Pinned, com2019, Thu Apr 4 14:45:53 2019, Valerian, OTHER, Hardware, N/A, Surveyor's measurements GANIL-0057_E744_I-Stefan.pdfE744_XYZ_détecteurs_Si.xlsx
Summary of the Surveyors. Joined the pdf and the Excel file
Fixed, com2019, Sun Apr 7 14:24:18 2019, Valerian, OFFLINE, Software, CD2, Ex and E gamma with all statistic (run 256+257) bilan_run256_and_257_Ex_Egamma.png
Fixed, e744s, Mon Apr 8 11:58:37 2019, Valerian, OTHER, General, N/A, Particle identification with CATS or HF  6x
After discussing about using the HF, I join as a reference what was obtained during experiment e744 of last year. We can see that using an external TAC the time resolution is much worse using HF.

I also join as a reference the same timing using MUVI.

Edit: Added simulation with 3 different time resolution.

Joined plot are :

  • TAC MUST2-CATS
  • TAC MUST2-HF
  • Calibrated ToF on MUVI
  • Simulated : CATS (500ps resolution)
  • Simulated : HF best case scenario (2ns)
  • Simulated : HF as in previous experiment (4ns)
Fixed, com2019, Tue Apr 9 10:44:40 2019, Valerian, MUST2, Software, CD2, CsI Calib files from the commissioning for proton Cal_CsI_E_MM1.calCal_CsI_E_MM2.calCal_CsI_E_MM3.calCal_CsI_E_MM4.cal
CsI calibration obtained from the proton data of the commissioning of mugast.
Fixed, e744s, Thu Apr 11 11:36:09 2019, Valerian, OFFLINE, Software, CH2, Comparison between the two e744 current_exp_old_e744.png
Current calibration is off, a shift of -1.5MeV was done to reproduce the previous data.
Fixed, e744s, Thu Apr 11 13:43:26 2019, Valerian, OFFLINE, Software, CH2, Target position for 14N runs XY_run14N.pngXshift_run14N.pngYshift_run14N.png
Fixed, e744s, Thu Apr 11 14:03:45 2019, Valerian, CATS, Hardware, N/A, In case of a CATS2 tripped 
Currently, CATS2 is inserted, if the window :

CATS2 has tripped appear, one should do the following :

  • Open a terminal
  • Type: HV
  • Username is : admin
  • Password is : admin
  • Press enter twice

The list of channels is going to appear.

  • Using the arrow of the keyboard, go down until you reach cats2
  • then go right until you reach Off
  • press space, the dectector should start ramping up
  • If it trips again, you can try to put voltage to cats1 (right above cats2)

Two possibility :

  • If cats1 trip it is the gas bottle that is empty call : 07 83 34 30 81 (Sebastien)
  • If cats1 doesn't trip try cats2 again, and if it trips lower the voltage of cats2

To lower the voltage of cats2 : go to the left until you reach the first cell after cats2 name and lower the voltage 10 volts and try to start cats2 again.

THEN CLOSE THE TERMINAL
Fixed, e744s, Fri Apr 12 10:12:45 2019, Valerian, OFFLINE, Software, CH2, Comparison between HF and CATS for ELab ThetaLab ELabThetaLabrun87_E744_HFvsCATS.png
Using the HF show noises that is not seen with CATS on the ToF spectra
Fixed, e744s, Fri Apr 12 10:36:09 2019, Valerian, RUN, General, N/A, Procedure to follow during runs 
At the beginning of a run :

  • Do screenshots of the scalers and add them to the elog.(Say it is at the beginning of the run)

During a run, one should :

  • Check regularly the scalers.
  • Look at the different pages on vigru.
  • Check if there are any problem concerning the Voltages, a window should be prompted to you if it is the case. If the window has FP written it is on the VAMOS side.
  • Do conversion, see corresponding entry (Or paper next to computer)
  • Check the last pending issues in the elog and if they are fixed, edit them to fix (any HV trip or over-current are pending and shoudl be switch to fix).

Every hours and before stopping the run:

  • Do screenshots of the scalers and add them to the elog.(Say at which time it is done)
  • Do screenshots of all the vigru pages of MUGAST and VAMOS and add them to the elog.
Fixed, e744s, Mon Apr 15 17:05:24 2019, Valerian, OFFLINE, Software, CH2, Evolution of deE DeEEvolutionWithIntensity.png
Evolution of dEE between run 122 (8e4pps) and 129 (2.5e5pps)
Fixed, e744s, Mon Apr 15 17:10:54 2019, Valerian, OFFLINE, Software, CH2, Evolution of deE DeEEvolutionWithIntensityIC1vsIC2.png
Evolution of dEE between run 122 (8e4pps) and 130 (2.5e5pps)
Fixed, e744s, Tue Apr 16 11:39:40 2019, Valerian, OFFLINE, Software, CH2, two proton kinematics and center of mass reconstruction with all stat All_stat_2p.png
Fixed, e744s, Thu Apr 18 15:40:57 2019, Valerian, RUN, Hardware, N/A, TAC calib runs 226-227 
After multiple acq issues, TAC calib done...

Calib done with 2.56us range and 0.02us period. using cable with 5+16ns lenght

run 226:
TAC1 Vamos AGATA
TAC2 Fag-HF
TAC4 VAMOS CATS1
TAC5 MUST2 VAMOS (!!! Warning Parameter is wrong, it is called Si AGATA)
TAC6 MUST2 AGATA (!!! Warning Parameter is wrong, it is called Si VAMOS) -> Explains the prompt in AGATA...
TAC7 MUST2 CATS1
TAC8 CATS2 CATS1
TAC9 AGATA HF

run 227
TAC3 CATS1 HF
Fixed, e744s, Thu Apr 18 17:45:45 2019, Valerian, OTHER, Hardware, N/A, Start heating and stopping Voltage + Acq  
HV+BT are off and heating is started for Arnaud to open the chamber
Fixed, e744s, Wed Apr 24 15:01:01 2019, Valerian, OTHER, Hardware, N/A, Electronic scheme on the MUGAST side after experiment e744 MUGAST_Electronic_scheme_afterE744.pdf
Attached are schematics of the electronics.
Fixed, e744s, Fri Apr 26 11:29:02 2019, Valerian, OTHER, Software, N/A, Notes about the electronic Signal_oscilloscope_coinc_module.jpeg
While doing the electronic scheme, I saw that the Inspection line was inducing a bad signal in the coincidence module between fag and CATS1. See joined image.

Also, I broke a cable between CATS1 Fifo and the delay box, the delay is now 2 ns longer.
Fixed, e786s, Mon Jul 1 13:27:44 2019, Valerian, OTHER, General, N/A, Time calibration didn't work 
Rerunning Time Calibration
Fixed, TEST, Fri Feb 21 17:45:54 2020, VAMOS, VAMOS, General, N/A, Logbook VAMOS setup preparation e775s LogBook_e775s-comprimido.pdf
Fixed, , Tue Jul 23 17:09:43 2019, UK refugees, OTHER, Social, , Kipper issues. 3948.jpg
    Fixed, , Tue Jul 23 23:22:56 2019, UK refugees, OTHER, Social, , Kipper issues. 
I would mail you some but the postage is nuts those days!