Energy resolution of the 8 MUST2 telescopes from central pic of run 3029 calibrated with 3 alpha source.
T1: mean = 5.389MeV , sigma = 15.0keV
T2: mean = 5.385MeV , sigma = 15.7keV
T3: mean = 5.391MeV , sigma = 15.2keV
T4: mean = 5.418MeV , sigma = 14.0keV
T5: mean = 5.390MeV , sigma = 16.3keV
T6: mean = 5.391MeV , sigma = 14.6keV
T7: mean = 5.389MeV , sigma = 17.5keV
T8: mean = 5.386MeV , sigma = 14.7keV
Total: mean = 5.391MeV , sigma = 17.7keV |