Fixed, e744, Sat May 26 16:31:53 2018, Jaromir, Giuseppe, EXL-PPT, Software, N/A, EXL calibration, light leak batchSelector.CEXL.pdfpairs0102_0910.pdfpairs1112_1718.pdf
Selector.C can be used to adjust energy bad reconstruction in the EXL detectors due to energy leak (see the PDF attached for more information about it) once we already have the coefficients for the energy on EXL crystals. To operate with it, for first calibration coefficient must be deduced. Once it is done the slopes and offset of the calibration have to be inserted in Selector.C file at lines 41 (a[19]) and at line 49 (b[19]). After thqt thresholds must be fixed (line 58, tr[19]) and also gains on the pairs. We deduced gains as the ratio between x-axis and y-axis of each uncalibrated matrix: for each pair (1-2,3-4,5-6,...,17-18) we took the ratio between x=2000;4000;6000 and the respective y value, and y=2000;4000;6000 and the respective x values. taking the averages of those ratios; those were than used as the coefficients R1ex[9] and R2ex[9] in Selector.C (all information on those values and how are used are in EXL.pdf).
Once the variables are set, the program can be executed using the batch file in attachement by typing the line root rootdir/run_number.root -b batch -q from terminal. Here "run_number" is the name of the run to use. It will generate then a .root file (in attachement some output file for run 0430.root). Now the name is "whatever_output.root", and can be changed from "Selector.C"
The program also provides randomization of bins for the operation of sym between contiguous crystals
Ongoing, e744, Sat May 26 20:20:55 2018, Valerian, OFFLINE, General, CH2, 15F observations for target 4 (100um) Ecm_Thetacm_target4_notes.pdfEx_15F_proj_0_4deg_Ecm_target4.pdfEx_0to40degEcm_15F.pdf
Analysis offline permits us to conclude that the current target has the good size to observe our resonances.
Ongoing, e744, Sat May 26 20:23:52 2018, Cristina Iulian, OFFLINE, General, , PPT gamma spectrum runs_target4_proton_prompt_PPT.pdf
Attached is the gamma spectrum from PPT detector with a condition on prompt (T_MM_PPT), proton detected in MUST and multiplicity event equal to 1
Pending, e744, Sun May 27 15:36:31 2018, François, BEAM, Social, , News updated 
19H
They succeeded to get 90 % transmission with CSS1 but the energy at the exit was changed by 4%, that is not compatible with the CSS2 limit of 1 %. After fitting the CSS1 energy they obtained 70-80 % transmission only, but it is enough. The problem with the HF of CSS2 was solved. They are trying to accelerate with CSS2. They are now at midway inside CSS2. They still need several hours to finish the work.


15H
The problem with CSS1 is not yet solved. They want 90% transmission of the beam in CSS1, they have 75%.
There is another problem, the HF of CSS2 has dysfonctionnements.
They don't expect beam before 20H.
Fixed, e744, Sun May 27 20:47:02 2018, e744- Valérie, DAQ, Hardware, N/A, Info to reboot the crates in case of problem IOLANmust2reboot.pngRunControlE744Ready.png
In case of problems in the run control, if it is not possible to reinstall the crates in the "Ready" mode status,
it is possible to reboot them remotely via the software commands.
this can be done using the iolan windows.
first, open a Terminal, then write:
iolan must2 all
Two windows are opened, for each of the VXI crates. see snapshot IOLANmust2reboot.png
then it is possible to reboot the crates by writing in each of the gansterm terminals the "reboot" command.
(Note that a password is asked, it corresponds to the first letter of all the words between (La cigale ...l'été) in the famous sentence of La Fontaine's fable "La Cigale et la Fourmi" ("The Grasshopper and the Ant").
Then after the operation the run Control Narval should be ready (see snapshot).
If not, it means that we need to enter in the D6 room to reboot the crates directly by pushing the "SYS RES" button (using the tiny yellow screw) on the two crates:
start with the Master crate (ganlx1, the one with the main GMT and MUVI1), then press the button of the second crate (ganlx14, with the GMTslave and the second MUVI).
The click-click on each crate means that the reset is done.
(NB: Simultaneously, the two gansterm terminals show the lines of the on-going reboot operations).
Ongoing, e744, Mon May 28 02:07:41 2018, Valerian, OFFLINE, General, CH2, New improved resolution for 14N et 15F Exc_15O_newpos.pdfExc_15F_newpos.pdf
After correcting position for the Must2 detectors by taking into account the real hit position on the target that was (0,15cm,0) and not (0,0,0). The way to do it was to artificially move every detector 15cm lower.

Joined are the results at 0deg, for 14N the resolution increases from 125keV to 102keV.
Fixed, e744, Mon May 28 11:00:07 2018, Valerian and Iulian and Phil, RUN, General, , two protons  e2p_15O.pdf
Fixed, e744, Mon May 28 12:22:42 2018, Francois & Marlete, OFFLINE, General, , superposition exp / theory for 15O and 15F sum.jpg15F_final.jpg
Pending, e744, Mon May 28 16:14:29 2018, François & Valérie, MUST2, General, N/A, Cooling system 
Monday 28th 16h10: we refill the cooling system for MUST2. The level should be ok for the rest of the experiment.
(Around 4 days).
Fixed, e744, Mon May 28 16:32:38 2018, E744- Valérie, DAQ, General, N/A, DAS configuration files-References 15x
Fixed, e744, Mon May 28 18:27:45 2018, E744- Valérie, DAQ, General, N/A, High Voltages and intensities on MUST2 HT_MUST2withoutBeam28may18.png
Check regularly the HT on the MUST2.
NB: without beam the current on Telescope 4 is high, between 9 and 11 µA. see snapshot.
NB: The Ioset was increased at 20 µA on Friday 25th (cf Id entry=490) to avoid trips on this telescope due to strong intensity fluctuations due to the beam.
Ongoing, e744, Mon May 28 20:28:50 2018, Valerian and Simon, OFFLINE, General, CH2, Explanation for the 0 degree low energy proton explanation_spotT5.pdf
The 0 degree energy proton are coming from the beam hitting the aluminum target. If you do a cut on the beam projection on T5 you obtain the black dot on the joined plot.
Fixed, e744, Mon May 28 20:33:39 2018, E744- evening shift LISE+e744 teams (edit Iulan, Valérie), RUN, General, CH2, BEAM AGAIN RUN 2100 Tuning beam_spot_2100_end.png
See plot of the Must2 impacts, Wien filter set at +/-0.8kV
Beam on CATS 1.2 10**5 pps
CH2 183 µm
Div as before
Fixed, e744, Mon May 28 20:38:21 2018, E744- evening shift LISE+MUST2 teams (edit Valérie), RUN, General, CH2, RUN 2101 DATA target 4 12x
run #2101
start @20h27
stop @22h20
target 4

Evening shift team:
LISE and E744 François, Iulan, Valérian, Louis, Jean-Charles, Marlete A., Pierre D., Simon G., Valérie L.
E744
Beam on CATS 1.2 10**5 pps
CH2 183 µm // ERROR ?? this should be 100um
Div as before
All Triggers in GMT (without LISE)
Fixed, e744, Mon May 28 22:11:52 2018, e748- Louis, EXL-PPT, Hardware, CH2, EXL4 trigger excess Scaler_EXLTriggerExeed_Run2101.png
During run 2101, Frequency of the EXL4 channel at 478Hz compared to values around 30Hz for other channels
Fixed, e744, Mon May 28 22:27:14 2018, Simon et Louis, RUN, General, CH2, RUN 2102 DATA target 4  Scaler_start_Run2102.pngScaler_stop_Run2102.png
run #2102
start @22h20
stop @23h30
target 4

Evening shift team:
LISE and E744 François, Iulan, Valérian, Louis, Jean-Charles, Marlete A., Pierre D., Simon G., Valérie L.
E744
Beam on CATS 1.3 10**5 pps
CH2 183 µm // ERROR THIS SHOULD BE 100um
Div as before
All Triggers in GMT (without LISE)
Ongoing, e744, Mon May 28 23:14:26 2018, Valerian and Iulian and Phil, OFFLINE, General, CH2, New hint of the 5/2- resonance New_resonance_hint.pdf
We see a new hint of a resonance
Fixed, e744, Mon May 28 23:52:19 2018, Simon and Louis (start), Franck (stop), RUN, General, Carbon, RUN 2103 DATA target 1 Carbon Scaler_start_Run2103.pngScaler_stop_Run2103.png
run #2103
start @23h49
stop @1h51
target 1 Carbon

Evening shift team:
LISE and E744 François, Iulan, Valérian, Louis, Jean-Charles, Marlete A., Pierre D., Simon G., Valérie L.
E744
Beam on CATS 1.3 10**5 pps
Carbon 75um
Div as before
All Triggers in GMT (without LISE)
Ongoing, e744, Tue May 29 00:17:18 2018, Night shifter, BEAM, General, Carbon, Beam intensity variation hv_run2103.png
Beam shut down at 00h03. Increase the intensity progressively up to now 00h18.
At 0:28: primary intensity dropped down to 0.
Fixed, e744, Tue May 29 01:08:22 2018, e748, BEAM, General, Carbon, Beam intensity 
At 0:46 primary beam intensity reaches 5 uA -> 1.5e5 pps on CATS.