Message ID: 735
Entry time: Thu Jun 28 10:38:58 2018
|
| Author: |
Freddy |
| ExpNbr: |
e755 |
| Type: |
MUST2 |
| Category: |
General |
| Status: |
Fixed |
| Target-Source: |
N/A |
| Subject: |
T4 Comparison of calibration runs |
|
|
Attached is a comparison of two calibrations performed before the experiment (2 weeks and one day before) for Telescope 4 (the one having a degraded resolution in Ex on the offline).
One can see that:- The T4 resolution for strips X has been degraded between the two source runs.
- The T4 resolution for strips Y remains ok and should be used for the analysis (slightly degraded only)
- From the pedestal extrapolation, it also seems that there was a global drift for some channels.
|
|
|
|
|