MUST2 SAMURAI_2018 S3-LEB-LPC SUPERNEMO MUGAST EXPAND SCALP GALATRON HiCARI VELO
  MUGAST, Page 13 of 132  Not logged in MUGAST
Status ExpNbr Date Authordown Type Category Target-Source Subject Text Attachments
Pinnedcom2019 Thu Apr 4 14:45:53 2019 ValerianOTHERHardwareN/ASurveyor's measurementsSummary of the Surveyors. Joined the pdf and
the Excel file
 GANIL-0057_E744_I-Stefan.pdfE744_XYZ_détecteurs_Si.xlsx 
Fixedcom2019 Sun Apr 7 14:24:18 2019 ValerianOFFLINESoftwareCD2Ex and E gamma with all statistic (run 256+257)   bilan_run256_and_257_Ex_Egamma.png 
Fixede744s Mon Apr 8 11:58:37 2019 ValerianOTHERGeneralN/AParticle identification with CATS or HF After discussing about using the HF, I join
as a reference what was obtained during experiment
e744 of last year. We can see that using
 6x 
Fixedcom2019 Tue Apr 9 10:44:40 2019 ValerianMUST2SoftwareCD2CsI Calib files from the commissioning for protonCsI calibration obtained from the proton data
of the commissioning of mugast.
 Cal_CsI_E_MM1.calCal_CsI_E_MM2.calCal_CsI_E_MM3.calCal_CsI_E_MM4.cal 
Fixede744s Thu Apr 11 11:36:09 2019 ValerianOFFLINESoftwareCH2Comparison between the two e744Current calibration is off, a shift of -1.5MeV
was done to reproduce the previous data.
 current_exp_old_e744.png 
Fixede744s Thu Apr 11 13:43:26 2019 ValerianOFFLINESoftwareCH2Target position for 14N runs   XY_run14N.pngXshift_run14N.pngYshift_run14N.png 
Fixede744s Thu Apr 11 14:03:45 2019 ValerianCATSHardwareN/AIn case of a CATS2 trippedCurrently, CATS2 is inserted, if the window
:
  
Fixede744s Fri Apr 12 10:12:45 2019 ValerianOFFLINESoftwareCH2Comparison between HF and CATS for ELab ThetaLabUsing the HF show noises that is not seen
with CATS on the ToF spectra
 ELabThetaLabrun87_E744_HFvsCATS.png 
Fixede744s Fri Apr 12 10:36:09 2019 ValerianRUNGeneralN/AProcedure to follow during runs[SIZE=4][B][SIZE=3]At the beginning of a run
:
[/B]
  
Fixede744s Mon Apr 15 17:05:24 2019 ValerianOFFLINESoftwareCH2Evolution of deEEvolution of dEE between run 122 (8e4pps)
and 129 (2.5e5pps)
 DeEEvolutionWithIntensity.png 
Fixede744s Mon Apr 15 17:10:54 2019 ValerianOFFLINESoftwareCH2Evolution of deEEvolution of dEE between run 122 (8e4pps)
and 130 (2.5e5pps)
 DeEEvolutionWithIntensityIC1vsIC2.png 
Fixede744s Tue Apr 16 11:39:40 2019 ValerianOFFLINESoftwareCH2two proton kinematics and center of mass reconstruction with all stat   All_stat_2p.png 
Fixede744s Thu Apr 18 15:40:57 2019 ValerianRUNHardwareN/ATAC calib runs 226-227After multiple acq issues, TAC calib done...

Calib done with 2.56us range and 0.02us period.
  
Fixede744s Thu Apr 18 17:45:45 2019 ValerianOTHERHardwareN/AStart heating and stopping Voltage + Acq HV+BT are off and heating is started for Arnaud
to open the chamber
  
Fixede744s Wed Apr 24 15:01:01 2019 ValerianOTHERHardwareN/AElectronic scheme on the MUGAST side after experiment e744Attached are schematics of the electronics.  MUGAST_Electronic_scheme_afterE744.pdf 
Fixede744s Fri Apr 26 11:29:02 2019 ValerianOTHERSoftwareN/ANotes about the electronicWhile doing the electronic scheme, I saw that
the Inspection line was inducing a bad signal
in the coincidence module between fag and
 Signal_oscilloscope_coinc_module.jpeg 
Fixede786s Mon Jul 1 13:27:44 2019 ValerianOTHERGeneralN/ATime calibration didn't workRerunning Time Calibration   
FixedTEST Fri Feb 21 17:45:54 2020 VAMOSVAMOSGeneralN/ALogbook VAMOS setup preparation e775s   LogBook_e775s-comprimido.pdf 
Fixed  Tue Jul 23 17:09:43 2019 UK refugeesOTHERSocial Kipper issues.   3948.jpg 
Fixed  Tue Jul 23 23:22:56 2019 UK refugeesOTHERSocial Kipper issues.I would mail you some but the postage is nuts
those days!
  

ELOG Home Page