MUST2 SAMURAI_2018 S3-LEB-LPC SUPERNEMO MUGAST EXPAND SCALP GALATRON HiCARI VELO
  MUST2, Page 24 of 48  Not logged in MUST2
Status ExpNbr Date Authordown Type Category Target-Source Subject Text Attachments
Pinnede748 Tue Jun 13 11:14:23 2017 Valérie LapouxMUST2General GMT configuration, FAGThe Gatconf parameter is now defined in the
action file.
The new configuration for GMT (see chronogramme
 GMTconfigOKMust2FAG100ns.pngGMTtrigParamActions.pngGMTparamDASgatconf.png 
Pinnede748 Thu Jun 15 10:16:03 2017 Valérie LapouxMUST2Hardware Electronics for GMT trigger and FAGThe GMT is used in a mode where we do not
make coincidence and decision, GMT entries
are single and all accepted as triggers of
 DaqElectronicsMUST2gmtforE748.pdf 
Fixede748 Wed Jun 28 10:52:15 2017 Valérie LapouxMUST2General Memo for CALIMERO MUST2 Time CalibrationTemporary file indicating the main steps for
the time calibration of MUST2 using the Calimero
DAQ package program.
 ConfigDAQDasCalimerov2017vlxTMP.pdf 
Fixede748 Wed Jun 28 10:54:34 2017 Valérie LapouxRUNSoftware DAQ problems due to TCP/IP errorsProblems occurred during the run #29 at 2h38
Wed. 28th June, with the following error
messages written in the Run Control GUI window:
 ProblemRun29TCPerror.png 
Fixede748 Wed Jun 28 11:02:38 2017 Valérie LapouxRUNGeneral3-alphaCalibration run #31 with Alpha source - MM5 and MM8Run #31 started at 11:01, Wednesday 28th June.
The alpha source is rotated towards the telescopes
5, 8.
 Run31startNarvalAlphaSource.png 
Fixede748 Wed Jun 28 15:17:24 2017 Valérie LapouxRUNGeneral End of run 31-statusRun #31 started at 11h01; stopped at 15h13
Problem on ganlx9 VME Chio
 TIMEOUT on the stop process.
 Run31stopProbTimeoutGanlx9Chio.png 
Fixede748 Thu Jun 29 23:26:10 2017 Valérie LapouxRUNGeneralEmptyALL CALIMERO Time Calibration MUVI129 JUNE 2017 
CONDITIONS FOR CALIMERO CALIBRATIONS
Time calibrator Period 20ns ; range 512 ns
 CalimeroRunControl.pngConfigCalimeroStat.pngCalibTmuvi1Yperiod20nsrange600ns.pngTimeMust2X_MUVI1_uncalib.pngTimeMust2Y_MUVI1_uncalib.png 
Fixede748 Fri Jun 30 00:33:36 2017 Valérie LapouxRUNGeneralEmptyALL CALIMERO Time Calibration MUVI230 JUNE 2017
CONDITIONS FOR CALIMERO CALIBRATIONS
Time calibrator Period 20ns ; range 512 ns
 CalibTimeMUVI2stripXT.pngSpectresMUVI2stripXT300ns.png 
Fixede748 Fri Jun 30 01:01:08 2017 Valérie LapouxRUNGeneralEmptyALL CALIMERO Time Calibration MUVI2 GOOD RUNS 600ns[B]RANGE TAC MUST2 600ns
CALIMERO TIME CALIBRATOR PERIOD 20ns; Range
512ns[/B]
 TimeMust2X_MUVI2_uncalib.pngTimeMust2Y_MUVI2_uncalib.png 
Fixede748 Fri Jun 30 12:24:17 2017 Valérie LapouxDAQGeneral DAQ Parameters: TAC, Caviar, LISE D4, D6, Caviar   DASParamVXIMust2CatsTACLise.pngDASParamVXIChioQDC.pngDASParamVXIChioADClisepar.png 
Fixede748 Fri Jun 30 16:49:49 2017 Valérie LapouxDAQGeneral GMT triggers and chronographConfigurations for the GMT inputs.
30 June 2017 pm
all Must2 telescopes,
 DASConfigTriggerGMT.pngChronogrammeGMT30juinFAG100nsCT5µs.png 
Fixede748 Fri Jun 30 16:55:04 2017 Valérie LapouxDAQGeneral Run control for global DAQ with NUMEXOThe run start/stop sequence for the global
DAQ including the NUMEXO:
 RunControlAll748ModeMonitoring.pngNarvalRunControlAllE748.png 
Fixede748 Fri Jun 30 17:53:35 2017 Valérie LapouxDAQGeneral Scalers parameters U2M   Actionse748ganlx11withU2M.pngScalerAnalyzerU2M.png 
Fixede748 Fri Jun 30 20:46:16 2017 Valérie LapouxTARGETGeneral TARGET configurationList of the reaction target in the chamber
from bottom to up:
  
Fixede748 Sat Jul 1 14:32:02 2017 Valérie LapouxDAQGeneral Scalers U2M, parametersU2M configuration +Actions ganlx11

The signals for the scalers are converted
 ScalersU2Mok.pngCompilActionsganlx11ScalerGr1to4..png 
Fixede748 Sat Jul 1 14:45:17 2017 Valérie LapouxDAQGeneral Parameters for correlation checkingsThe [B]parameters[/B] to be used for the checking
of the correlations between crates/VME/VXI
and good running of the merging
  
Fixede748 Sat Jul 1 15:12:01 2017 Valérie LapouxONLINEHardware Conditions for Bidiv and TAC 1st July 15pm(Update the file CONDTIONS748.txt during the
experiment)
  
Fixede748 Sun Jul 2 08:06:53 2017 Valérie LapouxRUNGeneralCD2Run 77 global checkConditions of Run 77

2 July - start at 6h29 - stop at 8h38 
 6x 
Fixede748 Sun Jul 2 08:40:25 2017 Valérie LapouxRUNGeneralCD2Run 78 - 2 July - 18O beam   Run #78 CD2 2.5 mg/cm2 position 3 - 18O
beam 
Same conditions as for Run 77
 Run78ConfdevDataTrig.pngRUN78tPLcats1CORREL.pngRun78CorrelWindowtimePlastCats1.pngRun78TACplastCats.pngRun78TrigMM2impact.png 
Fixede748 Sun Jul 2 08:46:13 2017 Valérie LapouxDAQHardwareCD2Summary of the TAC conditionsJULY 2nd 8pm

Bidiv =  1 for CATS1 and 2
  

ELOG Home Page