MUST2
SAMURAI_2018
S3-LEB-LPC
SUPERNEMO
MUGAST
EXPAND
SCALP
GALATRON
HiCARI
VELO
MUGAST
Not logged in
Back
|
Find
|
Login
Message ID:
4948
Entry time:
Fri Jul 12 17:51:06 2019
Author:
mugast (Sylvain, Chloe, François, Marlène, Nicolas)
ExpNbr:
e768s
Type:
OTHER
Category:
General
Status:
Fixed
Target-Source:
Subject:
Monitoring of beam profile with BAF2
In order to provide a efficient way for PCP to monitor beam profile a new TAC measurement has set up.
A BAF2 detector has been inserted close to the reaction chamber (in position 7 of AGATA structure frame). Operating voltage of BaF2 is 1800V and the typical signal amplitude is above 100 mV.
The output signal is sent to a TFA (with gain =1). No additional delay was used. The BAF2 trigger signal is made from CFD and used as a start a TAC. TAC range was chosen to 1us.
The trigger signal was also sent to GMT trough input nb 5.
ELOG Home Page