MUST2 SAMURAI_2018 S3-LEB-LPC SUPERNEMO MUGAST EXPAND SCALP GALATRON HiCARI VELO
  MUGAST, Page 67 of 132  Not logged in MUGAST
Status ExpNbr Date Author Type Category Target-Source Subjectdown Text Attachments
Fixede768s Fri Jul 12 17:51:06 2019 mugast (Sylvain, Chloe, François, Marlène, Nicolas)OTHERGeneral Monitoring of beam profile with BAF2In order to provide a efficient way for PCP
to monitor beam profile a new TAC measurement
has set up. 
  
FixedTEST Wed Feb 12 16:24:52 2020 Marlène, Freddy & FrancoMUST2Hardware Modifications on the electronic schemeWe cleaned the electronic scheme.
Now the HF CLEAN VAMOS is taken directly
from VAMOS and not re-cleaned by us.
  
FixedTEST Wed Dec 4 19:19:38 2019 MarlèneGRITHardware Modifications of set-upWe have installed the new MUFEE board in the
reaction chamber for MG2. This board is a
new version with ATHED.
  
Fixede744s Thu Apr 11 02:12:42 2019 FrançoisTARGETGeneral Measured target thicknesses
[COLOR=blue]CATS2[/COLOR]
Incident beam 14N7+  Brho 0.79 Tm
  
Fixede768s Mon Jul 15 14:29:47 2019 MarlèneMUST2General Masks for MUST2 & Kapton changed for MG5Masks are now correctly mounted

All 4 kaptons from MG5 have been changed
 IMG_20190715_121646_crop_flip.jpgIMG_20190715_122035.jpg 
Fixede768s Fri Jul 12 12:07:56 2019 mugastOTHERGeneralN/AMasks for MUST2Masks for MUST2 are now mounted. 

NOTE: one mask is not properly screwed (top
 masks_MUST2.jpgmasques_Must_pour_Mugast.pngmatrixImpact_shield_MM2.png 
Fixedcom2019 Fri Apr 5 11:26:09 2019 Adrien, Freddy and the worldOFFLINEGeneralN/AMask reconstruction CATS1-CATS2 run 223Mask between CATS1 and CATS2 in Z=-2190
Reconsctruction with nptool and attached
mugast.detector and ConfigCATS.dat attached.
 MaskCATS1CATS2_run223.pdfmugast.detectorConfigCATS.dat 
Fixedcom2019 Thu Apr 4 14:41:52 2019 JennTARGETGeneralN/AMask[IMG]elog:3696/1[/IMG]

Image of the Mask position in the target
 Target_Frame.jpg 
Fixede793s Wed Mar 3 17:18:42 2021 FrancoGRITGeneralN/AMUVI's scheme and connections on the chamber   connections_chamber_MUGAST.pngMUVI.PNG 
FixedTEST Mon Mar 4 14:34:31 2019 Valerian, Marlène, IulianMUST2GeneralN/AMUST2 positions (based on e755)Based on experiment e755, the MUST2 config
is the following :
 Screenshot_2019-03-04_at_14.39.02.png 
Fixede768s Wed Jul 17 09:50:44 2019 mugastMUST2GeneralN/AMUST2 CAEN alim burned downYesterday at around 23h the CAEN alim stopped
working.
 - The 48V power supply is blown up
  
Fixede793s Fri Jan 29 15:11:54 2021 AdrienVACUUMGeneralN/AMUGAST undervacuum and powered upMUGAST is undervacuum thanks to Arnaud and
Laurent. 
  
FixedTEST Tue Feb 2 16:41:17 2021 FrancoVACUUMHardwareN/AMUGAST undervacuum and powered upI put back the supply of MM1 to the channel
1 of the MUST2 box (as in the normal configuration)
and polarized the detectors to make a rapid
  
Fixede786s Mon Jun 3 18:00:10 2019 MarlèneGRITGeneral MUGAST detectors status --> MG4T The problem of MG4T ha sbeen solved by changing
the feedthrough.
The pulser on Y was not working and we were
  
Fixede775s Fri Feb 28 22:58:31 2020 Franck, SylvainONLINEGeneralCD2MUGAST Spectra Run 153Run 153 (target #2 CD2+Au)

MUGAST Spectra reset at start of run (~ 20h)
 RUN153_ElabThetaLab.pngRUN153_impact.pngRUN153_tac.pngRUN153_tof.png 
Fixede775s Thu Feb 27 19:29:24 2020 Chloé, FranckONLINEGeneralCD2MUGAST Spectra   RUN143_ana_Elab_Thetalab_19h23.pngRUN143_impact_19h24.pngRUN143_tac_19h25.pngRUN143_tof_19h26.png 
Fixede775s Fri Feb 28 00:35:02 2020 Chloé, Franck, MichalONLINEGeneralCD2MUGAST Spectra   RUN143_ana_Elab_Thetalab_0h32.pngRUN143_impact_0h30.pngRUN143_tac_0h32.pngRUN143_tof_0h31.png 
Fixede793s Fri Mar 12 09:55:30 2021 FrancoGRITGeneralCD2MM5The first plot attached has been done by artificially
removing 7 deg in theta lab, just to see
if the the kinematic lines of 47K(d,d) and
 square_shifted_theta.pngsquare_shifted_Z.png 
FixedTEST Tue May 28 11:08:10 2019 mugastMUST2Hardware3-alphaMG3 is trippingMG3T is tripping, tryied to set max current
to 20uA but it still trip around 30V.
  
FixedTEST Fri Feb 14 10:49:22 2020 Marlène & FrancoRUNGeneral3-alphaMG11 issuesWe analyzed the behaviour of the annular.
We realized that the 4-peaks behaviour on
X was already present last year but only
  

ELOG Home Page