OngoingTEST Tue Jan 22 18:52:23 2019 Valérian-MarlèneGRITGeneral3-alpha3 alpha source run on MG10C & MG3T3 alpha source (not centered)
MG10C and MG3T ON
MG10C is connected to MG5T HT and BT and
  
FixedTEST Wed Jan 23 09:11:26 2019 Valérian-MarlèneGRITGeneral3-alphaRUN9 : 3 alpha source run on MG10C & MG3T[B]RUN 9 (over night 22/01 ->23/01)[/B]
3 alpha source (not centered)
MG10C and MG3T ON
 3alpha_sourceMG10C-MM2_MG3T-MM3.pngHV_squares_plugged_on_MG4T-MG5T_andMG3T.pngHV_squares_plugged_on_MG4T-MG5T_andMG3T_afternightrun.pngMG10C_7missingstrips.png 
Fixed  Wed Jan 23 12:24:10 2019 MarlèneGRITGeneral Change cable MG1T - Exchange of detectors named MG9C & MG10C23/01/19

Change of the HT/BT cable of MG1T (the one
  
Fixed  Fri Jan 25 17:02:25 2019 MarlèneGRITGeneral Detectors numbering   MUGAST_detectors.pdf 
Fixed  Thu Feb 14 08:50:58 2019 MarleneGRITGeneral3-alphaRUN 27RUN 27 on 14th of feb.
3 alpha source (one in front of trap., one
in front of square detectors)
 8x 
Fixed  Thu Feb 14 15:30:14 2019 Marlene Daniele JennyGRITGeneral3-alphaRUN 28RUN 28 14th Feb
3 alpha source on MG11A, MG3T, MG1T
 3alpha_sourceMG03XY._OKpng3alpha_sourceMG11XY.png3alpha_sourceMG01XY_OK.png 
Fixed  Fri Feb 15 15:31:28 2019 MarlèneGRITGeneralN/ADetectors bias    
Fixedcom2019 Mon Feb 18 10:41:36 2019 MarleneGRITGeneralN/ABias & current of detectors   bias.png 
FixedTEST Thu Feb 21 14:37:26 2019 iulian, valerian &FreddyGRITHardware3-alphaResults Calibration 3-alpha RUN30 MG1, MG3,MG5 & MG7Results of the calibration run 30
In order:
 Telescope1 X,Y (MG5 X,Y)
 8x 
FixedTEST Thu Feb 21 16:43:45 2019 Valérian, Freddy, Iulian and MarlèneGRITHardwareN/ACurrent measurementsJoined are the the I(V).

Detectors seems to respond accordingly with
 Depletion_Mugast_MG.pdfDepletion_Mugast_M2.pdfDepletion_Mugast.xlsx 
FixedTEST Thu Feb 21 17:32:42 2019 iulian, freddyGRITHardware3-alpharun 33 3 alfa with MG1,MG3,MG4,MG5,MG7,MG9 short runFirst test of the repaired 2xcables. The repaired
cables were placed between MUVI1_2 <=> MG4
& MUVI1_4 <=>MG11 (longer cable)
 3alpha_sourceMG1MG4MG7MG9XY_run33.png3alpha_sourceMG3MG5XY_run33.png 
FixedTEST Thu Feb 21 18:14:48 2019 iulianGRITHardware3-alphaACQ problem run 34 3 alfa with MG1,MG3,MG4,MG5,MG7,MG9
Alpha source run for trapezoid detectors
+ anulaire detector with pedestals aligned
 acq_problem_21and22fevrier.png 
FixedTEST Fri Feb 22 11:56:53 2019 iulian GRITHardware3-alpharun 38 3-alpha with reordering of the MUGAST cables for the logic and analogic signals[U][B]Run 38
[/B][/U] 
 Alpha source run for trapezoid detectors
 run38_3alpha_mg1345.pngrun38_3alpha_mg711.png 
FixedTEST Fri Feb 22 15:54:30 2019 iulian GRITHardware3-alpharun 40 3-alpha night run [CENTER][/CENTER][U][B]Run 40
[/B][/U] 
 Alpha source run for trapezoid detectors
 run40_3alpha_mg1345.pngrun40_3alpha_mg79.png 
FixedTEST Tue Feb 26 13:21:32 2019 iulian freddy GRITSoftwareN/Arun 42 Calimero run MUVI 1 on X sectors[U][B]run #42
[/B][/U]Calimero MUVI1 X
  
FixedTEST Tue Feb 26 13:50:36 2019 iulian freddy GRITSoftwareN/Arun 43 Calimero run MUVI 1 on Y sectors[U][B]run #43
[/B][/U]Calimero MUVI1 Y
  
FixedTEST Tue Feb 26 14:52:49 2019 iulian freddy GRITSoftwareN/Arun 44 Calimero run MUVI 2 on X sectors (bad run)[U][B]run #44[/B][/U] bad run
Calimero MUVI2 X
 run44_error_calimero.png 
FixedTEST Fri Mar 1 12:07:44 2019 FreddyGRITGeneral3-alphaEfficiency reconstructionblue: reconstructed efficiency using alpha
run 40 (arbitratily normalized)
red: simulated efficiency
 comparison.pdfcomparison.Calpha_eff.root 
FixedTEST Mon Mar 4 13:54:24 2019 iulian & ValerianGRITHardware3-alpharun 50 3-alpha short run[U][B]RUN 50[/B][/U]
First 3a test with  all Must2 detectors.
HV on all set to 50 V for this test only
 run50_3a_Must.pngrun50_3a_Must2.png 
FixedTEST Mon Mar 4 16:24:51 2019 valerian marlèneGRITGeneral RUN52 3 alpha source 4MUST2 no pedestalRUN52 
3 alpha source run with only 4MUST2 (in MUVI3)
with pedestal suppression at 13 deg.
 HV_MUST2_during_run52.pngHV_MUST2_afternightrun52.png3alpha_source_4MUST2_nightrun52.png 