Status |
ExpNbr |
Date |
Author |
Type |
Category |
Target-Source |
Subject |
Fixed | e744 |
Mon Mar 26 11:46:43 2018 |
e744 : Valerian | RUN | General | 3-alpha | Tests runs Ganacq log |
(15:51:33 - 19/03/2018) INFO - rcc_gui - Started run #1
(16:22:01 - 19/03/2018) INFO - rcc_gui - Stopped run #1
(16:25:05 - 19/03/2018) INFO - rcc_gui - Started run #2
Pedestal aligned T1-4
(16:31:10 - 19/03/2018) INFO - rcc_gui - Stopped run #2
(16:32:26 - 19/03/2018) INFO - rcc_gui - Started run #3
realigned the pedestal on T1-4
(16:33:48 - 19/03/2018) INFO - rcc_gui - Stopped run #3
(16:41:52 - 19/03/2018) INFO - rcc_gui - Started run #4
Set polarity of T1-4 correctly and set again the pedestals
(16:49:34 - 19/03/2018) INFO - rcc_gui - Stopped run #4
(16:53:42 - 19/03/2018) INFO - rcc_gui - Started run #5
(16:55:20 - 19/03/2018) INFO - rcc_gui - Stopped run #5
(16:56:40 - 19/03/2018) INFO - rcc_gui - Started run #6
(16:57:11 - 19/03/2018) INFO - rcc_gui - Stopped run #6
(16:58:04 - 19/03/2018) INFO - rcc_gui - Started run #7
(16:58:27 - 19/03/2018) INFO - rcc_gui - Stopped run #7
(17:06:47 - 19/03/2018) INFO - rcc_gui - Stopped run #8
(17:08:44 - 19/03/2018) INFO - rcc_gui - Started run #9
(17:08:49 - 19/03/2018) INFO - rcc_gui - Started run #8
(17:13:44 - 19/03/2018) INFO - rcc_gui - Stopped run #9
(17:14:56 - 19/03/2018) INFO - rcc_gui - Started run #10
(17:15:44 - 19/03/2018) INFO - rcc_gui - Stopped run #10
(17:17:52 - 19/03/2018) INFO - rcc_gui - Started run #11
(17:18:16 - 19/03/2018) INFO - rcc_gui - Stopped run #11
(17:20:54 - 19/03/2018) INFO - rcc_gui - Started run #12
(17:26:05 - 19/03/2018) INFO - rcc_gui - Stopped run #12
(17:26:56 - 19/03/2018) INFO - rcc_gui - Started run #13
(17:32:16 - 19/03/2018) INFO - rcc_gui - Stopped run #13
(17:32:55 - 19/03/2018) INFO - rcc_gui - Started run #14
(17:34:05 - 19/03/2018) INFO - rcc_gui - Stopped run #14
(17:35:25 - 19/03/2018) INFO - rcc_gui - Started run #15
(18:20:40 - 19/03/2018) INFO - rcc_gui - Stopped run #15
(18:21:00 - 19/03/2018) INFO - rcc_gui - Started run #16
(18:23:36 - 19/03/2018) INFO - rcc_gui - Stopped run #16
(18:23:52 - 19/03/2018) INFO - rcc_gui - Started run #17
(18:30:52 - 19/03/2018) INFO - rcc_gui - Stopped run #17
(18:31:42 - 19/03/2018) INFO - rcc_gui - Started run #18
(18:37:11 - 19/03/2018) INFO - rcc_gui - Stopped run #18
(18:37:43 - 19/03/2018) INFO - rcc_gui - Started run #19
(18:39:44 - 19/03/2018) INFO - rcc_gui - Stopped run #19
(18:40:08 - 19/03/2018) INFO - rcc_gui - Started run #20
(18:43:10 - 19/03/2018) INFO - rcc_gui - Stopped run #20
(18:44:12 - 19/03/2018) INFO - rcc_gui - Started run #21
(18:45:31 - 19/03/2018) INFO - rcc_gui - Stopped run #21
(09:19:38 - 20/03/2018) INFO - rcc_gui - Stopped run #22
(09:19:46 - 20/03/2018) INFO - rcc_gui - Started run #22
(09:21:46 - 20/03/2018) INFO - rcc_gui - Started run #23
(09:23:22 - 20/03/2018) INFO - rcc_gui - Stopped run #23
(09:23:56 - 20/03/2018) INFO - rcc_gui - Started run #24
(09:25:31 - 20/03/2018) INFO - rcc_gui - Stopped run #24
(09:29:18 - 20/03/2018) INFO - rcc_gui - Stopped run #25
(09:31:08 - 20/03/2018) INFO - rcc_gui - Started run #26
(09:31:14 - 20/03/2018) INFO - rcc_gui - Started run #25
(09:32:23 - 20/03/2018) INFO - rcc_gui - Stopped run #26
(09:55:33 - 20/03/2018) ERROR - rcc_gui - Operation forbidden - Storage mode can be changed only in READY state
(09:56:20 - 20/03/2018) INFO - rcc_gui - Started run #26
source alpha, en G3, piedestaux suppression set
(14:24:46 - 20/03/2018) INFO - rcc_gui - Stopped run #26
(14:25:43 - 20/03/2018) INFO - rcc_gui - Started run #27
follow up, same condition as run 26, stopped daq for analysis.
(09:42:36 - 21/03/2018) INFO - rcc_gui - Stopped run #27
(18:06:46 - 21/03/2018) INFO - rcc_gui - Started run #28
(18:09:02 - 21/03/2018) INFO - rcc_gui - Stopped run #28
(18:10:33 - 21/03/2018) INFO - rcc_gui - Started run #29
(18:11:47 - 21/03/2018) INFO - rcc_gui - Stopped run #29
(18:16:59 - 21/03/2018) INFO - rcc_gui - Started run #30
(18:18:15 - 21/03/2018) INFO - rcc_gui - Stopped run #30
(18:25:10 - 21/03/2018) INFO - rcc_gui - Started run #31
(18:26:51 - 21/03/2018) INFO - rcc_gui - Stopped run #31
(18:37:07 - 21/03/2018) INFO - rcc_gui - Started run #32
(18:39:04 - 21/03/2018) INFO - rcc_gui - Stopped run #32
(18:41:49 - 21/03/2018) INFO - rcc_gui - Started run #33
(18:49:13 - 21/03/2018) INFO - rcc_gui - Stopped run #33
(18:53:08 - 21/03/2018) INFO - rcc_gui - Started run #34
(18:53:42 - 21/03/2018) INFO - rcc_gui - Stopped run #34
(18:56:02 - 21/03/2018) INFO - rcc_gui - Started run #35
(18:57:06 - 21/03/2018) INFO - rcc_gui - Stopped run #35
(10:22:37 - 22/03/2018) INFO - rcc_gui - Started run #37
Start source 3a MUVI2 1,2 (MM5 & MM6)
(11:24:12 - 22/03/2018) INFO - rcc_gui - Stopped run #37
(11:25:45 - 22/03/2018) INFO - rcc_gui - Started run #38
correction thresolds Muvi2
(12:26:38 - 22/03/2018) INFO - rcc_gui - Stopped run #38
(12:46:48 - 22/03/2018) INFO - rcc_gui - Started run #39
(12:49:47 - 22/03/2018) INFO - rcc_gui - Stopped run #39
(12:59:48 - 22/03/2018) INFO - rcc_gui - Started run #40
(15:44:26 - 22/03/2018) INFO - rcc_gui - Stopped run #40
(09:11:26 - 23/03/2018) INFO - rcc_gui - Started run #41
test
(09:13:41 - 23/03/2018) INFO - rcc_gui - Stopped run #41
(09:13:48 - 23/03/2018) ERROR - rcc_gui - Operation forbidden - Storage mode can be changed only in READY state
(09:41:06 - 23/03/2018) INFO - rcc_gui - Started run #41
MM5 connected to MM8
MM6 connected to MM5
correct piedestal
3 pics alpha source
(14:02:04 - 23/03/2018) INFO - rcc_gui - Stopped run #41
(14:25:59 - 23/03/2018) INFO - rcc_gui - Started run #42
channel MM5 = real MM7
channel MM6 = real MM6
alpha source
(14:31:33 - 23/03/2018) INFO - rcc_gui - Stopped run #42
(14:33:42 - 23/03/2018) INFO - rcc_gui - Started run #43
same after change in the piedestals
(15:54:10 - 23/03/2018) INFO - rcc_gui - Stopped run #43 |
Fixed | e744 |
Mon Mar 26 14:20:22 2018 |
e744 : Valerian | MUST2 | Hardware | 3-alpha | Bidimensional graphs for Telescope 7 and 6 |
WARNING T5 == T7
EDIT 27/03/18 : WARNING X and Y are inverted, this entry should be ignored see entry : Bidim for T1 T3 T4 T5 T6 T7 T8 |
Attachment 1: T6_3alpha.pdf
|
|
Attachment 2: T7_3alpha.pdf
|
|
Ongoing | e744 |
Mon Mar 26 15:07:06 2018 |
e744 : Valerian et Louis | MUST2 | Hardware | 3-alpha | Energy vs Strip for T6 et T7 |
Weird signal on StripX of T6... might be the copper oxyde projections making a short-circuit
EDIT 27/03/18 : The problem seen on Telescope 6 might be due to either some Asics Problem, or due to some dirt, the current reason in not known currently. The safest way would be to remove the telescope completely and replace it for Daisuke's and Olivier's experiments. |
Attachment 1: StripXE_vs_StripNbr_T6.pdf
|
|
Attachment 2: StripXE_vs_StripNbr_T6_zoom.pdf
|
|
Attachment 3: StripYE_vs_StripNbr_T6.pdf
|
|
Ongoing | TEST |
Tue Mar 27 10:47:39 2018 |
e744 : Valerian | MUST2 | Hardware | 3-alpha | Choice of DSSSDs |
|
Ongoing | e744 |
Tue Mar 27 11:11:04 2018 |
e744 : Valerian | MUST2 | Hardware | 3-alpha | Choice of DSSSDs |
Different plots of Strip channel vs Strip number for X and Y in order to make our choice |
Attachment 1: StripXE_vs_StripNbr_T1.pdf
|
|
Attachment 2: StripYE_vs_StripNbr_T1.pdf
|
|
Attachment 3: StripXE_vs_StripNbr_T3.pdf
|
|
Attachment 4: StripYE_vs_StripNbr_T3.pdf
|
|
Attachment 5: StripXE_vs_StripNbr_T4.pdf
|
|
Attachment 6: StripYE_vs_StripNbr_T4.pdf
|
|
Attachment 7: StripXE_vs_StripNbr_T5.pdf
|
|
Attachment 8: StripYE_vs_StripNbr_T5.pdf
|
|
Attachment 9: StripXE_vs_StripNbr_T6.pdf
|
|
Attachment 10: StripYE_vs_StripNbr_T6.pdf
|
|
Attachment 11: StripXE_vs_StripNbr_T7.pdf
|
|
Attachment 12: StripYE_vs_StripNbr_T7.pdf
|
|
Attachment 13: StripXE_vs_StripNbr_T8.pdf
|
|
Attachment 14: StripYE_vs_StripNbr_T8.pdf
|
|
Fixed | e744 |
Tue Mar 27 16:26:29 2018 |
e744 : Valerian | MUST2 | General | 3-alpha | Bidim for T1 T3 T4 T5 T6 T7 T8 |
|
Attachment 1: StripX_vs_StripY_T1.pdf
|
|
Attachment 2: StripX_vs_StripY_T3.pdf
|
|
Attachment 3: StripX_vs_StripY_T4.pdf
|
|
Attachment 4: StripX_vs_StripY_T5.pdf
|
|
Attachment 5: StripX_vs_StripY_T6.pdf
|
|
Attachment 6: StripX_vs_StripY_T7.pdf
|
|
Attachment 7: StripX_vs_StripY_T8.pdf
|
|
Ongoing | e744 |
Tue Mar 27 17:22:24 2018 |
e744 : Valerian | MUST2 | Hardware | 3-alpha | Choosing the DSSSDs |
We can see that the DSSSDs on T1(CsI) T3(CsI) T4(CsI) and T5(SiLi) are in good shape (cf :Bidim for T1 T3 T4 T5 T6 T7 T8 ).
As for now, we would like to keep the telescope as is. M2C should be moved tomorrow (28/03/18).
After that we would like to use the two spare DSSSDs available on both T2 and on the O degree telescope (now called T5(CsI)).
If the spare DSSSDs are judged not be good enough for the experiment, we will swap current T5(SiLi) DSSDs on the 0 degree Telescope T5(CsI). |
Ongoing | e744 |
Wed Mar 28 13:21:32 2018 |
e744 : Valerian | MUST2 | Hardware | 3-alpha | Choosing the DSSSDs |
e744 : Valerian wrote: | We can see that the DSSSDs on T1(CsI) T3(CsI) T4(CsI) and T5(SiLi) are in good shape (cf :Bidim for T1 T3 T4 T5 T6 T7 T8 ).
As for now, we would like to keep the telescope as is. M2C should be moved tomorrow (28/03/18).
After that we would like to use the two spare DSSSDs available on both T2 and on the O degree telescope (now called T5(CsI)).
If the spare DSSSDs are judged not be good enough for the experiment, we will swap current T5(SiLi) DSSDs on the 0 degree Telescope T5(CsI). |
Which telescope should we take for T5? T7 or T6? |
Ongoing | e744 |
Wed Mar 28 16:35:01 2018 |
e744 : Valerian | MUST2 | Hardware | 3-alpha | Choosing the DSSSDs |
e744 : Valerian wrote: |
e744 : Valerian wrote: | We can see that the DSSSDs on T1(CsI) T3(CsI) T4(CsI) and T5(SiLi) are in good shape (cf :Bidim for T1 T3 T4 T5 T6 T7 T8 ).
As for now, we would like to keep the telescope as is. M2C should be moved tomorrow (28/03/18).
After that we would like to use the two spare DSSSDs available on both T2 and on the O degree telescope (now called T5(CsI)).
If the spare DSSSDs are judged not be good enough for the experiment, we will swap current T5(SiLi) DSSDs on the 0 degree Telescope T5(CsI). |
Which telescope should we take for T5? T7 or T6? |
Looking at the previous Calimero tests Adrien did during his experiment, neither telescope are good as is... the T7 has no time calibration and T6 might have some Asics problem...
We can either use T6 and check if the problem we saw on it's DSSDs are due to some electronics problem or use a completly different telescope from the spare we have...
EDIT 29/03/18 : After discussion with Laurent the option to use a new DSSD on T6 seems more appropriate, still ongoing. |
Attachment 1: TimeMust2X_MUVI2_uncalib.png
|
|
Attachment 2: TimeMust2Y_MUVI2_uncalib.png
|
|
Attachment 3: SpectresMUVI2stripXT300ns.png
|
|
Fixed | e744 |
Tue Apr 3 10:44:38 2018 |
e744 : Valérian | MUST2 | Hardware | | Update on Must2 telescopes |
For the next day, the plan is to check the DSSD's connections of T6 & T7 followed by an alpha run, if the problem seen on T6 are not resolved, we will change the DSSD of T6 and re-do another alpha run, the last option would be to change the Muffee card.
Just for a quick visual update, telescopes 5&8 are dismounted and a new DSSD is mounted on T2 as you can see ! |
Attachment 1: 20180403_103732.jpg
|
|
Ongoing | e744 |
Tue Apr 3 15:30:49 2018 |
e744 : Valérian | MUST2 | Hardware | | Cabling of T6 & T7 |
After checking the cable of T6 & T7 it seems that there were in fact an inversion during experiment e748. T6 is the bottom one, and T7 the top one.
During e748, it seems that there were some difference regarding the position, between the hardware and software position...
If it was the case, it would mean that it is T6 that has the time calibration problem, not T7.
re-
In conclusion, it is T6 that has both the time calibration problem on X and Y, and the weird signal seen during the calibration.
EDIT 4/04/18 : For the experiment e748 and before the time calibration, the inversion had been made, as a conclusion what was previously stated is wrong, it is T7 that has no time calibration... Anyhow, wee will re-do the time calibration for T6 and T7 to check if the problem are still present. |
Attachment 1: Inversion_between_telescopes.pdf
|
|
Ongoing | e744 |
Wed Apr 4 17:49:05 2018 |
e744 : Valérian | EXL-PPT | Hardware | | Update on EXL |
The EXL detector has been mounted, it is very close to the target ~4cm and doesn't do any shadow on the M2 telescopes. But it isn't at 90° anymore. The cable have been plugged-in (cf photos). With a new mapping of the detector (cf drawing). |
Attachment 1: 20180404_113303.jpg
|
|
Attachment 2: 20180404_113312.jpg
|
|
Attachment 3: 20180404_135324.jpg
|
|
Attachment 4: 20180404_135334.jpg
|
|
Attachment 5: EXL_cables.pdf
|
|
Attachment 6: Mapping_of_EXL.pdf
|
|
Fixed | e744 |
Tue Apr 10 17:43:28 2018 |
Adrien Matta and FyFy | MUST2 | Hardware | N/A | login to CAEN SY1527 HV/BT mainframe |
To control the CAEN SY1527 HV/BT mainframe use the following command from a terminal:
telnet 172.16.139.10 1527
OR
Run the alias HV
Login: admin
Password: admin
Press enter, then select channel.
To switch between on/off press space. |
Fixed | e744 |
Tue Apr 17 09:07:55 2018 |
Adrien Matta | DAQ | General | N/A | DAQ issue summary |
Previously on e744:
After a chat with Charles Houarner we decided to have configuration with a Muvi card per VXI crate. This is supposed to reduce the dead time (DT) by parallelising the crate readout. This have a few implication:
- The slave crate need a GMT, because the MUVI read out was never tested using a GAMER
- The GMT is not designed to act as centrum receiver, so a centrum slave module is installed in the slave crate
- The action file compilation is not designed to have a centrum receiver, so the compilation has to be done by Fred. He is working on fixing the issue, in the mean time, we SHOULD NOT try to change ADC/QDC/TDC label and recompile throught DAS
- The slave GMT needs to be block correctly, Charles setup the VIS2/BLK signal with a logic FIFO.
- The slave GMT is triggered by the master GMT VAL
- The MUVI from the Slave VXI is still triggering the MASTER GMT, as usual.
After all this setup, we change the NARVAL topology to have two crate and a EBYEDAT merger. This merger prove to be no longer functional. Clotilde had a look but have no clue so far why it is not working.
We switch to MFM data format with a EBY2MFM merger. Because of this change of format we had to change both Online and Offline program in order to extract the EBYE buffer from the MFM frame.
This is no tested and working, for some reasons it worked only using ROOT6. |
Fixed | e744 |
Tue Apr 17 09:18:00 2018 |
Cherry MX team | MUST2 | Hardware | N/A | T5 is back and running strong |
T5 telescope appears to be fully functional again. It shows both Energy and Time with excellent resolution. |
Attachment 1: T5.pdf
|
|
Ongoing | e744 |
Tue Apr 17 17:36:37 2018 |
Cherry Mx Team | MUST2 | Software | N/A | GRU SUCKS |
Pretty much today's summary in the title.
https://i.pinimg.com/originals/a8/d4/12/a8d41206e8bd9c189397fb4907656981.jpg |
Fixed | e744 |
Mon Apr 23 11:33:53 2018 |
e744 Valerian | OTHER | Hardware | N/A | Inspection signals |
Inspection signals between D6 and LISE Acquisition room are connected as follow :
178: FAG
179: LT ADC 1
180: LI1 ADC 1
181: AI ADC 1
182: LT ADC 2
183: LI1 ADC 2
184: AI ADC 2
185: HF validated
186: CATS delayed
188: FAG delayed
201: IL1 MUVI 1
202: IA1 MUVI 1
203: IL1 MUVI 2
204: IA1 MUVI 2 |
Pending | e744 |
Tue Apr 24 14:18:22 2018 |
Valerian Freddy and Adrien aka Cherry Mx Team | MUST2 | Hardware | N/A | Current telescope-pcb layout |
Telescope 1 : PCB 5 (ref on pcb: 2274)
Telescope 2 : PCB 4
Telescope 3 : PCB 7
Telescope 4 : PCB 11 (ref on pcb: n°2834-9)
Telescope 5 : PCB 2 (ref on pcb: X2489-8) Backward angles
Telescope 6 : PCB 6 around 90d eg.
------------------------ ------------------------
/ \
/ \
4 / \ 1
/ \
/ \
/ \
/ \
----------------- -----------------
----------------- x -----------------
\ BEAM /
\ /
\ /
3 \ / 2
\ /
\ /
\ /
------------------------ ------------------------ |
Fixed | e744 |
Tue Apr 24 15:42:03 2018 |
e744 Valérie L. | DAQ | General | | GMT chronogramme |
Crate ganlx1 GMT master FAG 100ns ; VAL 2 µs ; CT 6 µs
Crate ganlx14 GMT slave FAG 100ns ; VAL 2 µs ; CT 3 µs |
Attachment 1: GMTmasterChrono24avril18.png
|
|
Attachment 2: GMTslaveChrono24avril18.png
|
|
Fixed | e744 |
Tue Apr 24 15:51:04 2018 |
e744 Valérie L. | MUST2 | General | | MUST2 configuration /parameters |
TAC range for all telescopes: 600ns
Thresholds: 402 keV |
Attachment 1: ConfigMuviPhysMust2Thr.png
|
|