MUST2
SAMURAI_2018
S3-LEB-LPC
SUPERNEMO
MUGAST
EXPAND
SCALP
GALATRON
HiCARI
VELO
|
| MUST2 |
Not logged in |
|
|
|
Message ID: 378
Entry time: Sun May 13 19:36:09 2018
|
| Author: |
Valérian + François |
| ExpNbr: |
e744 |
| Type: |
MUST2 |
| Category: |
General |
| Status: |
Ongoing |
| Target-Source: |
3-alpha |
| Subject: |
Run #116 #117 3 alpha run Calibration T1-T5 trigger |
|
|
run #116
start @12h07 13/05/18
stop @19h45 13/05/18
I refilled the 0 degree cooler.
run #117
start @19h47 13/05/18
stop @ 9H04 14/05/18
Joined is a screenshot of the HT of MUST2
Vacuum in M2C is 5.7e-6 mbar and ~4e-5 in the 0 degree chamber which is higher than before.
Joined spectra. |
|
|
|
|