Currently, CATS2 is inserted, if the window :
CATS2 has tripped appear, one should do the following :
- Open a terminal
- Type: HV
- Username is : admin
- Password is : admin
- Press enter twice
The list of channels is going to appear.
- Using the arrow of the keyboard, go down until you reach cats2
- then go right until you reach Off
- press space, the dectector should start ramping up
- If it trips again, you can try to put voltage to cats1 (right above cats2)
Two possibility :
- If cats1 trip it is the gas bottle that is empty call : 07 83 34 30 81 (Sebastien)
- If cats1 doesn't trip try cats2 again, and if it trips lower the voltage of cats2
To lower the voltage of cats2 : go to the left until you reach the first cell after cats2 name and lower the voltage 10 volts and try to start cats2 again.
THEN CLOSE THE TERMINAL |