MUST2
SAMURAI_2018
S3-LEB-LPC
SUPERNEMO
MUGAST
EXPAND
SCALP
GALATRON
HiCARI
VELO
|
MUST2 |
Not logged in |
|
|
Message ID: 110
Entry time: Sun Jul 2 10:31:48 2017
|
Author: |
Valérie Lapoux |
ExpNbr: |
e748 |
Type: |
ONLINE |
Category: |
Hardware |
Status: |
Fixed |
Target-Source: |
CD2 |
Subject: |
Change of the electronics for TAC Plast CATS1 |
|
|
Shift U (Béa +Vlx) entered in the room to check the electronics of Cats1 and of the TAC plast-Cats1
Now we have taken the output of one of the TAC : START=CATS1_FAG - STOP plast (no val)
We send it to a Linear Fifo from which it is duplicated for the ADC channels:
-ADC of VXI MUST2 parameter T_PL_CATS1 ; add 23
-ADC of VXI CHIO parameter T_PLch_CATS1; add 19
NB:
Now we put 50 Ohms everywhere in the logic Fifo for Cats1 Validated in the CATS electronics. |