MUST2 SAMURAI_2018 S3-LEB-LPC SUPERNEMO MUGAST EXPAND SCALP GALATRON HiCARI VELO
  MUST2  Not logged in MUST2
Message ID: 445     Entry time: Thu May 24 05:44:56 2018
Author: iulian 
ExpNbr: e744 
Type: RUN 
Category: General 
Status: Pinned 
Target-Source: N/A 
Subject: Important::analysis of the run 443/ things to do 
No stop on t5, but TAC MM_CATS ok.. So two possibilities:
1) Stop Cable not connected on Muvi2
2) Problem with the validation of CATS by FAG

HF validated by FAG was not working

Dead time scaler was not working

We need to erase all the scalers at the beginning of the run, including the part from Muvi (ORD, DESC, etc). Atm we erase only the U2M not the other part

The big problem is the transmission between LISE silicon detector D6 and T5 must with CATS on or off. 3000 counts/s on silicon d6 or CATS -> maximum 300 on T5. Is there a way to miss counting in Must2?


TAC_CATS_HF ok
TAC_MM_CATS ok
T_MM_EXLor present with prompt from cosmics?!
T_MM_HF ok
T_MM_HFv not present
Attachment 1: t_mm_PPT_run443.pdf  13 kB  Uploaded Thu May 24 06:56:13 2018  | Hide | Hide all
t_mm_PPT_run443.pdf
Attachment 2: t_mmhf_run443.pdf  14 kB  Uploaded Thu May 24 06:56:21 2018  | Hide | Hide all
t_mmhf_run443.pdf
Attachment 3: t_mm_exlor_run443.pdf  14 kB  Uploaded Thu May 24 06:56:29 2018  | Hide | Hide all
t_mm_exlor_run443.pdf
Attachment 4: correlation_run443.pdf  14 kB  Uploaded Thu May 24 06:56:38 2018  | Hide | Hide all
correlation_run443.pdf

ELOG Home Page