MUST2 SAMURAI_2018 S3-LEB-LPC SUPERNEMO MUGAST EXPAND SCALP GALATRON HiCARI VELO
  MUST2, Page 17 of 48  Not logged in MUST2
New entries since:Thu Jan 1 01:00:00 1970
Status ExpNbr Date Author Type Category Target-Source Subject Text Attachments
Fixede744 Mon Mar 26 11:46:43 2018 e744 : ValerianRUNGeneral3-alphaTests runs Ganacq log(15:51:33 - 19/03/2018) INFO - rcc_gui - Started
run #1
  
Fixede744 Mon Mar 26 14:20:22 2018 e744 : ValerianMUST2Hardware3-alphaBidimensional graphs for Telescope 7 and 6WARNING T5 == T7
[U][B]
EDIT 27/03/18 : WARNING X and Y are inverted,
 T6_3alpha.pdfT7_3alpha.pdf 
Ongoinge744 Mon Mar 26 15:07:06 2018 e744 : Valerian et LouisMUST2Hardware3-alphaEnergy vs Strip for T6 et T7Weird signal on StripX of T6... might be the
copper oxyde projections making a short-circuit
 StripXE_vs_StripNbr_T6.pdfStripXE_vs_StripNbr_T6_zoom.pdfStripYE_vs_StripNbr_T6.pdf 
OngoingTEST Tue Mar 27 10:47:39 2018 e744 : ValerianMUST2Hardware3-alphaChoice of DSSSDs    
Ongoinge744 Tue Mar 27 11:11:04 2018 e744 : ValerianMUST2Hardware3-alphaChoice of DSSSDsDifferent plots of Strip channel vs Strip
number for X and Y in order to make our choice
 14x 
Fixede744 Tue Mar 27 16:26:29 2018 e744 : ValerianMUST2General3-alphaBidim for T1 T3 T4 T5 T6 T7 T8   7x 
Ongoinge744 Tue Mar 27 17:22:24 2018 e744 : ValerianMUST2Hardware3-alphaChoosing the DSSSDsWe can see that the DSSSDs on T1(CsI) T3(CsI)
T4(CsI) and T5(SiLi) are in good shape (cf
:[URL=https://elog.lpc-caen.in2p3.fr/MUST2/333]Bidim
  
Ongoinge744 Wed Mar 28 13:21:32 2018 e744 : ValerianMUST2Hardware3-alphaChoosing the DSSSDs[quote="e744 : Valerian"]We can see that the
DSSSDs on T1(CsI) T3(CsI) T4(CsI) and T5(SiLi)
are in good shape (cf :[URL=https://elog.lpc-caen.in2p3.fr/MUST2/333]Bidim
  
Ongoinge744 Wed Mar 28 16:35:01 2018 e744 : ValerianMUST2Hardware3-alphaChoosing the DSSSDs[quote="e744 : Valerian"][quote="e744 : Valerian"]We
can see that the DSSSDs on T1(CsI) T3(CsI)
T4(CsI) and T5(SiLi) are in good shape (cf
 TimeMust2X_MUVI2_uncalib.pngTimeMust2Y_MUVI2_uncalib.pngSpectresMUVI2stripXT300ns.png 
Fixede744 Tue Apr 3 10:44:38 2018 e744 : ValérianMUST2Hardware Update on Must2 telescopesFor the next day, the plan is to check the
DSSD's connections of T6 & T7 followed by
an alpha run, if the problem seen on T6 are
 20180403_103732.jpg 
Ongoinge744 Tue Apr 3 15:30:49 2018 e744 : ValérianMUST2Hardware Cabling of T6 & T7After checking the cable of T6 & T7 it seems
that there were in fact an inversion during
experiment e748. T6 is the bottom one, and
 Inversion_between_telescopes.pdf 
Ongoinge744 Wed Apr 4 17:49:05 2018 e744 : ValérianEXL-PPTHardware Update on EXLThe EXL detector has been mounted, it is very
close to the target ~4cm and doesn't do any
shadow on the M2 telescopes. But it isn't
 6x 
Fixede744 Tue Apr 10 17:43:28 2018 Adrien Matta and FyFyMUST2HardwareN/Alogin to CAEN SY1527 HV/BT mainframeTo control the CAEN SY1527 HV/BT mainframe
use the following command from a terminal:
  
Fixede744 Tue Apr 17 09:07:55 2018 Adrien MattaDAQGeneralN/ADAQ issue summaryPreviously on e744:

After a chat with Charles Houarner we decided
  
Fixede744 Tue Apr 17 09:18:00 2018 Cherry MX teamMUST2HardwareN/AT5 is back and running strongT5 telescope appears to be fully functional
again. It shows both Energy and Time with
excellent resolution.
 T5.pdf 
Ongoinge744 Tue Apr 17 17:36:37 2018 Cherry Mx TeamMUST2SoftwareN/AGRU SUCKSPretty much today's summary in the title.

[URL=https://i.pinimg.com/originals/a8/d4/12/a8d41206e8bd9c189397fb4907656981.jpg]https://i.pinimg.com/originals/a8/d4/12/a8d41206e8bd9c189397fb4907656981.jpg[/URL]
  
Fixede744 Mon Apr 23 11:33:53 2018 e744 ValerianOTHERHardwareN/AInspection signalsInspection signals between D6 and LISE Acquisition
room are connected as follow : 
  
Pendinge744 Tue Apr 24 14:18:22 2018 Valerian Freddy and Adrien aka Cherry Mx TeamMUST2HardwareN/ACurrent telescope-pcb layoutTelescope 1 : PCB 5 (ref on pcb: 2274)
Telescope 2 : PCB 4
Telescope 3 : PCB 7
  
Fixede744 Tue Apr 24 15:42:03 2018 e744 Valérie L.DAQGeneral GMT chronogrammeCrate ganlx1 GMT master FAG 100ns ; VAL 2
µs ; CT 6 µs
Crate ganlx14 GMT slave FAG 100ns ; VAL 2
 GMTmasterChrono24avril18.pngGMTslaveChrono24avril18.png 
Fixede744 Tue Apr 24 15:51:04 2018 e744 Valérie L.MUST2General MUST2 configuration /parameters
TAC range for all telescopes: 600ns
Thresholds: 402 keV
 ConfigMuviPhysMust2Thr.png 

ELOG Home Page