MUST2 SAMURAI_2018 S3-LEB-LPC SUPERNEMO MUGAST EXPAND SCALP GALATRON HiCARI VELO
  MUST2, Page 6 of 48  Not logged in MUST2
Status ExpNbr Date Author Type Category Target-Source Subject
Fixede748 Sun Jul 2 05:45:18 2017 e748CATSSoftwareN/AMask CATS 2
Run 75

shift surveyor
CATS_SHIFT[2][2]={{0.3,0.8},{0.7,0.2}} ie (X1,Y1)(X2,Y2) -> to be adjusted in more detail
TARGET_SHIFT[2]={-0.2,0.0}

List of inversion : X -> -X on CATS1
Attachment 1: Run75MaskCATS2.png
Run75MaskCATS2.png
Fixede748 Sun Jul 2 05:56:12 2017 e748CATSSoftwareN/AMask on target
Run 74 Mask on target

With surveyor offsets and the inversion X -> -X on CATS1 (determined with run 75), 2 holes are observed. see figure Run74MaskTarget.png
Need to scan the mask with the beam.

Play with offset to fit with the hole which is at zero (we think at zero, has to be confirmed with a new run scanning all the target with beam) -> new set of offsets X1=+0.3mm Y1=+0.8mm X2=+0.3mm Y2=0mm
See figure Run74MaskTarget_Offsetmv.png
Attachment 1: Run74MaskTarget.png
Run74MaskTarget.png
Attachment 2: Run74MaskTarget_Offsetmv.png
Run74MaskTarget_Offsetmv.png
Fixede748 Sun Jul 2 08:06:53 2017 Valérie LapouxRUNGeneralCD2Run 77 global check
Conditions of Run 77

2 July - start at 6h29 - stop at 8h38
e748 Exxx 126943232 2 juil. 08:38 run_0077.dat.02-07-17_06h29m23s

N.B: The run is stopped at 8h38 to check the bidim of correlations TPL_CATS1 coded in VXI MUST2 and VXI CHIO.
GMT Triggers all MUST2 T1 to T8
----------------------------------------------------------------------------------------------------
CD2 2.5 mg/cm2 POSITION 3
CATS rate ~ 10 K/s to 12K /s; SCaler Analyzer see plot below
Muvi1 and Muvi2 counting rates: plots below

The merge configuration is checked with the parameters CONFDEV GATCONF etc...see plot below at 8h00
----------------------------------------------------------------------------------------------------
Attachment 1: CheckMergeConf2july17at8h00.png
CheckMergeConf2july17at8h00.png
Attachment 2: RunControlOnlineNarvalRun77.png
RunControlOnlineNarvalRun77.png
Attachment 3: ScalerAnalyzerRun77At8h15.png
ScalerAnalyzerRun77At8h15.png
Attachment 4: ScalerMuvi1Run77.png
ScalerMuvi1Run77.png
Attachment 5: ScalerMuvi2Run77.png
ScalerMuvi2Run77.png
Attachment 6: GMTConfigRun77.png
GMTConfigRun77.png
Fixede748 Sun Jul 2 08:40:25 2017 Valérie LapouxRUNGeneralCD2Run 78 - 2 July - 18O beam
Run #78 CD2 2.5 mg/cm2 position 3 - 18O beam
Same conditions as for Run 77
Triggers all MUST2, T1 to T8, GMT IN1 to IN8
RUN 78 2nd July START at 8h39, STOP 10h01
e748 Exxx 92405760 2 juil. 10:01 run_0078.dat.02-07-17_08h39m30s

STOP of the beam and of the run to enter in D6, to make some checking of electronics and of the timing.
Attachment 1: Run78ConfdevDataTrig.png
Run78ConfdevDataTrig.png
Attachment 2: RUN78tPLcats1CORREL.png
RUN78tPLcats1CORREL.png
Attachment 3: Run78CorrelWindowtimePlastCats1.png
Run78CorrelWindowtimePlastCats1.png
Attachment 4: Run78TACplastCats.png
Run78TACplastCats.png
Attachment 5: Run78TrigMM2impact.png
Run78TrigMM2impact.png
Fixede748 Sun Jul 2 08:46:13 2017 Valérie LapouxDAQHardwareCD2Summary of the TAC conditions
JULY 2nd 8pm

Bidiv = 1 for CATS1 and 2
Bidiv = 1 for Plast and ChIo

TAC ranges

START STOP RANGE PARAMETER ADDRESS VXI MODULE
-------------------------------------------------------------------------------------------
-------------------------------------------------------------------------------------------
CATS1_FAG CAVIAR (del 550ns) 1µs T_CATS1_CAV ADC VXI MUST2 add=19
CATS1_FAG CATS2_FAG (del 33ns) 200ns T_CATS1_2 ADC VXI MUST2 add=28
OU_MUVI CATS1_FAG 500ns T_MUVI_CATS1 ADC VXI MUST2 add=21
CATS2_FAG PLAST (val) 1µs T_PL_CATS2 ADC VXI MUST2 add=22
CATS1_FAG PLAST (no val) 1µs T_PL_CATS1 ADC VXI MUST2 add=23
-------------------------------------------------------------------------------------------
CATS1_FAG PLAST (no val) 1µs T_PLch_CATS1 ADC VXI CHIO add=19
PLAST val Gamer ChIo CHIO (no val) 2µs T_PL_CHIO ADC VXI CHIO add=3
-------------------------------------------------------------------------------------------
TAC EXL HF 100ns EXL_HF ADC VXI MUST2 add=18
-------------------------------------------------------------------------------------------
Fixede748 Sun Jul 2 09:54:05 2017 Valérie LapouxDAQGeneralCD2Test of the TAC plast CATS ADC VXI Run78
Attachment 1: InspectLinesVXIMust2Run78LTtAiCatsFAG.png
InspectLinesVXIMust2Run78LTtAiCatsFAG.png
Attachment 2: InspectVXIchioADCsignalTplchCATS1Run78.png
InspectVXIchioADCsignalTplchCATS1Run78.png
Fixede748 Sun Jul 2 10:31:48 2017 Valérie LapouxONLINEHardwareCD2Change of the electronics for TAC Plast CATS1
Shift U (Béa +Vlx) entered in the room to check the electronics of Cats1 and of the TAC plast-Cats1
Now we have taken the output of one of the TAC : START=CATS1_FAG - STOP plast (no val)
We send it to a Linear Fifo from which it is duplicated for the ADC channels:
-ADC of VXI MUST2 parameter T_PL_CATS1 ; add 23
-ADC of VXI CHIO parameter T_PLch_CATS1; add 19
NB:
Now we put 50 Ohms everywhere in the logic Fifo for Cats1 Validated in the CATS electronics.
Fixede748 Sun Jul 2 10:44:59 2017 Valérie LapouxRUNGeneralCD2Run 79 Beam of 18O
run_0079.dat.02-07-17_10h37m20s
Triggers : all MUST2 telescopes 1 to 8.
The pulser is ON again in the DAQ and Scaler Analyzer. See plots.
The pulser is taken from the GAMER of the VXI CHIO.
For the correlations the situation is now improved (no multiple peaks in Tplast_CATS1)
but a lot of 0 in the TAC spectra. Low counting rates in the spectra between Plast and CATS, which are obtained
with the MUST2 triggers. We need to check the correlations between CATS and MUST2.

For all the runs the validations of the CATS was taken using the signal from the GAMER of VXI CATS, the sequenceur.
See below the configuration of "SEQUENCEUR".
Attachment 1: ScalersAnalyzerRun79.png
ScalersAnalyzerRun79.png
Attachment 2: PulserONforRUn79GAMERchiO.png
PulserONforRUn79GAMERchiO.png
Attachment 3: ScalerAnalyzerRun79OkPulser.png
ScalerAnalyzerRun79OkPulser.png
Attachment 4: Run79CorrelPlastCATS1vxiChioAndMUST2.png
Run79CorrelPlastCATS1vxiChioAndMUST2.png
Attachment 5: Run79CorrelTPlastCats1.png
Run79CorrelTPlastCats1.png
Attachment 6: DASconfigSequenceurCATS.png
DASconfigSequenceurCATS.png
Attachment 7: ADCvxiMust2configPAramTAC.png
ADCvxiMust2configPAramTAC.png
Attachment 8: ConfigADCvxiChioRun79.png
ConfigADCvxiChioRun79.png
Fixede748 Sun Jul 2 11:38:03 2017 Valérie LapouxRUNGeneralCD2Runs 80-81 TESTS change of TAC and triggers
For runs 80- 81:
for test purposes, we changed the coincidence mode for CATS: we put FAG instead of the Seq1 coming from the gamer.
Run 80 start 11h33; stop 11h40
e748 Exxx 15794176 2 juil. 11:40 run_0080.dat.02-07-17_11h33m07s
CD2 2.5 mg/cm2
change of the trigger conditions only CATS1 div =100
(all bidiv at 100)
-------------------------------------------------------------------------------------------
Run 81 start 11h42; stop 11h48 triggers IN1 to IN8, all Must2
e748 Exxx 4128768 2 juil. 11:48 run_0081.dat.02-07-17_11h42m27s
-------------------------------------------------------------------------------------------

CONDITIONS FOR TAC:
START STOP RANGE PARAMETER ADDRESS VXI MODULE
-------------------------------------------------------------------------------------------
CATS1_FAG CAVIAR (del 550ns) 1µs T_CATS1_CAV ADC VXI MUST2 add=19
CATS1_FAG CATS2_FAG (del 33ns) 200ns T_CATS1_2 ADC VXI MUST2 add=28
OU_MUVI CATS1_FAG 500ns T_MUVI_CATS1 ADC VXI MUST2 add=21
CATS2_FAG PLAST (val) 200ns T_PL_CATS2 ADC VXI MUST2 add=22
CATS1_FAG PLAST (no val) 200 ns T_PL_CATS1 ADC VXI MUST2 add=23
-------------------------------------------------------------------------------------------
CATS1_FAG PLAST (no val) 200ns T_PLch_CATS1 ADC VXI CHIO add=19
PLAST val Gamer ChIo CHIO (no val) 2µs T_PL_CHIO ADC VXI CHIO add=3
-------------------------------------------------------------------------------------------
TAC EXL HF 200 ns EXL_HF ADC VXI MUST2 add=18
Fixede748 Sun Jul 2 12:08:56 2017 Valérie LapouxRUNGeneralCD2Run 82
Run82 start at 12h00 stop at 12h09
e748 Exxx 6684672 2 juil. 12:09 run_0082.dat.02-07-17_12h00m43s
Before starting run 82, we enter in the D6 room to restaure the electronics for CATS as it was before
(back to the situation of SEQ1 for the coincidence in CATS electronics, as before run 80)
triggers all MUST2
CD2 target
counting 10-12 k/s on CATS
CATS 1 and 2 : HV 650V
Fixede748 Sun Jul 2 12:16:57 2017 Valérie LapouxRUNGeneralCD2Runs 83-84 only CATS1 trigger
Run83 start at 12h11 stop at 12h16
e748 Exxx 21561344 2 juil. 12:16 run_0083.dat.02-07-17_12h11m16s

----------------------------------------------
Run83 start at 12h24 stop at

---------------------------------------------
trigger: ONLY Cats1 div (div=100)
CD2 target 2.5 mg/cm2
counting 10-12 k/s on CATS
CATS 1 and 2 : HV 650V
----------------------------------------------
Attachment 1: ScalersRun84Cats1trigg.png
ScalersRun84Cats1trigg.png
Attachment 2: ConfigTRigCorrelMErgeRun84.png
ConfigTRigCorrelMErgeRun84.png
Fixede748 Sun Jul 2 13:22:46 2017 Valérie LapouxRUNGeneralCD2Run 85+summary TAC
Started run #85 at 13h21 stop at 13h31
e748 Exxx 5439488 2 juil. 13:31 run_0085.dat.02-07-17_13h21m57s
Triggers all MUST2
CD2 target 2.5 mg/cm2
18O beam
NOW (Run 85) Plast DEL 200ns instead of 500 ns (for previous runs)
signal for all the TAC with CATS
Range of TAC 1µs for T_PL_CATS2, T_PL_CATS1
(same for signal T_PLch_CATS1 duplicated from T_PL_CATS1 and coded in VXI CHIO)
START STOP RANGE PARAMETER ADDRESS VXI MODULE
-------------------------------------------------------------------------------------------
-------------------------------------------------------------------------------------------
CATS1_FAG CAVIAR (del 550ns) 1µs T_CATS1_CAV ADC VXI MUST2 add=19
CATS1_FAG CATS2_FAG (del 33ns) 200ns T_CATS1_2 ADC VXI MUST2 add=28
OU_MUVI CATS1_FAG 500ns T_MUVI_CATS1 ADC VXI MUST2 add=21
CATS2_FAG PLAST (no val) 1µs T_PL_CATS2 ADC VXI MUST2 add=22
CATS1_FAG PLAST (no val) 1µs T_PL_CATS1 ADC VXI MUST2 add=23

!!!NOW PLAST is delayed with 200 ns (DGG) previously it was 500ns
-------------------------------------------------------------------------------------------
CATS1_FAG PLAST (no val) 1µs T_PLch_CATS1 ADC VXI CHIO add=19
PLAST val Gamer ChIo CHIO (no val) 2µs T_PL_CHIO ADC VXI CHIO add=3
-------------------------------------------------------------------------------------------
TAC EXL HF 100ns EXL_HF ADC VXI MUST2 add=18
Fixede748 Sun Jul 2 13:31:21 2017 Valérie LapouxRUNGeneralCD2Run 86 trigger CATS1
Run 86 start at 13h32 stop at 13h43
e748 Exxx 34340864 2 juil. 13:43 run_0086.dat.02-07-17_13h32m13s

TRIGGER CATS1 div =100 only

Summary of TAC plots
Attachment 1: CorrelRun86mergeOKruncontrol.png
CorrelRun86mergeOKruncontrol.png
Attachment 2: Run86TacPlastCATStrigCats1.png
Run86TacPlastCATStrigCats1.png
Fixede748 Sun Jul 2 13:49:20 2017 Valérie LapouxRUNGeneralCD2Run 88 trigger CATS1
Run 88 start at 13h48
TRIGGER CATS1 div =100 only
18O beam CD2 target

Fixede748 Sun Jul 2 14:49:41 2017 Valérie LapouxONLINEHardwareCD2TAC and signals for Plast-CATS1
Now we return to the situation of 2 TACs for the coding of the same time Plast-Cats1 in the 2 DAQs systems VXI_CHIO and VXI_MUST2.

TAC1 start CATS1val; STOP Plast (delayed by 200ns with DGG) range 1µs OUTPUT signal T_PL_CATS1 VXI_MUST2 ADC add=23
TAC1 start CATS1val; STOP Plast (delayed by 200ns with DGG) range 1µs OUTPUT signal T_PLch_CATS1 VXI_ChIo aADC_CHIO add=19
Fixede748 Sun Jul 2 14:54:39 2017 Valérie LapouxRUNGeneral Runs 90-92 Tuning of the cocktail beam in D4
Runs 90 to 92 Tuning of LISE
GMT Trigger IN13=Trig_LISE (D4)

e748 Exxx 15859712 2 juil. 14:51 run_0090.dat.02-07-17_14h47m41s
e748 Exxx 39714816 2 juil. 14:57 run_0091.dat.02-07-17_14h52m38s
e748 Exxx 4390912 2 juil. 15:04 run_0092.dat.02-07-17_14h59m19s
Fixede748 Sun Jul 2 15:13:49 2017 Valérie LapouxRUNGeneral Runs 93-94 Tuning of the cocktail beam in D6
Runs 93-94 Tuning of LISE
GMT Trigger IN13=Trig_LISE (=D6)
e748 Exxx 25493504 2 juil. 15:11 run_0093.dat.02-07-17_15h07m01s
e748 Exxx 15269888 2 juil. 15:14 run_0094.dat.02-07-17_15h11m38s

Attachment 1: ScalerAnalyzerCaviar15h29.png
ScalerAnalyzerCaviar15h29.png
Fixede748 Sun Jul 2 17:25:22 2017 Valérie LapouxDAQHardwareCD2ADC parameters-new ones for LISE +Pulser for Correlations
New parameter for LISE in VXI_CHIO ADC_CHIO
parameter TED4HF address=4 channel 5 (cable label "5" on connected, first branch of ADC_CHIO)
-------------------------------------------------------------------------------------
For the tests of the correlations between VXI CHIO and VXI MUST2:
- IN14 input GMT of the pulser
- Parameters
VXI_CHIO
ADC_CHIO_V15 channel 16, add=15

VXI_MUST2 ADC (4th part of 8 ch)
ADC_VOIE_29 channel 28, add=29
---------------------------------------------------------
Attachment 1: DASVXIchioAdcLiseParam.png
DASVXIchioAdcLiseParam.png
Attachment 2: ConfigDASnew2ndJulyVXIMust2.png
ConfigDASnew2ndJulyVXIMust2.png
Fixede748 Sun Jul 2 18:19:24 2017 Valérie LapouxDAQHardwareCD2Change of DAQ parameters for LISE
Attachment 1: DASparamLISEVXIChioNEW.png
DASparamLISEVXIChioNEW.png
Fixede748 Sun Jul 2 18:27:48 2017 Valérie LapouxONLINESoftware Narval configuration, Watcher ports
New file xml is loaded with various Watchers, the corresponding ports
Watcher 10205
Watcher WatRaw 10206
Watcher WatNPTool 10208
Watcher WatCalib 10207
Watcher WatLISE 10209
Attachment 1: RunControlRun101allwtch.png
RunControlRun101allwtch.png

ELOG Home Page