MUST2
SAMURAI_2018
S3-LEB-LPC
SUPERNEMO
MUGAST
EXPAND
SCALP
GALATRON
HiCARI
VELO
|
MUST2 |
Not logged in |
|
|
Message ID: 735
Entry time: Thu Jun 28 10:38:58 2018
|
Author: |
Freddy |
ExpNbr: |
e755 |
Type: |
MUST2 |
Category: |
General |
Status: |
Fixed |
Target-Source: |
N/A |
Subject: |
T4 Comparison of calibration runs |
|
|
Attached is a comparison of two calibrations performed before the experiment (2 weeks and one day before) for Telescope 4 (the one having a degraded resolution in Ex on the offline).
One can see that:- The T4 resolution for strips X has been degraded between the two source runs.
- The T4 resolution for strips Y remains ok and should be used for the analysis (slightly degraded only)
- From the pedestal extrapolation, it also seems that there was a global drift for some channels.
|
|
|
|
|