Fixed, e744, Mon Mar 26 14:20:22 2018, e744 : Valerian, MUST2, Hardware, 3-alpha, Bidimensional graphs for Telescope 7 and 6 
|
WARNING T5 == T7
[U][B]
EDIT 27/03/18 : WARNING X and Y are inverted, this entry should be ignored see entry : [URL=https://elog.lpc-caen.in2p3.fr/MUST2/333]Bidim for T1 T3 T4 |
Ongoing, e744, Mon Mar 26 15:07:06 2018, e744 : Valerian et Louis, MUST2, Hardware, 3-alpha, Energy vs Strip for T6 et T7  
|
Weird signal on StripX of T6... might be the copper oxyde projections making a short-circuit
[U][B]EDIT 27/03/18 : The problem seen on Telescope 6 might be due to either some Asics Problem, or due to some dirt, the current reason in not known currently. |
Ongoing, TEST, Tue Mar 27 10:47:39 2018, e744 : Valerian, MUST2, Hardware, 3-alpha, Choice of DSSSDs
|
|
Ongoing, e744, Tue Mar 27 11:11:04 2018, e744 : Valerian, MUST2, Hardware, 3-alpha, Choice of DSSSDs 14x
|
Different plots of Strip channel vs Strip number for X and Y in order to make our choice |
Fixed, e744, Tue Mar 27 16:26:29 2018, e744 : Valerian, MUST2, General, 3-alpha, Bidim for T1 T3 T4 T5 T6 T7 T8 7x
|
|
Ongoing, e744, Tue Mar 27 17:22:24 2018, e744 : Valerian, MUST2, Hardware, 3-alpha, Choosing the DSSSDs
|
We can see that the DSSSDs on T1(CsI) T3(CsI) T4(CsI) and T5(SiLi) are in good shape (cf :[URL=https://elog.lpc-caen.in2p3.fr/MUST2/333]Bidim for T1 T3
T4 T5 T6 T7 T8 [/URL]).
|
Ongoing, e744, Wed Mar 28 13:21:32 2018, e744 : Valerian, MUST2, Hardware, 3-alpha, Choosing the DSSSDs
|
[quote="e744 : Valerian"]We can see that the DSSSDs on T1(CsI) T3(CsI) T4(CsI) and T5(SiLi) are in good shape (cf :[URL=https://elog.lpc-caen.in2p3.fr/MUST2/333]Bidim
for T1 T3 T4 T5 T6 T7 T8 [/URL]).
|
Ongoing, e744, Wed Mar 28 16:35:01 2018, e744 : Valerian, MUST2, Hardware, 3-alpha, Choosing the DSSSDs  
|
[quote="e744 : Valerian"][quote="e744 : Valerian"]We can see that the DSSSDs on T1(CsI) T3(CsI) T4(CsI) and T5(SiLi) are in good shape (cf :[URL=https://elog.lpc-caen.in2p3.fr/MUST2/333]Bidim
for T1 T3 T4 T5 T6 T7 T8 [/URL]).
|
Fixed, e744, Tue Apr 3 10:44:38 2018, e744 : Valérian, MUST2, Hardware, , Update on Must2 telescopes
|
For the next day, the plan is to check the DSSD's connections of T6 & T7 followed by an alpha run, if the problem seen on T6 are not resolved, we will change
the DSSD of T6 and re-do another alpha run, the last option would be to change the Muffee card.
|
Ongoing, e744, Tue Apr 3 15:30:49 2018, e744 : Valérian, MUST2, Hardware, , Cabling of T6 & T7
|
After checking the cable of T6 & T7 it seems that there were in fact an inversion during experiment e748. T6 is the bottom one, and T7 the top one.
During e748, it seems that there were some difference regarding the position, between the hardware and software position...
|
Ongoing, e744, Wed Apr 4 17:49:05 2018, e744 : Valérian, EXL-PPT, Hardware, , Update on EXL 6x
|
The EXL detector has been mounted, it is very close to the target ~4cm and doesn't do any shadow on the M2 telescopes. But it isn't at 90° anymore. The
cable have been plugged-in (cf photos). With a new mapping of the detector (cf drawing). |
Fixed, e744, Tue Apr 10 17:43:28 2018, Adrien Matta and FyFy, MUST2, Hardware, N/A, login to CAEN SY1527 HV/BT mainframe
|
To control the CAEN SY1527 HV/BT mainframe use the following command from a terminal:
telnet 172.16.139.10 1527
|
Fixed, e744, Tue Apr 17 09:07:55 2018, Adrien Matta, DAQ, General, N/A, DAQ issue summary
|
Previously on e744:
After a chat with Charles Houarner we decided to have configuration with a Muvi card per VXI crate. This is supposed to reduce the dead time (DT) by parallelising |
Fixed, e744, Tue Apr 17 09:18:00 2018, Cherry MX team, MUST2, Hardware, N/A, T5 is back and running strong
|
T5 telescope appears to be fully functional again. It shows both Energy and Time with excellent resolution. |
Ongoing, e744, Tue Apr 17 17:36:37 2018, Cherry Mx Team, MUST2, Software, N/A, GRU SUCKS
|
Pretty much today's summary in the title.
[URL=https://i.pinimg.com/originals/a8/d4/12/a8d41206e8bd9c189397fb4907656981.jpg]https://i.pinimg.com/originals/a8/d4/12/a8d41206e8bd9c189397fb4907656981.jpg[/URL] |
Fixed, e744, Mon Apr 23 11:33:53 2018, e744 Valerian, OTHER, Hardware, N/A, Inspection signals
|
Inspection signals between D6 and LISE Acquisition room are connected as follow :
178: FAG
|
Pending, e744, Tue Apr 24 14:18:22 2018, Valerian Freddy and Adrien aka Cherry Mx Team, MUST2, Hardware, N/A, Current telescope-pcb layout
|
Telescope 1 : PCB 5 (ref on pcb: 2274)
Telescope 2 : PCB 4
Telescope 3 : PCB 7
|
Fixed, e744, Tue Apr 24 15:42:03 2018, e744 Valérie L., DAQ, General, , GMT chronogramme 
|
Crate ganlx1 GMT master FAG 100ns ; VAL 2 µs ; CT 6 µs
Crate ganlx14 GMT slave FAG 100ns ; VAL 2 µs ; CT 3 µs |
Fixed, e744, Tue Apr 24 15:51:04 2018, e744 Valérie L., MUST2, General, , MUST2 configuration /parameters
|
TAC range for all telescopes: 600ns
Thresholds: 402 keV |
Fixed, e744, Tue Apr 24 16:22:08 2018, e744 Valérie L., RUN, Software, , Narval Merger Parameters
|
|
Fixed, e744, Wed Apr 25 15:59:51 2018, Valérian, EXL-PPT, Software, N/A, Mesytech setup
|
To use the remote settings of the Mesytech modules :
[CODE]
- start GANP170
|
Fixed, e744, Wed Apr 25 20:47:59 2018, e744 Valérie, MUST2, General, , MUST2 mate Temperature on the DAS interface 
|
tests
The plot:ConfigDASMuvi1Tel4mufeeX.png shows the situation on the Mufee card when the cooling system was blocked (ice in the tube! there was water inside
instead of alcohol) on Wednesday 25th. cf there was 47°C on MufeeX mate of Telescope 4.
|
|