July The daq was unstable with several problems
of the merge from LISE.
We did tk kna (kill narval process) and na.
  
Fixed  Mon Jul 3 22:40:33 2017 e748RUNGeneral Plastic_Cats_TTac run_223
 
Range: T_CATS1_CAV 1 µs
 run_223_tac2.pdfrun_223_tac1.pdf 
Fixede748 Mon Jul 3 23:06:24 2017 Valérie LRUNGeneralCD2Problem during Run 230-231 with CATSr-- 1 e748 Exxx   34013184  3 juil. 22:52
run_0230.dat.03-07-17_22h46m03s
-rw-r--r-- 1 e748 Exxx   97910784  3 juil.
 ResetEventCounterAfterRun230Centrum.pngPRoblemRun230.pngScalersRUn230.png 
Fixede748 Mon Jul 3 23:16:33 2017 Valérie LDAQGeneralCD2Change of TAC RANGES (from run 230)TAC range changed after run 229.
NB: Change applied for Runs 230 and 231 
BUT these runs are not working due to problems
  
Fixed  Tue Jul 4 00:53:18 2017 Valérie LRUNGeneralCD2Runs with merge and time out problems-r--r--r-- 1 e748 Exxx     720896  3 juil.
21:55 run_0226.dat.03-07-17_21h54m51s
-rw-r--r-- 1 e748 Exxx     786432  3 juil.
  
Fixed  Tue Jul 4 00:54:41 2017 Valérie LRUNGeneralCD2Runs after reset of VXI
VXI CHIO was offline --> reset of ganlx9
 (entering in D6) on the VXI crate 
  
Fixede748 Tue Jul 4 01:00:26 2017 Valérie LRUNGeneralCD2Run 239 test with trigger MUST2Started run #239 4JULY start 00h59 stop 1h15
PROBLEM junk header checked with Ganil Tape
triggers only MUST2 IN1 to IN8
  
Fixede748 Tue Jul 4 01:17:35 2017 Valérie LRUNGeneralCD2Run 240Run #240 4JULY start  1h17
(run ok checked with Ganil_Tape)
 All triggers   MUST2 IN1 to IN8, 
  
Fixede748 Tue Jul 4 01:21:04 2017 e748OFFLINEGeneralCD2run 236cats > target
cats > target, gated by must
 Capture.png 
Fixede748 Tue Jul 4 02:33:36 2017 Valérie L, shifts U+S YorickRUNHardware GMT channel 8 Apparently Channel 8 of the GMT is not operational.
We put telescope T8 in entry 15.
This operation gives good results, see pictures
 GMTtriggerRun248.pngGMTtriggerGatconfRun248.png 
Fixede748 Tue Jul 4 04:01:50 2017 e748RUNHardware RUN249RUN249 

Same setting as RUN248. 
  
Fixede748 Tue Jul 4 05:01:21 2017 YorickRUNGeneralCD2Runs 248-250These are long runs with CD2
MUST8 is in GMT 15.
Missing Times: T7 X and Y; T5 Y.
  
Fixede748 Tue Jul 4 05:08:30 2017 e748MUST2Hardware MUST2 signql check in offline analysisRUN248
MUST2 DSSD XE signal for all telescopes T1-T8
confirmed.
  
Fixede748 Tue Jul 4 05:11:20 2017 e748MUST2Hardware MUST2 signql check in offline analysisRUN248
MUST2 DSSD XE signal for all telescopes T1-T8
confirmed.
 10x 
Fixede748 Tue Jul 4 05:20:31 2017 e748RUNHardware RUN250Same setting as RUN248. 
run_0250.dat.04-07-17_04h58m50s
start 4h58 Gtape ok
  
Fixede748 Tue Jul 4 05:23:20 2017 e748CATSSoftware CATS signalRUN248
The chqrge distribution of CATS1&2, X&Y
Gate:CATSdiv trigger
 run248_CATS1XQ.pngrun248_CATS1YQ.pngrun248_CAT2XQ.pngrun248_CATS2YQ.png 
Fixede748 Tue Jul 4 07:00:55 2017 e748RUNHardware RUN252Same setting as RUN248. 
RUN251 junk header
run_0252.dat.04-07-17_05h33m06s
  
Fixede748 Tue Jul 4 07:01:50 2017 e748RUNHardware RUN254Same setting as RUN248. 
RUN253 junk header
run_0254.dat.04-07-17_06h32m10s
  
Fixede748 Tue Jul 4 07:37:09 2017 e748RUNHardware RUN256Same setting as RUN248. 
RUN255 junk header
run_0256.dat.04-07-17_07h34m59s
  
Fixede748 Tue Jul 4 07:46:48 2017 e748RUNHardware RUN258Same setting as RUN248. 
RUN257 junk header
run_0258.dat.04-07-17_07h45m32s
  
Fixede748 Tue Jul 4 08:36:50 2017 e748RUNHardware RUN260Same setting as RUN248. 
RUN259 junk header
run_0260.dat.04-07-17_08h32m05s
  
Fixede748 Tue Jul 4 09:01:19 2017 e748RUNHardware RUN262Same setting as RUN248. 
RUN261 junk header
run_0262.dat.04-07-17_08h56m13s
  
Fixede748 Tue Jul 4 09:48:53 2017 e748RUNGeneralCD2Run 265Run 265: Trigger MUST Only   
Fixede748 Tue Jul 4 09:49:52 2017 e748RUNGeneralCD2Run 266Run 266: Trigger CATS1 Only   
Fixede748 Tue Jul 4 09:50:11 2017 7e748RUNGeneralCD2Run 267Run 267: Trigger CATS2 Only   
Fixede748 Tue Jul 4 09:51:55 2017 8e748RUNGeneralCD2Run 268Run 268: Trigger Plastic Only   
Fixede748 Tue Jul 4 09:57:48 2017 e748RUNGeneralCD2Run 269Run 269: Trigger MUST2 Only
Checking CHIO Digitization
  
Fixede748 Tue Jul 4 10:21:52 2017 e748RUNGeneralCD2Run 270Run 270: Trigger CATS1 or CATS2 Only
Checking CHIO Digitization
  
Fixede748 Tue Jul 4 10:34:06 2017 e748RUNGeneralCD2Run 271Run 271: Trigger [B]Telescope 2[/B]  Only
 (this one sees less beam particles)
Checking CHIO Digitization
  
Fixede748 Tue Jul 4 11:00:21 2017 e748RUNGeneralCD2Run 272Run 272: Trigger Plastic  Only 
Checking CHIO Digitization
  
Fixede748 Tue Jul 4 11:02:07 2017 e748RUNGeneralCD2Run 273Run 273: Trigger CATS1 and CATS2  Only  
Checking CHIO Digitization
  
Fixede748 Tue Jul 4 11:04:46 2017 e748RUNGeneralCD2Run 274Run 274: Trigger T5-8  Only  
Checking CHIO Digitization
  
Fixede748 Tue Jul 4 13:15:11 2017 e748OTHERGeneralN/Atroubles with CHIO digitizer 
1p.m.
CHIO Digitizer was fixed by Frederic and
  
Fixede748 Tue Jul 4 13:24:40 2017 e748RUNGeneralCD2Run 293Run 293  
restart with all triggers. Chio digit fixed
  
Fixede748 Tue Jul 4 14:24:37 2017 e748RUNGeneralCD2Run 296Run 296 
restart with all triggers. Chio digit fixed.
NUMEXO started
  
Fixede748 Tue Jul 4 14:48:11 2017 8e748RUNGeneralCD2Run 296Run 298
restart with all triggers MUST2
  
Fixed  Tue Jul 4 14:49:17 2017 e748MUST2General CsI THRESHOLD RUN 297By removing the threshold in the CsI we can
clearly see now the light particles well
separated in the DE/E 
 Pedestal_adjustment_CsIE.png 
Fixede748 Tue Jul 4 15:50:02 2017 8e748RUNGeneralCD2Run 296Run 299
restart with all triggers MUST2
  
Fixede748 Tue Jul 4 18:29:39 2017 e748MUST2SoftwareN/AStart a new acquisition with Numexo-To start a run :

  START  Acquisition NUMEXO
 CaptureSResetCHIODig.png 
Fixede748 Tue Jul 4 19:38:32 2017 e748RUNGeneralCD2Run #323 main daq - Run #322 NUMEXO2Run #323 on main daq <=> Run #322 NUMEXO2
Physics run: 12Be on CD2 #3 2.5 mg
On next run we will try to have the correct
  
Fixede748 Tue Jul 4 20:27:59 2017 Franck et ValérianCHIOHardwareN/AUPDATE : CHIO's electronic scheme (valid from run 217)Modification of the scheme to add the delay
on the CHIO's logic signal for stop of the
digitizer.
 Schema_electronique_CHIO.pdf 
Fixede748 Tue Jul 4 20:30:52 2017 e748RUNGeneralCD2Run #323 scaler captureRun #323 scaler capture at 20h30  Capture-SCALERANALYSER_run323.png 
Fixede748 Wed Jul 5 00:29:06 2017 Valérie LRUNGeneralCD2Summary of the evening runs 322(e748chio)-323 (e748)============================================================================================================
[e748'@ganp616 run]$ cd /data/e748chioX/e748chio/acquisition/run
-r--r--r-- 1 e748chio Exxx 4194288609  4
 RunControlChIoRun322.pngRunControlRun323.png 
Fixede748 Wed Jul 5 00:40:36 2017 Valérie LRUNGeneralCD2Overview of the evening config 322(e748chio)-323 (e748)BEAM conditions
CAVIAR LISE 35k
GMT TRiggers:
 ScalersRun323e748at0h48.pngConfigCEntrumE748.pngPulserChIoForScalerDAQ.png 
Fixede748 Wed Jul 5 00:58:43 2017 Valérie LRUNGeneralCD2Overview of the evening plots Run323-e748 (Run 322 chio)Run 323 DAQ E748; Run 322 DAQ CHIO see plots
below.
 19x 
Fixede748 Wed Jul 5 01:07:15 2017 e748OFFLINESoftware Merge of NUMEXO runs with standard runsHere are some plots after offline merge correlation
of the CHIO energy coded in NUMEX0 with other
parameters.
 run200_IC_numexodiff.pngrun300_ICENUMEXO_TCATSCAV.pngrun300_ICENUMEXO_TPLCATS1.pngrun300_TPLCHIO_LTSdiff.png 
Fixede748 Wed Jul 5 03:35:53 2017 Valérie LRUNGeneralCD2SiLi 1/T5 tripAT 3h30 we saw that the SiLi1 of T5 has tripped
during the run run_0323.dat.04-07-17_19h25m59s.2
(run started at 2h22)
  
Fixede748 Wed Jul 5 04:34:06 2017 e748 shift URUNGeneralCD2Overview of Config and TAC during run 323 (.3)The counting rates in the Gatconf, confdev,
dataTrig of the different branches are checked
in the first plot below.
 run_323_conf.pngrun_323_tac.png 
Fixede748 Wed Jul 5 04:55:06 2017 Valérie LRUNGeneralCD2Scalers and summary of Run 323 at 4h55-7h30E748 Runs (around 3h30 for 1Go)

Run 323_2  started at 2h22 stop 5h56
 ScalersRun323AT4h54.pngRunControlBuffer5julyAt6h.png 
Fixede748 Wed Jul 5 06:11:47 2017 e748 shift URUNGeneralCD2Overview of Config +QplastTcats1 during run 323 (.5)The counting rates in the Gatconf, confdev,
dataTrig of the different branches are checked
in the first plot below.
 Run323QplastTcatsconf.pngRunControlBuffer5julyAt6h.png 
Fixede748 Wed Jul 5 08:10:03 2017 Valérie LRUNGeneralCD2Run 325 DAQ 748- run 325 DAQ ChIoWe stopped the run 323 (DAQ E748) and run
322 (ChIo) when the beam was stopped by the
machine during 10 minutes to check the beam
 Run325DAQE748daqChIo.pngRun325ChioDigET.pngScalerAnalyserRun325at8h30.pngGMTtriggerAllRun325.png 
Fixede748 Wed Jul 5 09:13:17 2017 e748-A. LemassonCHIOHardwareCD2CHIO cathode problemsThomas and Julien going to D6, the cathode
power supply tripped.
Setting it back and we will check signal
  
Fixede748 Wed Jul 5 09:22:16 2017 Valérie LRUNGeneralCD2Run 326DAQ- Run 326ChioRun 325 start at 9h19
We are running again after some change -switching
on- of the power supply of the ChIo cathode).
 Run326controlBuffer.pngScalerAnalyzerRun326.pngProblemCaviarEndOfRun326.png 
Fixede748 Wed Jul 5 10:54:09 2017 Valérie LRUNGeneralCD2SiLi 1/T5 tripAT 10h50 we saw that the SiLi1 of T5 has tripped
during the run run_0328.dat.  (run started
at 10h15 )
  
Fixede748 Wed Jul 5 10:58:54 2017 Valérie LRUNGeneralCD2Run summary 327, 328run daq E748  (3h30 for 1Go)
-r--r--r-- 1 e748 Exxx  100M  5 juil. 10:14
run_0327.dat.05-07-17_09h55m11s
 Run328Scaler.pngCheckParamBytesEvtNumexoRun328.png 
Fixede748 Wed Jul 5 11:10:48 2017 Valérie L.MUST2GeneralCD2Geometrical configurations for MUST2   ConfigGEOMImpactMUST2e748.pdf 
Fixede748 Wed Jul 5 11:13:27 2017 Valérie L.MUST2GeneralCD2Power boards HV and LV for MUST2   ConfigHVetLVExpE748MUST2d6v2017.pdf 
Fixede748 Wed Jul 5 14:34:15 2017 MarleneRUNGeneralCD2Run summary 329,330We thought the merge was broken, but the issue
was the Raw Data gruscipt connecting to the
wrong port.
 Scalers_run330.pngTAC_run330.pngcorr_run330.pngexl_run330.pngCapture-1.png 
Fixede748 Wed Jul 5 15:30:03 2017 FreddyOFFLINEGeneral MUST2 offline analysis of run 323Here is a plot of the MUST2 PID for telescope
3:
- dE - TOF with a veto on CsI_ to remove
 PID_M2_run332.png 
Fixede748 Wed Jul 5 17:15:57 2017 e748 DBRUNGeneralCD2screenshot of scalers primary I = 6.9 uA

[IMG]elog:209/1[/IMG]
 Capture-run330.png 
Fixede748 Wed Jul 5 18:36:49 2017 FreddyOFFLINEGeneral PID from NUMEXO merged (partially)Here is the PID(dE Vs E)  plot obained using
the IC energy from NUMEXO Vs the plastic
energy coded in our acquisition for the long
 DeE_CHIO_NUMEXO_run323.png 
Fixede748 Wed Jul 5 18:47:25 2017 e748 DBRUNGeneralCD2new run - 331primary I = 6.9 uA

[IMG]elog:211/1[/IMG]
 Capture-run331.png 
Fixede748 Wed Jul 5 19:05:13 2017 e748 DB & ISRUNGeneralCD2tritons in the beamTritons can be seen from the following plot.
Some of them cross target frame and then
lose 1.5MeV in the DSSDs. 
 Capture-tritons.png 
Fixede748 Wed Jul 5 19:10:33 2017 e748 DB&ISONLINEGeneralCD2tritons in the beamTritons can be seen from the following plot.
Some of them cross target frame and then
lose 1.5MeV in the DSSDs. 
 Capture-tritons.png 
Fixede748 Wed Jul 5 21:13:14 2017 e748 ISRUNGeneralCD2RUN 331 scalers primary beam I = 6.5 uA unstable in the last
30 minutes (20h20 - 20h45)
  
Fixede748 Wed Jul 5 22:29:10 2017 e748 DBONLINEGeneralCD2trajectories[IMG]elog:215/1[/IMG]  Capture-traj.png 
Fixede748 Thu Jul 6 00:07:33 2017 e748 DBONLINEGeneralCD2Strips CATS raw [IMG]elog:216/1[/IMG]  Cats-raw-run331.png 
Fixede748 Thu Jul 6 01:00:21 2017 FreddyMUST2HardwareN/AProblem with position of télescopes 6&7Due to a cabling problem telescope 6 was seen
as telescope 7 and inversly.
 Nouveau_document_2017-07-06.pdf 
Fixede748 Thu Jul 6 01:26:15 2017 Franck et ValérianCHIOHardwareN/AUPDATE : CHIO's electronic scheme (valid from run 293)Since Run 293 inspection line of NUMEXO 2
has been used as input signal of CAEN digitizer.
 Schema_electronique_CHIO.pdf 
Fixede748 Thu Jul 6 01:34:42 2017 e748RUNGeneralCD2Run 331 (file change)1:25 file change -> run_0331.dat.05-07-17_18h45m44s.2
18O beam: 6.6 uAe
CATS1: 40000 cps
 Scalers_run331_2.png 
Fixed  Thu Jul 6 02:02:48 2017 AnastasiaOFFLINEGeneral First signs of the elasticHereby you can see some first signs of the
[U]elastic scattering[/U] [COLOR=purple]12Be(d,d)12Be[/COLOR]
with punch through events that will be later
 Elastic.png 
Pinnede748 Thu Jul 6 02:41:13 2017 Freddy and the night shiftersCHIOGeneral CHIO analog treatmentInterestingly, the bipolar output of the amplifier
of the CHIO sent (and splitted) to check
correlation on the channel 15 of the ADC
  
Fixede748 Thu Jul 6 03:03:39 2017 iFreddy and the night shiftersCHIOGeneral CHIO analog (unipolar and bipolar) and digitalInterestingly, the bipolar output of the amplifier
of the CHIO sent (and splitted) to check
correlation on the channel 15 of the ADC
 run323_chiocompare.pngDeE_CHIO_NUMEXO_run323.pngrun323_CHIO_DIGIVAEN.pdf 
Fixede748 Thu Jul 6 03:35:32 2017 e748RUNGeneralCD2Run 331 (file change)3:30 file change -> run_0331.dat.05-07-17_18h45m44s.3
18O: 7.2 uAe
CATS1: 40000 cps
 Scalers_run331_3.png 
Fixede748 Thu Jul 6 05:48:26 2017 e748MUST2GeneralCD2SiLi 1/T5 tripRestart SiLi 1/T5 that reached 26 uA and tripped
a few minutes ago.
Current = 21 uA when biased up again, and
  
Fixede748 Thu Jul 6 06:23:18 2017 e748RUNGeneralCD2Run 331 (file change)6:19 file change -> run_0331.dat.05-07-17_18h45m44s.4
18O 7.1 uAe
CATS1 : 36000 cps
 scalers_run331_4.png 
Fixede748 Thu Jul 6 07:17:05 2017 e748MUST2GeneralCD2SiLi 1/T5 tripSiLi 1/T5 tripped at 7:16.
We turn it on again immediately -> 23 uA
  
Fixede748 Thu Jul 6 07:23:22 2017 e748MUST2SoftwareCD2MUST2 : VMECOM : TCP error - TCP/IP : time-out"MUST2 : VMECOM : TCP error - TCP/IP : time-out"
at 7:19.
RRC shows MUST2 "OFFLINE"
  
Fixede748 Thu Jul 6 07:58:09 2017 e748MUST2GeneralCD2SiLi 1/T5 tripedSiLi 1/T5 triped at 7:30. We leave it off
while trying to restart the run control.
We turn it on again at 8:03.
  
Fixede748 Thu Jul 6 08:22:08 2017 Valérie LRUNGeneralCD2Problem of DAQ and networks problemsAfter the problem on Run control and the offline
of MUST2, the DAQ on numexo could not be
started again.
 ProblèmeDAQ6juillet8h.pngProblemGUIRuncontrolNumExo.pngRebootsoftCrateAGAVApassword.pngProblemDAQ5juilletNumExo.png 
Fixede748 Thu Jul 6 09:54:53 2017 Valérie LRUNGeneralCD2Run control OK -Runs 339 and 340After the reboot soft of the AGAVA NUMEXO
and realignment of the TGS (time stamp Agava)
the situation is stabilized.
 CorrelBidimChioandplotCatsRun340.pngRun339scalers.pngScalersAnalyserRun340.png 
Fixede748 Thu Jul 6 10:35:50 2017 Valérie LMUST2GeneralCD2Sili-1/T5 tripThe Sili1 of T5 has tripped (again) at 10h30.
On again. 400V; I ~ 20µA at 10h40, still
increasing 25.6µA at 11h30.(The limit Io
  
Fixede748 Thu Jul 6 10:39:09 2017 BéaF & ValérieL - shift UOFFLINEGeneral Test of merge on Run 339spectra of the merge checked by Béa.
The merge is correct on the correlations
cf plots below.
 run339_ICEdiff_LTSDIff_TrigM2.pngrun339_TAC_LTSDIff_TrigM2.png 
Fixede748 Thu Jul 6 11:41:20 2017 Valérie LDAQGeneralN/AMEMO START/STOP + AGAVA configuration in DASSTART/STOP of the DAQ
---------------------------------------------------------------------
START acquisition e748chio  (DAQ Numexo;
 AfterRebootAgavaRegStatusOx80entréeAfterAlignt.pngDasVMEChioCheckSamplesNumber.pngRunControl748On.pngRunControlNarvalE748chioON.pngParameterListWatcherOnRCe748chio.png 
Fixede748 Thu Jul 6 13:21:49 2017 DaisukeRUNHardwareCD2RUN340During RUN340 

13h10 The beam intensity dropped from 35
  
Fixede748 Thu Jul 6 13:33:23 2017 e748RUNGeneralCD2Stop of Run 340 due to beam problemsThe run 340 was stopped at 13h32. Before the
stop fo the run the beam was unstable and
stopped by machine team.
  
Fixede748 Thu Jul 6 13:37:22 2017 Valérie LRUNGeneralCD2Run 341Run 341 START AT 13h37 STOP AT 15h21 (beam
stop see the ELOG entry Id=239)
At the beginning (cf scalers below) the beam
 ScalersRun341Beginning.pngScalersRun341At14h35.png 
Fixede748 Thu Jul 6 13:37:37 2017 Adrien MattaOFFLINEGeneralN/AEnergy loss table for d, 3He and 3HHere it is!  20x 
Fixede748 Thu Jul 6 14:07:20 2017 BeaOFFLINEGeneralCD2Convert & Merge Numexo Runs in ganp166To Convert run DAQ_general

offline
  
Fixede748 Thu Jul 6 15:24:25 2017 e748-shift UBEAMGeneralCD2Beam stop, end of run 341The beam was stopped due to vacuum problem
at CSS2.
End of run 341 at 15h21.
  
Fixede748 Thu Jul 6 17:05:51 2017 Valérie LRUNGeneralCD2run 342 +overview of TACBeam recovered after shut down due to the
vacuum problem (power supply on the vacuum
system was off). Beam was off during 1h30.
 ScalersRun342at17h10.pngCorrelDatTrigGatconfRun342.pngCorrelTpulserChiorun342.pngRun342_Tcats12.pngRun342_TACcats1.png 
Fixede748 Thu Jul 6 19:17:12 2017 e748 MARUNGeneralCD2RUN 343Run 343
same conditions as run 342
beam=7uA
  
Fixede748 Thu Jul 6 20:07:46 2017 Daisuke + Thomas R+ Valérie LCHIOGeneralCD2Problem of counting on ChioIt seems that the ChIo is counting at a very
low rate, around 100/s instead of 5000/s
in previous runs.
 ScalersRun343At20h09.png 
Fixede748 Thu Jul 6 20:46:15 2017 Thomas R + shift U (Valérie L) +S (Daisuke, Marlène, Shumpei) RUNGeneralCD2Run 345 DAQ+344 chio after chio problemThe HV on cathode is now set again at 2900V.
At 20h50 we get back the beam for the run.
(Grid still ok at 920V)
 ScalersRun345at20h54.png 
Fixede748 Thu Jul 6 21:27:52 2017 Thomas R + shift U (Valérie L) +S (Daisuke, Marlène, Shumpei) RUNGeneralCD2Run 345 beam intensity lossThe beam rate was low after intervention with
the CHIO HV. 
  
Fixede748 Thu Jul 6 21:35:29 2017 Thomas R + shift U (Valérie L) +S (Daisuke, Marlène, Shumpei) RUNGeneralCD2Run 346 e748+e748chio
Run 346 started at 21h37
Beam around 7µA CATS 2 46k
 6x 
Fixede748 Fri Jul 7 00:25:16 2017 e748MUST2SoftwareCD2Calibration for CsIRough energy calibration for CsI was performed
(only for MM1 and MM4).
  
Fixede748 Fri Jul 7 00:42:50 2017 ShumpeiMUST2SoftwareCD2Calibration for CsIRough energy calibration for CsI was performed
(only for MM1 and MM4).
Method;
 CsICalib.pdf 
Fixede748 Fri Jul 7 00:43:51 2017 ShumpeiMUST2SoftwareCD2Calibration for CsIRough energy calibration for CsI was performed
(only for MM1 and MM4).
Method;
 CsICalib.pdf 
Fixede748 Fri Jul 7 02:28:26 2017 Anastasia + DidierONLINEGeneralCD2scalers during run 346[IMG]elog:249/1[/IMG]  Capture-scalers-run346.png 
Fixede748 Fri Jul 7 02:56:25 2017 Anastasia + DidierONLINEGeneralCD2CATS raw 2D charge during run 346[IMG]elog:250/1[/IMG]  Capture-catsraw-run346.png 
Fixed  Fri Jul 7 06:55:54 2017 Anastasia + IulianONLINEGeneralCD2Run 346 Run 346 still running.
We see in the Must2 impact also the other
two telescopes T6T7
 MUST2_7am.pngScalers7am_run346.png 
Fixed  Fri Jul 7 07:34:54 2017 Anastasia + IulianONLINEGeneralCD2Run 346Beam intensity: 7.433 µA   
Fixede748 Fri Jul 7 08:17:38 2017 e748RUNGeneralCD2Run 347Stop run 346 and start run 347 after rewriting
parameters onto CHIO digitizer.
Digitized CHIO signals back in Vigru.
 Scalers_run347.png 
Fixede748 Fri Jul 7 08:56:43 2017 e748MUST2GeneralCD2SiLi 1/T5 currentSiLi 1/T5: 380V, 24.9 uA (current limit set
at 26 uA)
Apparently, it hasn't triped since July 6,
  
Fixede748 Fri Jul 7 10:46:21 2017 e748MUST2GeneralCD2SiLi 1/T5 currentHV = 380 V
I = 25.1 uA
  
Fixede748 Fri Jul 7 10:56:12 2017 e748RUNGeneralCD2Run 347 (file change)10:50 data file change -> run_0347.dat.07-07-17_08h15m06s.1
18O: 7.5 uAe
CATS1 42 kcps
 scalers_run347_1.png 
Fixede748 Fri Jul 7 12:00:29 2017 e748RUNGeneralCD2Stop run 347 - Short machine problemShort beam interruption due to a machine problem.
We stop run 347.
  
Fixede748 Fri Jul 7 12:02:02 2017 e748RUNGeneralCD2Run 348Restart after short machine problem.
Same conditions as previous run.
 Scalers_run348.pngcorrel_run348.png 
Fixede748 Fri Jul 7 15:15:42 2017 e748BEAMGeneralCD2Runs 349 to 364 - 10Be LISE tuningSlits F31 = +/- 20 mm
Momentum acceptance = 2.5 % total
  
Fixede748 Fri Jul 7 15:20:45 2017 e748BEAMGeneralCD2Run 365 - 10Be LISE tuningSlits F31 = +/- 4 mm
Momentum acceptance = 0.5 % total
2 uAe -> 80000 cps on CATS
  
Fixede748 Fri Jul 7 15:55:29 2017 e748BEAMGeneralCD2Run 366 - 10Be tuning - Centering on TARGETFH31 slits = +/- 4 mm
DAC2 = -3 A
 run_366_dac2_moins3A_CATS.pngrun_366_dac2_moins3A_Target.png 
Fixede748 Fri Jul 7 16:01:22 2017 e748RUNGeneralCD2Run 367FH31 slits (momentum slits) = +/- 4 mm
DAC2 = -3 A
Triggers = MUST2, CATS1div, CATS2div
 run_367_scalers.png 
Fixede748 Fri Jul 7 16:32:24 2017 e748RUNGeneralCD2summary of runs 363-367
/data/e748X/e748/acquisition/run
r--r--r-- 1 e748 Exxx   19M  7 juil. 15:01
  
Fixede748 Fri Jul 7 16:34:34 2017 e748 shift U Simon G. + Valérie LRUNGeneralCD2Overview of Run 368 - 10Be beamRun 368 started at 16h30 stop at 18h01 (automatic
stop)
 12x 
Fixed  Fri Jul 7 18:04:58 2017 e748OFFLINEGeneral Kinematic Line12Be(d,3He)11Li    kine.pdf 
Fixede748 Fri Jul 7 18:18:13 2017 e748 Thomas R. +shift U Simon G +Valérie LCHIOGeneralCD2Thomas's explanations on Agava
AGAVA
cf drawing below done by Thomas.
 P2280829explanationChIo.JPG 
Fixede748 Fri Jul 7 20:26:30 2017 e748 shift U Simon G +Valérie LRUNGeneralCD2Runs 368, 368_1, 368_2, 368_3 10Be runsRUNS 368, 368_1, 368_2
I ~ 3.7 eµA (on D6 TI22)
General DAQ 1Go during 1h30 
 ScalersRun368At20h30.pngTAC_Run368_At21h.pngScalersRun368at21h52.png 
Fixede748 Fri Jul 7 22:02:39 2017 e748 shift U Simon G +Valérie LCATSGeneralCD2CATS situationDuring Run 368, we note at 21h30 that now
CATS2 has continuously a 0.1 µA current.
Maybe there are some microtrips on some central
 StrQRawCats_Run368At22h.pngPosCats_Run368At22h.png 
Fixede748 Fri Jul 7 22:36:12 2017 e748 shift U Simon G +Valérie LBEAMGeneralCD2Beam during Run368, situation at 22h30After strong instabilities the beam intensity
is better and we have recovered nearly 93kcps
on CATS2 as during the beginning
 ScalersRun368_At22h35.pngP2280831ecranTIat22h40.JPG 
Fixede748 Fri Jul 7 23:45:52 2017 e748 shift U Simon G +Valérie L (-12 am)/ Daisuke+Shumpei+Marlene (12 am -)RUNGeneralCD2Run 368 (following) at 23h45At 23h45; Beam stable I  3.95 µA
CATS2 ~ 94 kcps (2 pictures of CATS at 4
hours difference attached)
 6x 
Fixede748 Sat Jul 8 06:02:18 2017 e748 Marlène Daisuke ShumpeiRUNGeneralCD2RUN369Caviar has tripped during the night (before
4a.m.)
We have put the bias again but we could not
 scalers_e369.png 
Fixede748 Sat Jul 8 08:39:10 2017 Daisuke, ShumpeiOFFLINEGeneralCD212Be(d,t)?/home/e748/analysis/nptool/Outputs/Analysis/runtest.root
(run_0331_0.root?) was analyzed.
DSSD energy vs theta_lab plot for triton
 2017-07-08_run331_12be_dt_likely.png2017_07_08_gate.png2017_07_08_chio_gate.png 
Fixede748 Sat Jul 8 08:39:37 2017 Shumpei, DaisukeOFFLINEGeneralCD212Be(d,t)?/home/e748/analysis/nptool/Outputs/Analysis/runtest.root
(run_0331_0.root?) was analyzed.
DSSD energy vs theta_lab plot for triton
 2017-07-08_run331_12be_dt_likely.png2017_07_08_gate.png2017_07_08_chio_gate.png 
Fixede748 Sat Jul 8 09:00:32 2017 e748 Anastasia, D. RUNGeneralN/Aerror msg on run control (run 369)[IMG]elog:274/1[/IMG]  Capture-error-msg-run369.png 
Fixede748 Sat Jul 8 09:19:37 2017 e748 Anastasia, D. RUNGeneralN/ACATS raw 2D charge during run 369[IMG]elog:275/1[/IMG]  Capture-catsraw-run369.png 
Fixede748 Sat Jul 8 09:43:17 2017 e748 Anastasia, D. RUNGeneralN/ADAQ Parameters: Caviar new run 370Threshold of CAVIAR_DDM changed from ~ -800
to ~ -510
run 370 Triggering only in Cats to check
  
Fixede748 Sat Jul 8 11:42:43 2017 e748 Anastasia, D. RUNGeneralN/Anew Run 371Triggers MUST2 and CATS as before, 
-rw-r--r-- 1 e748 Exxx   16M  8 juil. 11:43
run_0371.dat.08-07-17_11h41m39s
 Caviar_DDM4_thresholds.pngCaviar_DDM3_thresholds.png 
Fixede748 Sat Jul 8 12:53:09 2017 e748 Anastasia, D. RUNGeneralN/ARUN 371Changing Caviar thresholds during the run
to find the optimum efficiency 
More specific The DDM16_4 the channels 4
 caviar_par.pngcaviar_par2.png 
Fixede748 Sat Jul 8 14:00:06 2017 e748 Anastasia, D. RUNGeneralCD2Run 372
Start new run with final CAVIAR threshold
settings 
 DDM16_3.pngDDM_4.pngDDM16_5.png 
Fixede748 Sat Jul 8 16:39:00 2017 e748RUNGeneralCD2Stop run 372 - CHIO digitizer lost parameters    
Fixede748 Sat Jul 8 16:41:57 2017 e748RUNGeneralCD2Run 373Restart after rewriting parameters onto CHIO
digitizer
Same conditions as previous run
 scalers_run373.pngCATS_raw_run373.pngCATS_run373.png 
Fixede748 Sat Jul 8 17:04:25 2017 e748RUNGeneralCD2RUN 374We stopped run 373 to check the MUVI scalers
on the DAS.
We wrote the parameters onto the CHIO digitizer
 scalers_run374.png 
Fixede748 Sat Jul 8 19:41:25 2017 e748RUNGeneralCD2Run 374 (file change)19:29 data file change -> run_0374.dat.08-07-17_16h57m56s.2
Previous change : 18:11 -> run_0374.dat.08-07-17_16h57m56s.1
 scalers_run374_2.pngCATS_run374_2.pngCATS_raw_run374_2.png 
Fixede748 Sat Jul 8 20:49:48 2017 e748BEAMGeneralCD2Beam position on targetThe beam spot on target has shifted upwards
by ~3 mm since beginning of 10Be run (see
run 366 and attached plot from run 374, July
 TargetPos_run374_20h48.png 
Fixede748 Sat Jul 8 21:04:07 2017 e748BEAMGeneralCD2Update: beam position on targetAt ~21:00 we called PCP: They adjusted the
DC41VE steerer to center vertically the beam
on our target.
 TargetPos_run374_afterDC41VE.png 
Fixede748 Sat Jul 8 21:53:58 2017 e748BEAMGeneralCD2Scalers run 374Counting rates keep decreasing though the
primary is still ~4.5 uAe and the beam is
centered on our target.
 scalers_run374_21h49.png 
Fixede748 Sat Jul 8 23:22:18 2017 e748BEAMGeneralCD2Update on counting rates + scalers18O 4.5 uAe
CATS1 91 kcps
CATS2 114 kcps
 scalers_run374_23h21.png 
Fixede748 Sat Jul 8 23:51:06 2017 e748 shift U Valérie L. RUNGeneralCD2Overview of run 374 around midnight
Run at 23h48 with stable conditions, 18O
beam: 4.5 µA
 7x 
Fixede748 Sun Jul 9 02:41:49 2017 e748- shift U:Valérie L +Beyhan B.RUNGeneralCD2Run 374 - good beam conditions The 18O was very stable between 0h and 2h
around I ~ 4.5 µA. 
CAVIAR ~ 161k; CATS1 ~ 88kcps;  CATS2~ 111kcps
 10x 
Fixede748 Sun Jul 9 04:21:42 2017 e748- Beyhan B + Valérie LBEAMGeneralCD2BEAM SURVEY AT PCPThe beam Survey from the PCP can be seen in
the attached picture below.
Our spokesperson Adrien M can go and ask
 GraphicBeame748july17.jpg 
Fixede748 Sun Jul 9 04:43:40 2017 e748 shift U Valérie L. +Beyhan B.RUNGeneralCD2Run 374- 9 JULY - stop at 5hBetween 4h-4h50 the 18O has decreased from
4.5 down to 4 µA.
CAVIAR ~ 161k; CATS1 ~ 90kcps; CATS2~ 114kcps
 ScalersRun374At4h46.pngRun374CatsRawStrQ_4h40.png 
Fixede748 Sun Jul 9 05:21:29 2017 e748 shift U Valérie L. +Beyhan B.(-8h)/ Daisuke Shumpei (8h-)RUNGeneralCD2Run 375 after beam stopThe tuning of the beam alignment was checked
by the PCP. cf beam profiles at 5h15 in LISE
notebook and below.
 11x 
Fixede748 Sun Jul 9 09:07:31 2017 Jean-Charles Valérie L. Daisuke ShumpeiRUNGeneralCD2Run 376 LISE tuning (underway)RUN 376 short run to check beam profile during
LISE tuning.
 run376_must2.pngrun376_cats.png 
Fixede748 Sun Jul 9 09:13:54 2017 Jean-Charles Valérie L. Daisuke ShumpeiRUNGeneralCD2Run 377 LISE tuning (underway)RUN 376. LISE tuning to optimize the beam
spot on target. During this run, the beam
stop will be moved. 
 run377_cats_final.pngrun377_must2_final.pngrun377_echelles_final.png 
Fixede748 Sun Jul 9 09:36:42 2017 Daisuke Shumpei Valérie L. (-9h50) Jean-Charles (-10h)RUNGeneralCD2Run 378Data accumulation after LISE tuning. 

9h36 started. 
 10x 
Fixede748 Sun Jul 9 09:43:29 2017 Daisuke Shumpei Jean-CharlesOTHERHardwareCD2Printer Magnolia is out of cartridge..Use "iris-color". The printer is in the room
ACS17 (found left in the corridor from the
LISE counting room to the building exit).
  
Fixede748 Sun Jul 9 10:27:09 2017 Shumpei, DaisukeOFFLINEGeneralCD212Be(d,t)More statistics for (d,t) channel.
Same gate as before () but MM1&MM4 are used.
 E_Theta_dt2.pdfE_Theta_dt3.pdf 
Fixede748 Sun Jul 9 10:29:20 2017 Shumpei, DaisukeOFFLINEGeneralCD212Be(d,t)More statistics for (d,t) channel.
Same gate as before (Message ID: 273    
Entry time: Sat Jul 8 08:39:37 2017) but
 E_Theta_dt2.pdfE_Theta_dt3.pdf 
Fixede748 Sun Jul 9 11:56:31 2017 Shumpei, DaisukeOFFLINEGeneralCD212Be(d,t)[quote="Shumpei, Daisuke"]More statistics
for (d,t) channel.
Same gate as before (Message ID: 273    
  
Fixede748 Sun Jul 9 13:42:26 2017 Shumpei, DaisukeOFFLINEGeneralCD210Be(d,t) & 10Be(d,3He)?Same analysis for (d,t) channel but 10 Be
beam. run368-374.4 (-1 day data)
Separation for Z=3&4 with CHIO(analog) +
 dE_E_CHIO_pla.pdfE_Theta_dt_10Be2.pdfE_Theta_dt_10Be3.pdf 
Fixede748 Sun Jul 9 16:37:21 2017 e748CHIOGeneralCD2CHIO digitizer outCHIO digitizer was out in run 378   
Fixede748 Sun Jul 9 16:43:40 2017 e748OFFLINEGeneralCD2Elastic scattering MM6&7Telescope 6&7 are analyzed for 12Be beam.
The energy of Si vs angle plot for each telescope
(CsI veto and beam position gate) are shown.
 12Be_elastic.pdf 
Fixede748 Sun Jul 9 16:44:58 2017 Shumpei, DaisukeOFFLINEGeneralCD2Elastic scattering MM6&7Telescope 6&7 are analyzed for 12Be beam.
The energy of Si vs angle plot for each telescope
(CsI veto and beam position gate on target)
 12Be_elastic.pdf 
Fixede748 Sun Jul 9 18:44:38 2017 e748BEAMGeneralCD2LOST BEAM around 18h30Vaccuum was lost after CSS2, so we lost the
primary beam (18:30)
Operators called the astreinte.
  
Fixede748 Sun Jul 9 20:38:11 2017 e748BEAMGeneralCD2LOST BEAM around 18h30Seems there is a water leak in the beam line
near alpha. 
Astreinte is at work. 
  
Fixede748 Sun Jul 9 23:57:34 2017 e748BEAMGeneralCD2NUMBER FOR PCPThere is a major shutdown of the GANIL phone
system. 
You can not call the pcp via the usual extention;
  
Fixede748 Mon Jul 10 00:23:41 2017 e748RUNGeneralCD2Return of the beam (Jedi not available)Beam is back.
JCT refoc the beam on target (attached pictures).
All god beside that. Still no phone on campus!
 CATS.png 
Fixede748 Mon Jul 10 00:35:29 2017 e748RUNGeneralCD2Run 380Started at 0:02

18O 4.3 uAe, fluctuating
 scalers_run380.png 
Fixede748 Mon Jul 10 01:34:12 2017 e748RUNGeneralCD2Run 380 (file change)Data file change at 1:18 -> run_0380.dat.10-07-17_00h02m00s.1

18O 4.5 uAe
 scalers_run380_1.png 
Fixede748 Mon Jul 10 02:04:33 2017 e748CATSGeneralCD2CATS raw strips and trajectories   CATS_raw_run380_1.pngCATS_run380_1.png 
Fixede748 Mon Jul 10 02:47:27 2017 e748RUNGeneralCD2Run 380 (file change)2:26 data file change -> run_0380.dat.10-07-17_00h02m00s.2

18O 4.5 uAe
 Scalers_run_380_2.png 
Fixede748 Mon Jul 10 03:34:24 2017 e748RUNGeneralCD2Run 380 (file change)3:33 data file change -> run_0380.dat.10-07-17_00h02m00s.3

18O 4.5 uAe
 scalers_run380_3.png 
Fixede748 Mon Jul 10 03:45:19 2017 e748MUST2GeneralCD2SiLi 1/T5 tripedTriped between 3:34 and 3:44.

We turn it back on at 3:46 (370 V) -> 22.5
  
Fixede748 Mon Jul 10 04:33:32 2017 e748CATSGeneralCD2 CATS: raw strips, trajectories and pos on target   CATS_raw_run_380_3_4h35.pngCATS_run380_3_4h35.pngtarget_pos_run380_3.png 
Fixede748 Mon Jul 10 04:46:11 2017 e748RUNGeneralCD2Run 380 (file change)4:42 data file change -> run_0380.dat.10-07-17_00h02m00s.4

18O 4.4 uAe
 scalers_run380_4.png 
Fixede748 Mon Jul 10 05:47:40 2017 e748RUNGeneralCD2Run 380 (file change)5:46 data file change -> run_0380.dat.10-07-17_00h02m00s.5

18O 4.5 uAe
 scalers_run380_5.png 
Fixede748 Mon Jul 10 06:53:40 2017 e748MUST2GeneralCD2SiLi 1/T5 tripedTriped at ~ 6:50.
We turn it back on at 370 V -> I = 18 uA
but increases rapidly to 21 uA
  
Fixede748 Mon Jul 10 06:55:58 2017 e748RUNGeneralCD2Run 380 (file change)6:48 data file change -> run_0380.dat.10-07-17_00h02m00s.6

18O 4.5 uAe
 scalers_run380_6.png 
Fixede748 Mon Jul 10 08:19:45 2017 e748RUNGeneralCD2End run 380We stop run 380 to enter D6 and select mask
target for CATS reference data.
  
Fixede748 Mon Jul 10 08:37:23 2017 e748RUNGeneralN/ARun 381 mask targetTrigger = Plastic, div=1
Primary intensity = 1 uAe
CATS rate = 35 kcps
  
Fixede748 Mon Jul 10 10:06:27 2017 e748RUNGeneralN/Aend of experiment10h06 we stop the experiment.   
Fixede748 Mon Jul 10 17:43:37 2017 AdrienMUST2GeneralN/ACalimero time calibratorRun 1000 : MUVI 1 X

Time calibrator 20ns period, 512ns range
  
Fixede748 Thu Jul 20 11:07:54 2017 e748CATSGeneralN/ARun 1035 calib cats1After experiment CATS 1 pulser calibration
Pulser Quality: very bad - to be analyzed
with care
  
Fixede748 Thu Jul 20 11:09:34 2017 e748CATSGeneralN/ARun 1036 calib cats 2 YAfter experiment pulser calibration CATS 2
(Y)
1st amplitude slightly saturates CATS 2 X.
  
Pinnede744 Thu Mar 15 10:50:00 2018 Adrien MattaMUST2Hardware  Distribution box LPCCabling file by Fabien Naury (LPC).  8x 
Pinnede744 Mon Mar 26 11:44:27 2018 e744 : ValerianOTHERGeneral Current Elog stateCurrently we do not have access to the experiment
number. To write an entry for e744 the nickname
TEST will be used.
  
Fixede744 Mon Mar 26 11:46:43 2018 e744 : ValerianRUNGeneral3-alphaTests runs Ganacq log(15:51:33 - 19/03/2018) INFO - rcc_gui - Started
run #1
  
Fixede744 Mon Mar 26 14:20:22 2018 e744 : ValerianMUST2Hardware3-alphaBidimensional graphs for Telescope 7 and 6WARNING T5 == T7
[U][B]
EDIT 27/03/18 : WARNING X and Y are inverted,
 T6_3alpha.pdfT7_3alpha.pdf 
Ongoinge744 Mon Mar 26 15:07:06 2018 e744 : Valerian et LouisMUST2Hardware3-alphaEnergy vs Strip for T6 et T7Weird signal on StripX of T6... might be the
copper oxyde projections making a short-circuit
 StripXE_vs_StripNbr_T6.pdfStripXE_vs_StripNbr_T6_zoom.pdfStripYE_vs_StripNbr_T6.pdf 
OngoingTEST Tue Mar 27 10:47:39 2018 e744 : ValerianMUST2Hardware3-alphaChoice of DSSSDs    
Ongoinge744 Tue Mar 27 11:11:04 2018 e744 : ValerianMUST2Hardware3-alphaChoice of DSSSDsDifferent plots of Strip channel vs Strip
number for X and Y in order to make our choice
 14x 
Fixede744 Tue Mar 27 16:26:29 2018 e744 : ValerianMUST2General3-alphaBidim for T1 T3 T4 T5 T6 T7 T8   7x 
Ongoinge744 Tue Mar 27 17:22:24 2018 e744 : ValerianMUST2Hardware3-alphaChoosing the DSSSDsWe can see that the DSSSDs on T1(CsI) T3(CsI)
T4(CsI) and T5(SiLi) are in good shape (cf
:[URL=https://elog.lpc-caen.in2p3.fr/MUST2/333]Bidim
  
Ongoinge744 Wed Mar 28 13:21:32 2018 e744 : ValerianMUST2Hardware3-alphaChoosing the DSSSDs[quote="e744 : Valerian"]We can see that the
DSSSDs on T1(CsI) T3(CsI) T4(CsI) and T5(SiLi)
are in good shape (cf :[URL=https://elog.lpc-caen.in2p3.fr/MUST2/333]Bidim
  
Ongoinge744 Wed Mar 28 16:35:01 2018 e744 : ValerianMUST2Hardware3-alphaChoosing the DSSSDs[quote="e744 : Valerian"][quote="e744 : Valerian"]We
can see that the DSSSDs on T1(CsI) T3(CsI)
T4(CsI) and T5(SiLi) are in good shape (cf
 TimeMust2X_MUVI2_uncalib.pngTimeMust2Y_MUVI2_uncalib.pngSpectresMUVI2stripXT300ns.png 
Fixede744 Tue Apr 3 10:44:38 2018 e744 : ValérianMUST2Hardware Update on Must2 telescopesFor the next day, the plan is to check the
DSSD's connections of T6 & T7 followed by
an alpha run, if the problem seen on T6 are
 20180403_103732.jpg 
Ongoinge744 Tue Apr 3 15:30:49 2018 e744 : ValérianMUST2Hardware Cabling of T6 & T7After checking the cable of T6 & T7 it seems
that there were in fact an inversion during
experiment e748. T6 is the bottom one, and
 Inversion_between_telescopes.pdf 
Ongoinge744 Wed Apr 4 17:49:05 2018 e744 : ValérianEXL-PPTHardware Update on EXLThe EXL detector has been mounted, it is very
close to the target ~4cm and doesn't do any
shadow on the M2 telescopes. But it isn't
 6x 
Fixede744 Tue Apr 10 17:43:28 2018 Adrien Matta and FyFyMUST2HardwareN/Alogin to CAEN SY1527 HV/BT mainframeTo control the CAEN SY1527 HV/BT mainframe
use the following command from a terminal:
  
Fixede744 Tue Apr 17 09:07:55 2018 Adrien MattaDAQGeneralN/ADAQ issue summaryPreviously on e744:

After a chat with Charles Houarner we decided
  
Fixede744 Tue Apr 17 09:18:00 2018 Cherry MX teamMUST2HardwareN/AT5 is back and running strongT5 telescope appears to be fully functional
again. It shows both Energy and Time with
excellent resolution.
 T5.pdf 
Ongoinge744 Tue Apr 17 17:36:37 2018 Cherry Mx TeamMUST2SoftwareN/AGRU SUCKSPretty much today's summary in the title.

[URL=https://i.pinimg.com/originals/a8/d4/12/a8d41206e8bd9c189397fb4907656981.jpg]https://i.pinimg.com/originals/a8/d4/12/a8d41206e8bd9c189397fb4907656981.jpg[/URL]
  
Fixede744 Mon Apr 23 11:33:53 2018 e744 ValerianOTHERHardwareN/AInspection signalsInspection signals between D6 and LISE Acquisition
room are connected as follow : 
  
Pendinge744 Tue Apr 24 14:18:22 2018 Valerian Freddy and Adrien aka Cherry Mx TeamMUST2HardwareN/ACurrent telescope-pcb layoutTelescope 1 : PCB 5 (ref on pcb: 2274)
Telescope 2 : PCB 4
Telescope 3 : PCB 7
  
Fixede744 Tue Apr 24 15:42:03 2018 e744 Valérie L.DAQGeneral GMT chronogrammeCrate ganlx1 GMT master FAG 100ns ; VAL 2
µs ; CT 6 µs
Crate ganlx14 GMT slave FAG 100ns ; VAL 2
 GMTmasterChrono24avril18.pngGMTslaveChrono24avril18.png 
Fixede744 Tue Apr 24 15:51:04 2018 e744 Valérie L.MUST2General MUST2 configuration /parameters
TAC range for all telescopes: 600ns
Thresholds: 402 keV
 ConfigMuviPhysMust2Thr.png 
Fixede744 Tue Apr 24 16:22:08 2018 e744 Valérie L.RUNSoftware Narval Merger Parameters   ConfigRunControlNArvalMerger.png 
Fixede744 Wed Apr 25 15:59:51 2018 ValérianEXL-PPTSoftwareN/AMesytech setupTo use the remote settings of the Mesytech
modules : 
[CODE]
  
Fixede744 Wed Apr 25 20:47:59 2018 e744 ValérieMUST2General MUST2 mate Temperature on the DAS interfacetests
The plot:ConfigDASMuvi1Tel4mufeeX.png shows
the situation on the Mufee card when the
 ConfigDASMuvi1Tel4mufeeX.pngConfigDASMuvi1Tel5mufeeXokTemp.png 
Ongoinge744 Thu Apr 26 19:37:24 2018 Valerian and FreddyMUST2HardwareN/ADebug of Must2 telescopesToday we found different problems, no time
on T4 and no signal on T5 and T6 CsI tripping.
After changing cables and a lot of tests
 debug_mm_telescopes.pdfProblem_on_T5_T6_cable_swap.pdf 
Ongoinge744 Sat Apr 28 10:17:22 2018 Valerian and IulianMUST2Hardware3-alphaStatus of MUST2 telescopes 27/04/18Yesterday night we tested T1, T2, T3, T4,
T5, and T6.
 Third_good_run_Energy_and_time_for_5_Must2_detectors_2018-04-27_19-32-23.png 
Ongoinge744 Sat Apr 28 13:27:26 2018 Valerian and IulianMUST2Hardware3-alphamultiples runs with two alpha source on T1 T2 T3 T4 and T5For the following runs we will use two 3 alpha
sources :
  
Ongoinge744 Sat Apr 28 13:36:52 2018 Valerian and IulianOTHERSoftwareN/AProblem with DAS configurationThere is an issue with the GMT slave. When
you click Online and then apply write hardware,
if the message : [U]register error on GMTSlave[/U]
  
Pending  Sat Apr 28 21:12:55 2018 iulianMUST2General3-alphaRun 182 end ; problem with coolingRUN 182 start @13h
Stoped @ 21H
 end_result_alpha_source_28_April.png 
Fixede744 Sun Apr 29 20:37:58 2018 FrancoisCATSGeneral Chronogram   chrono.jpg 
Fixede744 Mon Apr 30 20:38:21 2018 iulianRUNGeneral3-alpha6run 186 3a source 60 co source[U][B]run #186[/B][/U] 

start@11h 30/04
  
Fixede744 Mon Apr 30 20:42:34 2018 iulianRUNGeneral60Corun 188 60 co source[U][B]run #188[/B][/U] [I]60co source EXL+PPT
trigger[/I]
  
Fixede744 Tue May 1 10:47:41 2018 iulian & valerianRUNGeneral60Corun 189 60 Co source[U][B]run #189[/B][/U] [I]60co source EXL+PPT
trigger[/I]
Changed the way of codding PPT. for this
  
Fixede744 Wed May 2 18:11:59 2018 Iulian Marlenne LouisRUNGeneral3-alpha3 alpha testRun191 start 18h stop 18h10 
Run ok, saved config DAS with good polarity
on the thresholds CSI & piedestal allignement
 run_191.png 
Ongoinge744 Thu May 3 08:50:17 2018 FrançoisVACUUMGeneral stop vacuumturbo pumps stopped at 8H45

problem with the cooling system of the M2C
  
Fixede744 Thu May 3 15:04:03 2018 Iulian MarlenneMUST2HardwareN/AX and Y possitions on the detectorTelescope 1 : PCB 5 (ref on pcb: 2274)
Telescope 2 : PCB 4
Telescope 3 : PCB 7
  
Fixede744 Thu May 3 15:20:16 2018 Iulian MarlenneMUST2HardwareN/AX and Y possitions on the detectorTelescope 1 : PCB 5 (ref on pcb: 2274)
Telescope 2 : PCB 4
Telescope 3 : PCB 7
  
Ongoinge744 Thu May 3 19:11:14 2018 Valerian Marlene and FrancoisMUST2General3-alpharun 3 alpha T5[U][B]run #196[/B][/U]
3 alpha source and pedestals supressed
Trigger T5 only
  
Fixede744 Fri May 4 11:20:50 2018 ValerianMUST2General3-alphaT5 : 0° telescope After increasing the voltage on T5 DSSD from
80V to 100V the resolution increased from
~40keV sigma to 20keV sigma.
 StripX_vs_StripY_T5_100V.pdfSi_E_T5_100V.pdf 
Fixede744 Fri May 4 23:58:07 2018 iulianRUNGeneral3-alphaRun #203 alfa source & 60co source[U][B]run #203[/B][/U]

alpha source & 60co source
 run_203_ht.pngrun_203_ppt.pngrun_203_EXL.pngrun_203_must2.png 
Ongoinge744 Sat May 5 11:14:33 2018 Iulian & ValerianRUNGeneral3-alpharun #204 alpha source[U][B]run #204[/B][/U]
alpha source
start @11h15 05/05
 BT_tripped_T5.pngT5_das_no_cooling.pngT5_das_no_cooling2.pngT5_vigru_no_cooling.png20180506_152338.jpg 
Fixede744 Sun May 6 17:52:21 2018 ValerianRUNGeneral3-alphaRun #205 T1 T2 T3 & T4[U][B]run #205[/B][/U]
T1 T2 T3 & T4 T5 still in trigger but no
HT/BT
  
Ongoinge744 Sun May 6 17:55:05 2018 ValerianOTHERGeneral3-alphaUpdate on the cooling system problemAfter waiting for 1-2 hours. I started the
BT again on T5 no tripping BT anymore. While
writing hardware with DAS T5 responded well.
 6x 
Fixede744 Mon May 7 10:42:58 2018 ValerianOTHERGeneral60CoUpdate on the cooling system and run gamma (signal problem on PPT)Today, both cooling system are working fine
and their levels are OK.
The vacuum are 7e-6 mbar in M2C and 2e-5
 HT_EXL.pngEXL_spectra_run207.pngPPT_spectra_run207.pngEXL_spectra_run209.png 
Fixede744 Thu May 10 14:38:41 2018 iulianEXL-PPTHardwareN/AMesytech settingsTried unsuccessfully to change the gain on
EXL energies.
Looking at the inspection lines noticed that
  
Fixede744 Thu May 10 15:06:44 2018 iulian & valerianEXL-PPTGeneral60CoStarted run #211 & #212 #213 60Co source efficiency run[U][B]Run #211[/B][/U] BAD Run no BT
60Co source efficiency run EXL & PPT trigger.
Gains and thresholds set
 PPT_spectra_run213.pngEXL_spectra_run213.png 
Pinnede744 Fri May 11 14:00:48 2018 iulianEXL-PPTHardwareN/AMesytech settings[quote="iulian"]Tried unsuccessfully to change
the gain on EXL energies.
Looking at the inspection lines noticed that
 20180511_132034_1526038557770.jpg 
Fixede744 Sat May 12 08:54:42 2018 iulianMUST2General3-alphaRun #114 T1-T5 trigger 3alpha source calib run [U][B]Run  #214[/B][/U]
Trigger T1-T5 & gamma 
3 alpha run 
 Must_spectra_start_run214.pngMust_spectra_stop_run214.png 
Ongoinge744 Sat May 12 20:09:09 2018 iulianMUST2General3-alphaRun #115 3 alfa run Calibration T1-T5 trigger[U][B]Run #215[/B][/U]
Trigger T1-T5 
3 alpha run
 Must_spectra_start_run215.pngMust_spectra_stop_run215.pngerror_run215.pngerror_1minlater_run215.png 
Ongoinge744 Sun May 13 12:09:55 2018 iulianMUST2General3-alphaRun #116 3 alfa run Calibration T1-T5 trigger [U][B]Run #115[/B][/U] 
3 alfa run Calibration T1-T5 trigger 
start 12h10
 Must_spectra_start_run216.png 
Pendinge744 Sun May 13 13:32:44 2018 iulianONLINESoftwareN/Astart_online ilegal instruction faultFrom time to time the start_online code gives
an ilegal instruction error and crash...
Once the ctrl-c pressed it shows:
  
Ongoinge744 Sun May 13 19:36:09 2018 Valérian + FrançoisMUST2General3-alphaRun #116 #117 3 alpha run Calibration T1-T5 trigger[U][B]run #116[/B][/U]
start @12h07 13/05/18
stop  @19h45 13/05/18
 MUST2_HT_BT_13-05-2018.pngspectra_14-05-2018.png 
Fixede748 Mon May 14 10:49:47 2018 FrançoisDAQSoftware TDC problemProblem :
TDC labels are 0. After the compilation of
action list the labels seem to appear (if
  
Ongoinge744 Mon May 14 12:57:44 2018 FrancoisTARGETGeneral target preparation

Cibles Georges
  
Fixede744 Mon May 14 13:52:25 2018 FrançoisBEAMGeneral Profileur

 Sans_titre.jpg 
Ongoinge744 Tue May 15 09:58:52 2018 ValérianRUNGeneral3-alpharun #118 3 alpha[U][B]run #118[/B][/U]
Trigger T1 T2 T3 T4 & T5 
EXL + BT not biased but still in trigger
  
Fixede744 Tue May 15 19:42:40 2018 ValérianEXL-PPTGeneral60CoCalibration of EXL from e748 dataThe EXL calibration program is ready. Join
is the result with previous data from e748.
 EXL_Sum_e748_Calib.png 
Ongoinge744 Tue May 15 21:32:46 2018 Iulian ValerianEXL-PPTGeneral60Corun #300 60 Co source EXL & PPT trigger[U][B]Run #300[/B][/U]
PPT EXL trigger 60Co
  
Ongoinge744 Wed May 16 09:49:26 2018 Francois & Ruxandra & ArmelBEAMHardware E1D6 and E2D6 detectorsAfter struggling with the noise, we achieved
~  50 keV resolution.
 essai.jpg 
Fixede744 Wed May 16 11:30:21 2018 iulianEXL-PPTGeneral60Corun 302 60Co source[U][B]Run #302[/B][/U]

60 Co source PPT & EXL trigger
  
Fixede744 Wed May 16 16:27:29 2018 e744, lama, beyhanOTHERHardware TACS from LISE to the exp DAQThe following TACs signals from LISE have
been cabled and send to the experimental
acquisition (->D6)
  
Fixede744 Wed May 16 17:03:18 2018 RuxandraEXL-PPTSoftware calibrationCalibration coefficients for the PPT detector
using the 60Co and 40K lines: E=0.5514*ch
- 83.322. Internal radioactivity at 1444
  
Fixede744 Wed May 16 17:14:20 2018 e744, lama, beyhanOTHERHardware setting of signals from LISE to exp. DAQWe put the following signals from LISE on
DAQ :
  
Fixede744 Wed May 16 18:01:47 2018 Iulian ValerianMUST2SoftwareN/AFirst successful Calimero run on T1 voie X run 316 First successful Calimero run on T1 voie X
run 316 
 Calimerro_run316.png 
Fixede744 Wed May 16 19:06:39 2018 Iulian ValerianMUST2SoftwareN/Arun #318 Calimero T1-T4 strips X[U][B]Run #318[/B][/U]

Calimero T1-T4 X
 Calimerro_run318.pngCalimerro_T4_X_strip2_run319.pngCalimerro_T2_X_strip12_run319.png 
Fixede744 Wed May 16 20:26:48 2018 Iulian ValerianMUST2SoftwareN/Arun #319 calimero T1-T4 Y strips [U][B]run #319[/B][/U] 
calimero T1-T4 Y strips 
 Calimerro_run319.png 
Fixede744 Wed May 16 20:36:40 2018 Iulian ValerianMUST2SoftwareN/Arun #320 Calimero T5 X[U][B]run #320[/B][/U] 

Calimero T5 X
 Calimerro_run320.png 
Fixede744 Wed May 16 20:42:43 2018 Iulian ValerianMUST2SoftwareN/ARun #321 Calimero T5 Y strips [U][B]Run #321[/B][/U] 

Calimero T5 Y strips 
 Calimerro_run321.png 
Ongoinge744 Wed May 16 21:27:55 2018 Iulian ValerianMUST2SoftwareN/ARun #322 Calimero T1-T4 X stripsRun #322 
Calimero T1-T4 X strips
 1 channel per mate not done (Valerian too
 Calimerro_run322.png 
Ongoinge744 Wed May 16 21:45:47 2018 iulianMUST2SoftwareN/A Run #323 Calimero T1-T4 strips X[U][B] Run #323[/B][/U] 
Calimero T1-T4 strips X
 Calimerro_run323.png 
Ongoinge744 Wed May 16 22:03:13 2018 iulianEXL-PPTSoftware60Corun #324 60Co [U][B]run #324[/B][/U] 
60Co 
 EXL_run324.pngPPT_run324.png 
Pinnede744 Thu May 17 12:04:45 2018 iulianEXL-PPTHardware60CoEXL delay of ADC Codding window EXL timing delays for Mesitech module 2 &
1
 Exl_timing_mesy_1_2.pdf 
Ongoinge744 Thu May 17 18:29:16 2018 iulianMUST2General3-alphaRun #333 Last alfa run before experimentRun #333 
3a source 
T1-T5 trigger
 Must_spectra_start_run333.png 
Ongoinge744 Thu May 17 19:15:21 2018 Iulian EXL-PPTGeneral60CoRun #334 Last 60 co run for eficacity before experiment[U][B]Run #334[/B][/U]

start 19h15
  
Fixede744 Fri May 18 10:47:37 2018 RuxandraEXL-PPTHardware60CoCo60 caliibrationThu  May 17 . run 339
60Co calibration
start @19h26 17/05/18
  
Fixede744 Fri May 18 13:28:24 2018 RuxandraEXL-PPTSoftware60Conew calibrationThe new calibration coefficients for the PPT
detector (run 339): E=0.526*ch+32.57
  
Fixed  Fri May 18 15:17:19 2018 Megumi and LouisOTHERHardware TAC range setting and time calibration1 mm-cats 1us
2 mm-hf 200ns
3 mm-exl 1us
  
Fixede744 Fri May 18 17:38:21 2018 Francois and PedjaOTHERHardware translator nim to eclThere was a problem with the scalers, not
all of them were working. After several tests,
we changed the electronic module, from Lecroy
 scalers.png 
Ongoinge744 Sat May 19 15:32:51 2018 Iulian ValerianEXL-PPTGeneralN/AStarted run #346 Room background PPT EXL[U][B]run #346[/B][/U]

Room background PPT EXL
  
Fixede744 Sun May 20 18:53:44 2018 iulian Jaromir CristinaEXL-PPTGeneralN/ARun #347 Background run EXL & PPT triggers[U][B]Run #347[/B][/U]
Background run EXL & PPT triggers
start @18h53 20/05
 PPT_run347.pngEXL_T_run347.pngEXL_E_run347.png 
Ongoinge744 Sun May 20 19:11:57 2018 ruxandra cristina giuseppe jaromirEXL-PPTHardware60CoEXL-addback and light leak considerationsin the middle of the considerations - pdf
file 
 EXL.pdf 
Pinnede744 Mon May 21 11:26:55 2018 iulian cristinaOTHERHardwareN/Aerror stop run/ run still goingAfter stopping run 347, the narval looks like
everything was stopped (blue ready), but
I still had FAGs, VALs, and the leds on the
 error_Narval.pngerror_DAS1.png 
Fixede744 Mon May 21 11:42:16 2018 iulian cristinaMUST2SoftwareN/Arun 348 test comunication with Must. Pulser T1 channel 11   MUST2_Pulser_T1_E11_run348.png 
Fixede744 Tue May 22 10:44:30 2018 iulian Fairouz ValerieMUST2SoftwareN/ARun #353 Calimero T1-T4 strips XStarted run #353
Calimero T1-T4 strips X
Time Calibrator 320 ns ; period 20 ns
  
Fixede744 Tue May 22 11:37:50 2018 ValérieMUST2SoftwareN/Acalimero config GECO 
Calimero T1-T4 strips X
Time Calibrator 640 ns ; period 20 ns
 PlotTel1strxT101calimOk.pngPlotXTel1tXT88Probmate.pngPlotXTel1to4XTProbmate.pngPlotXTel1to4XTcalimero.png 
Fixede744 Tue May 22 11:59:06 2018 ValérieMUST2SoftwareN/A Calimero config GECO   8x 
Fixede744 Tue May 22 12:09:27 2018 iulian Fairouz ValerieMUST2SoftwareN/ARun #354 Calimero T1-T4 strips XStarted run #353
Calimero T1-T4 strips X
Time Calibrator 640 ns ; period 20 ns
 PlotXTel1tXT88Probmate.pngPlotTel1strxT101calimOk.pngPlotXTel1to4XTProbmate.pngPlotXTel1to4XTcalimero.pngPlotXTel1to4XTcalimero.png 
Fixede744 Tue May 22 12:10:27 2018 iulian Fairouz ValerieMUST2SoftwareN/ARun #355 Calimero T1-T4 strips YStarted run #353
Calimero T1-T4 strips Y
Time Calibrator 640 ns ; period 20 ns
  
Fixede744 Tue May 22 13:40:03 2018 Fairouz ValerieMUST2SoftwareN/ARun #356 Calimero T1-T4 strips YRun #356 
Time Calibrator Calimero stripYT T1to T4
640ns period 20ns
  
Pinnede744 Tue May 22 18:33:32 2018 Nicolas, AdrienOTHERSoftware How to run Online and Convert raw data files[U][B]1. How to run Basic_online or Online
code[/B][/U]
[LIST]
  
Fixede744 Tue May 22 19:36:50 2018 Fairouz, ValérieMUST2SoftwareN/ARun #359 Calimero T1-T4 strips XRun #359
amplitude of the gene is now set to 10 MeV
Time Calibrator Calimero stripsX  T1to T4
 8x 
Pendinge744 Tue May 22 19:40:26 2018 FrancoisBEAMGeneral Planning of the experiment   Chekclist_E744_4.pdf 
Fixede744 Tue May 22 19:40:48 2018 Valérie MUST2GeneralN/Asnapshots for steps in Calimero calibration   11x 
Fixede744 Tue May 22 20:06:15 2018 Fairouz, Valerie, edited by Iulian MegumiMUST2SoftwareN/ARun #360 Calimero T1-T4 strips YStarted run #360
calimero stripsYT T1 to T4 640 ns; period
20 ns
 Calimero_GMTproblem.png 
Fixede744 Tue May 22 21:43:31 2018 iulian & megumiMUST2SoftwareN/Arun 361 [U][B]run 361[/B][/U] 
start 20h38
stop 21h41
 T1T4_Xstrips_run361.png 
Fixede744 Tue May 22 21:57:51 2018 iulian & megumiMUST2SoftwareN/ARun #362 Calimero T1-T4 strips X again with solutionChanged the Thresholds on T4. No longer double
peaks...
I guess there is some noise on the T4 and
 Thresolds_T4_run362.pngT1T4_Xstrips_run362.png 
Fixede744 Tue May 22 22:33:09 2018 e744MUST2SoftwareN/ARun #363 Calimero T1-T4 strips X again with threshold at 50% (same as 40%)[U][B]Run #363[/B][/U]
NB: Thresholds  are the ones on the pulser
amplitude to inject the signal (Calimero
 T1T4_Xstrips_run363.png 
Fixede744 Tue May 22 22:50:47 2018 e744MUST2SoftwareN/ARun #365 Calimero strips Y T1-T4 100% Threshold[U][B]Run #365[/B][/U]
Calimero strips Y T1-T4
100% threshold on T4
 T1T4_Ystrips_run364.png 
Pinnede744 Tue May 22 23:42:02 2018 iulian & megumiEXL-PPTSoftwareN/ARun #366 NARVAL FATAL ERRORRUN #366
EXL PPT triggers
Falat error from NARAVL
 error_narval_run366.png 
Fixede744 Wed May 23 00:30:34 2018 e744EXL-PPTGeneralN/ARUN #402, EXL PPT triggerRUN #402
start exl ppt run for checking coincidence
exl ch 17 put in to adc2 ch 32
  
Fixede744 Wed May 23 10:52:28 2018 ValérieEXL-PPTGeneralN/ARUN #405, EXL PPT triggerRUN #405
start exl ppt run for checking coincidence
 
 GMTtriggerRun405.pngScalersRun405.pngEXL_E1to16duringRun405.pngEXL_T1to16duringRun405.png 
Fixede744 Wed May 23 10:57:13 2018 megumiDAQGeneral Photo of the circuitphoto of the circuit area  15x 
Fixede744 Wed May 23 13:15:47 2018 megumiDAQGeneralN/ATime calibration for TDC (EXL)Time calibrator setting
Range 5.12 us
Period 160 ns
 Capture_du_2018-05-23_13-29-34.png 
Pinnede744 Wed May 23 15:10:15 2018 iulianOTHERGeneralN/AProcedure to start stop RUNS   fiche_start_run.pdffiche_stop_run.pdf 
Fixede744 Wed May 23 15:40:58 2018 Charles Houarner ValérieMUST2SoftwareN/ATESTS MUVI internal pulser T1-T4 strips XTo check the signals and trigger on T4 we
used the internal pulser on T4, 
from the SYNCHROTEST pulser at 150 Hz.
 testGeneInterneT4plotCh5Mate.pngtestGeneInterneT4plotCh5MatePied.pngTestMust2tel4matePulserInt.pngTestMust2tel4PulserIntMate5.pngTestMust2telT1toT3andT5mat5onT4.png 
Fixede744 Wed May 23 15:42:04 2018 Louis, ValérieMUST2GeneralN/ACALIMERO run 409 - XT - T1 to T4Started run #409 
GMT triggers: GecoCalimeroMuviGMT.png 
HT on cf : HTtelescopesMustsWed23May.png
 GecoCalimeroMuviTel4Thr1MeV.pngconfigCalim1.pngconfigCalimOKexterne.pngGecoCalimeroMuviGMT.pngHTtelescopesMustsWed23May.png 
Pinnede744 Wed May 23 17:23:34 2018 FreddyMUST2General Pedestals and alpha calibration analysisCalibration coefficients for MUST2 telescopes
were extracted from run_0118 (acquired during
night from 13th to 14th of May). They are
 Plot_DiffPedestals.png 
Fixede744 Wed May 23 18:20:59 2018 e744MUST2GeneralN/ATest Calimero Run 425Started run #425 Calimero  NEW XT  T1toT4
ampl Pulser 12MEV; thr for T4 1 MeV
problem on data taking
  
Ongoinge744 Wed May 23 20:52:01 2018 Iulian ValerianRUNGeneral run #429 and #430 and config gmtChanged VAL from 2 us to 2.5 us

run #429 exl ppt triggers
 config_gmt.pngconfig_gmtslave.pngrun430_E_PPT_vs_T_PPT.pdfrun430_E_PPT_proj.pdfrun430_E_EXL_vs_T_EXLor.pdf 
Fixede744 Wed May 23 23:37:14 2018 Valérie L.MUST2GeneralN/AMUVI configuration DASConfigurations files of the MUVI 1 MufeeX
and Y.
On MufeeX and Y configuration:
 13x 
Fixede744 Thu May 24 01:02:19 2018 ValerianOTHERGeneralN/Arun #431[U][B]run #431[/B][/U]
Trigger LISE
Beam in D4
 ED4_TD4_23-05-18_00-58.png 
Fixede744 Thu May 24 01:10:30 2018 iuliqnRUNGeneral run 433 Lise triggerShould be ql foil in front of the si detector....   
Fixede744 Thu May 24 01:11:26 2018 Iulian ValerianRUNGeneralN/Arun 434 Lise triggerCats inserted

efficiency is around ~99%
 ED4_TCATSHF_23-05-18_1-18.pngED4_TCATSHF_23-05-18_1-18_efficiency.png 
Fixede744 Thu May 24 01:53:28 2018 iulianRUNGeneralN/Arun 438 Lise trigger run 438  D4 in
run 439 d6 in.
 run_439_d6_cats_hf.pngrun_439_transmisiond4_d6.png 
Ongoinge744 Thu May 24 02:30:52 2018 iulianRUNGeneralN/ARUN 440 14N no target trigger Must & cats %100RUN 440 
14N beam  \
no target 
 run440.png 
Fixede744 Thu May 24 03:21:37 2018 iulianOTHERHardwareN/APosition measurements Must2/PPT/EXL   STP-2018-E054_E744_I-Stefan_J-Mrazek_LISE-D6.pdf 
Fixede744 Thu May 24 03:57:34 2018 iulianRUNGeneralN/Arun 442 Lise trigger d6   cats_e1_d6_t1_D6_run442.png 
Fixede744 Thu May 24 04:20:34 2018 iulianRUNGeneralN/Arun 443 T1-t5 triggersrun 443
T1-T5 triggers
 mm5_run443.pngmm5_2_run443.pngmm5_3_run443.pngmm5_4_run443.png 
Pinnede744 Thu May 24 05:44:56 2018 iulianRUNGeneralN/AImportant::analysis of the run 443/ things to do[I][B]No stop on t5, but TAC MM_CATS ok..
So two possibilities: 
1) Stop Cable not connected on Muvi2 
 t_mm_PPT_run443.pdft_mmhf_run443.pdft_mm_exlor_run443.pdfcorrelation_run443.pdf 
Ongoinge744 Thu May 24 05:59:44 2018 iulianMUST2GeneralN/Arun 444 t1-t4 trigger no target 14n beamrun 444 
t1-t4 trigger
 no target 
 mm5_8_run445.pngmm5_7_run444.pngmm5_6_run444.pngmm5_5_run444.pngmm5_4_run444.png 
Ongoinge744 Thu May 24 06:30:27 2018 iulianRUNGeneralN/Arun 445 t1-t4 trigger no target 14n beam no CATS   mm5_run445_3.pngmm5_run445_2.pngmm5_run445_1.png 
Pendinge744 Thu May 24 06:53:12 2018 e748RUNGeneralN/Arun #446 CATS trigger    
Fixede744 Thu May 24 12:01:54 2018 Valérie for Calimero team - Valérian, FairouzMUST2General Bug server Calimero Run #447-449Calimero on ganp161
Telscopes T1 to T4 thresholds for T1 to T3
400 keV ; T4 : 1 MeV
 12x 
Fixede744 Thu May 24 17:18:58 2018 Valérie for Calimero team : Valérian, LouisMUST2GeneralN/ACALIMERO Tests Runs 451 to 452 (done in D6) Test runs for telescopes T1 to t4 
Nb: The applied threshold is now 1 MeV but
it seems to be not the case when we look
 configCalimeroRun452YTthr1MeV.pngPlotCalimrun452YTthr1MeVmaybe.pngPlotCalimrun453YT5thr1MeVtrue.png 
Fixede744 Thu May 24 17:23:57 2018 Valérie for Calimero team: Valérian, LouisMUST2GeneralN/ACALIMERO Runs 453-457 OKCalimero on ganp1074 in D6
Time Calibrator Range: 640 ns; Period 20
ns Amplitude of pulser (test generator) 9MeV
 16x 
Fixede744 Thu May 24 19:05:41 2018 e744 team (shift Valérie)RUNGeneralN/ARuns 458 and 459Run 458
focalisation on D4 LISE
100Hz on Si Lise triggers: Cats1div (3 10**4)
 ScalersEndofRun458_beam14N.pngRun458plotED4TD4HF.pngRun458plotED4_TcatsLise.png 
Fixede744 Thu May 24 19:20:40 2018 e744 Must2 + Lise teams (edit: Valérie)RUNGeneralN/ARun 460-Spiral 14N testRun 460
Triggers Must2 Telescopes T1 to T5
( MM1 to MM5)
 Run460plotMM5impact.pngScalersDuringRun460.pngMM5impactendofRun460beam.png 
Fixede744 Thu May 24 19:32:31 2018 e744 Must2 + Lise teams (edit: Valérie)RUNGeneralN/ARun 461-Spiral 14N testRun 461
Triggers Must2 Telescopes T1 to T5
(MM1 to MM5)
 ScalersDuringRun461.pngMM5impactDuringRun461.png 
Fixede744 Thu May 24 19:34:40 2018 e744 Must2 + Lise teamsRUNGeneralN/Atuning Wien-Spiral 14N testRun 24May at 19h50-20pm no storage
Triggers Must2 Telescopes T1 to T5
(MM1 to MM5)
  
Fixede744 Thu May 24 20:18:00 2018 e744 Must2 + Lise teamsRUNGeneralN/Atuning Wien-Spiral 14N testRun 24May at 19h50-20pm no storage
Triggers Must2 Telescopes T1 to T5
(MM1 to MM5)
 MM5impactWien2kV.pngMM5stripsXYWien0.4et0.4kV.pngMM5impactXYWien0.4et0.4kV.pngscalersduringTuningwithWienfilter.png 
Fixede744 Thu May 24 20:19:19 2018 e744 Must2 + Lise teamsRUNGeneralN/ABeam centering on MM5-Spiral 14N testRun 24May at 19h50-20pm no storage
Triggers Must2 Telescopes T1 to T5
(MM1 to MM5)
 MM5strXYWienCenter.pngMM5impactXYWienCenter.pngScalersTuningWienCenterMM5.png 
Fixede744 Thu May 24 20:21:57 2018 e744 Must2 + Lise teamsRUNGeneralN/ARun 462 Beam centering on MM5-Spiral 14N testRun 462 focalisation on target  24May at 20h15pm
storage
Triggers Must2 Telescopes T1 to T5
 MM5ImpactsRun462.pngBeamOnMM5stripsXYRun462.pngScalersduringRun462.pngScalersEndRun462.png 
Fixede744 Thu May 24 20:31:30 2018 e744 Must2 + Lise teams (edit: Valérie)RUNGeneralN/ARun 463 Beam Spiral 14N testRun 463 focalisation on Must2  24May at 20h30pm
storage
Triggers Must2 Telescopes T1 to T5
 scalersendOfRun463.pngMM5stripsXYERun463.pngMM5ImpactsRun463.png 
Fixede744 Thu May 24 21:05:05 2018 Valerian Iulian and FrancoisTARGETHardware 14N - 187um target placementFor resonant elastic scattering of 14N(pp)14N
the 187um target is placed.
 20180524_210115.jpg 
Fixede744 Thu May 24 21:05:19 2018 e744 Must2 + Lise teams (edit: Valérie)MUST2HardwareN/Achange of STOP MUST2The STOP of MUST2 with CATS is now directly
the CATS "delayed signal".
No coincidence with FAG.
  
Pinnede744 Thu May 24 21:13:29 2018 ValerianTARGETGeneralN/AList of target present in M2CThe list of target present in M2C chamber
are the following : 
  
Fixede744 Thu May 24 21:28:45 2018 e744 Must2 + Lise teams (edit: Valérie)RUNGeneralCH2Run 464 Beam Spiral 14N testRun 464 focalisation on Must2  24May at 21h25m
storage
Triggers Must2 Telescopes T1 to T5
 TAcrefGAGmmCatsHFRun464.pngScalersDuringRun464at21h30.pngMM5ImpactsRun464.pngMM5ImpactsRun464zoom.pngMM5StrXYRun464.png 
Fixede744 Thu May 24 22:38:44 2018 e744 Must2+lise teams RUNGeneralCH2Run Beam Spiral 14N centered  Shift teams: François, Iulan, Freddy, Fairouz,
Valérian, Valérie, Mihail
LISE teams during evening: Omar Kamalou,
 7x 
Fixede744 Thu May 24 22:45:36 2018 e744 Must2+lise teams RUNGeneralCH2Run 1001 Beam Spiral 14N centeredRun 1001, cible n°6: 183 um
start @ 21h55; Stop @ 22h36 
 Shift teams: François, Iulan, Freddy, Fairouz,
 scaler-run1001.pngMMimpactsRun1001.png 
Pinnede744 Thu May 24 23:20:40 2018 e744 Must2 + Lise teamsRUNGeneralCH2run 1002[I][B]run 1002:[/B][/I] 14N beam, CH2 187
um
trigger: MM1,2,3,4,5 + CATS%, EXL%, PPT%
 scaler-stop-run1002.png 
Fixede744 Fri May 25 00:09:03 2018 Cristina RUNGeneralCH2run 1003 (target no 6 -CH2 180 um)run 1003

14N
 6x 
Fixede744 Fri May 25 00:41:19 2018 Cristina and PedjaRUNGeneralCH2run 1004 (target no 6 -CH2 180 um)
Run 1004
the thresholds of MUST detector changed to
 7x 
Fixede744 Fri May 25 01:05:05 2018 Cristina and PedjaRUNGeneralCH2run 1005 (target no 6 -CH2 180 um)
Run 1005
 scalers_start_run1005.pngscalers_stop_run1005.pngMMimpactsRun1005.pngTAC_run1005.png 
Fixede744 Fri May 25 02:35:13 2018 Cristina and PedjaRUNGeneralCH2run 1006 (target no 6 -CH2 180 um)
Run 1006
 scalers_start_run1006.pngscaler-stop-run1006.pngtac-run1006.pngmm_impact_run1006.png 
Ongoinge744 Fri May 25 02:57:44 2018 Valerian and IulianOFFLINEGeneralCH2Simulation Azure 1H(14N,p) and measurement We can know conclude without any doubt that
we were able to measure resonances from 1H(14N,p)14N.
The resolution with a rough analysis and
 o15.pdfE_cm_ThetaLab_0_4.pdfE_cm_ThetaLab_2_3.pdfE_cm_ThetaLab_10_12.pdfE_cm_ThetaLab_15_20.pdf 
Pinnede744 Fri May 25 04:02:29 2018 Valerian Iulian Pedja and CristinaOTHERGeneralN/AGeneral power cut of ganilDue to some water leaks, CSS1 CSS2 and CIME
as the vacuum system of GANIL are down...
Time loss could be considerable... They talked
  
Fixede744 Fri May 25 08:37:04 2018 Adrien MattaMUST2HardwareN/AMUST2 coolingVacuum at 1.7 e-5 torr this morning.
I started the MUST2 cooling, setting both
coolers at -5C.
  
Pendinge744 Fri May 25 13:08:44 2018 e748OTHERSoftwareN/ACSH is brokenSomeone change the .cshrc file this morning
this morning at 11h30. None of the alias
worked anymore.
  
Pinnede744 Fri May 25 14:10:34 2018 Freddy, Louis, LamaOFFLINEGeneral MM - dE_TOF diagnostic with/without CsI_EAttached are some plot of dE-TOF PID for run
1004 after the threshold change ( 1 MeV -->
500 keV).
 T2_diag_dETOF.pdfMM12345_diag_dETOF.pdf 
Ongoinge744 Fri May 25 15:10:23 2018 FrançoisCATSHardware delay on CATSCATS delay 500 ns was changed to 300 ns, in
order to get the stop signal on MUST2 TAC
well within the range of the TAC. 
  
Ongoinge744 Fri May 25 15:19:38 2018 CristinaOFFLINEGeneral isomers present in the 14N runs
Attached are some pictures of the energy_vs_time
spectra for EXL and PPT detectors recorded
 14Nruns_isomers_EXL.pdf14Nruns_isomers_PPT.pdf 
Fixede744 Fri May 25 15:29:51 2018 FrançoisOTHERHardware tac FAG_HFA delay of 500 ns was put to the start signal
= FAG of the FAG_HF_validated TAC
and a range of 500 ns selected
  
Fixede744 Fri May 25 15:41:50 2018 e744- ValérieDAQGeneralN/ADAS configurations for E744-SummaryList of the saved files of the DAS configuration
used on May-24h and May-25th:
[Note that:
 11x 
Ongoinge744 Fri May 25 16:49:53 2018 e748MUST2GeneralCH2Raw dE-TOF bidim for the 5 telescopes zith the 14N beamPedestal seems not to be aligned for T4  Example1.png 
Fixede744 Fri May 25 17:42:21 2018 shumpeiMUST2SoftwareN/ACsI energy calibration for MUST2 telescopesCalibration parameter for MUST2 CsI crystals
are calculated with run1001-1006.
Attached are dE(DSSD)-E(CsI) plot of all
 6x 
Fixede744 Fri May 25 17:51:45 2018 e744 Iulan ValérieRUNGeneralCH2Run 1008 starting runs with 14N[U][B]BAD RUN FOR 14N[/B][/U]

Run 1008 BAD RUN 0 Events
 BTdssdMust2_25May18.pngHTdssdMust2_25May.pngHV_EXL_25May.png 
Fixede744 Fri May 25 18:15:41 2018 PierreMUST2GeneralCD2new dEE PID using Shumpei calibration   PID_dEE_newCsIcalib.png 
Fixede744 Fri May 25 21:34:19 2018 e744 Iulan, Fairouz, ValerieDAQHardwareCH2Problem of FAG validation +change of ADC parameter for TAC signalThe Time signal  MM_PPT is now in the parameter
MCP_HF of the ADC.
The problem was on the FAG signal validation
 ProblemGMTtriggerVXI.png 
Fixede744 Fri May 25 21:37:45 2018 iulian nicolas valérieRUNGeneralN/Arun 1012 all triggers 14O beam in D4   Run1012_E1_TcatsHF.pngRun1012_E1_T1HF_D6.png 
Fixede744 Fri May 25 22:03:12 2018 e744 Must2+LISE teamsRUNGeneralCH2RUNS 2000, 2001Beam 14O
Run 2000 start with all triggers, including
LISE trigger, during 1 minute for tests.
 6x 
Fixede744 Fri May 25 22:46:26 2018 e744 Must2+LISE teams, edit ValérieRUNGeneralCH2RUN 2004
 Run 2004   Date : 25/05/2018 22:45:58  25/05/2018
23:17:39 
 6x 
Ongoinge744 Fri May 25 23:15:57 2018 ValerianOFFLINEGeneralCH2Oxygen 15 excitation functionAfter a quick analysis, we can conclude that
MUST2 is working for resonant elastic scattering.
We find a resolution of 125 keV which will
 Exc_15O.pdf 
Fixede744 Fri May 25 23:26:23 2018 e744 Must2+LISE teams, edit ValérieRUNGeneralCH2START NARVAL, new versionthe NARVAL version is now reinstalled by Fred
Saillant.
see plot  ResolProblemNarvalEndofRun25may23h.png
 ResolProblemNarvalEndofRun25may23h.png 
Fixede744 Fri May 25 23:33:27 2018 iulianOTHERHardwareCH2changes after run 2004Target was changed to target number 5 (CH2
100 um GOODFELLOW)
The T_EXL1718 ok
 impact_matrix_run_2005.png 
Fixede744 Sat May 26 00:24:50 2018 e744 valérieDAQGeneralCH2change of trigger and divisionTRIGGER entry for CATS, presently IN8, seems
to be not running. We put it on IN7(Gatconf
=64).
 TRIGGERgmtChangedForCATS.png 
Fixede744 Sat May 26 00:38:38 2018 e744 all night shift teams, (edit) ValérieRUNGeneralCH2Run 2010RUN 2010
Date : 26/05/2018 00:37:44
 Target n°5  CH2 100 µm 
 20x 
Fixede744 Sat May 26 01:41:19 2018 e744 night shift team, edit valérieDAQGeneralCH2summary GMT, DAS config since run2005 Triggers: all except LISE

IN1 to IN5 : MM1 to MM5;
 chronoGMTmasterRun14O.pngchronogramGMTslave_Run14O.pngConfigDASThresMUST2at1MeV.pngConfigTDCTestParamInpectch0.png 
Fixede744 Sat May 26 02:39:18 2018 Mihai, PedjaRUNGeneralCH2Run 2011, Target n°5RUN 2011
Date : 26/05/2018 
Beam used: 14O
 9x 
Ongoinge744 Sat May 26 03:23:46 2018 ValerianOFFLINEGeneralCH215O excitation function as a function of ThetaLab   Ex_15O.pdf 
Fixede744 Sat May 26 04:24:16 2018 Mihai, PedjaRUNGeneralCH2Run 2012, Target n°5RUN 2012
Date : 26/05/2018 
Beam used: 14O
 Scalers_start_run2012.png 
Ongoinge744 Sat May 26 05:02:42 2018 ValerianTARGETGeneralCH2Problem with the target for run 2004 and 2010The target seems identical between the two
runs... We will have to investigate to understand
what is going on.
 Target_Issue_2004_2010.pngtargets.jpg 
Fixede744 Sat May 26 05:16:12 2018 Iulian Valerian, pedja, mihaiTARGETGeneralCH2Target changeTarget No 4 is placed in-beam 5h15.   
Fixede744 Sat May 26 05:23:12 2018 Mihai, PedjaRUNGeneralCH2Run 2013, Target n°4RUN 2013
Date : 26/05/2018 
Beam used: 14O
 Scalers_start_run2013.pngRun2014MM_IMPACT5h53.png 
Fixede744 Sat May 26 05:51:00 2018 Mihai, PedjaRUNGeneralCH2Run 2014, Target n°4[U][B]RUN 2014[/B][/U]
Date : 26/05/2018 
Beam used: 14O
 9x 
Fixede744 Sat May 26 06:05:13 2018 Iulian Valerian PedjiaTARGETGeneralCH2Target issue No5In the attached figure it could be seen that
with the target No5 (run 2010) lower energy
resonances were visible (lower part of the
 Target_Issue_2004_2010_corrected_withGeorgesTarget.png 
Ongoinge744 Sat May 26 07:03:12 2018 Valerian and Iulian and FrancoisOFFLINEGeneralCH2First picture for 15F from runs 2010-2013Target -> Thick target (Sandwich target ??)

The predictions are different....  
 Ex_15F_proj_0_8deg_17_27.pdfsig_theo.jpg 
   Sat May 26 07:20:03 2018          
Ongoinge744 Sat May 26 07:32:17 2018 Valerian, pedja, mihaiRUNGeneralCH2Run 2015 after vacuum problem of Css1 target 4[U][B]Run 2015[/B][/U] 
26/05/18
start @7h29
 Scalers_start_run2015.png 
Fixede744 Sat May 26 09:33:48 2018 Giuseppe, DieterRUNGeneralCH2run 2015Missing strips:
x
MM1 1
 MUST2_end_run2016.pngScalers_stop_run2015.pngMUST2_y_ongoing_run2018.png 
Fixede744 Sat May 26 10:03:37 2018 Giuseppe, DieterRUNGeneral run 2016& 2017

Problems with the ACQ (merger), they were
  
Ongoinge744 Sat May 26 10:06:06 2018 Giuseppe, DieterRUNGeneralCH2Run 2018Run 2018 
Start @ 09:56 26/05/2018
Stop @ 10:51
 Scalers_start_run2018.pngScalers_stop_run2018.pngMM_IMPACT_stop_run2018.png 
Fixede744 Sat May 26 11:19:37 2018 Giuseppe, DieterRUNGeneralCH2interruption due to to waterleak ibn CSS110:50 h Water leak at CSS1
        cryo pumps have to be regenerated
        continuation until the expert intervention
  
Fixede744 Sat May 26 11:42:03 2018 Giuseppe, DieterRUNGeneralCH2run 2019Run 2019

 scalers_start_run2019.pngscalers_stop_run2019.png 
Pendinge744 Sat May 26 13:22:08 2018 FrancoisBEAMGeneral problem with CSS1There is a leakage of water in cavity north
of cyclotron CSS1. I discussed with Alain
Savale. 
  
Fixede744 Sat May 26 14:03:27 2018 Giuseppe, Dieter, J. EricRUNGeneralCH2Run 2020run 2020

 9x 
Fixede744 Sat May 26 16:31:53 2018 Jaromir, GiuseppeEXL-PPTSoftwareN/AEXL calibration, light leakSelector.C can be used to adjust energy bad
reconstruction in the EXL detectors due to
energy leak (see the PDF attached for more
 batchSelector.CEXL.pdfpairs0102_0910.pdfpairs1112_1718.pdf 
Ongoinge744 Sat May 26 20:20:55 2018 ValerianOFFLINEGeneralCH215F observations for target 4 (100um)Analysis offline permits us to conclude that
the current target has the good size to observe
our resonances.
 Ecm_Thetacm_target4_notes.pdfEx_15F_proj_0_4deg_Ecm_target4.pdfEx_0to40degEcm_15F.pdf 
Ongoinge744 Sat May 26 20:23:52 2018 Cristina IulianOFFLINEGeneral  PPT gamma spectrum Attached is the gamma spectrum from PPT detector
with a condition on prompt (T_MM_PPT), proton
detected in MUST and multiplicity event equal
 runs_target4_proton_prompt_PPT.pdf 
Pendinge744 Sun May 27 15:36:31 2018 FrançoisBEAMSocial News updated19H
They succeeded to get 90 % transmission with
CSS1 but the energy at the exit was changed
  
Fixede744 Sun May 27 20:47:02 2018 e744- ValérieDAQHardwareN/AInfo to reboot the crates in case of problemIn case of problems in the run control, if
it is not possible to reinstall the crates
in the  "Ready" mode status, 
 IOLANmust2reboot.pngRunControlE744Ready.png 
Ongoinge744 Mon May 28 02:07:41 2018 ValerianOFFLINEGeneralCH2New improved resolution for 14N et 15FAfter correcting position for the Must2 detectors
by taking into account the real hit position
on the target that was (0,15cm,0) and not
 Exc_15O_newpos.pdfExc_15F_newpos.pdf 
Fixede744 Mon May 28 11:00:07 2018 Valerian and Iulian and PhilRUNGeneral two protons    e2p_15O.pdf 
Fixede744 Mon May 28 12:22:42 2018 Francois & MarleteOFFLINEGeneral superposition exp / theory for 15O and 15F   sum.jpg15F_final.jpg 
Pendinge744 Mon May 28 16:14:29 2018 François & ValérieMUST2GeneralN/ACooling system Monday 28th 16h10: we refill the cooling
system for MUST2. The level should be ok
for the rest of the experiment.
  
Fixede744 Mon May 28 16:32:38 2018 E744- ValérieDAQGeneralN/ADAS configuration files-References   15x 
Fixede744 Mon May 28 18:27:45 2018 E744- ValérieDAQGeneralN/AHigh Voltages and intensities on MUST2 Check regularly the HT on the MUST2.
NB: without beam the current on Telescope
4 is high, between 9 and 11 µA. see snapshot.
 HT_MUST2withoutBeam28may18.png 
Ongoinge744 Mon May 28 20:28:50 2018 Valerian and SimonOFFLINEGeneralCH2Explanation for the 0 degree low energy protonThe 0 degree energy proton are coming from
the beam hitting the aluminum target. If
you do a cut on the beam projection on T5
 explanation_spotT5.pdf 
Fixede744 Mon May 28 20:33:39 2018 E744- evening shift LISE+e744 teams (edit Iulan, Valérie)RUNGeneralCH2BEAM AGAIN RUN 2100 Tuning  See plot of the Must2 impacts, Wien filter
set at +/-0.8kV  
Beam on CATS 1.2 10**5 pps
 beam_spot_2100_end.png 
Fixede744 Mon May 28 20:38:21 2018 E744- evening shift LISE+MUST2 teams (edit Valérie)RUNGeneralCH2 RUN 2101 DATA target 4[U][B]run #2101[/B][/U]
start @20h27
stop  @22h20
 12x 
Fixede744 Mon May 28 22:11:52 2018 e748- LouisEXL-PPTHardwareCH2EXL4 trigger excessDuring run 2101, Frequency of the EXL4 channel
at 478Hz compared to values around 30Hz for
other channels
 Scaler_EXLTriggerExeed_Run2101.png 
Fixede744 Mon May 28 22:27:14 2018 Simon et LouisRUNGeneralCH2RUN 2102 DATA target 4 run #2102
start @22h20
stop @23h30
 Scaler_start_Run2102.pngScaler_stop_Run2102.png 
Ongoinge744 Mon May 28 23:14:26 2018 Valerian and Iulian and PhilOFFLINEGeneralCH2New hint of the 5/2- resonanceWe see a new hint of a resonance  New_resonance_hint.pdf 
Fixede744 Mon May 28 23:52:19 2018 Simon and Louis (start), Franck (stop)RUNGeneralCarbonRUN 2103 DATA target 1 Carbonrun #2103
start @23h49
stop @1h51
 Scaler_start_Run2103.pngScaler_stop_Run2103.png 
Ongoinge744 Tue May 29 00:17:18 2018 Night shifterBEAMGeneralCarbonBeam intensity variationBeam shut down at 00h03. Increase the intensity
progressively up to now 00h18.
At 0:28: primary intensity dropped down to
 hv_run2103.png 
Fixede744 Tue May 29 01:08:22 2018 e748BEAMGeneralCarbonBeam intensityAt 0:46 primary beam intensity reaches 5 uA
-> 1.5e5 pps on CATS.
  
Ongoinge744 Tue May 29 01:25:07 2018 Phil, Valerian, FranckEXL-PPTGeneralN/ALaBr3 Background for referenceFor reference, we include a picture of the
LaBr3 background found by some random Google
searching until we found a nice happy looking
 Response-of-3-3-CeBr-3-and-LaBr-3-Ce-to-natural-background-radiation-Events-due-to.jpg.png 
Ongoinge744 Tue May 29 01:33:58 2018 Phil, Valerian, FranckEXL-PPTGeneral Gamma-ray spectra from PPTAttached are plots of gamma-ray energy spectra
from PPT.
 PPTSpectrum2D.pngPPTSpectrumPromptPeak1.pngPPTSpectrumPromptPeak2.png 
Fixede744 Tue May 29 02:01:58 2018 Franck, Valerian, PhilRUNGeneralCarbonRun 2104 - Target 1 CarbonRun 2104
Beam: 14O
Start @ 2:00  29 May 2018
 Scaler_start_Run2104.pngScaler_stop_Run2104.png 
Fixede744 Tue May 29 02:38:21 2018 Franck, Valerian, PhilONLINEGeneralCarbonSpectra and HV Start Run 2104   11x 
Fixede744 Tue May 29 04:08:52 2018 Franck, Valerian, PhilRUNGeneralCarbonRun 2105 - Target 1 CarbonRun 2105

Beam 14O
 Scaler_start_Run2105.pngScaler_stop_Run2105.png 
Ongoinge744 Tue May 29 05:07:33 2018 ValerianOFFLINEGeneralCH2Fusion evaporation background and physics dataComparison between the fusion evaporation
background and the data for the telescope
5 ~0degree angles and rough subtraction of
 Ex_15F_and_Carbon.pdfrough_backgrund_suppresion.pdf 
Ongoinge744 Tue May 29 05:24:34 2018 ValerianEXL-PPTGeneralCH2Comparison of background and physics data for gamma ray detectorsGamma ray spectrum for both targets seem extremely
close. Subtraction is then made but no gamma
show. More statistic is needed
 gamma_comparison_CH2_and_Carbon.pdfbackground_substr_gamma_ray.pdf 
Ongoinge744 Tue May 29 06:09:58 2018 Franck, PhilRUNGeneralCarbonRun 2106 - Target 1 CarbonRun 2106

Beam 14O (CATS: 1.4e5 pps)
 Scaler_start_Run2106.pngScaler_stop_Run2106.png 
Ongoinge744 Tue May 29 07:39:53 2018 Franck, PhilRUNGeneralCarbonRun 2107 - Restart after Reboot of VXI - Same problemLow FAG rate: 30 /s
Low ORD rates: 0-10 /s, 15 /s on MM5
  
Ongoinge744 Tue May 29 08:09:02 2018 e748RUNGeneralCarbonRUN 2108 - Target1 Carbon[B][U]Run 2019[/U][/B]

Acq recovered
 11x 
Ongoinge744 Tue May 29 08:59:18 2018 Freddy, Iulian, PierreRUNGeneralCH2Run 2109 - Target 4 CH2 - Tuning run[U][B]Run 2109[/B][/U]

Beam: 14O (1.5*10^5 pps)
 impact_Run2109_end.png 
Pinnede744 Tue May 29 09:30:42 2018 Freddy, Iulian, PierreRUNGeneralCH2Run 2110 - Target 4 CH2Run 2110

Beam: 14O (1.5*10^5 pps)
 Scaler_start_Run2110.pngimpact_Run2110.png 
Pinnede744 Tue May 29 09:54:01 2018 Freddy, Iulian, PierreRUNGeneralCH2Run 2111 - Target 4 CH2Run 2111

Beam: 14O (1.5*10^5 pps)
 8x 
Fixede744 Tue May 29 10:42:00 2018 FrancoisBEAMGeneral effective beam timeIn total, there were:

1.4 UT of 14O, average 2x10^5 pps
 os1.jpeg 
Fixede744 Tue May 29 12:06:05 2018 iulian maurycyOFFLINEGeneralCH2Carbon substraction The scale factor for Carbon subtraction is
1.224. This means: RealCountsHidrogen=Counts
CH2 - 1.224 * CountsCarbon 
 Carbon_runs_scale_factor.pdf 
Ongoinge744 Tue May 29 15:53:17 2018 Iulian MaurycyOFFLINEGeneralCH2 Angular distribution per telescope 1-4Angular distribution per telescope 1-4  ang_dist.pdf 
Ongoinge744 Tue May 29 16:23:56 2018 Valerian Iulian MaurycyOFFLINEGeneralCH2Hint of a 2p peaks    Two_proton_14Ovs12C.pdf 
Fixede744 Tue May 29 17:14:41 2018 François, Freddy, ValérieBEAMGeneral Beam back in D6At 17pm Tuesday 29th we have beam coming back
in D6. Run (without storage)
NB: The current on T4 (with Hv set 60V) is
 HV_MUST2_29mayAt17h20.pngHV_EXL_29mayAt17h21.png 
Ongoinge744 Tue May 29 18:49:21 2018 ValerianOFFLINEGeneralCH2Comparison of gamma prompt spectrum for EXL and PPTQuick note, there is less data seen in EXL
prompt. This should be investigated
 prompt_energyEXL_12C_14O.pdfprompt_energyPPT_12C_14O.pdf 
Fixede744 Tue May 29 19:03:38 2018 Valérie (edit)RUNGeneralCH2Run #2112 BEAM BACKrun #2112
Beam back given by the machine. 
Beam+ Lise team: Omar K and Pierre Delahaye,
 14x 
Fixede744 Tue May 29 21:41:55 2018 ValérieRUNGeneralCH2Run #2113[B]RUN #2113 start @ 21h37; stop 0h14 @ [/B]

 target n°4. CH2 104 µm
 9x 
Fixede744 Wed May 30 00:17:02 2018 Valérie, LouisRUNGeneralCH2Run #2114[B]RUN #2114 30 mai 2018 start @ 0h15; stop
 @ 2h58[/B] 
BEAM STOP DUE to the shut down of the deviation
 8x 
Fixede744 Wed May 30 01:15:58 2018 Valérie, LouisRUNGeneralCH2Run #2115[B]Since 2h58 : WAITING FOR THE BEAM TO START
THE RUN[/B]
 6x 
Ongoinge744 Wed May 30 01:30:34 2018 ValerianOFFLINEGeneralCH2Checking the T4After checking every telescope, it seems that
the calibration of T4 is very off, I think
that for now, it might be a good thing to
  
Ongoinge744 Wed May 30 03:09:09 2018 Valérie, LouisRUNGeneralCH2BEAM LINE PROBLEMS, alarm "water leakage" The beam was stopped at 4h56, due to the
alarm of "water leakage" on equipments, indicated
in the D6 area.
 HT_MUST2_Wed30May_At5h52.png 
Fixede744 Wed May 30 07:06:58 2018 François, Louis, ValérieRUNGeneralCH2Run 2116Wednesday 30th May.
Beam back at 6h59. Run 2116 started at 6h69.
STOP at 8h33 (beam rate decreasing at the
 Run2116_initScalers.pngHT_MUST2_Wed30May_At7h05.pngRun2116_MMimpacts_At7h23.png 
Fixede744 Wed May 30 08:34:51 2018 ValérieRUNGeneralCH2Run 2117- low or no beamWednesday 30th May.
Run 2117 started at 8h36 ; Stop at 8h45:
  BEAM STOP.  
 Run2117_scalersINIT.png 
Ongoinge744 Wed May 30 09:24:28 2018 Valérie, Adrien, FreddyRUNGeneralCH2Run 2118Wednesday 30th May.
Run 2118 started at 9h23 ; 
Stop at: 11h25
 Run2118_scalersINIT.pngRun2118_HT_EXLat9h29.pngRun2118_scalersIat9h33.pngRun2118_MMimpacts.pngRun2118_CsIExlandETimePPT.png 
Ongoinge744 Wed May 30 11:17:00 2018 Valerian maurycy and OlivierOFFLINEGeneralCH2Condition on time T5One should add the condition Si_TX < 150 for
T5, problem from Ta?
 SiEvsSiT_T5_issue.png 
Ongoinge744 Wed May 30 11:22:16 2018 Valerian maurycy and OlivierOFFLINEGeneralCH2Checking the cutscut_DeE : 

- T1 OK
  
Fixede744 Wed May 30 11:42:41 2018 Freddy, AdrienRUNGeneralCH2Run 2119 - Target 4 CH2[U][B]Run 2119[/B][/U]

Start: 11h25m53s  
 Run2119_initScalers.pngRun2119_MMimpacts_At11h25.pngRun2119_TAC.pngRun2119_Trigger.pngRun2119_HT.png 
Fixede744 Wed May 30 13:25:57 2018 Phil, FreddyRUNGeneralCH2Run 2020 - target number 4Run 2020

Start: 13h24m
 Scalers_start_run2020.pngRun2020_leakages_13h35.png 
Ongoinge744 Wed May 30 13:48:02 2018 PhilDAQGeneralN/APulser valuesHere's a list of pulser values taken from
the last few runs as a reference for what
rate the "100-Hz" pulser is actually seeing.
  
Pendinge744 Wed May 30 15:26:40 2018 Francois MarleteBEAMGeneral problem with beam14H00 : The primary beam was misaligned, they
are still working on it. No 14O beam.
  
Ongoinge744 Thu May 31 16:42:30 2018 iulianEXL-PPTGeneral60CoRun 3000 60Co source PPT & EXL triggersRun 3000 
60Co source 
PPT & EXL triggers
 Scalers_start_run3001.pngScalers_stop_run3000.pngppt_exl_run3000.pngppt_exl2_run3000.png 
Ongoinge744 Thu May 31 17:45:21 2018 iulianMUST2General3-alpharun #3001 alpha source T1-T5run #3001 
alpha source T1-T5
 scaler_startt_run3001.pngscaler_stop_run3002.png 
Fixed  Fri Jun 1 11:11:13 2018 e748EXL-PPTGeneral60Corun #3002 ppt triggerrun #3002 ppt trigger

start 11h11
 scaler_start_run3003.png 
Fixede744 Fri Jun 1 12:06:47 2018 ValerianEXL-PPTGeneral60Corun #3002 Calib PPTrun #3002
start @11h09
stop  @12h04
 scaler_start_run3002.pngscaler_stop_run3002.png 
Fixede744 Fri Jun 1 13:12:16 2018 ValérianEXL-PPTGeneral60Corun #3003 Calib EXLrun #3003
start @12h06
stop @13h08
 scaler_start_run3003_ok.pngscaler_stop_run3003_ok.png 
Fixede744 Fri Jun 1 19:19:20 2018 ValerianMUST2GeneralN/ACalimero endFor the end of e744 the calimero runs are
the following ones :
  
Fixede744 Fri Jun 1 19:29:17 2018 ValérianEXL-PPTGeneral60CoEXL efficiencyTo test EXL efficiency I start a long run
with the source in front of it.
run 3014
  
Fixede744 Mon Jun 4 18:45:15 2018 shumpeiMUST2HardwareN/ANo data for two CsIs of MUST2During E744 experiment, two CsI crystals (number
1 & 10) of MM2 had no data as attached.
When the M2C chamber was opened today, I
 csin.pdfcsip.pdf 
Fixede755 Thu Jun 7 09:46:47 2018 LouisMUST2Hardware3-alphaMUST2 testThe 8 MUST2 telescopes have been installed
and pluged in final configuration. Energy
signal have been tested with 2 3-alphas sources,
 MMYE1_4_3alpha_06062022.pngMMYE1_4_3alpha_06062021.pngMMYE1_4_3alpha_06062020.pngMMYE1_4_3alpha_06062019.png 
Fixede755 Thu Jun 7 12:19:18 2018 LouisMUST2HardwareN/AMUST2 configuration, X1Y1 and PCB numbers Pictures showing the MUST2 configuration.

X and Y positions and PCB numbers are as
 Config_MUST2.jpgConfig_MUST2_T56.jpg 
Fixede755 Thu Jun 7 13:54:59 2018 LouisMUST2HardwareN/AMUST2 TimeMUST2 time have been tested for T1 to T8 for
X and Y. FAG has been delayed with 150ns
and 500ns and sent as stop in MUVI 1 and
 MMXT1_4_07062018.pngMMYT1_4_07062018.pngMMYT5_8_07062018.pngMMXT5_8_07062018.png 
Fixede755 Thu Jun 7 17:12:38 2018 LouisMUST2Hardware3-alphaCalib AlphaRun 3028: Calib 3 alpha started at 17h05
Run3029:Calib 3 alpha started at 17h54
 MMYE1_4_08062020.pngMMYE5_8_08062020.pngMMXE1_4_08062019.pngMMXE5_8_08062019.png 
Fixede755 Sun Jun 10 19:51:52 2018 LouisMUST2Hardware3-alphaCalibration runRun3031:Calib 3 alpha started at 19h49
stoped at 9h15
 MMYE1_4_11062020.pngMMYE5_8_11062020.pngMMXE1_4_11062020.pngMMXE5_8_projection_11062020.png 
Fixede755 Mon Jun 11 12:50:56 2018 LouisMUST2Hardware3-alphaImpact matrixImpact matrix for Telescope 1 to 8 from the
run 3029 calibrated
 8x 
Fixede755 Mon Jun 11 13:28:21 2018 LouisONLINEHardware3-alphaEnergy resolution MUST2Energy resolution of the 8 MUST2 telescopes
from central pic of run 3029 calibrated with
3 alpha source.
  
Fixede755 Wed Jun 13 11:40:18 2018 Adrien Valérian LouisMUST2Hardware3-alphaRUN Source 4 Alpharun 1: 4 alpha source, without heat shield,
1 bar air inside the target cell.
 EY_T1_4_13062020.pngEY_T5_8_13062020.pngEX_T1_4_13062020.pngEX_T5_8_13062020.png 
Fixede755 Wed Jun 13 18:20:02 2018 linda, Daisuke, LouisMUST2Hardware3-alphaRun test with heat shieldRun 9: heat shield added with two MYLAR foil
of 2um, vacuum has been made inside target
cell. 4 alpha source
 EX_T5_8_14062021.pngEX_T1_4_14062021.pngEY_T1_4_14062021.pngEY_T5_8_14062021.png 
Fixede755 Thu Jun 14 19:03:06 2018 shumpeiOFFLINEHardwareN/A8MeV alpha source T5&T68MeV alpha source image on the target was
reconstructed from T5&T6 hit position and
target position information.
 tqrget.pdf 
Ongoinge755 Sun Jun 17 11:12:57 2018 takeshiTARGETHardwareN/Acurrent cooling status of the cryogenic targetThis is the temperature log of the cryogenic
hydrogen target during the cooling test done
last night.
  
Fixede755 Sun Jun 17 11:25:31 2018 takeshiTARGETHardwareN/Acurrent cooling status of the cryogenic targetThis is the temperature log (temperaturelog.png)
of the cryogenic hydrogen target during the
cooling test done last night.
 temperaturelog.pngpeak.pngboilingpoint.png 
Fixede755 Wed Jun 20 16:45:00 2018 Cherry Mx TeamTARGETHardware3-alphaCryo target is filledWe dont see any alpha in detector 1-6m but
we see them in detector 7-8.
 targefilled.png 
Fixede755 Thu Jun 21 07:16:51 2018 daisukeRUNGeneral3-alphaRUN0039 three alpha source calibration MUST2alpha source run over night.   2018-06-21_run0039_01.png2018-06-21_run0039_02.png2018-06-21_run0039_03.png2018-06-21_run0039_04.png 
Fixede755 Thu Jun 21 08:30:48 2018 shumpei, daisukeRUNGeneral3-alphaRUN0040 8 MeV source run with liquid hydrogenRUN 0040. started 7:23 am 
 
8 MeV source was set upstream at 0 degrees
 2018-06-21_run0040_04.png2018-06-21_run0040_03.png2018-06-21_run0040_02.png2018-06-21_run0040_01.png 
Fixede755 Thu Jun 21 09:16:01 2018 KitamuraTARGETHardwareN/AVapor pressureVapor pressures of isobutane and CF4. The
paramateres of the Antoine equation were
taken from the following NIST
 isobutane.pngcf4.png 
Fixede755 Thu Jun 21 16:08:32 2018 ValerianMUST2HardwareN/ARun 3045 : Calimero T1-4 X[U][B]run 3045 :[/B][/U] 
start @ 15h35
stop  @ 15h59
 Calimero_T14X.png 
Pinnede755 Thu Jun 21 17:07:14 2018 ValerianMUST2GeneralN/Arun 3046 : Calimero T1-4Y[U]run 3046 :[/U] Calimero T1-4Y
start @ 16h28
stop @ 16h53
 Calimero_T14Y.png 
Fixede755 Thu Jun 21 18:52:42 2018 takeshiTARGETHardwareN/Atarget temperatureHere I attached the temperature graph which
is daisuke discussed at the meeting.
Both the heat shield (blue line) and the
 entire.pngheatshield.pngcell.png 
Fixede755 Fri Jun 22 03:11:51 2018 takeshiVACUUMHardwareN/Athe vacuum pressure and the target temperatureWe suspect the impurity of the vacuum causes
the rebound behavior of the target temperature.
Here is the data of the vacuum level v.s.
 M2CvsE.pngM2CvsA.png 
Fixede755 Fri Jun 22 11:14:01 2018 Valerian Pierre Adrien Fred FairouzMUST2GeneralN/Arun 3061 Calimero T5-8X[U][B]run 3061[/B][/U] Calimero T5-8X  Calimero_T56X.png 
Fixede755 Fri Jun 22 13:06:29 2018 daisukeMECHHardwareN/AAlpha source ladder instructionInstruction to translate and rotate the ladder
for the alpha sources (by Vincent).
 Photos_porte_source_E755_et_E738.pdf 
Fixede755 Fri Jun 22 13:22:19 2018 daisukeBEAMSocialN/ABeam time schedule   Planning_Faisceaux_2018_run2.1.0.pdf 
Fixede755 Fri Jun 22 13:45:22 2018 takeshiTARGETHardwareN/Atarget temperature statusI would report the temperature log of the
target up to now.
We started cooling the target at 2:21 last
 tamperature2206.pngheatshield2206.pngcell2206.png 
Fixede755 Fri Jun 22 16:57:27 2018 daisukeOTHERSocialN/AWho is in GANIL?https://docs.google.com/spreadsheets/d/1jyfw2LT3RGfDY4bcN-9zepZoe2uE89b5CnF4tXzqzj8/edit#gid=0   
Fixede755 Fri Jun 22 21:29:05 2018 Pierre Fairouz Adrien ValerianCATSGeneralN/AMask reconstructionMask reconstruction after adding the correction
shift for CATS.
CATS1X = +1.2
 mask1.pdfmask2.pdf 
Fixede755 Fri Jun 22 22:12:43 2018 daisukeMECHHardwareN/Apositions, geometry and dimensions   2018-06-13_e738_setup.pdf 
Fixede755 Fri Jun 22 22:14:30 2018 daisukeMECHHardwareN/AHow to open the gate valve   Procedure_d'ouverture_de_vanne_d'isolement.pdf 
Pinnede755 Fri Jun 22 22:15:45 2018 daisukeMECHHardwareN/AGeometrical survey report   E755_Listing_DSSD_X_Y_Z_repère_géomètre.xlsxSTP-2018-E071_D6_E755_Sorlin_Suzuki.pdfE755_Listing_DSSD_X_Y_Z.pdf 
Fixede755 Sat Jun 23 00:39:33 2018 takeshiTARGETHardwareN/Atarget temperatureHere is the target temperature up to now.
The black lines are today, and bule lines
are reference.
 temperature2306.pngheatshield2306.pngcell2306.png 
Fixede755 Sat Jun 23 00:47:19 2018 Pierre ValerianCATSGeneralN/ANEW GOOD mask reconstructionNew shift for CATS after fixing Z position
of CATS detectors.
 mask1.pdfmask2.pdf 
Fixede755 Sat Jun 23 01:55:34 2018 LouisBEAMGeneralCH2Beam tunning Beam was tunned around 1am to center the beam
on the target. On the picture, up is after
tunning and down is before tunning
 beam_tunnig23_06_1h_24m.png 
Fixede755 Sat Jun 23 02:28:03 2018 daisuke, louis, jean-charlesRUNSoftwareN/ARUN 63 (40Ca + liq. H) scalers and histograms   10x 
Fixede755 Sat Jun 23 04:16:03 2018 LouisOFFLINEGeneralCH2deltaE Ehere a plot of energy loss in DSSD in function
of energy in CsI. One can distinguish proton
and alpha. Energy calibration of the CsI
 dEE_noCalib_run0063.jpg 
Fixede755 Sat Jun 23 06:10:01 2018 shumpeiOFFLINESoftwareCH240Ca(p,d)39 reactionQuick off-line analysis for 40Ca(p,d)39 reaction
is shown.
Used gate - target image and dE-E cut for
 tgt.pdfd.pdf40capd.pdfspectre_en_energie_39Ca.PNG 
 e755 Sat Jun 23 06:16:25 2018 LouisOFFLINE       
Fixede755 Sat Jun 23 14:07:40 2018 Valerian FairouzRUNGeneralN/AProblem from NarvalNarval actors went crazy see joined images

run 75 must be trash as it created 18 files
 narval_crash_problem.png 
Fixede755 Sat Jun 23 17:07:35 2018 daisukeRUNSocialCryo-LH2Run summary 40Ca(p,t)
total 698 min or 11.6 hours (RUN 63 to 81)
  
Fixede755 Sat Jun 23 18:01:58 2018 Valerian MaurycyOFFLINEGeneralCryo-LH2Kinematics LineUsing Run 63-64 we drew Kinematics Lines.  ELab_Theta_2H.pdfELab_Theta_3H.pdfELab_Theta_3He.pdfELab_Theta_4He.pdf 
Ongoinge755 Sat Jun 23 19:10:23 2018 NoriOFFLINEGeneralCryo-LH2     
Fixede755 Sat Jun 23 19:23:19 2018 NoriOFFLINEGeneralCryo-LH240Ca(p,d) analysisAnalysis of run0072 (two hours accumulation).

Attachment 1:
 7x 
Fixede755 Sat Jun 23 19:41:49 2018 ValerianOFFLINEGeneralCryo-LH2De_E plot   DE_E_run81.pdf 
Fixede755 Sat Jun 23 20:07:49 2018 NoriOFFLINEGeneralCryo-LH240Ca(p,d) analysisOrder-of-magnitude estimate of the target
thickness.
  
Fixede755 Sat Jun 23 20:09:00 2018 e755 - Louis, ValérieRUNGeneralCryo-LH2Summary of Runs 83 and 84CATS beam 1.7 10**5 /s
Run 84 start @ 19h54 stop @ 20h16 on problem
with CO. 
 19x 
Fixede755 Sat Jun 23 20:56:15 2018 e755 - Louis, ValérieRUNGeneralCryo-LH2Run 85CATS beam 1.8 10**5 /s
Beam back at 20h46
Run 85 start @ 20h46 stop @21h02
 Run85_scalersAt20h48.pngrun85Catsprofiles.png 
Fixede755 Sat Jun 23 21:17:03 2018 e755 - Louis, Daisuke,Shumpei, Takeshi, Nori, ValérieRUNGeneralCryo-LH2Run 86-Test measurement, (heating target)CATS beam 1.8 10**5 /s
 Target temperature increasing during the
run. 
 run86Scalersat21h17.png 
Fixede755 Sat Jun 23 21:51:04 2018 e755 - Louis, Daisuke,Shumpei, Takeshi, Nori, ValérieRUNGeneralCryo-LH2Run 87-Empty target measurementCATS beam 1.8 10**5 /s
  
Run 87 start @ 21h42 stop @ 21h52
 run87Scalersat21h50.png 
Fixede755 Sat Jun 23 21:52:01 2018 takeshiTARGETHardwareN/Aevaporation of the H2 targetHere is the temperature log when we evaporate
the liquid H2 by heater.
The black line is what we call sensor A,
 evapolation2206.png 
Fixede755 Sat Jun 23 22:03:17 2018 takeshiTARGETHardware temperasure log of the targetI would report the temperature log of the
liquid hydrogen target.
We consider it was liquid during the beam
 temperature2306_3.pngheatshield2306_3.pngcell2306_3.png 
Fixede755 Sat Jun 23 22:08:43 2018 e755 - Louis, Daisuke,Shumpei, Takeshi, Nori, ValérieRUNGeneralCryo-LH2Run 88-Empty target measurementCATS beam 1.8 10**5 /s
  
Run 88 start @ 22h07 
 10x 
Fixede755 Sat Jun 23 22:42:33 2018 e755 - Louis, ValérieMUST2HardwareCryo-LH2Status of MUST2 telescopes 23 june 18Temperatures on the Mufeex-Y cards:

Temperature set on the regulator -4°C. 
 TempMuvi1MufeeXTel1.pngTempMuvi2MufeeYTel6.pngTempMuvi2MufeeXTel6.pngTempMuvi2MufeeXTel8.pngRun88scalersAt23hORDI.png 
Fixede755 Sat Jun 23 22:47:21 2018 daisukeTARGETHardwareN/ATarget images for Liq. H and empty targetAttachment 1: RUN 85 with liq.H
Attachment 2: RUN 88 with empty cell. 
 run85Catsprofiles.pngRun88emptyT_CATSbeam.png 
Fixede755 Sat Jun 23 23:31:12 2018 e755 - Louis, ValérieRUNGeneralCryo-LH2Run 89-Empty target measurementCATS beam 1.5 10**5 /s
  
Run 89 start @ 23h29 
 Run89CatsProfile.png 
Fixede755 Sat Jun 23 23:38:32 2018 takeshiRUNSocialCryo-LH2run summary for 40Ca beamrun 63-85, total 15.4 hours data was taken

start                           stop    
  
Fixede755 Sun Jun 24 09:48:46 2018 FrancoisBEAMGeneral No beam9H10: D5 stopped at 9H10. We asked for the
beam.
  
Fixede755 Sun Jun 24 12:10:35 2018 takeshiTARGETHardware target temperature during warming upI would share the temperature log of the target
cell during warming it up.
The thick line is the heat shield and the
 warming2406.png 
Fixede755 Sun Jun 24 13:20:51 2018 takeshiTARGETHardwareN/Astart target coolingWe started cooling the target cell at 13h03.
It would take 4 - 5 hours to cooling the
cell enough low for liquifying hydrogen.
 cooling.png 
Fixede755 Sun Jun 24 13:25:30 2018 Fred, valérian, daisukeDAQSoftwareN/AScalers to be recordedThe U2M scaler data cannot be recorded to
the data in the current version of Narval.
So there were no scaler data during the measurement
 run_0092_scaler_record_check.png 
Fixede755 Sun Jun 24 16:11:38 2018 daisukeONLINESoftwareN/AAutomated screenshotautomated screenshot commands

shot: Capture the window selected by the
 screenshot_2018-06-24_18h06m11s.pngscreenshot.sh 
Fixede755 Sun Jun 24 19:51:09 2018 Faïrouz, daisukeMUST2SoftwareN/AMUST2 threshold from 1 MeV to 0.4 MeVIt was found that the energy threshold of
the strips was as high as 1 MeV. This threshold
setting was chosen for the calibration and
 das_2018-06-24_19h33m36s.pngdas_2018-06-24_19h49m33s.png 
Fixede755 Sun Jun 24 21:12:25 2018 e755- ValérieDAQHardware GMT chronogrammesSince Run 103 
MASTER FAG 100 ns
SLAVE 100 ns
 GMTchronoMaster.pngGMTslaveChrono.png 
Fixede755 Sun Jun 24 21:36:24 2018 e755-Daisuke, Louis, Olivier, Pierre, ValérieRUNGeneralCryo-emptySummary of the evening runs 101-105 run 101 Trigger Cats 2 div; div =1
Run 102
run 103
  
Fixede755 Sun Jun 24 22:13:29 2018 takeshiTARGETHardwareN/Atemperature log of the targetWe liquefied the hydrogen target around 22:00.

I would share the temperature log of the
 temperature2406.pngliquiefy2406.png 
Fixede755 Sun Jun 24 22:13:30 2018 valérie, daisukeRUNGeneralCryo-emptyRUN 106 empty target run Run 106 Trigger: MUST2 all
start @ 21h31; stop @21h44
e748 Exxx 168154021 24 juin 21:44 /data/e748X/e755/acquisition/run/run_0106.dat.24-06-18_21h31m26s
 6x 
Fixede755 Sun Jun 24 22:18:50 2018 Daisuke, Louis, Olivier, ValérieRUNGeneralCryo-LH2RUN 107 run 107 start@ 22h15; stop@ 22h37
e748 Exxx  689278077 24 juin  22:37 /data/e748X/e755/acquisition/run/run_0107.dat.24-06-18_22h15m16s
 Run107scalersAt22h17.pngTriggerMMandCATSRun107.pngTriggerMM2Run107ProfileCats.png 
Fixede755 Sun Jun 24 22:25:53 2018 shumpeiMUST2HardwareCryo-emptythreshold for DSSDs of MUST2Threshold of DSSDs were changed (see entry
638).
Run 106 was analyzed and now the threshold
 pid_threshold.pdf 
Fixede755 Sun Jun 24 22:26:46 2018 Daisuke, Louis, Olivier, ValérieMUST2GeneralCryo-LH2HV on MUST2, currents, CATSNB: SiLi1 T7 has tripped during the evening
(since Run 104)
It should be put on again (better to increase
 HVonMUST2mm1to4.pngHVonMUST2mm5to8andCats.png 
Fixede755 Sun Jun 24 22:45:41 2018 Daisuke, Fairouz, Louis, Olivier, ValérieRUNGeneralCryo-LH2RUN 108Run 108 24 june start @ 22h38 stop@ 22h54
e748 Exxx   10068143 24 juin  22:54 /data/e748X/e755/acquisition/run/run_0108.dat.24-06-18_22h38m15s
Triggers Cats 1 and Cats2 DIV= 3.10**4
 Run108_TriggersCats.pngRun108_scalersAt22h43.png 
Fixede755 Sun Jun 24 23:07:45 2018 Daisuke, Fairouz, Louis, Olivier, ValérieRUNGeneralCryo-LH2RUN 109Run 109 24 june start @ 22h54 stop@ 23h14
e748 Exxx    9723325 24 juin  23:14 /data/e748X/e755/acquisition/run/run_0109.dat.24-06-18_22h54m26s
Triggers Cats 1 and Cats2 DIV= 3.10**4
 Run109_ScalersAt22h56.pngRun109_CaTSslits10-8.5mm_W1.4kV.pngRun109_CaTSslitsFV43at10mm_W2kV.pngRun109filledTargetOKat16KprojBinY.pngRun109_ProfileCatsTargetFrame.png 
Fixede755 Sun Jun 24 23:17:01 2018 Daisuke, Fairouz, Olivier, ValérieRUNGeneralCryo-LH2RUN 110Run 110 24 june start @  23h15 stop @ 23h56

e748 Exxx 1339167483 24 juin  23:56 /data/e748X/e755/acquisition/run/run_0110.dat.24-06-18_23h14m53s
 Run110-projCATS2-CAVIAR.png 
Fixede755 Mon Jun 25 00:13:48 2018 Valérie +Lise team: Jean-Charles+Target: TakeshiRUNGeneralCryo-LH2RUN 111Run 111  25th June start @ 0h stop @1h13

e748 Exxx 3590054877 25 juin  01:14 /data/e748X/e755/acquisition/run/run_0111.dat.24-06-18_23h59m52s
 16x 
Fixede755 Mon Jun 25 01:17:47 2018 Valérie +Lise team: Jean-Charles+Target: TakeshiRUNGeneralCryo-LH2RUN 112Run 112 25th June start @ 1h14 stop @ 2h34
e748 Exxx 3769078581 25 juin  02:34 /data/e748X/e755/acquisition/run/run_0112.dat.25-06-18_01h14m03s
 Target seems to be still ok and filled!
 6x 
Fixede755 Mon Jun 25 02:34:30 2018 Valérie +Target: TakeshiRUNGeneralCryo-LH2RUN 113Run 113 25th June start @ 2h34 stop @ 3h57
e748 Exxx 3957925525 25 juin  03:57 /data/e748X/e755/acquisition/run/run_0113.dat.25-06-18_02h34m10s
  
 11x 
Fixede755 Mon Jun 25 03:14:38 2018 e755- ValérieMUST2HardwareCryo-LH2HV on MUST2, currents, CATSHV and currents on DSSDs and SiLi at 3h20

Telescopes T
  
Fixede755 Mon Jun 25 03:58:22 2018 Valérie +Target: TakeshiRUNGeneralCryo-LH2RUN 114Run 114 25th June start @ 3h57 stop @5h34

 e748 Exxx 4015552641 25 juin  05:34 /data/e748X/e755/acquisition/run/run_0114.dat.25-06-18_03h57m34s
 9x 
Fixede755 Mon Jun 25 05:34:53 2018 Valérie +Target: TakeshiRUNGeneralCryo-LH2RUNS 115, 115.1 and 116Run 115 25th June start @ 5h34 stop@6h57
 e748 Exxx 4190541701 25 juin  06:57 /data/e748X/e755/acquisition/run/run_0115.dat.25-06-18_05h34m33s
  -------------------------------------------------------------------------------
 Run115ScalersAt5h37.pngRun115ScalersAt6h39.pngRun115b1_scalersAt7h20.pngRun115b1BeamProfile.pngRun116_ScalersAt8h26.png 
Fixede755 Mon Jun 25 07:41:54 2018 e755- ValérieMUST2HardwareCryo-LH2SUMMARY-Missing strips MUST2 
Telescopes MM1 to MM8, XY ET missing strips
----------------------------------------------------
 Run116_MMimpacts.pngRun116_Mm1toMM4StrXYE.png 
Fixede755 Mon Jun 25 08:38:34 2018 e755 Louis Francois MarleneOFFLINEGeneralCryo-LH238Ca(p,d)37CaHere Kine;atics line for th 38Ca(p,d)37Ca
Channel from run 110
 38Capd_Run_110.pdf38Capd_Run_110_deE.pdf 
Fixede755 Mon Jun 25 10:42:47 2018 daisukeTARGETSoftwareCryo-emptyRUN 106 (38Ca with empty target) S/Np, d, t ratio of events from the target frame
and from the target aperture (covered by
Havar 6.5 micron and Mylar 2 micron)
 7x 
Fixede755 Mon Jun 25 11:12:40 2018 Marlene, LouisRUNGeneralCryo-LH2Run 117, 38Ca, Target filledRun 117 started @ 10h54
stopped @ 11h32 to enter in the room
 6x 
Fixede755 Mon Jun 25 11:23:03 2018 daisukeTARGETHardwareCryo-LH2Liquid hydrogen target logger   screenshot_2018-06-25_11.17.25.png 
Fixede755 Mon Jun 25 13:40:41 2018 Marlene, LouisRUNGeneralCryo-LH2RUN 118 CAVIAR settingrun 118 started @ 13h13
stopped @ 13h43
  
Fixede755 Mon Jun 25 14:04:17 2018 Marlene, LouisRUNGeneralCryo-LH2RUN 119 to 121Test for CAVIAR and CHIO:

run 119 with two 1/15 reductors   13h54 to
  
Fixede755 Mon Jun 25 16:09:24 2018 Benoit LouisRUNGeneralCryo-LH2Beam and CHIO tunningDuring run 126:
Beam was tuned to be centered and homogeneous
in the target
 Run126_deTOF.pngRun127_BeamProfil.pngRun126_Scaler.png 
Fixede755 Mon Jun 25 17:23:49 2018 Cherry Mx TeamCATSGeneralCryo-LH2Confirmation of CATS reconsctructionGating on T1 to T4 and looking at the the
CATS reconstruction on Target we confirm
that;
  
Fixede755 Mon Jun 25 17:27:59 2018 Cherry Mx TeamCATSGeneralCryo-LH2Confirmation of CATS reconsctructionGating on T1 to T4 and looking at the the
CATS reconstruction on Target we confirm
that;
 CATS_TELESCOPE.pngCATS_TELESCOPE_M2.png 
Fixede755 Mon Jun 25 17:28:03 2018 NoriTARGETHardwareCryo-LH2Liquid hydrogen target loggerThere was a sudden increase of the temperature
at 4pm. Fortunately the cell is still filled
with liquid.
 screenshot_2018-06-25_17.20.41.pngscreenshot_2018-06-25_20.56.47.png 
Fixede755 Mon Jun 25 17:31:27 2018 daisuke, benoïtTARGETSoftwareCryo-LH2RUN 127 (Still filled with liquid? test with Wien filter)Beam spot is now too small to see the cell
frames. Moved the beam spot down and up by
using the Wien filter. It is not conclusive,
 Run127_BeamProfil.pngvisu_2018-06-25_16h55m02s.pngvisu_2018-06-25_17h01m48s.pngRun127_target_frame_projection.png 
Fixede755 Mon Jun 25 17:40:44 2018 daisuke, benoïtRUNSoftwareCryo-LH2RUN 128 (Still filled with liquid?)   10x 
Fixede755 Mon Jun 25 19:59:55 2018 marlène, freddy, benoïtMUST2SoftwareN/ACsI disabled in MUVI/DAS   das_2018-06-25_19h59m19s.png 
Fixede755 Mon Jun 25 20:01:07 2018 olivier, daisukeDAQHardwareN/AAnother deat time scalerAfter RUN 128, we entered D6 to add another
deat time measuring pulses to U2M channel
9. This is supposed to correctly measure
  
Fixede755 Mon Jun 25 20:10:21 2018 daisuke, benoîtRUNGeneralCryo-LH2RUN 130   scaler_2018-06-25_20h10m53s.pngvisu_2018-06-25_20h09m45s.png 
Fixede755 Mon Jun 25 20:51:28 2018 daisuke, benoîtRUNGeneralCryo-LH2RUN 132Previous run (RUN 131), MUST2 threshold was
raised to 500 keV to see how the dead time
(DESC/ORDD) is improved (found not effective).
 9x 
Fixede755 Mon Jun 25 22:40:38 2018 daisuke, benoîtRUNGeneralCryo-LH2Run 133same as before  6x 
Fixede755 Mon Jun 25 23:17:56 2018 NoriMUST2GeneralCryo-LH2DSSD multiplicityData analysis of run0128 (38Ca physics run).
The multiplicity distribution of DSSD x-strips
is shown in
 run0128_simult.pngrun0128_allstripsfired.pngAll_strip_fired.png 
Fixede755 Tue Jun 26 00:09:40 2018 Pierre, ShumpeiRUNGeneralCryo-LH2RUN 134Start: 26/06 at 00h05
Stop: 00h40 for PCP to improve beam tuning
 scaler_2018-06-26_00h08m24s.pngvisu_2018-06-26_00h08m16s.pngvisu_2018-06-26_00h42m01s.png 
Fixede755 Tue Jun 26 01:02:42 2018 Pierre, ShumpeiRUNGeneralCryo-LH2RUN 135start: 01h02
stop: 01h39
Trigger: MM+CATS DIV
 scaler_2018-06-26_01h02m13s.pngvisu_2018-06-26_01h04m13s.png 
Fixede755 Tue Jun 26 01:46:56 2018 Pierre, ShumpeiRUNGeneralCryo-LH2RUN 136start: 01h41
stop: 03h37
 scaler_2018-06-26_01h44m28s.pngvisu_2018-06-26_01h47m09s.pngvisu_2018-06-26_02h53m01s.png 
Fixede755 Tue Jun 26 03:40:15 2018 Pierre, ShumpeiRUNGeneralCryo-LH2RUN 137start: 03h38
stop: 05h30
 scaler_2018-06-26_03h38m54s.pngvisu_2018-06-26_03h42m20s.pngvisu_2018-06-26_04h05m34s.png 
Fixede755 Tue Jun 26 05:32:35 2018 Pierre, ShumpeiRUNGeneralCryo-LH2RUN 138start: 05h31
stop: 07h23
 scaler_2018-06-26_05h32m06s.pngvisu_2018-06-26_05h36m34s.png 
Fixede755 Tue Jun 26 05:40:52 2018 PierreOFFLINEGeneralCryo-LH2Kinematic lines and Ex for 38Ca(p,d) and 38Ca(p,t)-> 38Ca(p,d) uses run 128 

-> 38Ca(p,t) uses run from 128 up to 136
 kine_38Capd_run0128.pdfkine_38Capt.pdf 
Fixede755 Tue Jun 26 06:55:51 2018 shumpeiOFFLINESoftwareCryo-LH238Ca (p,t) 36CaRun 111 - 134 (almost 1 day) data was analyzed.

38Ca (p,t) 36 Ca was reconstructed
only from Telescope 3.
 fig.pdf 
Fixede755 Tue Jun 26 07:24:59 2018 Pierre, Shumpei, Yorick, ValérieRUNGeneralCryo-LH2RUN 139start: 07h23
stop: 8h52
748 Exxx 3003199093 26 juin  08:52 /data/e748X/e755/acquisition/run/run_0139.dat.26-06-18_07h22m11s
 scaler_2018-06-26_07h23m43s.pngvisu_2018-06-26_07h26m37s.pngrun139_scalersAt8h42.png 
Fixede755 Tue Jun 26 08:13:02 2018 e755-ValérieRUNGeneralCryo-LH2data Run summary 25+26 june RUNS since 25 june pm till 26th june 7am

r--r--r-- 1 e748 Exxx  626093289 25 juin
  
Fixede755 Tue Jun 26 08:15:29 2018 e755 ValérieMUST2GeneralCryo-LH2MUST2 configuration summaryMUST2 summary 26th June 8am

MUST2 thresholds 402 keV
 MUST2thresholdsMUVI1tel1.pngrun139_CsIpadsMUST2.png 
Fixede755 Tue Jun 26 09:01:23 2018 Daisuke, Takeshi, Yorick, ValérieRUNGeneralCryo-LH2RUN 140Run 140 start: 8h55 ;  stop: 9h59
748 Exxx 3348489513 26 juin  09:59 /data/e748X/e755/acquisition/run/run_0140.dat.26-06-18_08h55m32s
 The Csi Mate were put on the trigger again
 21x 
Fixede755 Tue Jun 26 09:28:10 2018 takeshiTARGETHardwareN/Atemperature of the target<p>Here is the temperature log of the target
up to now. The black lines are today, The
thicker line is the cell window this time,
 temperature2606.pngcell2606.pngscreenshot_2018-06-26_10.30.45.pngscreenshot_2018-06-26_17.49.29.png 
Fixede755 Tue Jun 26 09:52:35 2018 e755 ValérieRUNGeneralCryo-LH2Summary of MUST2 plots during RUN 140 
 The Csi Mate were put on the trigger again
"ENABLE" in the DAS. see snapshot DAS_MUVIenableMateCsI.png
 Run140_MM1toMM4StrXYE.pngRun140_MM5toMM8StrXYE.pngRun140_MMCsiIpads.pngRun140_MMImpacts.png 
Fixede755 Tue Jun 26 10:03:29 2018 Adrien, Freddy, Marlène, Yorick, Valérie, target team: Daisuke, TakeshiRUNGeneralCryo-LH2RUN 141Run 141 26th june start@ 9h59 stop: 11h12
e748 Exxx 3449521865 26 juin  11:12 /data/e748X/e755/acquisition/run/run_0141.dat.26-06-18_09h59m57s
 Run141_scalersAt10h.pngRun141_SiLiT7andT8.pngRun141_scalersAt10h55.pngRun141_scalersAt10h56.pngRun141_scalersAt10h57.png 
Fixede755 Tue Jun 26 11:47:16 2018 NPTeamOFFLINEGeneralCryo-LH238Ca(p,d)37Ca with zero deg gateWe did a gate on IC_E:PlasticQ on particle
with same Z as the beam.
 37Ca_getzero.pdf 
Fixede755 Tue Jun 26 12:06:41 2018 Marlène, Yorick, ValérieRUNGeneralCryo-LH2RUN 142Run 142 26th june start@ 11h15 ; stop @11h27
 
e748 Exxx  492791993 26 juin  11:27 /data/e748X/e755/acquisition/run/run_0142.dat.26-06-18_11h15m39s
 ChangeThresholdsAllMateXMUST2.pngRun142_scalersAt11h17.png 
Fixede755 Tue Jun 26 12:18:08 2018 Marlène, Yorick, ValérieRUNGeneralCryo-LH2RUN 143Run 143 -- 26th June start@ 11h30  ; stop
@ 11h39 
 e748 Exxx  306151179 26 juin  11:39 /data/e748X/e755/acquisition/run/run_0143.dat.26-06-18_11h30m28s
 Run143scalersAt11h32.pngMMthresSilIMate17beforRun143PROB.pngMMthresSilIMate17run143okSiLi.png 
Fixede755 Tue Jun 26 12:22:56 2018 Marlène, Yorick, ValérieRUNGeneralCryo-LH2RUN 144Run 144 -- 26th June start@ 11h51  ; stop
@ 13h04
  
 6x 
Fixede755 Tue Jun 26 13:37:44 2018 LouisRUNGeneralCryo-LH2RUN 145Run 145 -- 26th June start@ 13h07 ; stop @
13h53
e748 Exxx 2239017281 26 juin  13:53 /data/e748X/e755/acquisition/run/run_0145.dat.26-06-18_13h06m30s
 run_145_CATScal.pngrun_145_CATSpos.pngrun_145_scalers.png 
Fixede755 Tue Jun 26 13:41:02 2018 NPTeamMUST2GeneralN/AChange in strips firingAfter changing in the MUST2 the range of pedestal
suppression (Ex < 8192 are rejected and Ey
> 8192 are rejected). we can clearly see
 firing_strips_X.pngfiring_strips_Y.png 
Fixede755 Tue Jun 26 14:29:36 2018 Daisuke, Louis, Olivier, ValérieRUNGeneralCryo-LH2RUN 146Run 146 -- 26th June start@ 14h07 stop @14h26
e748 Exxx  844513015 26 juin  14:26 /data/e748X/e755/acquisition/run/run_0146.dat.26-06-18_14h07m32s
 The run was stopped: no beam due to the
  
Pinnede755 Tue Jun 26 14:36:55 2018 NPteamOFFLINEGeneral 36Ca(p,t)36Ca analysis with zero degree gateHere is an update of the analysis results
for the 36Ca(p,t)36Ca channel.
 kine_38Capt_zrodeg_gated.pdfkine_38Capt_zrodeg_gatedZ19.pdf 
Fixede755 Tue Jun 26 14:39:52 2018 Charles H, Fred, ValérieRUNGeneralCryo-LH2RUN 147Run 147 -- 26th June start@ 14h34

We make some checkings of DT during the beam
 6x 
Fixede755 Tue Jun 26 16:38:31 2018 daisuke, thomasRUNGeneralCryo-LH2RUN 148 38Ca beam is back   20x 
Fixede755 Tue Jun 26 16:40:12 2018 AdrienOFFLINESoftwareCryo-LH2Offline analysis of p,tLooking at the pt reaction and gating on z
= 20 (solid blue) and z=19 (hashed orange)
we see a total of three states.
 38Capt_zero.pdf 
Fixede755 Tue Jun 26 18:28:45 2018 NoriOFFLINEGeneralCryo-LH237K,36Ar(p,t)Offline analysis of run0128-0146.

Cuts: tritons, target aperture, beam, and
 run0128_0146_37K.pngrun0128_0146_36Ar.png 
Fixede755 Tue Jun 26 18:41:59 2018 e755 ValérieRUNGeneralCryo-LH2DAS save Summary of MUST2 conditions -26 juneThe DAS configuration was saved under e755.das
and .e755_Exp26juin18Mmthreshd.das at 18h40
on June 26th.
  
Pinnede755 Tue Jun 26 18:43:35 2018 ValerianDAQGeneralN/AIn case of a data acqusition crash [CODE]If the acquisition crash :  

In the acquisition terminal : 
  
Fixede755 Tue Jun 26 20:33:17 2018 daisukeTARGETHardwareCryo-LH2Liquid hydrogen target logger   screenshot_2018-06-26_20.31.03.png 
Fixede755 Wed Jun 27 01:14:54 2018 takeshiRUNGeneralN/Aonline run summaryI made a run summary up to now in the google
spreadsheet.
You can see it on
  
Fixede755 Wed Jun 27 01:34:28 2018 Freddy LouisRUNGeneralCryo-LH2run153RUN 153

started @ 1h26  stopped at 3h20
 12x 
Fixede755 Wed Jun 27 03:29:32 2018 Freddy LouisRUNGeneralCryo-LH2RUN 154RUN 154
started @ 3h21    stopped @ 5h16
 Run154_scaler.pngRun154_scaler_stop.pngRun154_pos.png 
Fixede755 Wed Jun 27 04:12:57 2018 Freddy, LouisOFFLINEGeneral Summary 38Ca(p,t) - Run 128 to 152Here are some plot summarizing the analysis
status on the 38Ca(p,t) channel.
 kine_38Capt_zrodeg_r128_152.pdfkine_38Capt_zrodeg_r128_152_T1.pdfkine_38Capt_zrodeg_r128_152_T3.pdfkine_38Capt_zrodeg_r128_152_telescope.pdf 
Fixede755 Wed Jun 27 04:46:00 2018 Freddy, LouisOFFLINEGeneral Summary 36Ar(p,t) - Run 128 to 152 Here are some plot summarizing the analysis
status on the 36Ar(p,t) similar method as
in the previous entry:
 kine_36Arpt_zrodeg_gatedZ18.pdfkine_36Arpt_zrodeg_gatedZ18_r128_152_T1.pdfkine_36Arpt_zrodeg_gatedZ18_r128_152_telescope.pdf 
Fixede755 Wed Jun 27 05:20:22 2018 Freddy LouisRUNGeneralCryo-LH2RUN 155RUN 155
started @ 5h16
stoppd at 7h19
 Run155_pos.pngRun155_scaler.pngRun155_scaler_stop.png 
Fixede755 Wed Jun 27 05:57:49 2018 takeshiTARGETGeneralCryo-LH2liquid hydrogen target temperature logIt seems stable.
We liquified H2 around 22:00 24th (10 hours
in the plot), so the target kept liquefying
 target2706.png 
Fixede755 Wed Jun 27 07:23:41 2018 Freddy LouisRUNGeneralCryo-LH2RUN 156RUN 156
start at 7h17
 Run156_scaler.png 
Fixede755 Wed Jun 27 09:26:32 2018 daisukeTARGETHardwareCryo-LH2Liquid hydrogen target logger (end of 38Ca)   screenshot_2018-06-27_09.24.30.png 
Fixede755 Wed Jun 27 09:49:37 2018 gibRUNGeneralCryo-LH2RUN 155run 156 for ref. LH2 empty next  visu_2018-06-27_09h50m34s.pngvisu_2018-06-27_09h51m31s.pngvisu_2018-06-27_09h51m43s.png 
Fixede755 Wed Jun 27 10:08:41 2018 gibRUNGeneralCryo-emptyRUN 157 : empty target   17x 
Fixede755 Wed Jun 27 10:15:36 2018 takeshiRUNGeneralCryo-LH2run summary for 38Ca + liq. H2Here is the summary of the measurement of
38Ca + liq H2 setting.
We obtain 49.45-hour data acquisition time,
  
Fixede755 Wed Jun 27 11:38:09 2018 gibOTHERGeneralCryo-emptyend of empty run. Now tuning for 36Ca   visu_2018-06-27_12h28m12s.pngvisu_2018-06-27_12h28m20s.png 
Fixede755 Wed Jun 27 16:03:26 2018 ValerianOFFLINEGeneralCryo-LH2Status on 38Ca pt   38Ca_pt_Canvas.pdf 
Ongoinge738 Wed Jun 27 17:42:14 2018 daisukeOTHERGeneralN/AToward E738   2018-06-13_e738_setup.pdf2018-06-27_e738_chio_energy_loss.pdf2018-06-27_e738_light_tof.pdfpt_sim_phase_plot_7c_t.pdfpt_sim_phase_plot_8c_d.pdf 
Fixede755 Wed Jun 27 17:46:55 2018 e755 Daisuke, Olivier, Nori, ValérieRUNGeneralCryo-emptyRun 158 Beam tuning for 36CaAt 17h20 the beam was back in D6, ready for
tuning by the LISE team, Christelle, Jean-Charles
and Omar.
 Run158TrajectoriesWien1kV.pngRun158_trajectoriesWien5kV.pngRun158_trajectoriesWien5kVok.png 
Fixede755 Wed Jun 27 18:34:31 2018 NPTeamOFFLINEGeneralCryo-LH2Update on the offline Analysis for 38Ca pt run 128-156   kine_38Capt_zrodeg_r128_156_telescope.pdfkine_38Capt_zrodeg_r128_156.pdf 
Fixede755 Wed Jun 27 18:42:09 2018 e755 Daisuke, Olivier, ValérieRUNGeneralCryo-emptyRUN 159 New input in GMT trigger +scalersChange of the GMT Trigger:
New entry IN12 corresponding to the CATS2
with no division.
 Run159ICDEtacatsHf.pngRun159ICDEtacatsHf_Wien100kV74kV.pngRun159Trajectories_Wien100kV74kV.pngRun159_GMTalltriggerINewN12.png 
Fixede755 Wed Jun 27 19:26:13 2018 e755 Daisuke, Olivier, Pierre, ValérieRUNGeneralCryo-emptyRun 160 -36CaTAC Cats2 HF delay of 30 ns is removed

Run 160 start@ 19h13 stop@ 19h23
 Run160CDEtacatsHf_Wien100kV74kV.png 
Fixede755 Wed Jun 27 19:27:52 2018 e755 Daisuke, Olivier, ValérieRUNGeneralCryo-emptyRun 161 -36Ca
Run 161 start @ 19h25 stop@20h30
e748 Exxx 2411656699 27 juin  20:30 /data/e748X/e755/acquisition/run/run_0161.dat.27-06-18_19h25m55s
 Run161CDEtacatsHf_noWienSlitsFH31at8mm.pngRun161_scalers.pngRun161EdeTOFcats.pngRun161TrajectoriesCats.pngRun161mmImpactswithSourcealpha.png 
Fixede755 Wed Jun 27 20:39:44 2018 e755 Daisuke, Olivier, ValérieRUNGeneralCryo-emptyRun 162-38CaTUNING OF 38Ca to be ready during the night,
after the change of CO source at the machine
between 21h ans 2am (maybe). 
 run162_Trajectories38Ca.pngRun162_ICDE1TcatsHF.png 
Fixede755 Wed Jun 27 21:13:40 2018 e755 Daisuke, Olivier, edited by ValérieRUNGeneralCryo-emptyConditions for the night runsTUNING OF 38Ca to be ready during the night,
after the change of CO source at the machine
between 21h and 2am (maybe). 
 GECOIonizationTrigger.png 
Fixede755 Wed Jun 27 21:25:41 2018 takeshi, shunpeiTARGETHardwareCryo-LH2temperature log during liquefyingHere is the temperature log before and during
liquefying.
The temperature trend seems almost same as
 temperature2706.png 
Fixede755 Wed Jun 27 22:11:20 2018 Pierre LouisOFFLINEGeneralCryo-LH2Offline analysis of 38Ca(p,d)37CaHere are some plot summarizing the analysis
status on the 38Ca(p,d) channel.
 38Ca_pd_128_156.png38Ca_pd_128_156_kinematics.png 
Fixede755 Thu Jun 28 00:59:55 2018 e755 Daisuke, Olivier, Pierre, Marlène, Valérie (: edit)RUNGeneralCryo-emptyRun 162-38Ca 38Ca  Beam at ~ 2. 10**5 pps
F62 +/- 11 mm
SLITS FH31 +/-5mm (moment)
 8x 
Fixede755 Thu Jun 28 01:22:43 2018 NoriBEAMGeneralN/ABrho2 scan for 36CaWe scanned Brho2 values so that the yield
for 36Ca is maximized (LISE run0282-0284,
F31 = +/-6 mm). 
 brho2scan_1.pngbrho2scan_2.pngbrho2scan_3.png 
Fixede755 Thu Jun 28 01:26:34 2018 daisukeTARGETHardwareCryo-LH2Liquid hydrogen target logger   6x 
Fixede755 Thu Jun 28 01:39:58 2018 NoriBEAMGeneralN/ADiscussion for 37CaOffline analysis of LISE run0281. F31 = +/-0.6
mm, without wedge.
The number of counts of 37Ca was 900/50 =
 run281.pngScreen_Shot_2018-06-27_at_23.16.13.png 
Fixede755 Thu Jun 28 08:20:54 2018 e755RUNGeneralCryo-LH2Summary Run 165-169[U][B]Run 165[/B][/U][LIST]
[*] start: 2h12
[*] start: 3h11
 run169_tof_cats_hf.png 
Fixede755 Thu Jun 28 08:29:08 2018 LouisCATSGeneralCryo-LH2CATS-HF time shift during the night noticed that the time between CATS and HF
shifted during the night.
run 164 to 169
 6x 
Fixede755 Thu Jun 28 08:34:30 2018 Fredd, Louis, DaisukeRUNGeneralCryo-LH2Run 170[U][B]Run 170[/B][/U][LIST]
[*] start: 8h07
[*] start: 
 Run170_TCATSHF.pngscreenshot_2018-06-28_08h15m27s.pngRun170_Target.pngRun170_CATSRAWt.pngRun171_M2Energy.png 
Fixede755 Thu Jun 28 10:38:58 2018 FreddyMUST2GeneralN/AT4 Comparison of calibration runsAttached is a comparison of two calibrations
performed before the experiment (2 weeks
and one day before) for Telescope 4 (the
 Comparison_T4X_calibbeforeexp.pngComparison_T4Y_calibbeforeexp.png 
Fixede755 Thu Jun 28 11:08:54 2018 Freddy, Louis, ValerianRUNGeneral3-alphaIntermediate alpha calibration T1234Run 171

3alpha source in front of T1234
 Run171_M2_EX_RAW.pngRun171_M2_EY_RAW.png 
Fixede755 Thu Jun 28 13:53:28 2018 Freddy, Louis, ValerianRUNGeneral3-alphaPedestal run + Alpha calibration with T4 at 80V[U][B]Run 172[/B][/U][LIST]
[*] Pedestal run on T1234
[/LIST]
  
Fixede755 Thu Jun 28 19:58:12 2018 shumpeiOTHERGeneralN/AKinematics for the reactionsKinematics for the reaction of E755 and E738
experiments are shown.
 kinematics.pdf 
Fixede755 Fri Jun 29 00:59:15 2018 LouisOFFLINEGeneralCryo-LH2Analysis of run 63 to 89 in 40Ca(p,t)38ca reactionHere are some plot summarizing the analysis
status on the [B]40Ca(p,t)38Ca[/B] channel.
 Chek_cuts_40Capt.pngKinematics_40Ca.pngTelescope_Ex_40Capt.pngKinematics_38Capt_withoutCutZero.png 
Fixede755 Fri Jun 29 07:04:35 2018 takeshiTARGETHardwareCryo-LH2target temperature   temperature2906.png 
Fixede755 Fri Jun 29 08:06:42 2018 e755 ValérieRUNGeneralCryo-LH2Summary- Runs 175 to 184 RUNS during night Friday 29th June From 3am
to 8am
RUNS 175 to 184 - cf Scalers, and TcatsHF
 29x 
Fixede755 Fri Jun 29 08:08:47 2018 e755 Valérie, target :TakeshiRUNGeneralCryo-LH2Run 185 Run 185 Friday 29th June start@ 8h stop@
All triggers Must2+cats1 and Cats 2 div.
Mean intensity of the beam between 1.5 to
 13x 
Fixede755 Fri Jun 29 09:08:50 2018 e755 ValérieRUNGeneralCryo-LH2Plots of Must2 and Cats during run 185 Run 185 Friday 29th June see entry 742
 
Mean intensity of the beam between 1.5 to
 14x 
Fixede755 Fri Jun 29 09:53:33 2018 e755 Valérie, target :TakeshiRUNGeneralCryo-LH2Run 186 problem Numexo-Narval Run 186 Friday 29th June start@ 9h48 stop@9h51
e748 Exxx        551 29 juin  09:50 run_0186.dat.29-06-18_09h48m46s
All triggers Must2+cats1 and Cats 2 div.
 Run186_TNarvalECCerror.png 
Fixede755 Fri Jun 29 10:03:06 2018 e755 Valérie, target :TakeshiRUNGeneralCryo-LH2Run 187 Run 187 Friday 29th June start@ 9h53 stop@10h05
e748 Exxx   95976431 29 juin  10:05 run_0187.dat.29-06-18_09h53m29s
All triggers Must2+cats1 and Cats 2 div.
  
Fixede755 Fri Jun 29 10:41:00 2018 e755 Olivier, Omar, Valérie, target : Nori, TakeshiRUNGeneralCryo-LH2Run 188 Run 188 Friday 29th June start@10h36 stop@10h40
   
e748 Exxx   18078991 29 juin  10:40 run_0188.dat.29-06-18_10h35m20s
  
Fixede755 Fri Jun 29 12:39:34 2018 LouisOFFLINEGeneralCryo-LH2Analysis 38Ar(p,t)36Ar run 128 to 156Here are some plot summarizing the analysis
status on the [B]36Ar(p,t)34Ar[/B] channel.
 Kinematics_36Arpt_FIT1Ex.pngKinematics_36Arpt_FITGS.pngTelescope_Ex_36Arpt.pngTelescope_Ex_36Arpt_fit_exited.png36Ar_level_list.png 
Fixede755 Fri Jun 29 14:00:51 2018 takeshiRUNGeneralCryo-LH2summary for 38Ca + liq. H2, updatedI would summarize the statistics of 38Ca +
liq H2 target measurement.
We have obtained the data of 49.5 hours until
  
Fixede755 Fri Jun 29 17:43:46 2018 daisukeTARGETGeneralCryo-LH2liquid hydrogen target loggerlast update at 6 am on June 30  10x 
Fixede755 Sat Jun 30 01:55:35 2018 Daisuke LouisRUNGeneralCryo-LH2RUN 191 FIRST PHYSICS RUN IN 37Ca [B]RUN 191:[/B]

started at 1:20
 21x 
Fixede755 Sat Jun 30 03:18:14 2018 louis, DaisukeRUNSoftwareCryo-LH2Beam image after retracting CHIOthe beam seems more focused after retracting
CHIO upstream of M2C. Screenshot taken during
RUN 193.
 visu_2018-06-30_03h12m04s.png 
Fixede755 Sat Jun 30 04:54:55 2018 louis, daisukeRUNGeneralCryo-LH2RUN 194Continue 37ca data taking.  45x 
Fixede755 Sat Jun 30 08:24:07 2018 Freddy, Shumpei, LouisRUNGeneralCryo-LH2Run 195[U][B]Run 195[/B][/U]

start: 8h08
 r195_scalers.pngr195_CATS.pngr195_TOFin.pngr195_PIDout.pngr195_CATS_cal.png 
Fixede755 Sat Jun 30 09:17:25 2018 LouisOFFLINEGeneralCryo-LH2Offline analysis of 37Ca(p,d)36Ca reactionHere are some plot summarizing the analysis
status on the [B]37Ca(p,d)36Ca[/B] channel.
Plots shows performance of the setup in optimal
 7x 
Fixede755 Sat Jun 30 11:25:59 2018 Freddy, ShumpeiBEAMGeneralCryo-LH2Beam profile on the targetThe beam profile on the target for each isotope
is shown for all trigger and CATS trigger
(GATCONF > 500).
 target_profile_triggerall.pngtarget_profile_catstrigger.png 
Fixede755 Sat Jun 30 13:25:27 2018 Freddy, Shumpei RUNGeneralCryo-LH2Run 196[U][B]Run 196[/B][/U]

start: 12h09
 r196_TOF.pngr196_CATS.pngr196_MMImpact.png 
Fixede755 Sat Jun 30 15:07:43 2018 ShumpeiOFFLINEGeneralCryo-LH237Ca (p,d) 36Ca37Ca (p,d) 36Ca reaction is analyzed (run
191-195).
 37Capd.pdf37Capd_ex.pdf 
Fixede755 Sat Jun 30 16:04:11 2018 takeshi, DidierRUNGeneralCryo-LH2run 197RUN 197
start: 15h39
stop : 17h38
 scaler_2018-06-30_16h01m46s.pngvisu_2018-06-30_16h03m02s.pngvisu_2018-06-30_16h03m18s.pngvisu_2018-06-30_16h03m32s.png 
Fixede755 Sat Jun 30 17:45:17 2018 takeshi, DidierRUNGeneralCryo-LH2run 198RUN 198

start: 17h40
 scaler_2018-06-30_17h46m13s.pngvisu_2018-06-30_17h46m31s.pngvisu_2018-06-30_17h47m00s.pngvisu_2018-06-30_17h47m36s.png 
Fixede755 Sat Jun 30 18:09:53 2018 NoriBEAMGeneralN/ABrho1 and Brho2 optimization for 37CaSummary of Brho1 and Brho2 optimizations for
the 37Ca beam.
 ca37brho1.pngca37brho2_yield.pngca37brho2_purity.png 
Fixede755 Sat Jun 30 20:08:27 2018 didier, takeshi, daisukeRUNGeneralCryo-LH2RUN 199 37C + liq. H   28x 
Ongoinge755 Sat Jun 30 20:12:33 2018 takeshiTARGETHardwareCryo-LH2target temperatureLatest update at 12:15 pm on July 1
Here are the temperature logs up to 7:00pm
30/6.
 cell3006.pngheatshield3006.pngscreenshot_2018-07-01_12.17.48.pngscreenshot_2018-07-01_12.22.29.png 
Fixede755 Sat Jun 30 23:47:45 2018 takeshi, daisuke, didierBEAMGeneralCryo-LH2result of Wien filter tuningHere is the summary of the tuning of the Wien
filter.
We tried to tune the correction magnet of
 wienfiltertuning3006.pngposition37Carun200.pdfposition37Carun201.pdfposition37Carun202.pdf 
Fixede755 Sun Jul 1 01:48:23 2018 AdrienRUNGeneralCryo-LH2Run 205Starting Run 205 at 1:45.
CATS 1 : 22kps
CATS 2 : 23kps
 run205.psr205_HT.pngr205_HT0.pngr205_scaler.png 
Fixede755 Sun Jul 1 06:42:28 2018 e748OFFLINEGeneralCryo-LH2offline of run 192-206Here are te plot of analysis of all the available
statistic on 37Ca pd.
 37Capd_192-207.pdf 
Fixede755 Sun Jul 1 09:43:58 2018 Benoit, FranckRUNGeneralCryo-LH2Run 209Sun 1 July 2018

Start 9:26
 9x 
Fixede755 Sun Jul 1 09:48:44 2018 Benoit, FranckRUNGeneralCryo-LH2Run 208 (ENTRY POSTPONED)Sun 1 July

Start 7:17
 7x 
Fixede755 Sun Jul 1 11:52:27 2018 Benoit, FranckRUNGeneralCryo-LH2Run 210Sun 1 July 2018

Start 11:46
 7x 
Fixede755 Sun Jul 1 17:11:55 2018 freddy, shumpei, daisukeRUNGeneralCryo-LH2RUN 214from 17h01 to 18h39 (anoxia alert. PCP to
inspect). 40ca ~3.0 micro A. 
 scaler_2018-07-01_17h10m39s.pngvisu_2018-07-01_17h10m37s.pngvisu_2018-07-01_17h13m30s.png 
Fixede755 Sun Jul 1 18:44:24 2018 freddy, shumpei, daisukeRUNGeneralN/ARUN 215Alpha source run. no beam
18h41 started 
19h22 ended
 scaler_2018-07-01_19h19m32s.pngvisu_2018-07-01_19h19m32s.pngvisu_2018-07-01_19h19m40s.png 
Fixede755 Sun Jul 1 19:22:32 2018 freddy, shumpei, daisukeRUNGeneralCryo-LH2RUN 21637Ca + liquid H 
19h20 started 
21h18 ended.
 scaler_2018-07-01_20h05m43s.pngvisu_2018-07-01_20h05m44s.pngvisu_2018-07-01_20h05m52s.pngscaler_2018-07-01_21h18m41s.pngvisu_2018-07-01_21h15m41s.png 
Fixede755 Sun Jul 1 21:21:18 2018 LouisRUNGeneralCryo-LH2RUN 217RUN 217

37Ca beam at 2.104pps
 scaler_2018-07-01_21h22m54s.pngvisu_2018-07-01_21h24m16s.pngvisu_2018-07-01_21h24m33s.pngvisu_2018-07-01_21h24m48s.png 
Fixede755 Sun Jul 1 23:18:21 2018 freddy, louis, shumpei, daisukeRUNGeneralCryo-LH2RUN 21937ca + liq.H
22h50 started
1h00 ended
 7x 
Fixede755 Mon Jul 2 00:53:49 2018 LouisOFFLINEGeneralCryo-LH2Update offline analysis on 37Ca(p,d)36CaUpdate of offline Analysis of the [B]37Ca(p,d)36Ca[/B]
reaction:
 37Capd_All_192_0277.pdf37Capd_Z19_FIT_192_0277.pdf37Capd_Z20_192_0277.pdf 
Fixede755 Mon Jul 2 01:01:15 2018 daisukeTARGETGeneralCryo-LH2cryogenic target logger   screenshot_2018-07-02_01.00.23.pngscreenshot_2018-07-02_02.57.45.pngscreenshot_2018-07-02_02.58.23.png 
Fixede755 Mon Jul 2 01:05:12 2018 louis, daisukeRUNGeneralCryo-LH2RUN 22037ca + liq.H. 
1h01 started
1h56 ended.Beam stopped due to the trouble
 scaler_2018-07-02_01h04m12s.pngvisu_2018-07-02_01h13m50s.pngvisu_2018-07-02_01h14m29s.pngscaler_2018-07-02_01h56m52s.png 
Ongoinge738 Mon Jul 2 01:29:58 2018 daisukeOTHERGeneralN/AQ values 
 Reaction : 9C(1H,3H)7C  Q -29.5098 MeV 
Ex 0 MeV
 ensdf_10c_via_pt.pdfPhysRevC.3.543.pdf 
Fixede755 Mon Jul 2 01:58:07 2018 louis daisukeRUNGeneralCryo-LH2RUN 22137ca + liquid H (no beam at times)
1h56 started. no beam
2h03 beam came back with low intensity. Keep
 6x 
Fixede755 Mon Jul 2 04:02:15 2018 louis daisukeRUNGeneralCryo-LH2RUN 22237Ca + liq. H
4h00 started
5h57 ended.
 8x 
Fixede755 Mon Jul 2 06:01:35 2018 louis daisukeRUNGeneralCryo-LH2RUN 22337Ca + liquid H. 
5h59 started.40Ca  3.3 micro A.
6h17 beam stop. A few issues including RF.
 7x 
Fixede755 Mon Jul 2 07:23:34 2018 LouisOFFLINEGeneralCryo-LH236Ca and 37Ca(p,t)35Ca reactionRUN 192 to 222
Beam time: 62h
 Ex_all_pd_192to222.pngKinematics_pt_192to222_Z20.pngpt36Ca_Ex_all_192_222.pngKinematics_pt_192to222_Z20_without_line.png 
Fixede755 Mon Jul 2 07:37:50 2018 LouisOFFLINEGeneralCryo-LH2BackgroundRUN: 86 to 89
Beam time: 2h30
Beam 40Ca with empty target
 Background.pdf 
Fixede755 Mon Jul 2 08:02:52 2018 louis, daisuke, takeshi, didierRUNGeneralCryo-LH2RUN 22437ca + liquid H
8h00 started
10h01 stopped
 scaler_2018-07-02_08h49m14s.pngvisu_2018-07-02_08h49m47s.pngvisu_2018-07-02_08h50m07s.png 
Fixede755 Mon Jul 2 10:12:29 2018 takeshi, didierRUNGeneralCryo-LH2RUN225RUN 225
37Ca beam + liq. H2 target 
10h02 started
 6x 
Fixede755 Mon Jul 2 12:09:56 2018 takeshi, Didier, olivierRUNGeneralCryo-LH2RUN 226RUN 226
37Ca + liq. H2
12h03 started
 6x 
Fixede755 Mon Jul 2 14:37:44 2018 takeshi, didier, olivierRUNGeneralCryo-LH2RUN 229RUN 229
Ca37 + liq. H2
Beam is back 14h30
  
Fixede755 Mon Jul 2 15:18:21 2018 takeshi, didier, olivierRUNGeneralCryo-LH2RUN 230RUN 230
37Ca + liq. H2
15h11 started
 7x 
Ongoinge755 Mon Jul 2 16:16:40 2018 takeshiTARGETHardwareCryo-LH2temperature log of the target   heatshield0207.pngcell0207.png 
Fixede755 Mon Jul 2 16:50:10 2018 Pierre, LouisRUNGeneralCryo-LH2Run 231start: 16h47
stop: 18h48
 scaler_2018-07-02_16h51m31s.pngvisu_2018-07-02_16h52m17s.png 
Fixede755 Mon Jul 2 18:51:11 2018 PierreRUNGeneralCryo-LH2Run 232start: 18h49
stop: 20h41
 31x 
Fixede755 Mon Jul 2 20:44:46 2018 pierre, takeshi, louis, nori, daisukeRUNGeneralCryo-LH2RUN 23337ca + liquid H 
20h43 started.
21h31 stop
 scaler_2018-07-02_20h45m47s.pngscaler_2018-07-02_21h31m38s.png 
Fixede755 Mon Jul 2 22:07:27 2018 takeshiRUNGeneralCryo-LH2run summary for 37CaHere is the run summary for the 37Ca beam
+ liq. H2 target setting.
We have obtained totally 65.7 hours data
  
Fixede755 Mon Jul 2 23:09:55 2018 Pierre LouisOFFLINEGeneralCryo-LH237Ca(p,d)36Ca   37Ca_pd_192_231.pdf 
Fixede755 Mon Jul 2 23:29:32 2018 Louis PierreOTHERHardwareN/ANew division for CATS and plasticDivision for CATS and Plastic: 3x10^4

Plastic => Chanel 16 of GMT
  
Fixede755 Tue Jul 3 10:47:01 2018 LouisRUNGeneralCryo-LH2Run beam tuningRUN 236:
started 10h46,  stop at 11h09
  
Fixede755 Tue Jul 3 11:48:30 2018 LouisRUNGeneralCryo-LH2Run 238 40CaRUN 238:

START 11h23    STOP 12h50
 18x 
Fixede755 Tue Jul 3 13:37:19 2018 LouisRUNGeneralCryo-LH2RUN 239Run 239
 start at  13h30,  stop at 15h18
 scaler_2018-07-03_13h39m44s.pngvisu_2018-07-03_13h39m45s.pngvisu_2018-07-03_13h41m15s.png 
Fixede755 Tue Jul 3 15:22:26 2018 LouisRUNGeneralCryo-LH2RUN 240Run 240
start at 15h17, stop at 
 scaler_2018-07-03_15h23m18s.png 
Fixede755 Tue Jul 3 17:54:44 2018 LouisRUNGeneralCryo-LH2Run 241, 242RUN 241

start at 17h12, stop at 17h35
 22x 
Fixede755 Tue Jul 3 17:55:01 2018 takeshiRUNGeneralCryo-LH2run summary for 40Ca + liq. H2We have obtained 6 hours data for the 40Ca
+ liq. hydogen setting today. 
Note that we have already taken 15.6 hours
  
Fixede755 Tue Jul 3 18:19:30 2018 daisukeOTHERGeneralN/AData saved after experimentDAS:    /home/e748/ganacq_manip/e755/das-save/2018-07-03_fin_e755
VIGRU:  /home/e748/analysis/GRU/GRU_18_03_12/vigru_e755_final_2018-07-03_18h15.xml
Online: /home/e748/analysis/nptool/Projects/e755/online/2018-07-03_18h24
 7x 
Fixede755 Tue Jul 3 18:48:52 2018 takeshiTARGETHardwareCryo-LH2temperature log of the targetHere is the temperature log of the liquid
hydrogen target,
We kept it about 6 days. 
 cell0307.pngheatshield0307.png 
Fixede755 Wed Jul 4 15:34:34 2018 NoriBEAMGeneralN/A37Ca PID at D6A particle identification plot of 37Ca from
LISE run 332.
 run332_pid.pdf 
Pinnede738 Wed Jul 4 17:10:28 2018 daisukeDAQSoftwareN/ASecond VXI for CATS and syncrhonisation diagnosisSecond VXI crate was addedd for CATS. If the
synchronisation is not working, there will
be error messages in the display of the chassis
 8x 
Fixede738 Wed Jul 4 18:11:44 2018 louis, daisukeRUNGeneral3-alphaRUN 243 MUST2 data taking with the 3 alpha source
(positioned downstream of the target to shoot
T1 to T8). 
 scaler_2018-07-04_18h07m35s.pngvisu_2018-07-04_18h12m36s.pngvisu_2018-07-04_18h12m43s.pngvisu_2018-07-04_18h12m49s.png 
Fixede738 Wed Jul 4 18:43:29 2018 daisukeRUNGeneral3-alphaRUN 2443 alpha source at 'off-beam-axis' position
(right center view from  upstream). FAG ~40
Hz (mostly from T1-4)
 scaler_2018-07-04_18h38m39s.pngvisu_2018-07-04_18h39m26s.pngvisu_2018-07-04_18h42m32s.pngvisu_2018-07-04_18h43m12s.png 
Fixede738 Thu Jul 5 10:47:42 2018 louis, serge, pierre, daisukeRUNGeneral3-alphaRUn 2453 alpha source, off beam, low down position
. FAG 40 Hz. Likely due to beta rays (see
the attachment)
 scaler_2018-07-05_10h26m15s.pngvisu_2018-07-05_10h26m16s.pngvisu_2018-07-05_10h26m37s.pngvisu_2018-07-05_10h26m46s.pngvisu_2018-07-05_10h27m13s.png 
Fixede738 Thu Jul 5 10:49:40 2018 louis, serge, pierre, daisukeRUNGeneral3-alphaRUn 2453 alpha source, off beam, low down position
. FAG 40 Hz. Likely due to beta rays (see
the attachment)
 scaler_2018-07-05_10h26m15s.pngvisu_2018-07-05_10h26m16s.pngvisu_2018-07-05_10h26m37s.pngvisu_2018-07-05_10h26m46s.pngvisu_2018-07-05_10h27m13s.png 
Fixede738 Thu Jul 5 11:08:04 2018 pierre, louis daisukeRUNGeneralN/ARUN 249cats 1 pulser calibration. trigger cats1%1  scaler_2018-07-05_11h04m06s.pngvisu_2018-07-05_11h04m05s.png 
Fixede738 Thu Jul 5 14:12:18 2018 LouisMUST2General3-alphaRUN 253 calib alpha   scaler_2018-07-05_13h24m41s.pngvisu_2018-07-05_13h24m40s.pngvisu_2018-07-05_13h24m48s.pngvisu_2018-07-05_13h25m03s.pngvisu_2018-07-05_13h26m10s.png 
Pinnede738 Thu Jul 5 18:33:54 2018 gilles,fred,J-C,daisukeDAQHardwareN/ACaviar DAQ merged.The VME crate of CAVIAR was integraded to
the DAQ controlled by the GMT master of MUST2.
 rc_2018-07-05_18h32m56s.pngdas_2018-07-05_18h32m46s.pngdas_2018-07-05_18h32m18s.pngscaler_2018-07-05_18h32m06s.pngdesktop_2018-07-05_18h25m17s.png 
Fixede738 Thu Jul 5 19:22:18 2018 daisukeRUNGeneral3-alphaRUN 2563 alpha measurement after merging CAVIAR vme
to the DAQ
 scaler_2018-07-05_19h21m55s.pngvisu_2018-07-05_19h21m38s.pngvisu_2018-07-05_19h21m41s.pngvisu_2018-07-05_19h21m44s.png 
Pinnede738 Thu Jul 5 19:26:31 2018 daisukeMUST2HardwareN/AMUST2 BT/HT   screenshot_2018-07-05_19h23m54s.pngscreenshot_2018-07-05_19h24m07s.pngscreenshot_2018-07-05_19h24m17s.pngscreenshot_2018-07-05_19h24m24s.png 
Pinnede738 Fri Jul 6 09:40:04 2018 daisukeOTHERGeneralN/Akinematics plots   pt_kinema_30deg_1050um.pdfpt_kinema_30deg_2250um.pdfpt_kinema_30deg_750um_7be.pdf 
Fixede738 Fri Jul 6 17:14:50 2018 daisukeONLINESoftwareN/AOnline analysis code: e755.detectorThe file e755.detector was updated to include
the following parameters to ModularLeaf.
 screenshot_2018-07-06_17h14m09s.pngscreenshot_2018-07-06_17h14m25s.pngscreenshot_2018-07-06_17h16m43s.png 
Fixede738 Fri Jul 6 18:43:39 2018 NoriONLINESoftwareN/AOnline analysis code: e755.detectorThe modular label file has to be modified.

lpcgrit:/home/e748/analysis/nptool/Projects/e755
 Capture_du_2018-07-06_18-46-31.png 
Fixede738 Sat Jul 7 08:16:50 2018 daisukeMUST2HardwareN/AMUST2 BT/HT   screenshot_2018-07-07_08h15m39s.pngscreenshot_2018-07-07_08h15m54s.pngscreenshot_2018-07-07_08h16m04s.pngscreenshot_2018-07-07_08h16m13s.png 
Pinnede738 Sat Jul 7 08:26:27 2018 NoriBEAMGeneralN/ABeam production procedure   e738beamproduction.pdf 
Pinnede738 Sat Jul 7 08:29:53 2018 daisukeOTHERSocialN/AShift tablehttps://docs.google.com/spreadsheets/d/1jyfw2LT3RGfDY4bcN-9zepZoe2uE89b5CnF4tXzqzj8/edit#gid=716299251  2018-05-08_lise_must2_crypta_schedule_-_Shift_(July_7_to_15).pdf 
Fixede738 Sat Jul 7 14:33:45 2018 shumpei takeshi megumiTARGETHardwareCryo-LH2Liquid H2 statusWe start liquifying H2 at 14:09 and seems
to be filled at 14:29.
 liquefying0707.png 
Fixede738 Sat Jul 7 19:47:21 2018 daisukeMUST2HardwareN/AMUST2 BT/HT   screenshot_2018-07-07_19h46m29s.pngscreenshot_2018-07-07_19h46m19s.pngscreenshot_2018-07-07_19h46m00s.pngscreenshot_2018-07-07_19h45m41s.pngscreenshot_2018-07-07_19h45m13s.png 
Fixede738 Sun Jul 8 00:45:44 2018 takeshiRUNGeneralN/Arun summaryThe runsummary for this E738 experiment is
available at the same place as previous e755
experiment.
  
Fixede738 Sun Jul 8 01:24:16 2018 PierreRUNGeneralCryo-LH2Run 284start: 1h15 
stop: 3h11
 scaler_2018-07-08_01h25m36s.pngvisu_2018-07-08_01h26m12s.pngscaler_2018-07-08_03h10m42s.png 
Fixede738 Sun Jul 8 03:13:35 2018 PierreRUNGeneralCryo-LH2Run 285start: 03h12
stop: 05h10
 scaler_2018-07-08_03h13m06s.pngvisu_2018-07-08_03h19m50s.pngvisu_2018-07-08_03h19m13s.pngvisu_2018-07-08_03h19m27s.pngvisu_2018-07-08_03h20m19s.png 
Fixede738 Sun Jul 8 05:09:11 2018 PierreRUNGeneralCryo-LH2Run 286start: 05h11
stop: 07h00
 13x 
Fixede738 Sun Jul 8 07:00:46 2018 PierreRUNGeneralCryo-LH2Run 287start: 07h01
stop: 08h55
 8x 
Fixede738 Sun Jul 8 07:19:45 2018 ShumpeiOFFLINEGeneralCryo-LH2Analysis for 10C(p,d)9C and 10C(p,t)8CUsed data : Run 0284 &  0285 (about 4 hours
data)
CsI energy calibration parameters are from
 pd_fig1.pdfpd_fig2.pdfpt_fig1.pdfpt_fig2.pdf 
Ongoinge738 Sun Jul 8 08:53:44 2018 takeshiTARGETHardwareCryo-LH2temperature log of the target   cell0807.pngheatshield0807.pngtempepature0807.png 
Fixede738 Sun Jul 8 09:14:18 2018 takeshi, Didier, ValérieRUNGeneralCryo-LH2RUNS 288-290RUN 288
10C + liq H2
start 08h56 stop 10h56
 11x 
Fixede738 Sun Jul 8 14:23:43 2018 ValérieRUNGeneralCryo-LH2RUN Summary and MUST2, DAQ conditionsMUST2 
Physical thresholds ~ 400 keV see DASmust2T4Thres400keV.png
 12x 
Fixede738 Sun Jul 8 14:57:40 2018 Daisuke, Didier, ValérieRUNGeneralCryo-LH2RUN 291 overviewBeam 10C between 9.10**4 and 10**5/s 
dead time between 7 and 12%
 Run 291 start 13h47 stop 15h51
 14x 
Fixede738 Sun Jul 8 16:03:06 2018 Daisuke, Didier, ValérieRUNGeneralCryo-LH2RUN 292 -still 10CBeam 10C around 1. and 1.1 10**5/s 
dead time between 6 and 12%
 Run 292 start 15h51 stop 16h19 to change
 run292ScalersAt16h02.png 
Fixede738 Sun Jul 8 16:28:56 2018 Daisuke, Didier, ValérieRUNGeneralCryo-LH2RUN 293From Run 293  CATS1 HV = 640V I ~ 0.16 µA
; CATS2 660V I ~ 0.10 µA 
 Run 293 start 16h22; stop at 17h30   
 run293ScalersAt16h31.png 
Fixede738 Sun Jul 8 17:36:47 2018 Daisuke, Didier, ValérieRUNGeneralCryo-LH2RUN 294Run 294 start 17h31  stop 17h50 due to the
trip of CATS1.
748 Exxx 403M  8 juil. 17:50 run_0294.dat.08-07-18_17h31m47s
 run294ScalersAt17h35.png 
Fixede738 Sun Jul 8 17:59:12 2018 Daisuke, Takeshi, ValérieRUNGeneralCryo-LH2RUN 295Run 295 start 17h52 stop at 19h52
Beam stop occurred during 19h39 and 19h42
(alarm at the machine)
 26x 
Fixede738 Sun Jul 8 20:00:42 2018 Daisuke, Takeshi, ValérieRUNGeneralCryo-LH2RUN 296 +(on-line) eff of CATSRun 296 start 19h52 stop at 21h51

e748 Exxx 3,1G  8 juil. 21:51 run_0296.dat.08-07-18_19h52m46s
 11x 
Pinnede738 Sun Jul 8 21:05:50 2018 daisukeOTHERGeneralN/APhase space simulation: 10C(p,t)8C and 9C(p,d)8C   pt_sim_phase_plot_8c_t.pdfpt_sim_phase_plot_8c_d.pdf 
Fixede738 Sun Jul 8 21:50:58 2018 Daisuke, Takeshi, ValérieRUNGeneralCryo-LH2RUN 297Run 297 start 21h51 (8th July)  stop at 0h
(9th July)
e748 Exxx 3,0G  9 juil. 00:00 run_0297.dat.08-07-18_21h50m55s
 15x 
Fixede738 Mon Jul 9 00:03:24 2018 Benoit, Megumi, Takeshi, ValérieRUNGeneralCryo-LH2RUN 298Run 298 start 0h (9th July) stop at 02h02
e748 Exxx 2,9G  9 juil. 02:02 run_0298.dat.08-07-18_23h59m40s
Beam stop occurred from 01h06 to 01h09 (rotation
 run298_ScalersAt0h10.pngscaler_2018-07-09_02h01m34s.pngvisu_2018-07-09_02h01m41s.pngvisu_2018-07-09_02h01m44s.pngvisu_2018-07-09_02h01m52s.png 
Fixede738 Mon Jul 9 00:25:30 2018 NoriCATSGeneralN/ACATS tracking efficiencyWe noticed that the tracking efficiency was
only 70% while the timing efficiencies were
around 95%.
 8x 
Fixede738 Mon Jul 9 02:16:19 2018 Benoit, MegumiRUNGeneralCryo-LH2RUN 299Run 299 start 02h01 (9th July) stop at 04h03
e748 Exxx 2,9G  9 juil. 04:03 run_0299.dat.09-07-18_02h01m29s
 scaler_2018-07-09_02h20m00s.pngvisu_2018-07-09_04h04m37s.pngvisu_2018-07-09_04h04m53s.png 
Fixede738 Mon Jul 9 04:22:21 2018 Benoit, MegumiRUNGeneralCryo-LH2RUN 300Run 300 start 04h02 (9th July) stop at 06h04
e748 Exxx 2,8G  9 juil. 06:05 run_0300.dat.09-07-18_04h02m08s
 scaler_2018-07-09_04h17m03s.pngvisu_2018-07-09_05h55m54s.pngvisu_2018-07-09_05h56m57s.pngvisu_2018-07-09_06h00m30s.png 
Fixede738 Mon Jul 9 06:21:54 2018 Benoit, MegumiRUNGeneralCryo-LH2RUN 301Run 301 start 06h04 (9th July) stop at 7h51

 scaler_2018-07-09_06h13m13s.pngscaler_2018-07-09_07h51m37s.png 
Fixede738 Mon Jul 9 07:55:05 2018 PierreRUNGeneralCryo-LH2Run 302start: 07h53
stop: 08h19
 scaler_2018-07-09_07h57m43s.pngvisu_2018-07-09_07h59m57s.pngvisu_2018-07-09_08h00m04s.pngvisu_2018-07-09_08h04m11s.png 
Fixede738 Mon Jul 9 08:00:11 2018 takeshiTARGETHardwareCryo-LH2temperature log of the targetHere are the temperature log of the cyo H2
target during the 10C setting.
 heatshield0907.pngcell0907.png 
Fixede738 Mon Jul 9 08:55:13 2018 Pierre DaisukeRUNGeneralEmptyRun 303start: 08h27
stop: 08h50
  
Fixede738 Mon Jul 9 08:56:50 2018 Pierre DaisukeRUNGeneralCryo-emptyRun 304start: 08h54
stop: 10h00
 scaler_2018-07-09_08h57m21s.pngvisu_2018-07-09_09h03m24s.pngvisu_2018-07-09_09h03m30s.pngvisu_2018-07-09_09h04m05s.pngvisu_2018-07-09_09h04m21s.png 
Fixede738 Mon Jul 9 15:43:50 2018 Pierre ValérieRUNGeneralCryo-emptyRun 309start: 15h36
stop: 16h10
 scaler_2018-07-09_15h38m55s.pngscaler_2018-07-09_16h10m56s.png 
Fixede738 Mon Jul 9 17:18:55 2018 e748RUNGeneralCryo-emptyRUN 311Run 311
Beam/Detectors Tuning
 13x 
Fixede738 Mon Jul 9 18:03:48 2018 Valérie LouisRUNGeneralCryo-LH2RUN 312Run 312

First RUN of 9C on filled LH2 target
 25x 
Fixede738 Mon Jul 9 19:16:13 2018 Freddy LouisRUNGeneralCryo-LH2RUN 313 to 316From RUN 313 to RUN 316:

Tuning dor the Drift Chamber:
 Run317_discri_GECO.png 
Fixede738 Mon Jul 9 19:24:06 2018 Freddy Valérie LouisRUNGeneralCryo-LH2RUN 317Run 317

start: 19h08 stop: 19h30
 scaler_2018-07-09_19h26m05s.pngvisu_2018-07-09_19h26m13s.pngvisu_2018-07-09_19h28m52s.pngvisu_2018-07-09_19h29m09s.pngscaler_STOP_2018-07-09_19h32m05s.png 
Fixede738 Mon Jul 9 20:28:09 2018 Daisuke, Freddy, Louis, ValérieRUNGeneralCryo-LH2RUNS 318 to 320From RUN 318 to RUN 320:
initial problem was that the stop process
was not correctly taken into account on Narval.
 ProblemDAQcrateCATS1RCAt20h.pngProblemDAQcrateCATS1iolan.pngProblemMergerCATSbranch.png 
Fixede738 Mon Jul 9 20:33:00 2018 Daisuke Louis Valérie RUNGeneralCryo-LH2RUN 321- ok again !Run 321 start 20:09 
Total beam on CATS ~ 1.7 10**5 /s 
Beam of 9C, 7Be, 8B and 6Li
 20x 
Fixede738 Mon Jul 9 22:15:04 2018 DIdier Daisuke Nori Louis ValérieRUNGeneralCryo-LH2RUN 322, strip CATS2 X 19 offRun 322 start: 22h00 stop: 0h21
e748 Exxx 3,7G 10 juil. 00:21 run_0322.dat.09-07-18_21h58m34s
Beam: 9C
 7x 
Ongoinge738 Mon Jul 9 23:21:55 2018 takeshiTARGETHardwareCryo-LH2temperature log of the targetHere are the temperature logs of the liquid
H2 target up to now.
I would share the log during the regeneration.
 heatshield0907-2.pngcell0907-2.pngregenerate0907.png 
Fixede738 Mon Jul 9 23:25:10 2018 takeshiRUNGeneralCryo-LH2run summary for 10C beamWith 10C beam, we have obtained 
31.7 hours of liq. H2 target and
1.2  hours of empty target.
  
Fixede738 Mon Jul 9 23:29:18 2018 NoriCATSGeneralCryo-LH2CATS tracking efficiencyAnalyzed run 312 (9C physics run). CATS calibration
files with lowered threshold values
(/home/e748/analysis/nptool/Projects/e755/Calibration/cats/CATS1_X.e738.npt,
 run312_plasticq.pngrun312_effexample.png 
Fixede738 Tue Jul 10 00:46:26 2018 Shumpei, ValérieRUNGeneralCryo-LH2RUN 323 + Brho line settingsRun 323 start: 0h20  stop: 2h45
e748 Exxx 3,5G 10 juil. 02:45 run_0323.dat.10-07-18_00h20m04s
 Run323_scalersAt0h36.pngRun323_TOFcatsHF.pngRun323_ScalersEndRun.png 
Fixede738 Tue Jul 10 01:26:21 2018 ShumpeiOFFLINEGeneralCryo-LH2Analysis for 10C(p,d)9C and 10C(p,t)8C updatedUsed data run run0284-0302 (almost all for
10C beam)
CsI calibration parameters are the same as
 6x 
Fixede738 Tue Jul 10 02:47:10 2018 Shumpei, ValérieRUNGeneralCryo-LH2RUN 324Run 324 start: 2h45 stop 5h10
 e748 Exxx 3,8G 10 juil. 05:10 run_0324.dat.10-07-18_02h44m52s
Beam: 9C
 24x 
Fixede738 Tue Jul 10 05:12:31 2018 Shumpei, ValérieRUNGeneralCryo-LH2RUN 325Run 325 start: 5h10 stop 6h53 due to CATS1
-->tripping
 e748 Exxx 2,5G 10 juil. 06:53 run_0325.dat.10-07-18_05h09m53s
 Run325_scalersAt5h12.pngRun325_EDECsIMMpidZoomZ3Z4.pngRun325_scalersEndRunCats1TRIP.png 
Fixede738 Tue Jul 10 05:23:19 2018 ShumpeiOFFLINEGeneralCryo-LH2Analysis for 9C beam settingUsed data: run0312 (about 80 minutes data
taking).
Analysis for (p,d) channel and invariant
 be7li6pd.pngc9b8pd.pngmissing_invariant_mass.png 
Fixede738 Tue Jul 10 05:25:42 2018 ShumpeiBEAMGeneralCryo-LH2triton contamination in the 9C beam settingIn MUST2 dE-E plot, strong locus for triton
E ~ 40 MeV exists.
Kinematics for triton bellow 45 MeV matches
 triton_contaminant0.pngtriton_contaminant1.pngtriton_contaminant2.png 
Fixede738 Tue Jul 10 07:16:16 2018 Shumpei, ValérieRUNGeneralCryo-LH2RUN 326Run 326 start: 6h59
  
Beam: 9C
 9x 
Fixede738 Tue Jul 10 12:51:37 2018 marlène, olivier, J-C, daisukeRUNGeneralCryo-LH2RUN 3309C + liquid H. Data taking after replacing
CATS1. 
12h03 started
 28x 
Fixede738 Tue Jul 10 13:59:09 2018 jacques, daisuke, freddyRUNGeneralCryo-LH2RUN 3319C + liquid H 
13h56 start. 12C 1.7 electric micro A. 
16h51 stop
 6x 
Fixede738 Tue Jul 10 14:22:22 2018 NoriBEAMGeneralCryo-LH2Beam image on targetReconstructed beam images from run0330.  run330_imagemust.pngrun330_imagepla.pngrun330_imagecats.png 
Fixede738 Tue Jul 10 15:17:21 2018 NoriCATSGeneralCryo-LH2CATS tracking efficiencyTracking efficiencies deduced from run 330
(after replacing CATS1).
 run330_catstracking.png 
Fixede738 Tue Jul 10 16:28:53 2018 daisukeOTHERSocialN/A6Li(p,d)5LiThe peak near 16 MeV was observed in a measurement
in the 1960s. The resonance is close to the
3He-d threshold. The narrow width indicates
 be7li6pd.png2018-07-10_5li_via_pd_spectrum.png2018-07-10_5li_via_pd_level_scheme.png5li_via_pd_msu_kull_pr_163_1066_1967.pdf 
Fixede738 Tue Jul 10 17:02:02 2018 jacques, daisuke, freddy, louisRUNGeneralCryo-LH2RUN 33212C primary: 1.7 uA
start: 16h52
stop: 17h52
 run_332_ScalerStartRun.pngrun_332_ScalerEndRun.pngrun_332_TargetImpact.png 
Fixede738 Tue Jul 10 18:23:01 2018 daisukeOTHERSocialN/Amirror reaction of 7Be(p,d)6Be   2018-07-10_6he_via_nd_spectrum.png6he_via_nd_ucdavis_pr_16_31_1977.pdf 
Fixede738 Tue Jul 10 21:26:48 2018 didier, daisukeRUNGeneralCryo-LH2RUN 3349C + liquid H. Probably nobody cares this
post right now, but the beam is back. 
 17x 
Fixede738 Tue Jul 10 23:14:31 2018 DDRUNGeneralCryo-LH2RUN 3359C + Liquid H run. 
23h11 start. 1.7 electric micro A. Wien filter
1 kV.
 scaler_2018-07-10_23h16m13s.png 
Fixede738 Wed Jul 11 02:52:54 2018 MarleneRUNGeneralCryo-LH2RUN336run336 start 1:07 / stop 2:48
fluctuations of beam intensity
 6x 
Fixede738 Wed Jul 11 02:55:44 2018 e748RUNGeneralCryo-LH2RUN 337RUN 337 start 2:49 /stop 4:09

Beam  intensity CATS 200k pps
 run_337_HT_BT_cats.pngrun_337_cats_pos.pngrun_337_scaler.png 
Ongoinge738 Wed Jul 11 03:57:50 2018 e748OFFLINEGeneralCryo-LH2Conversion problemConversion of runs 335 gives the following
error (attached)
Size of raw file =3.5 Go 
 conversion_problem.png 
Fixede738 Wed Jul 11 04:16:04 2018 NoriBEAMGeneralCryo-LH2Caviar statusCaviar is now operating at 700 V. TOF-CaviarX
position correlations are clearly seen (Attachment
1).
 run335_cavxcats1hf.pngrun335_cavmult.png 
Fixede738 Wed Jul 11 05:11:19 2018 e748RUNGeneralCryo-LH2RUN 338RUN 338
9C beam 200 k pps
start 4:09 /stop 5:10
 run_338_scaler.png 
Fixede738 Wed Jul 11 06:11:50 2018 e748RUNGeneralCryo-LH2RUN 339RUN 339
9C 210 k pps
start 5:11 / stop 6:08 (beam stopped)
 9x 
Fixede738 Wed Jul 11 06:17:19 2018 e748RUNGeneralCryo-LH2RUn 340RUn 340
start 6:10 / stop 7:43
  
Pendinge738 Wed Jul 11 06:29:00 2018 e748CATSGeneralCryo-LH2CATS2 statusCATS2 current fluctuates a lot and has increased
up to 0.26uA in run 340
In run 340 I observe that a second wire is
 run_340_cats.pngrun_339_cats.png 
Fixede738 Wed Jul 11 07:49:12 2018 Marlène, Franck, Jacques, DaisukeRUNGeneralCryo-LH2RUN 341RUn 341
start 7:44 /stop 9h29
 run_341_scaler.pngrun_341_cats.pngrun_341-ht_bt_cats.png 
Fixede738 Wed Jul 11 08:18:00 2018 NoriCATSGeneralCryo-LH2CATS efficiencyThe tracking efficiency is diminishing probably
due to damaged CATS2.
 run341_catstracking.png 
Ongoinge738 Wed Jul 11 08:32:42 2018 takeshiTARGETHardwareCryo-LH2temperature log of the target   heatshield1107.pngcell1107.png 
Fixede738 Wed Jul 11 09:32:37 2018 Franck, Jacques, DaisukeRUNGeneralCryo-LH2Run 342Wed 11 July 2018

Start @ 9:19
 10x 
Fixede738 Wed Jul 11 12:02:14 2018 NPTeamOFFLINEGeneralCryo-LH2Lauching parrallel analysis of multiple runWe made  a parralel conversion script in the
conversion folder.
You can run up to 5 run in parallel using
  
Fixede738 Wed Jul 11 12:04:29 2018 Franck, Jacques + CATS TeamCATSGeneralN/ARun 343 - CATS2 pulser calibrationStart 11:53
Stop 11:59
 visu_2018-07-11_12h07m59s.png 
Fixede738 Wed Jul 11 13:24:34 2018 Daisuke,CATS team: Séb, Thom ; Franck, Pierre, ValérieRUNGeneralCryo-LH2Runs 344-345Wed 11 July 2018
run 344 start 12h43 ; stop 13h03
e748 Exxx 900M 11 juil. 13:03 /data/e748X/e755/acquisition/run/run_0344.dat.11-07-18_12h43m33s
 12x 
Fixede738 Wed Jul 11 14:51:13 2018 Daisuke, Franck, Shumpei, ValérieRUNGeneralCryo-LH2Run 346Wed 11 July 2018
run 346 start 14h50; stop 15h14
748 Exxx 777M 11 juil. 15:14 /data/e748X/e755/acquisition/run/run_0346.dat.11-07-18_14h48m47s
 run346_ScalersAt14h57.png 
Ongoinge738 Wed Jul 11 15:41:33 2018 Franck, Jacques, DaisukeRUNGeneralCryo-LH2Run 347Run 347 Start 15:28 stop 17h17
e748 Exxx 3,6G 11 juil. 17:17 /data/e748X/e755/acquisition/run/run_0347.dat.11-07-18_15h28m32s
CATS thresholds lowered from 40 to 24 mV
 run347_HT_CATS.pngscaler_2018-07-11_15h36m47s.pngvisu_2018-07-11_15h37m23s.pngvisu_2018-07-11_15h37m37s.png 
Fixede738 Wed Jul 11 16:51:50 2018 NoriCATSGeneralCryo-LH2CATS efficiencyThe tracking efficiency recovered after replacing
CATS2.
 run347_catseff.png 
Ongoinge738 Wed Jul 11 17:25:11 2018 Louis, Daisuke, Nori, ValérieRUNGeneralCryo-LH2Run 348Run 348 Start 17h17

Beam: 9C at 2.3*10⁵pps in CATS1, 2uA
 run348_scalersAt17h26.pngvisu_2018-07-11_17h39m20s.pngvisu_2018-07-11_17h39m25s.pngscaler_STOP_2018-07-11_18h09m53s.png 
Fixede738 Wed Jul 11 18:31:23 2018 Jacques, Didier, LouisRUNGeneralCryo-LH2Run 349Run 349 

Start 18h22  Stop 18h47
 6x 
Fixede738 Wed Jul 11 19:08:40 2018 Didier, LouisRUNGeneralCryo-LH2RUN 350Run 350

Start 19h00 Stop 19h11
 10x 
Fixede738 Wed Jul 11 19:17:20 2018 Didier, LouisRUNGeneralCryo-LH2RUN 351Run 351

Start 19h14 Stop 19h53
 scaler_2018-07-11_19h20m33s.pngvisu_2018-07-11_19h21m57s.pngvisu_2018-07-11_19h22m00s.pngvisu_2018-07-11_19h22m58s.pngscaler_STOP_2018-07-11_19h57m00s.png 
Fixede738 Wed Jul 11 20:00:01 2018 Didier, Valérie, LouisRUNGeneralCryo-LH2RUN 352Run 352

Start 19h57 Stop 21h51
 7x 
Fixede738 Wed Jul 11 21:55:23 2018 Daisuke, Didier, Valérie, LouisRUNGeneralCryo-LH2RUN 353Run 353

Start 21h50 Stop 
 6x 
Fixede738 Wed Jul 11 23:36:27 2018 Didier, Louis, Freddy, ShumpeiRUNGeneralCryo-LH2RUN 354Run 354

Start 23h33 Stop 1h37
 scaler_2018-07-11_23h42m26s.pngvisu_2018-07-11_23h40m27s.pngvisu_2018-07-11_23h42m12s.pngvisu_2018-07-11_23h42m15s.pngvisu_2018-07-11_23h43m02s.png 
Fixede738 Thu Jul 12 02:12:20 2018 Shumpei, FreddyRUNGeneralCryo-LH2RUN 355[U][B]RUN 355[/B][/U]

start: 2h04
 7x 
Fixede738 Thu Jul 12 04:40:32 2018 ShumpeiCATSGeneralCryo-LH2Analysis with only one of the CATSsAnalyzed data : run 0331
Used reaction : 7Be(p,d)6Be (ground state
and 2+ state 1.6 MeV are mainly populated).
 cats1.pngcats2.pngcats3.pngcats4.png 
Fixede738 Thu Jul 12 05:03:49 2018 Shumpei, FreddyRUNGeneralCryo-LH2RUN 356[U][B]RUN 356[/B][/U]

start: 4h24
 7x 
Fixede738 Thu Jul 12 07:05:53 2018 Shumpei, Freddy, Megumi, ValérieRUNGeneralCryo-LH2RUN 357[U][B]RUN 357[/B][/U]

start: 6h39
 7x 
Fixede738 Thu Jul 12 07:43:25 2018 ShumpeiOFFLINEGeneralCryo-LH2Analysis of 9C beam settingUsed data : run0312-0355

For (p,t) reaction, the value of the mass
 c9b8pd2.pngbe7li6pt.pngc9b8pt.png 
Fixede738 Thu Jul 12 08:16:21 2018 Daisuke, Megumi, ValérieRUNGeneralCryo-LH2RUN 358[U][B]RUN 358[/B][/U]
start: 8h14 stop: 9h48
 20x 
Fixede738 Thu Jul 12 09:52:07 2018 Daisuke, Megumi, ValérieRUNGeneralCryo-LH2RUN 359RUN 359

start: 9h48 stop: 11h25
 6x 
Fixede738 Thu Jul 12 11:28:46 2018 Daisuke, Megumi, ValérieRUNGeneralCryo-LH2RUN 360RUN 360
e748 Exxx 3,4G 12 juil. 13:08 /data/e748X/e755/acquisition/run/run_0360.dat.12-07-18_11h24m44s
start: 11h25 stop: 13h08
 10x 
Fixede738 Thu Jul 12 13:09:14 2018 Daisuke, Megumi, ValérieRUNGeneralCryo-LH2RUN 361RUN 361
e748 Exxx 3,5G 12 juil. 15:01 /data/e748X/e755/acquisition/run/run_0361.dat.12-07-18_13h07m40s
start: 13h08 stop: 14h59
 6x 
Ongoinge738 Thu Jul 12 13:49:38 2018 takeshiTARGETHardwareCryo-LH2temperature log of the target   heatshield1207.pngcell1207.png 
Fixede738 Thu Jul 12 15:07:43 2018 Daisuke, Louis, Marlène, Megumi, ValérieRUNGeneralCryo-LH2RUN 362-TRIGGER TESTSRUN 362 WITH CHANGE OF TRIGGERS FOR TESTS
see below
e748 Exxx 1,7G 12 juil. 16:09 /data/e748X/e755/acquisition/run/run_0362.dat.12-07-18_15h08m50s
 11x 
Fixede738 Thu Jul 12 15:27:56 2018 daiuskeMUST2GeneralN/AMUST2*CATS coincidence triggerMessage from Adrien.  2018-07-09_message_adrien.pdfScreenshot_2018-07-09_()_-_document.png 
Fixede738 Thu Jul 12 16:22:43 2018 Frank, LouisRUNGeneralCryo-LH2RUN 363RUN 363
 
start: 16h09  stop: 18h10
 10x 
Fixede738 Thu Jul 12 18:12:33 2018 Frank, LouisRUNGeneralCryo-LH2RUN 364RUN 364
 
start: 18h11  stop: 20h45
 12x 
Pinnede738 Thu Jul 12 18:47:32 2018 NoriBEAMGeneralN/ASummary of Brho2 values


Beam
Brho2
  
Fixede738 Thu Jul 12 20:47:32 2018 Frank, LouisRUNGeneralCryo-LH2RUN 365RUN 365
 
start: 20h46  stop: 22h43
 10x 
Fixede738 Thu Jul 12 22:45:50 2018 Frank, LouisRUNGeneralCryo-LH2RUN 366RUN 366
 
start: 22h44  stop: 00h42
 15x 
Fixede738 Fri Jul 13 00:45:45 2018 PierreRUNGeneralCryo-LH2Run 367start: 00h44   
Fixede738 Fri Jul 13 00:47:23 2018 PierreRUNGeneralCryo-LH2Run 367RUN 367

start: 00h44
 scaler_2018-07-13_00h46m09s.pngvisu_2018-07-13_00h51m01s.pngvisu_2018-07-13_00h50m40s.png 
Fixede738 Fri Jul 13 02:45:44 2018 PierreRUNGeneralCryo-LH2Run 368RUN 368

start: 02h44
 7x 
Fixede738 Fri Jul 13 04:09:10 2018 takeshiTARGETHardwareCryo-LH2temperature of the targetWe found the target temperature suddenly increased
at around 3:40 13th.
The online beam image is still indicating
 cell1307.png 
Fixede738 Fri Jul 13 04:48:19 2018 PierreRUNGeneralCryo-LH2Run 369RUN 369

start: 04h47
 7x 
Fixede738 Fri Jul 13 06:02:31 2018 takeshi, pierreTARGETGeneralCryo-LH2target is half filledWe found the liquid hydrogen target is filled
only half during the run 368, around 4:10.
We will continue the measurement for now.
 target1307.pdfcell1307_2.png 
Fixede738 Fri Jul 13 09:03:11 2018 takeshiRUNGeneralCryo-LH2run summary 9C + H2Here is the run summary of the setting 9C
beam + liquid H2 target.
Up to now, we have obtained 78 hours data,
  
Fixede738 Fri Jul 13 09:35:26 2018 Olivier, Serge, Jacques, J-C, thomas, shumpei, daisukeRUNGeneralCryo-emptyRUN 370Empty target + 9C. 
9h29 The beam is back. start 9h28 stop 11h21
e748 Exxx 2,1G 13 juil. 11:21 /data/e748X/e755/acquisition/run/run_0370.dat.13-07-18_09h28m21s
 visu_2018-07-13_09h36m18s.pngscaler_2018-07-13_09h36m16s.png 
Fixede738 Fri Jul 13 13:41:48 2018 Daisuke, Jacques, Megumi, Olivier, Otsu, Takeshi, ValérieRUNGeneralCryo-emptyRUNS 371-372Empty target + 9C. 
Run 371 start 11h20 stop 13h20
 e748 Exxx 2,1G 13 juil. 13:20 /data/e748X/e755/acquisition/run/run_0371.dat.13-07-18_11h20m03s
 32x 
Fixede738 Fri Jul 13 16:41:49 2018 Antoine, MaurycyOFFLINESocialCryo-LH2Multiparticles from (p,t) up to 5 Light Charged Particles in Must !   15x 
Fixede738 Fri Jul 13 18:07:04 2018 J-C, cristelle, nori, olivier, daisukeRUNGeneralCryo-emptyRUN 37312C 5+ beam. CATS2 trigger. 
17h29 start 
17h41 stop
 scaler_2018-07-13_17h43m06s.pngvisu_2018-07-13_17h43m05s.png 
Fixede738 Fri Jul 13 18:08:31 2018 J-C, cristelle, nori, olivier, daisukeRUNGeneralCryo-emptyRUN 37412C5+ beam. Trigger T1-8 + CATS1%30000 + CATS2%30000
+ Plastic%10000. 
Target empty (under cooling). 
 6x 
Fixede738 Fri Jul 13 21:15:53 2018 J-C, nori, shumpei, takeshi, hide, didier, serge, daisukeRUNGeneralCryo-emptyRUN 375: 12C pilot run12C beam + empty target after measurement
of 9C momentum distribution. 
trigger t1-8 + cats 1 2 %30000 + plastic
 8x 
Fixede738 Fri Jul 13 22:00:59 2018 didier, nori, takeshi, shumpei, hide, daisukeRUNGeneralCryo-LH2RUN 376: 12C first run12C + liquid H. 
start 21h44
stop 22h43
 22x 
Fixede738 Fri Jul 13 23:10:55 2018 e748RUNGeneralCryo-LH2run 377count rate in Cats : 100 kHz
Start time: 22:44
Stop: 00:42
 13x 
Fixede738 Fri Jul 13 23:22:26 2018 ShumpeiOFFLINEGeneralCryo-LH212C beam analysisUsed data run0376 (about 1 hour)

Only target position and recoil pid (deuteron
 12c.pdf 
Fixede738 Sat Jul 14 00:32:55 2018 ShumpeiOFFLINEGeneralCryo-emptyempty target analysis for 9C beam settingH2 filled: run0312-0355
empty: run0370-0372
 pd.pdfpd_60deg.pdfpt.pdfpt_60deg.pdf 
Fixede738 Sat Jul 14 01:24:51 2018 e748RUNGeneralCryo-LH2run 378count rate in Cats : 100 kHz
Start time: 00:42
 scaler_2018-07-14_00h46m48s.pngvisu_2018-07-14_00h48m31s.pngvisu_2018-07-14_00h52m28s.png 
Fixede738 Sat Jul 14 02:46:34 2018 e748RUNGeneralCryo-LH2run 379count rate in Cats : 100 kHz
Start time: 02:41
 scaler_2018-07-14_02h44m19s.pngvisu_2018-07-14_02h45m36s.png 
Fixede738 Sat Jul 14 05:00:37 2018 e748RUNGeneralCryo-LH2run 380count rate in Cats : 100 kHz
Start time: 04:38
 scaler_2018-07-14_04h41m20s.pngvisu_2018-07-14_04h49m58s.png 
Fixede738 Sat Jul 14 06:49:03 2018 e748RUNGeneralCryo-LH2run 380count rate in Cats : 100 kHz
Start time: 06:39
 scaler_2018-07-14_06h43m26s.pngscreenshot_2018-07-14_06h42m28s.pngvisu_2018-07-14_06h44m35s.pngvisu_2018-07-14_06h48m04s.pngvisu_2018-07-14_06h50m58s.png 
Fixede738 Sat Jul 14 09:57:36 2018 takeshiTARGETHardwareCryo-LH2target temperature during nightI would share the temperature log of the target
as  it was liquefied 21:00 13th.
 liquiefy1307.png 
Fixede738 Sat Jul 14 10:14:00 2018 marlène, jacques, daisukeRUNGeneralCryo-LH2RUN 38212C + liquid H. Before the run, the beam stopped
for 2 hours due to the gate valve of E9 shut
out the beam. This is because the manually
 9x 
Fixede738 Sat Jul 14 11:37:58 2018 marlène, daisukeRUNGeneralCryo-LH2RUN 38312C + liquid H. 
11h33 start. 
12h00 Asked PCP to increase the intensity.
 25x 
Fixede738 Sat Jul 14 14:01:55 2018 Daisuke, Jacques, Shumpei, ValérieRUNGeneralCryo-LH2RUNS 384-38512C + liquid H. 

RUN 384 start 13h58 stop 15h54
 16x 
Fixede738 Sat Jul 14 18:18:33 2018 Daisuke, Jean-Charles, Megumi, Nori, Shumpei, ValérieRUNGeneralCryo-LH2RUN 386 TUNING 9C9C + liquid H. 
SLITS FH31 +/- 2mm then +/-5 mm at the beginning
of the run 
 8x 
Fixede738 Sat Jul 14 18:36:38 2018 Daisuke, Jean-Charles, Megumi, Nori, Shumpei, ValérieRUNGeneralCryo-LH2RUN 387 TUNING 9C9C + liquid H. 
SLITS FH31 +/- 5 mm F43 opened
I ~ 1.6µA  TRIGGERS CATS2, PLASTIC div
  
Fixede738 Sat Jul 14 18:37:29 2018 Daisuke, Jean-Charles, Megumi, Nori, Shumpei, ValérieRUNGeneralCryo-LH2RUN 388 TUNING 9C9C + liquid H. 
SLITS FH31 +/- 5 mm F43 opened
I ~ 1.6µA  normal TRIGGERS  MUST2 CATS div
 Run388_ScalersCaviarFH31_5mm_I1.6µA.pngRun388_Scalers.pngRun388_TOF.png 
Fixede738 Sat Jul 14 18:54:01 2018 Daisuke, Jean-Charles, Megumi, Nori, Shumpei, ValérieRUNGeneralCryo-LH2RUN 389 TUNING 9C9C + liquid H. 
SLITS FH31 +/- 30 mm F43 opened
I ~ 1.6µA  normal TRIGGERS  MUST2 CATS div
 7x 
Fixede738 Sat Jul 14 19:01:22 2018 Daisuke, Jean-Charles, Megumi, Nori, Shumpei, ValérieRUNGeneralCryo-LH2RUN 390 TUNING 9C9C + liquid H. 
SLITS FH31 +/- 20 mm F43 V 10 mm; FH43 10
mm  I~1.9 µA
 7x 
Fixede738 Sat Jul 14 19:38:25 2018 Daisuke, Jean-Charles, Shumpei, ValérieRUNGeneralCryo-LH2RUN 391 TUNING 9C9C + liquid H. 
SLITS FH31 +/- 20 mm F43 V 10 mm; FH43 10
mm  I~1.9 µA
 Run391_TOF.pngRun391_scalers.png 
Fixede738 Sat Jul 14 20:16:50 2018 Daisuke, Shumpei, ValérieRUNGeneralCryo-LH2RUN 392 9C9C + liquid H. 
RUN 392 start 19h46 stop 21h47
e748 Exxx 3,5G 14 juil. 21:49 /data/e748X/e755/acquisition/run/run_0392.dat.14-07-18_19h45m58s
 21x 
Fixede738 Sat Jul 14 21:53:48 2018 Daisuke, Shumpei, ValérieRUNGeneralCryo-LH2RUN 393 9C9C + liquid H. 
RUN 393 start 21h48  stop  23h49
e748 Exxx 4,0G 14 juil. 23:49 /data/e748X/e755/acquisition/run/run_0393.dat.14-07-18_21h48m17s
 9x 
Fixede738 Sun Jul 15 00:48:03 2018 takeshiRUNGeneralN/Arun summaryHere are the run summary of E738 experiment
up to 1:00 am 15th.
31.7 hours for 10C + H2
  
Fixede738 Sun Jul 15 00:49:52 2018 Megumi, Shumpei, Takeshi, Benoît, ValérieRUNGeneralCryo-emptyRUN 394 9C9C + Empty target  15th JULY
RUN 394 start 0h43  stop   03h20
e748 Exxx 2,9G 15 juil.  2018 /data/e748X/e755/acquisition/run/run_0394.dat.15-07-18_00h43m25s
 Run394_scalersAt0h45.pngRun394_CatsStrRaw.pngRun394_BeamTracks.pngRun394_TOF.pngvisu_2018-07-15_03h20m53s.png 
Fixede738 Sun Jul 15 01:26:49 2018 takeshiTARGETHardwareCryo-LH2temperature log of the targetHere are the temperature log of the target
before it was evaporated.
 heatshiled1507.pngcell1507.png 
Fixede738 Sun Jul 15 01:35:09 2018 takeshi, ValérieONLINEGeneralN/Aproblem of the beam image on vigruThe beam image at the target is not clear
on online monitor.
We have confirmed that the target is correctly
 run394.pngtarget_run393.pdftarget_run394.pdftargetY_run393.pdftargetY_run394.pdf 
Fixede738 Sun Jul 15 03:27:25 2018 Megumi, BenoîtRUNGeneralCryo-emptyRUN 395 9C9C + Empty target  15th JULY
RUN 395 start 03h20  stop  05h59
e748 Exxx 2,8G 15 juil. 06:01 /data/e748X/e755/acquisition/run/run_0395.dat.15-07-18_03h20m21s
 scaler_2018-07-15_03h30m58s.pngvisu_2018-07-15_06h00m44s.pngvisu_2018-07-15_06h00m56s.pngvisu_2018-07-15_06h01m02s.pngvisu_2018-07-15_06h01m11s.png 
Fixede738 Sun Jul 15 06:17:04 2018 Megumi, BenoîtRUNGeneralCryo-emptyRUN 396 9C9C + Empty target  15th JULY
RUN 396 start 06h00  stop  7h59
 8x 
Fixede738 Sun Jul 15 08:04:12 2018 daisukeRUNGeneralCryo-emptyRUN 3979C + empty. 

08h00 start. 12C 1.8 micro A. 
 32x 
Fixede738 Sun Jul 15 10:04:13 2018 daisukeRUNGeneralCryo-emptyRUN 3989Cc + empty 

10h01 start 
 scaler_2018-07-15_10h05m09s.png 
Fixede738 Sun Jul 15 13:57:59 2018 daisukeRUNGeneralCryo-emptyRUN 4159C + empty run. 

From 10 to 13h45, we had several issues with
 34x 
Fixede738 Sun Jul 15 19:05:20 2018 takeshiRUNGeneralN/Arun summary of E738Runsmmary of the E738 experiment.
31.7 hours for 10C + H2
1.2 hours for 10C + empty
  
Fixede738 Mon Jul 16 11:22:26 2018 thomas, daisukeRUNGeneralN/ARUN 416CATS1 pulser calibration. trigger CATS1.
Attenuation 1/1, 1/2, 1/4, 1/5, 1/10 and
pedestal.v
 visu_2018-07-16_11h14m17s.png 
Fixede738 Mon Jul 16 11:23:55 2018 thomas, daisukeRUNGeneralN/ARUN 417CATS2 pulser calibration. trigger CATS2. 
Attenuation 1/1, 1/2, 1/4, 1/5, 1/10 and
pedestal.
 visu_2018-07-16_11h25m58s.pngscaler_2018-07-16_11h25m58s.png 
Fixede738 Mon Jul 16 15:13:21 2018 daisukeRUNGeneral3-alphaRUN 419MUST2 3 alpha source at center 
13h22 start 
15h05 stop. 
 visu_2018-07-16_14h58m37s.pngvisu_2018-07-16_14h58m45s.pngvisu_2018-07-16_14h58m55s.png 
Fixede738 Mon Jul 16 16:13:00 2018 daisukeRUNGeneral3-alphaRUN 420MUST2 3 alpha shooting a bit left

15h10-16h08 
 scaler_2018-07-16_16h10m10s.pngvisu_2018-07-16_16h09m59s.pngvisu_2018-07-16_16h10m02s.pngvisu_2018-07-16_16h10m05s.png 
Fixede738 Mon Jul 16 16:44:43 2018 daisukeRUNGeneral3-alphaRUN 422MUST2 + 3 alpha shooting left   visu_2018-07-16_16h32m59s.png 
Fixede738 Mon Jul 16 17:25:22 2018 daisukeRUNGeneral3-alphaRUN 422MUST2 + 3 alpha shooting right.   visu_2018-07-16_17h22m02s.pngvisu_2018-07-16_17h05m26s.pngscaler_2018-07-16_17h22m04s.png 
Fixede738 Mon Jul 16 18:46:34 2018 daisukeRUNGeneral3-alphaRUN 423MUST2 + 3 alpha shooting center.  scaler_2018-07-16_18h40m56s.pngvisu_2018-07-16_18h03m27s.pngvisu_2018-07-16_18h39m46s.pngvisu_2018-07-16_18h39m49s.png 
Fixed  Tue Feb 26 12:16:10 2019 iulian freddy francoisMUST2SoftwareN/Arun 41 Calimero run MUVI 1 sector1X[U][B]Run 41[/B][/U]

Calimero WORKS!!
  

ELOG Home Page