Message ID: 19     Entry time: Thu Feb 21 18:14:48 2019
Author: iulian 
ExpNbr: TEST 
Type: GRIT 
Category: Hardware 
Status: Fixed 
Target-Source: 3-alpha 
Subject: ACQ problem run 34 3 alfa with MG1,MG3,MG4,MG5,MG7,MG9 

Alpha source run for trapezoid detectors + anulaire detector with pedestals aligned (with HV Frown ) and zero suppression. Cooling at 15°C.
Thresholds at 500 keV.

On MUVI1:

1--> MG5 (150V)
2--> MG4 (95V)
3--> MG3 (160V)
4--> MG11 (60V)

On MUVI2:
5--> MG7 (120V)
6--> MG1 (140V)
7--> --
8--> --

TRIGGERS are:
GMT ch1 --> OR MUVI1
GMT ch2 --> OR MUVI2
GMT ch3 --> OR MUVI3 (nothing on it)

PROBLEM ACQ , MUVI2 has no validations (DECS). MUVI1 OK!. Looks like a problem on the catchslave actor of Narval as no events seems to pass trough

The problem was caused by the cable used to triger the GMTSlave from the FanIn FanOut.