Message ID: 638
Entry time: Sun Jun 24 19:51:09 2018
|
Author: |
Faïrouz, daisuke |
ExpNbr: |
e755 |
Type: |
MUST2 |
Category: |
Software |
Status: |
Fixed |
Target-Source: |
N/A |
Subject: |
MUST2 threshold from 1 MeV to 0.4 MeV |
|
|
It was found that the energy threshold of the strips was as high as 1 MeV. This threshold setting was chosen for the calibration and was never set back to the nominal threshold (0.4 MeV) for experiments.
Lowered to 0.4 MeV.
Note that due to the low energy threshold, triton of energies higher than 70 MeV were partially rejected during the measurement of 40Ca(p,t). |
|
|
|
|