Ongoing, e744, Sun May 13 19:36:09 2018, Valérian + François, MUST2, General, 3-alpha, Run #116 #117 3 alpha run Calibration T1-T5 trigger 
|
[U][B]run #116[/B][/U]
start @12h07 13/05/18
stop @19h45 13/05/18
|
Fixed, e748, Thu Jun 22 11:10:35 2017, Valérian, EXL, Software, , EXL Mapping
|
Mapping of the EXL detector in the analysis program versus how it is written on the cables. |
Fixed, e748, Mon Jun 26 10:26:16 2017, Valérian, EXL, Software, 60Co, Add-back
|
We implemented the add-back correction with cross-talk subtraction.
On the joined graph, RawEnergy correspond to the sum of the energy in every crystals and Energy correspond to the sum of energy in every crystals with |
Fixed, e744, Wed Apr 25 15:59:51 2018, Valérian, EXL-PPT, Software, N/A, Mesytech setup
|
To use the remote settings of the Mesytech modules :
[CODE]
- start GANP170
|
Ongoing, e744, Tue May 15 09:58:52 2018, Valérian, RUN, General, 3-alpha, run #118 3 alpha
|
[U][B]run #118[/B][/U]
Trigger T1 T2 T3 T4 & T5
EXL + BT not biased but still in trigger
|
Fixed, e744, Tue May 15 19:42:40 2018, Valérian, EXL-PPT, General, 60Co, Calibration of EXL from e748 data
|
The EXL calibration program is ready. Join is the result with previous data from e748. |
Fixed, e744, Fri Jun 1 13:12:16 2018, Valérian, EXL-PPT, General, 60Co, run #3003 Calib EXL 
|
run #3003
start @12h06
stop @13h08 |
Fixed, e744, Fri Jun 1 19:29:17 2018, Valérian, EXL-PPT, General, 60Co, EXL efficiency
|
To test EXL efficiency I start a long run with the source in front of it.
run 3014
start @19h25
|
Ongoing, e748, Tue Jun 6 17:57:07 2017, Valérian et François, EXL, Hardware, , Hardware coupling of EXL with MUST2
|
EXL seems to fit in Adrien's setup. We still have to check if it fits in the chamber. |
Ongoing, e744, Sat May 26 07:32:17 2018, Valerian, pedja, mihai, RUN, General, CH2, Run 2015 after vacuum problem of Css1 target 4
|
[U][B]Run 2015[/B][/U]
26/05/18
start @7h29
|
Ongoing, e744, Wed May 30 11:17:00 2018, Valerian maurycy and Olivier, OFFLINE, General, CH2, Condition on time T5
|
One should add the condition Si_TX < 150 for T5, problem from Ta? |
Ongoing, e744, Wed May 30 11:22:16 2018, Valerian maurycy and Olivier, OFFLINE, General, CH2, Checking the cuts
|
cut_DeE :
- T1 OK
|
Ongoing, e744, Mon May 28 20:28:50 2018, Valerian and Simon, OFFLINE, General, CH2, Explanation for the 0 degree low energy proton
|
The 0 degree energy proton are coming from the beam hitting the aluminum target. If you do a cut on the beam projection on T5 you obtain the black dot on
the joined plot. |
Fixed, e744, Mon May 28 11:00:07 2018, Valerian and Iulian and Phil, RUN, General, , two protons
|
|
Ongoing, e744, Mon May 28 23:14:26 2018, Valerian and Iulian and Phil, OFFLINE, General, CH2, New hint of the 5/2- resonance
|
We see a new hint of a resonance |
Ongoing, e744, Sat May 26 07:03:12 2018, Valerian and Iulian and Francois, OFFLINE, General, CH2, First picture for 15F from runs 2010-2013 
|
Target -> Thick target (Sandwich target ??)
The predictions are different....
|
Ongoing, e744, Sat Apr 28 10:17:22 2018, Valerian and Iulian, MUST2, Hardware, 3-alpha, Status of MUST2 telescopes 27/04/18
|
Yesterday night we tested T1, T2, T3, T4, T5, and T6.
All telescopes (except T6 which still have a "tripping" CsI) have an Energy and Time signal on X and Y!
|
Ongoing, e744, Sat Apr 28 13:27:26 2018, Valerian and Iulian, MUST2, Hardware, 3-alpha, multiples runs with two alpha source on T1 T2 T3 T4 and T5
|
For the following runs we will use two 3 alpha sources :
On T1, T2, T3, and T4 : [B]11615[/B] : Flux = 1496 alpha in 2 pi
|
Ongoing, e744, Sat Apr 28 13:36:52 2018, Valerian and Iulian, OTHER, Software, N/A, Problem with DAS configuration
|
There is an issue with the GMT slave. When you click Online and then apply write hardware, if the message : [U]register error on GMTSlave[/U] appear, [B]reload
a functioning DAS config[/B] and [B]do not write hardware twice[/B].
|
Ongoing, e744, Fri May 25 02:57:44 2018, Valerian and Iulian, OFFLINE, General, CH2, Simulation Azure 1H(14N,p) and measurement    
|
We can know conclude without any doubt that we were able to measure resonances from 1H(14N,p)14N. The resolution with a rough analysis and the current bad
calibration is around 180keV in the com.
Identification of resonances is still difficult but some structure seems consistent with previous works. Also, at higher energy there are only few measurement |
|