Message ID: 4016     Entry time: Thu Apr 11 14:03:45 2019
Author: Valerian 
ExpNbr: e744s 
Type: CATS 
Category: Hardware 
Status: Fixed 
Target-Source: N/A 
Subject: In case of a CATS2 tripped 
Currently, CATS2 is inserted, if the window :

CATS2 has tripped appear, one should do the following :

  • Open a terminal
  • Type: HV
  • Username is : admin
  • Password is : admin
  • Press enter twice

The list of channels is going to appear.

  • Using the arrow of the keyboard, go down until you reach cats2
  • then go right until you reach Off
  • press space, the dectector should start ramping up
  • If it trips again, you can try to put voltage to cats1 (right above cats2)

Two possibility :

  • If cats1 trip it is the gas bottle that is empty call : 07 83 34 30 81 (Sebastien)
  • If cats1 doesn't trip try cats2 again, and if it trips lower the voltage of cats2

To lower the voltage of cats2 : go to the left until you reach the first cell after cats2 name and lower the voltage 10 volts and try to start cats2 again.

THEN CLOSE THE TERMINAL