MUST2 SAMURAI_2018 S3-LEB-LPC SUPERNEMO MUGAST EXPAND SCALP GALATRON HiCARI VELO
  MUGAST  Not logged in MUGAST
Entry  Wed Feb 12 09:32:17 2020, mugast, TEST, RUN, General, Fixed, 3-alpha, Run 500 - 3alpha - mugast standalone MG_all_X_alpha_source_run500MG_all_Y_alpha_source_run500
    Reply  Fri Feb 14 10:49:22 2020, Marlène & Franco, TEST, RUN, General, Fixed, 3-alpha, MG11 issues 
Message ID: 5564     Entry time: Fri Feb 14 10:49:22 2020     In reply to: 5561
Author: Marlène & Franco 
ExpNbr: TEST 
Type: RUN 
Category: General 
Status: Fixed 
Target-Source: 3-alpha 
Subject: MG11 issues 
We analyzed the behaviour of the annular. We realized that the 4-peaks behaviour on X was already present last year but only on one sector, now it's on two sectors. We noticed that there is a correlation between the malfunctioning sectors and the missing strips on Y. Last year there was only a missing strip on Y, affecting only one sector on X; now 3 strips are missing on Y, affecting 2 sectors on X. We believe that the fact that the boundings are missing have an impact on the grounding, therefore the sectors are not always biased properly, resulting in an oscillation of the pedestal. We estimate that this will cause a loss of efficiency of ~25% for the annular (from ~90% to ~67%).



mugast wrote:
RUN 500

start: 11/02/2020 @ 19h00
stop: 12/02/2020 @ 10h00

alpha source test - cooling 14°C

Remarks:
- The two new detectors (MG2 and 6) seem functional;
- MG11 annular, ok for sectors (Y) but many rings have double peaks;
- MM1 current was oscillating yesterday evening with values around 3 uA, this morning it was stable at ~0.55 uA.