Fixed, e744, Wed May 23 15:42:04 2018, Louis, Valérie, MUST2, General, N/A, CALIMERO run 409 - XT - T1 to T4    
|
Started run #409
GMT triggers: GecoCalimeroMuviGMT.png
HT on cf : HTtelescopesMustsWed23May.png
Calimero TC 640ns; period 20ns
XT for T1 to T4 ampli pulser 12MeV
thresholds for telescopes T1 to T3, 400 keV except T4: Threshold 1MeV
cf GecoCalimeroMuviTel4Thr1MeV.png
problem of configuration:
cf configCalim1.png trigger must be external on the calimero pulser interface.
cf good one on the configCalimOkexterne.png (to be done for the next runs) |
Pinned, e744, Wed May 23 17:23:34 2018, Freddy, MUST2, General, , Pedestals and alpha calibration analysis
|
Calibration coefficients for MUST2 telescopes were extracted from run_0118 (acquired during night from 13th to 14th of May). They are now located in the folder:
/home/e748/analysis/nptool/Projects/e744/Calibration/Energy/Calibration/Summary_run_0118
Based on the analysis of this run we realized that:
- T4_StripX had pedestal normalization issue (for MATE 1,3,5,7) and alpha peaks were enlarged or doubled.
- Pedestals used until today (23rd of May) were based on the pedestal measurement of the 28th of April !
Due to that, we remeasured the raw values of the pedestals today and compared them with the one used up to now.
The differences (in channels) between these two pedestal values for each telescope are plotted on the attached figure (X from 1-128 and Y after).
The drifts can be partially understood by the fact that the conditions evolved a lot during one month (cooling system temperature, etc).
For the online analysis, calibration of run_0118 (13/05) will be used but they somewhat lie in time between the two pedestal measurement (28/04 and 23/05).
However it is important right after the experiment to perform a new alpha calibration in conditions much closer to the measurement. |
Fixed, e744, Wed May 23 18:20:59 2018, e744, MUST2, General, N/A, Test Calimero Run 425
|
Started run #425 Calimero NEW XT T1toT4
ampl Pulser 12MEV; thr for T4 1 MeV
problem on data taking |
Ongoing, e744, Wed May 23 20:52:01 2018, Iulian Valerian, RUN, General, , run #429 and #430 and config gmt    
|
Changed VAL from 2 us to 2.5 us
run #429 exl ppt triggers
run #430 exl ppt triggers. Changed val on gmtslave also
The correlation between MUST and PPT (run430_E_PPT_vs_T_PPT.pdf) is due to the beta decay of 138La (internal contamination in the LaBr crystal). PPT detects the beta energy with 100% eff, while the gamma rays - emitted in the beta decay process- interacted in MUST detectors. The EXL bidim shows only random events (no internal contamination). |
Fixed, e744, Wed May 23 23:37:14 2018, Valérie L., MUST2, General, N/A, MUVI configuration DAS 13x
|
Configurations files of the MUVI 1 MufeeX and Y.
On MufeeX and Y configuration:
Note that the thresholds on Must2 T1 to T3 is at 400 keV on May 23th Midnight (same as before); and for T4, now it is set at 1 MeV (not 400 keV).
On Muvi/CAS configuration :
+ suppression of Energy and time pedestals, see the page in two snapshot files, ConfigMANIPMUVI2CASTel1A.png ; ConfigMANIPMUVI1CASTelB.png
(all green; in contrast with time calib configuration :red squares cf for instance : configCalimMUVI2CASTel8.png)
+ ORDi
checking temperature on the Mufee MUST2 cards:
example of good temperature on card (around 4 to 8°C when the cooling sytem is set at -4°C) : ConfigDASMuvi1Tel5mufeeXokTemp.png
Example of temperature problem on card (cooling system not running correctly, bad situation on April 25th) : Problem25AprilTemptTel4mufeeX.png
Configuration of the modules of the second Slave crate which includes MUVI2, GMT slave and Centrum slave:
Must2Crate2GMTslave.png ; MUVIcrate2ConfigCentrumSlave.png |
Fixed, e744, Thu May 24 01:02:19 2018, Valerian, OTHER, General, N/A, run #431
|
run #431
Trigger LISE
Beam in D4
start @ 00h36
stop @ 1h34 |
Fixed, e744, Thu May 24 01:10:30 2018, iuliqn, RUN, General, , run 433 Lise trigger
|
Should be ql foil in front of the si detector.... |
Fixed, e744, Thu May 24 01:11:26 2018, Iulian Valerian, RUN, General, N/A, run 434 Lise trigger 
|
Cats inserted
efficiency is around ~99% |
Fixed, e744, Thu May 24 01:53:28 2018, iulian, RUN, General, N/A, run 438 Lise trigger 
|
run 438 D4 in
run 439 d6 in.
D4 D6 transmision with CATS in:
83425./106149.=7.85923560278476430e-01 |
Ongoing, e744, Thu May 24 02:30:52 2018, iulian, RUN, General, N/A, RUN 440 14N no target trigger Must & cats %100
|
RUN 440
14N beam \
no target
trigger t1-t5
cats %100
some counts in t5. Very low transmission . max 30 while in cats we had 500 |
Fixed, e744, Thu May 24 03:21:37 2018, iulian, OTHER, Hardware, N/A, Position measurements Must2/PPT/EXL
|
|
Fixed, e744, Thu May 24 03:57:34 2018, iulian, RUN, General, N/A, run 442 Lise trigger d6
|
|
Fixed, e744, Thu May 24 04:20:34 2018, iulian, RUN, General, N/A, run 443 T1-t5 triggers   
|
run 443
T1-T5 triggers
still problems to recover the counting on must2. Cats: 5000 pps , T5: 50 pps max
Below changes in tuning with little change in counting...
The energy for 14N7+ at the entrance of must2 should be 28.44 meV if the beam dump thickness is correct (34 um). If a dead layer of 0.5 um of Al is included, the energy became 27.849. This looks very close to what we see on X strip, and about 1 MeV off on Y strips. |
Pinned, e744, Thu May 24 05:44:56 2018, iulian, RUN, General, N/A, Important::analysis of the run 443/ things to do   
|
No stop on t5, but TAC MM_CATS ok.. So two possibilities:
1) Stop Cable not connected on Muvi2
2) Problem with the validation of CATS by FAG
HF validated by FAG was not working
Dead time scaler was not working
We need to erase all the scalers at the beginning of the run, including the part from Muvi (ORD, DESC, etc). Atm we erase only the U2M not the other part
The big problem is the transmission between LISE silicon detector D6 and T5 must with CATS on or off. 3000 counts/s on silicon d6 or CATS -> maximum 300 on T5. Is there a way to miss counting in Must2?
TAC_CATS_HF ok
TAC_MM_CATS ok
T_MM_EXLor present with prompt from cosmics?!
T_MM_HF ok
T_MM_HFv not present |
Ongoing, e744, Thu May 24 05:59:44 2018, iulian, MUST2, General, N/A, run 444 t1-t4 trigger no target 14n beam    
|
run 444
t1-t4 trigger
no target
14n beam |
Ongoing, e744, Thu May 24 06:30:27 2018, iulian, RUN, General, N/A, run 445 t1-t4 trigger no target 14n beam no CATS  
|
|
Pending, e744, Thu May 24 06:53:12 2018, e748, RUN, General, N/A, run #446 CATS trigger
|
|
Fixed, e744, Thu May 24 12:01:54 2018, Valérie for Calimero team - Valérian, Fairouz, MUST2, General, , Bug server Calimero Run #447-449 12x
|
Calimero on ganp161
Telscopes T1 to T4 thresholds for T1 to T3 400 keV ; T4 : 1 MeV
ampli pulser 12 MeV
problem: stop calimero at 6%
Bad connection. |
Fixed, e744, Thu May 24 17:18:58 2018, Valérie for Calimero team : Valérian, Louis, MUST2, General, N/A, CALIMERO Tests Runs 451 to 452 (done in D6)  
|
Test runs for telescopes T1 to t4
Nb: The applied threshold is now 1 MeV but it seems to be not the case when we look at the spectra.
Maybe a problem of the read configuration in GECO.
Run 451 XT Pulser amplitude 9 MeV
Run 452 YT
NB: good configuration files are given in the next entry
See Notes signed by Valérie, checked by the team ! |
Fixed, e744, Thu May 24 17:23:57 2018, Valérie for Calimero team: Valérian, Louis, MUST2, General, N/A, CALIMERO Runs 453-457 OK 16x
|
Calimero on ganp1074 in D6
Time Calibrator Range: 640 ns; Period 20 ns Amplitude of pulser (test generator) 9MeV
Run 453 Telescope T5 (MUVI2-slave) XT see plot
Run 454 Telescope T5 (MUVI2-slave) YT blocked at 6% -->
NB: stop calimero, stop run, stop geco, load das (to have good config for Mate), reload Geco
(usual steps: Restart Calimero programs, start Run control, start Calimero in geco)
Run 455 Telescope T5 (MUVI2-slave) YT ok
Then stop Run, exit from Geco, load DAS, change configuration for T1 to T4.
Change configuration of Port in RCC Narval. see plot
Run 456 Telescopes T1 to T4 (MUVI1-Master) XT ok see plot
Run 457 Telescopes T1 to T4 (MUVI1-Master) YT ok see plot |
|