MUST2 SAMURAI_2018 S3-LEB-LPC SUPERNEMO MUGAST EXPAND SCALP GALATRON HiCARI VELO
  MUST2, Page 20 of 48  Not logged in MUST2
Status ExpNbr Date Authordown Type Category Target-Source Subject
Ongoinge744 Mon Mar 26 15:07:06 2018 e744 : Valerian et LouisMUST2Hardware3-alphaEnergy vs Strip for T6 et T7
Weird signal on StripX of T6... might be the copper oxyde projections making a short-circuit

EDIT 27/03/18 : The problem seen on Telescope 6 might be due to either some Asics Problem, or due to some dirt, the current reason in not known currently. The safest way would be to remove the telescope completely and replace it for Daisuke's and Olivier's experiments.
Attachment 1: StripXE_vs_StripNbr_T6.pdf
StripXE_vs_StripNbr_T6.pdf
Attachment 2: StripXE_vs_StripNbr_T6_zoom.pdf
StripXE_vs_StripNbr_T6_zoom.pdf
Attachment 3: StripYE_vs_StripNbr_T6.pdf
StripYE_vs_StripNbr_T6.pdf
Pinnede744 Mon Mar 26 11:44:27 2018 e744 : ValerianOTHERGeneral Current Elog state
Currently we do not have access to the experiment number. To write an entry for e744 the nickname TEST will be used.
Fixede744 Mon Mar 26 11:46:43 2018 e744 : ValerianRUNGeneral3-alphaTests runs Ganacq log
(15:51:33 - 19/03/2018) INFO - rcc_gui - Started run #1

(16:22:01 - 19/03/2018) INFO - rcc_gui - Stopped run #1

(16:25:05 - 19/03/2018) INFO - rcc_gui - Started run #2
Pedestal aligned T1-4
(16:31:10 - 19/03/2018) INFO - rcc_gui - Stopped run #2

(16:32:26 - 19/03/2018) INFO - rcc_gui - Started run #3
realigned the pedestal on T1-4
(16:33:48 - 19/03/2018) INFO - rcc_gui - Stopped run #3

(16:41:52 - 19/03/2018) INFO - rcc_gui - Started run #4
Set polarity of T1-4 correctly and set again the pedestals
(16:49:34 - 19/03/2018) INFO - rcc_gui - Stopped run #4

(16:53:42 - 19/03/2018) INFO - rcc_gui - Started run #5

(16:55:20 - 19/03/2018) INFO - rcc_gui - Stopped run #5

(16:56:40 - 19/03/2018) INFO - rcc_gui - Started run #6

(16:57:11 - 19/03/2018) INFO - rcc_gui - Stopped run #6

(16:58:04 - 19/03/2018) INFO - rcc_gui - Started run #7

(16:58:27 - 19/03/2018) INFO - rcc_gui - Stopped run #7

(17:06:47 - 19/03/2018) INFO - rcc_gui - Stopped run #8

(17:08:44 - 19/03/2018) INFO - rcc_gui - Started run #9

(17:08:49 - 19/03/2018) INFO - rcc_gui - Started run #8

(17:13:44 - 19/03/2018) INFO - rcc_gui - Stopped run #9

(17:14:56 - 19/03/2018) INFO - rcc_gui - Started run #10

(17:15:44 - 19/03/2018) INFO - rcc_gui - Stopped run #10

(17:17:52 - 19/03/2018) INFO - rcc_gui - Started run #11

(17:18:16 - 19/03/2018) INFO - rcc_gui - Stopped run #11

(17:20:54 - 19/03/2018) INFO - rcc_gui - Started run #12

(17:26:05 - 19/03/2018) INFO - rcc_gui - Stopped run #12

(17:26:56 - 19/03/2018) INFO - rcc_gui - Started run #13

(17:32:16 - 19/03/2018) INFO - rcc_gui - Stopped run #13

(17:32:55 - 19/03/2018) INFO - rcc_gui - Started run #14

(17:34:05 - 19/03/2018) INFO - rcc_gui - Stopped run #14

(17:35:25 - 19/03/2018) INFO - rcc_gui - Started run #15

(18:20:40 - 19/03/2018) INFO - rcc_gui - Stopped run #15

(18:21:00 - 19/03/2018) INFO - rcc_gui - Started run #16

(18:23:36 - 19/03/2018) INFO - rcc_gui - Stopped run #16

(18:23:52 - 19/03/2018) INFO - rcc_gui - Started run #17

(18:30:52 - 19/03/2018) INFO - rcc_gui - Stopped run #17

(18:31:42 - 19/03/2018) INFO - rcc_gui - Started run #18

(18:37:11 - 19/03/2018) INFO - rcc_gui - Stopped run #18

(18:37:43 - 19/03/2018) INFO - rcc_gui - Started run #19

(18:39:44 - 19/03/2018) INFO - rcc_gui - Stopped run #19

(18:40:08 - 19/03/2018) INFO - rcc_gui - Started run #20

(18:43:10 - 19/03/2018) INFO - rcc_gui - Stopped run #20

(18:44:12 - 19/03/2018) INFO - rcc_gui - Started run #21

(18:45:31 - 19/03/2018) INFO - rcc_gui - Stopped run #21

(09:19:38 - 20/03/2018) INFO - rcc_gui - Stopped run #22

(09:19:46 - 20/03/2018) INFO - rcc_gui - Started run #22

(09:21:46 - 20/03/2018) INFO - rcc_gui - Started run #23

(09:23:22 - 20/03/2018) INFO - rcc_gui - Stopped run #23

(09:23:56 - 20/03/2018) INFO - rcc_gui - Started run #24

(09:25:31 - 20/03/2018) INFO - rcc_gui - Stopped run #24

(09:29:18 - 20/03/2018) INFO - rcc_gui - Stopped run #25

(09:31:08 - 20/03/2018) INFO - rcc_gui - Started run #26

(09:31:14 - 20/03/2018) INFO - rcc_gui - Started run #25

(09:32:23 - 20/03/2018) INFO - rcc_gui - Stopped run #26

(09:55:33 - 20/03/2018) ERROR - rcc_gui - Operation forbidden - Storage mode can be changed only in READY state
(09:56:20 - 20/03/2018) INFO - rcc_gui - Started run #26
source alpha, en G3, piedestaux suppression set
(14:24:46 - 20/03/2018) INFO - rcc_gui - Stopped run #26

(14:25:43 - 20/03/2018) INFO - rcc_gui - Started run #27
follow up, same condition as run 26, stopped daq for analysis.
(09:42:36 - 21/03/2018) INFO - rcc_gui - Stopped run #27

(18:06:46 - 21/03/2018) INFO - rcc_gui - Started run #28

(18:09:02 - 21/03/2018) INFO - rcc_gui - Stopped run #28

(18:10:33 - 21/03/2018) INFO - rcc_gui - Started run #29

(18:11:47 - 21/03/2018) INFO - rcc_gui - Stopped run #29

(18:16:59 - 21/03/2018) INFO - rcc_gui - Started run #30

(18:18:15 - 21/03/2018) INFO - rcc_gui - Stopped run #30

(18:25:10 - 21/03/2018) INFO - rcc_gui - Started run #31

(18:26:51 - 21/03/2018) INFO - rcc_gui - Stopped run #31

(18:37:07 - 21/03/2018) INFO - rcc_gui - Started run #32

(18:39:04 - 21/03/2018) INFO - rcc_gui - Stopped run #32

(18:41:49 - 21/03/2018) INFO - rcc_gui - Started run #33

(18:49:13 - 21/03/2018) INFO - rcc_gui - Stopped run #33

(18:53:08 - 21/03/2018) INFO - rcc_gui - Started run #34

(18:53:42 - 21/03/2018) INFO - rcc_gui - Stopped run #34

(18:56:02 - 21/03/2018) INFO - rcc_gui - Started run #35

(18:57:06 - 21/03/2018) INFO - rcc_gui - Stopped run #35

(10:22:37 - 22/03/2018) INFO - rcc_gui - Started run #37
Start source 3a MUVI2 1,2 (MM5 & MM6)
(11:24:12 - 22/03/2018) INFO - rcc_gui - Stopped run #37

(11:25:45 - 22/03/2018) INFO - rcc_gui - Started run #38
correction thresolds Muvi2
(12:26:38 - 22/03/2018) INFO - rcc_gui - Stopped run #38

(12:46:48 - 22/03/2018) INFO - rcc_gui - Started run #39

(12:49:47 - 22/03/2018) INFO - rcc_gui - Stopped run #39

(12:59:48 - 22/03/2018) INFO - rcc_gui - Started run #40

(15:44:26 - 22/03/2018) INFO - rcc_gui - Stopped run #40

(09:11:26 - 23/03/2018) INFO - rcc_gui - Started run #41
test
(09:13:41 - 23/03/2018) INFO - rcc_gui - Stopped run #41

(09:13:48 - 23/03/2018) ERROR - rcc_gui - Operation forbidden - Storage mode can be changed only in READY state
(09:41:06 - 23/03/2018) INFO - rcc_gui - Started run #41
MM5 connected to MM8
MM6 connected to MM5
correct piedestal
3 pics alpha source
(14:02:04 - 23/03/2018) INFO - rcc_gui - Stopped run #41

(14:25:59 - 23/03/2018) INFO - rcc_gui - Started run #42
channel MM5 = real MM7
channel MM6 = real MM6
alpha source
(14:31:33 - 23/03/2018) INFO - rcc_gui - Stopped run #42

(14:33:42 - 23/03/2018) INFO - rcc_gui - Started run #43
same after change in the piedestals
(15:54:10 - 23/03/2018) INFO - rcc_gui - Stopped run #43
Fixede744 Mon Mar 26 14:20:22 2018 e744 : ValerianMUST2Hardware3-alphaBidimensional graphs for Telescope 7 and 6
WARNING T5 == T7

EDIT 27/03/18 : WARNING X and Y are inverted, this entry should be ignored see entry : Bidim for T1 T3 T4 T5 T6 T7 T8
Attachment 1: T6_3alpha.pdf
T6_3alpha.pdf
Attachment 2: T7_3alpha.pdf
T7_3alpha.pdf
OngoingTEST Tue Mar 27 10:47:39 2018 e744 : ValerianMUST2Hardware3-alphaChoice of DSSSDs
Ongoinge744 Tue Mar 27 11:11:04 2018 e744 : ValerianMUST2Hardware3-alphaChoice of DSSSDs
Different plots of Strip channel vs Strip number for X and Y in order to make our choice
Attachment 1: StripXE_vs_StripNbr_T1.pdf
StripXE_vs_StripNbr_T1.pdf
Attachment 2: StripYE_vs_StripNbr_T1.pdf
StripYE_vs_StripNbr_T1.pdf
Attachment 3: StripXE_vs_StripNbr_T3.pdf
StripXE_vs_StripNbr_T3.pdf
Attachment 4: StripYE_vs_StripNbr_T3.pdf
StripYE_vs_StripNbr_T3.pdf
Attachment 5: StripXE_vs_StripNbr_T4.pdf
StripXE_vs_StripNbr_T4.pdf
Attachment 6: StripYE_vs_StripNbr_T4.pdf
StripYE_vs_StripNbr_T4.pdf
Attachment 7: StripXE_vs_StripNbr_T5.pdf
StripXE_vs_StripNbr_T5.pdf
Attachment 8: StripYE_vs_StripNbr_T5.pdf
StripYE_vs_StripNbr_T5.pdf
Attachment 9: StripXE_vs_StripNbr_T6.pdf
StripXE_vs_StripNbr_T6.pdf
Attachment 10: StripYE_vs_StripNbr_T6.pdf
StripYE_vs_StripNbr_T6.pdf
Attachment 11: StripXE_vs_StripNbr_T7.pdf
StripXE_vs_StripNbr_T7.pdf
Attachment 12: StripYE_vs_StripNbr_T7.pdf
StripYE_vs_StripNbr_T7.pdf
Attachment 13: StripXE_vs_StripNbr_T8.pdf
StripXE_vs_StripNbr_T8.pdf
Attachment 14: StripYE_vs_StripNbr_T8.pdf
StripYE_vs_StripNbr_T8.pdf
Fixede744 Tue Mar 27 16:26:29 2018 e744 : ValerianMUST2General3-alphaBidim for T1 T3 T4 T5 T6 T7 T8
Attachment 1: StripX_vs_StripY_T1.pdf
StripX_vs_StripY_T1.pdf
Attachment 2: StripX_vs_StripY_T3.pdf
StripX_vs_StripY_T3.pdf
Attachment 3: StripX_vs_StripY_T4.pdf
StripX_vs_StripY_T4.pdf
Attachment 4: StripX_vs_StripY_T5.pdf
StripX_vs_StripY_T5.pdf
Attachment 5: StripX_vs_StripY_T6.pdf
StripX_vs_StripY_T6.pdf
Attachment 6: StripX_vs_StripY_T7.pdf
StripX_vs_StripY_T7.pdf
Attachment 7: StripX_vs_StripY_T8.pdf
StripX_vs_StripY_T8.pdf
Ongoinge744 Tue Mar 27 17:22:24 2018 e744 : ValerianMUST2Hardware3-alphaChoosing the DSSSDs
We can see that the DSSSDs on T1(CsI) T3(CsI) T4(CsI) and T5(SiLi) are in good shape (cf :Bidim for T1 T3 T4 T5 T6 T7 T8 ).

As for now, we would like to keep the telescope as is. M2C should be moved tomorrow (28/03/18).

After that we would like to use the two spare DSSSDs available on both T2 and on the O degree telescope (now called T5(CsI)).

If the spare DSSSDs are judged not be good enough for the experiment, we will swap current T5(SiLi) DSSDs on the 0 degree Telescope T5(CsI).
Ongoinge744 Wed Mar 28 13:21:32 2018 e744 : ValerianMUST2Hardware3-alphaChoosing the DSSSDs

e744 : Valerian wrote:
We can see that the DSSSDs on T1(CsI) T3(CsI) T4(CsI) and T5(SiLi) are in good shape (cf :Bidim for T1 T3 T4 T5 T6 T7 T8 ).

As for now, we would like to keep the telescope as is. M2C should be moved tomorrow (28/03/18).

After that we would like to use the two spare DSSSDs available on both T2 and on the O degree telescope (now called T5(CsI)).

If the spare DSSSDs are judged not be good enough for the experiment, we will swap current T5(SiLi) DSSDs on the 0 degree Telescope T5(CsI).


Which telescope should we take for T5? T7 or T6?
Ongoinge744 Wed Mar 28 16:35:01 2018 e744 : ValerianMUST2Hardware3-alphaChoosing the DSSSDs

e744 : Valerian wrote:

e744 : Valerian wrote:
We can see that the DSSSDs on T1(CsI) T3(CsI) T4(CsI) and T5(SiLi) are in good shape (cf :Bidim for T1 T3 T4 T5 T6 T7 T8 ).

As for now, we would like to keep the telescope as is. M2C should be moved tomorrow (28/03/18).

After that we would like to use the two spare DSSSDs available on both T2 and on the O degree telescope (now called T5(CsI)).

If the spare DSSSDs are judged not be good enough for the experiment, we will swap current T5(SiLi) DSSDs on the 0 degree Telescope T5(CsI).


Which telescope should we take for T5? T7 or T6?


Looking at the previous Calimero tests Adrien did during his experiment, neither telescope are good as is... the T7 has no time calibration and T6 might have some Asics problem...

We can either use T6 and check if the problem we saw on it's DSSDs are due to some electronics problem or use a completly different telescope from the spare we have...

EDIT 29/03/18 : After discussion with Laurent the option to use a new DSSD on T6 seems more appropriate, still ongoing.
Attachment 1: TimeMust2X_MUVI2_uncalib.png
TimeMust2X_MUVI2_uncalib.png
Attachment 2: TimeMust2Y_MUVI2_uncalib.png
TimeMust2Y_MUVI2_uncalib.png
Attachment 3: SpectresMUVI2stripXT300ns.png
SpectresMUVI2stripXT300ns.png
Fixede744 Tue Apr 3 10:44:38 2018 e744 : ValérianMUST2Hardware Update on Must2 telescopes
For the next day, the plan is to check the DSSD's connections of T6 & T7 followed by an alpha run, if the problem seen on T6 are not resolved, we will change the DSSD of T6 and re-do another alpha run, the last option would be to change the Muffee card.

Just for a quick visual update, telescopes 5&8 are dismounted and a new DSSD is mounted on T2 as you can see !
Attachment 1: 20180403_103732.jpg
20180403_103732.jpg
Ongoinge744 Tue Apr 3 15:30:49 2018 e744 : ValérianMUST2Hardware Cabling of T6 & T7
After checking the cable of T6 & T7 it seems that there were in fact an inversion during experiment e748. T6 is the bottom one, and T7 the top one.

During e748, it seems that there were some difference regarding the position, between the hardware and software position...

If it was the case, it would mean that it is T6 that has the time calibration problem, not T7.
re-
In conclusion, it is T6 that has both the time calibration problem on X and Y, and the weird signal seen during the calibration.

EDIT 4/04/18 : For the experiment e748 and before the time calibration, the inversion had been made, as a conclusion what was previously stated is wrong, it is T7 that has no time calibration... Anyhow, wee will re-do the time calibration for T6 and T7 to check if the problem are still present.
Attachment 1: Inversion_between_telescopes.pdf
Inversion_between_telescopes.pdf
Ongoinge744 Wed Apr 4 17:49:05 2018 e744 : ValérianEXL-PPTHardware Update on EXL
The EXL detector has been mounted, it is very close to the target ~4cm and doesn't do any shadow on the M2 telescopes. But it isn't at 90° anymore. The cable have been plugged-in (cf photos). With a new mapping of the detector (cf drawing).
Attachment 1: 20180404_113303.jpg
20180404_113303.jpg
Attachment 2: 20180404_113312.jpg
20180404_113312.jpg
Attachment 3: 20180404_135324.jpg
20180404_135324.jpg
Attachment 4: 20180404_135334.jpg
20180404_135334.jpg
Attachment 5: EXL_cables.pdf
EXL_cables.pdf
Attachment 6: Mapping_of_EXL.pdf
Mapping_of_EXL.pdf
Fixede744 Tue May 22 22:33:09 2018 e744MUST2SoftwareN/ARun #363 Calimero T1-T4 strips X again with threshold at 50% (same as 40%)
Run #363
NB: Thresholds are the ones on the pulser amplitude to inject the signal (Calimero Pulser interface on Geco). see snapshots on the Calimero summary.
Calimero with 50% thresholds on T4. Same spectra as with the thresholds @40%. I think T4 is triggering on cross-talk, so when the thresholds it's high it's ok.
Attachment 1: T1T4_Xstrips_run363.png
T1T4_Xstrips_run363.png
Fixede744 Tue May 22 22:50:47 2018 e744MUST2SoftwareN/ARun #365 Calimero strips Y T1-T4 100% Threshold
Run #365
Calimero strips Y T1-T4
100% threshold on T4
Something is wrong with the masking?!
Attachment 1: T1T4_Ystrips_run364.png
T1T4_Ystrips_run364.png
Fixede744 Wed May 23 00:30:34 2018 e744EXL-PPTGeneralN/ARUN #402, EXL PPT trigger
RUN #402
start exl ppt run for checking coincidence
exl ch 17 put in to adc2 ch 32
Fixede744 Wed May 23 18:20:59 2018 e744MUST2GeneralN/ATest Calimero Run 425
Started run #425 Calimero NEW XT T1toT4
ampl Pulser 12MEV; thr for T4 1 MeV
problem on data taking
Fixede755 Sat Jun 30 20:08:27 2018 didier, takeshi, daisukeRUNGeneralCryo-LH2RUN 199 37C + liq. H
Attachment 1: scaler_2018-06-30_19h41m59s.png
scaler_2018-06-30_19h41m59s.png
Attachment 2: visu_2018-06-30_19h42m32s.png
visu_2018-06-30_19h42m32s.png
Attachment 3: visu_2018-06-30_20h04m35s.png
visu_2018-06-30_20h04m35s.png
Attachment 4: visu_2018-06-30_20h04m46s.png
visu_2018-06-30_20h04m46s.png
Attachment 5: visu_2018-06-30_20h04m51s.png
visu_2018-06-30_20h04m51s.png
Attachment 6: visu_2018-06-30_20h04m59s.png
visu_2018-06-30_20h04m59s.png
Attachment 7: visu_2018-06-30_20h05m12s.png
visu_2018-06-30_20h05m12s.png
Attachment 8: visu_2018-06-30_20h05m17s.png
visu_2018-06-30_20h05m17s.png
Attachment 9: visu_2018-06-30_20h05m23s.png
visu_2018-06-30_20h05m23s.png
Attachment 10: visu_2018-06-30_20h05m29s.png
visu_2018-06-30_20h05m29s.png
Attachment 11: visu_2018-06-30_20h05m35s.png
visu_2018-06-30_20h05m35s.png
Attachment 12: visu_2018-06-30_20h05m44s.png
visu_2018-06-30_20h05m44s.png
Attachment 13: visu_2018-06-30_20h05m50s.png
visu_2018-06-30_20h05m50s.png
Attachment 14: visu_2018-06-30_20h06m05s.png
visu_2018-06-30_20h06m05s.png
Attachment 15: visu_2018-06-30_20h06m13s.png
visu_2018-06-30_20h06m13s.png
Attachment 16: visu_2018-06-30_20h06m20s.png
visu_2018-06-30_20h06m20s.png
Attachment 17: visu_2018-06-30_20h06m26s.png
visu_2018-06-30_20h06m26s.png
Attachment 18: visu_2018-06-30_20h06m31s.png
visu_2018-06-30_20h06m31s.png
Attachment 19: visu_2018-06-30_20h06m39s.png
visu_2018-06-30_20h06m39s.png
Attachment 20: visu_2018-06-30_20h06m45s.png
visu_2018-06-30_20h06m45s.png
Attachment 21: visu_2018-06-30_20h06m54s.png
visu_2018-06-30_20h06m54s.png
Attachment 22: visu_2018-06-30_20h07m04s.png
visu_2018-06-30_20h07m04s.png
Attachment 23: visu_2018-06-30_20h07m10s.png
visu_2018-06-30_20h07m10s.png
Attachment 24: visu_2018-06-30_20h07m20s.png
visu_2018-06-30_20h07m20s.png
Attachment 25: visu_2018-06-30_20h07m25s.png
visu_2018-06-30_20h07m25s.png
Attachment 26: visu_2018-06-30_20h07m33s.png
visu_2018-06-30_20h07m33s.png
Attachment 27: visu_2018-06-30_20h07m45s.png
visu_2018-06-30_20h07m45s.png
Attachment 28: visu_2018-06-30_20h07m55s.png
visu_2018-06-30_20h07m55s.png
Fixede738 Fri Jul 13 22:00:59 2018 didier, nori, takeshi, shumpei, hide, daisukeRUNGeneralCryo-LH2RUN 376: 12C first run
12C + liquid H.
start 21h44
stop 22h43
Attachment 1: scaler_2018-07-13_21h59m16s.png
scaler_2018-07-13_21h59m16s.png
Attachment 2: screenshot_2018-07-13_21h54m55s.png
screenshot_2018-07-13_21h54m55s.png
Attachment 3: visu_2018-07-13_21h54m41s.png
visu_2018-07-13_21h54m41s.png
Attachment 4: visu_2018-07-13_21h58m06s.png
visu_2018-07-13_21h58m06s.png
Attachment 5: visu_2018-07-13_21h58m08s.png
visu_2018-07-13_21h58m08s.png
Attachment 6: visu_2018-07-13_21h58m10s.png
visu_2018-07-13_21h58m10s.png
Attachment 7: visu_2018-07-13_21h58m13s.png
visu_2018-07-13_21h58m13s.png
Attachment 8: visu_2018-07-13_21h58m15s.png
visu_2018-07-13_21h58m15s.png
Attachment 9: visu_2018-07-13_21h58m17s.png
visu_2018-07-13_21h58m17s.png
Attachment 10: visu_2018-07-13_21h58m23s.png
visu_2018-07-13_21h58m23s.png
Attachment 11: visu_2018-07-13_21h58m28s.png
visu_2018-07-13_21h58m28s.png
Attachment 12: visu_2018-07-13_21h58m32s.png
visu_2018-07-13_21h58m32s.png
Attachment 13: visu_2018-07-13_21h58m36s.png
visu_2018-07-13_21h58m36s.png
Attachment 14: visu_2018-07-13_21h58m41s.png
visu_2018-07-13_21h58m41s.png
Attachment 15: visu_2018-07-13_21h58m23s.png
visu_2018-07-13_21h58m23s.png
Attachment 16: visu_2018-07-13_21h58m28s.png
visu_2018-07-13_21h58m28s.png
Attachment 17: visu_2018-07-13_21h58m32s.png
visu_2018-07-13_21h58m32s.png
Attachment 18: visu_2018-07-13_21h58m36s.png
visu_2018-07-13_21h58m36s.png
Attachment 19: visu_2018-07-13_21h58m41s.png
visu_2018-07-13_21h58m41s.png
Attachment 20: visu_2018-07-13_21h58m45s.png
visu_2018-07-13_21h58m45s.png
Attachment 21: visu_2018-07-13_21h58m49s.png
visu_2018-07-13_21h58m49s.png
Attachment 22: visu_2018-07-13_21h58m54s.png
visu_2018-07-13_21h58m54s.png
Fixede738 Tue Jul 10 21:26:48 2018 didier, daisukeRUNGeneralCryo-LH2RUN 334
9C + liquid H. Probably nobody cares this post right now, but the beam is back.

21h00 the beam back to LISE.
21h03 start. Tune the wien filter 3.0 to 0.0 kV to optimize the transmission from CATS to PLASTIC (with visual scaler). TOF spectrum is the same as before.
21h32 1.7 electric micro A.
22h20 Wien filter 0 to 1.0 kV. Scaler Plastic/CATS2 ratio is the highest (93%) with this setting.
23h11 stop
Attachment 1: scaler_2018-07-10_21h21m21s.png
scaler_2018-07-10_21h21m21s.png
Attachment 2: visu_2018-07-10_21h23m07s.png
visu_2018-07-10_21h23m07s.png
Attachment 3: visu_2018-07-10_21h23m10s.png
visu_2018-07-10_21h23m10s.png
Attachment 4: visu_2018-07-10_21h23m12s.png
visu_2018-07-10_21h23m12s.png
Attachment 5: visu_2018-07-10_21h23m18s.png
visu_2018-07-10_21h23m18s.png
Attachment 6: visu_2018-07-10_21h23m21s.png
visu_2018-07-10_21h23m21s.png
Attachment 7: visu_2018-07-10_21h23m23s.png
visu_2018-07-10_21h23m23s.png
Attachment 8: visu_2018-07-10_21h23m28s.png
visu_2018-07-10_21h23m28s.png
Attachment 9: visu_2018-07-10_21h23m38s.png
visu_2018-07-10_21h23m38s.png
Attachment 10: visu_2018-07-10_21h23m50s.png
visu_2018-07-10_21h23m50s.png
Attachment 11: visu_2018-07-10_21h23m53s.png
visu_2018-07-10_21h23m53s.png
Attachment 12: visu_2018-07-10_21h23m56s.png
visu_2018-07-10_21h23m56s.png
Attachment 13: visu_2018-07-10_21h23m59s.png
visu_2018-07-10_21h23m59s.png
Attachment 14: visu_2018-07-10_21h24m08s.png
visu_2018-07-10_21h24m08s.png
Attachment 15: visu_2018-07-10_21h24m11s.png
visu_2018-07-10_21h24m11s.png
Attachment 16: screenshot_2018-07-10_22h26m19s.png
screenshot_2018-07-10_22h26m19s.png
Attachment 17: scaler_2018-07-10_23h13m13s.png
scaler_2018-07-10_23h13m13s.png

ELOG Home Page