MUST2 SAMURAI_2018 S3-LEB-LPC SUPERNEMO MUGAST EXPAND SCALP GALATRON HiCARI VELO
  MUST2  Not logged in MUST2
Message ID: 365     Entry time: Fri May 4 11:20:50 2018
Author: Valerian 
ExpNbr: e744 
Type: MUST2 
Category: General 
Status: Fixed 
Target-Source: 3-alpha 
Subject: T5 : 0° telescope  
After increasing the voltage on T5 DSSD from 80V to 100V the resolution increased from ~40keV sigma to 20keV sigma.

StripX vs StripY bidim and the 3 alpha peaks are joined to this entry.
Attachment 1: StripX_vs_StripY_T5_100V.pdf  79 kB  Uploaded Fri May 4 12:29:37 2018  | Hide | Hide all
StripX_vs_StripY_T5_100V.pdf
Attachment 2: Si_E_T5_100V.pdf  15 kB  Uploaded Fri May 4 12:29:44 2018  | Hide | Hide all
Si_E_T5_100V.pdf

ELOG Home Page