MUST2 SAMURAI_2018 S3-LEB-LPC SUPERNEMO MUGAST EXPAND SCALP GALATRON HiCARI VELO
  MUST2  Not logged in MUST2
Message ID: 902     Entry time: Thu Jul 12 04:40:32 2018
Author: Shumpei 
ExpNbr: e738 
Type: CATS 
Category: General 
Status: Fixed 
Target-Source: Cryo-LH2 
Subject: Analysis with only one of the CATSs 
Analyzed data : run 0331
Used reaction : 7Be(p,d)6Be (ground state and 2+ state 1.6 MeV are mainly populated).

Method;
X and Y position on each CATS is compared with position on the target deduced by both CATSs. Assuming the beam profile is not changing, the correlation of these positions can by fit by linear function.
The position on the target can reconstructed with only one of the CATSs, with worse resolution. Beam angular spread seems rather small than angular deduced by only 1 CATS.
In terms of the excitation energy resolution,
sigma of the Subtraction of (both CATS Ex)-(only CATS1 Ex) ~ 0.53 MeV
sigma of the Subtraction of (both CATS Ex)-(only CATS2 Ex) ~ 0.32 MeV
Original estimated Ex resolution is ~ 1 MeV, so not so bad.
In terms of the background from the frame of the target cell, roughly can be removed by target position. Near the center part gate should be more secure.
Attachment 1: cats1.png  269 kB  | Hide | Hide all
cats1.png
Attachment 2: cats2.png  227 kB  Uploaded Thu Jul 12 05:40:46 2018  | Hide | Hide all
cats2.png
Attachment 3: cats3.png  367 kB  Uploaded Thu Jul 12 05:41:01 2018  | Hide | Hide all
cats3.png
Attachment 4: cats4.png  214 kB  Uploaded Thu Jul 12 05:41:09 2018  | Hide | Hide all
cats4.png