Author: Valerian 
ExpNbr: e744s 
Type: RUN 
Category: Hardware 
Status: Fixed 
Target-Source: N/A 
Subject: TAC calib runs 226-227 
After multiple acq issues, TAC calib done...

Calib done with 2.56us range and 0.02us period. using cable with 5+16ns lenght

run 226:
TAC1 Vamos AGATA
TAC2 Fag-HF
TAC4 VAMOS CATS1
TAC5 MUST2 VAMOS (!!! Warning Parameter is wrong, it is called Si AGATA)
TAC6 MUST2 AGATA (!!! Warning Parameter is wrong, it is called Si VAMOS) -> Explains the prompt in AGATA...
TAC7 MUST2 CATS1
TAC8 CATS2 CATS1
TAC9 AGATA HF

run 227
TAC3 CATS1 HF