Target-Source: CD2 
Subject: Sili-1/T5 trip 
The Sili1 of T5 has tripped (again) at 10h30.
On again. 400V; I ~ 20ľA at 10h40, still increasing 25.6ľA at 11h30.(The limit Io is set at 26ľA)
To regulate this detector the Vo set is put at 380V instead of 400V. Now I ~ 24.4ľA.
We wil check that the current is stabilized below 24ľA, and after the V will be set again at 400V.