Message ID: 638     Entry time: Sun Jun 24 19:51:09 2018
Author: Faïrouz, daisuke 
ExpNbr: e755 
Type: MUST2 
Category: Software 
Status: Fixed 
Target-Source: N/A 
Subject: MUST2 threshold from 1 MeV to 0.4 MeV 
It was found that the energy threshold of the strips was as high as 1 MeV. This threshold setting was chosen for the calibration and was never set back to the nominal threshold (0.4 MeV) for experiments.

Lowered to 0.4 MeV.

Note that due to the low energy threshold, triton of energies higher than 70 MeV were partially rejected during the measurement of 40Ca(p,t).
Attachment 1: das_2018-06-24_19h33m36s.png  59 kB  Uploaded Sun Jun 24 20:52:52 2018  | Hide | Hide all
das_2018-06-24_19h33m36s.png
Attachment 2: das_2018-06-24_19h49m33s.png  60 kB  Uploaded Sun Jun 24 20:52:56 2018  | Hide | Hide all
das_2018-06-24_19h49m33s.png