MUST2 SAMURAI_2018 S3-LEB-LPC SUPERNEMO MUGAST EXPAND SCALP GALATRON HiCARI VELO
  MUGAST, Page 120 of 132  Not logged in MUGAST
Status ExpNbr Date Author Type Category Target-Sourcedown Subject Text Attachments
Fixede768s Thu Aug 1 15:46:43 2019 OttoRunRUNGeneral3-alphaRun318*************************
  [SIZE=3][B]Run  : 318[/B][/SIZE]
*************************
  
Fixede768s Thu Aug 1 15:50:41 2019 OttoRun (ou pas)RUNGeneral3-alpharun 317Run 317 no pedestal supp alpha run    
Fixede768s Thu Aug 1 16:01:16 2019 OttoRunRUNGeneral3-alphaRun 319 (Alpha calibration with pedestal suppression)*************************
  [SIZE=3][B]Run  : 319[/B][/SIZE]
*************************
  
FixedTEST Thu Oct 17 12:01:09 2019 Franco & MarlèneGRITHardware3-alphaTest of configuration with 4 MUVI and 3 GAMERSThe electronics set-up for MUGAST2020 has
been tested with 4 MUVI on 4 VXI crates,
1GMT for MASTER and 3 GAMERS for slaves.
 Mugast_alpha_17102019Mugast_alpha_17102020_2 
FixedTEST Thu Jan 23 14:06:51 2020 Franco&MarlèneGRITHardware3-alphaStatus of detectors on 23-01-20Short alpha source run (not saved on disk)
to test the detectors.
 6x 
FixedTEST Wed Feb 12 09:32:17 2020 mugastRUNGeneral3-alphaRun 500 - 3alpha - mugast standaloneRUN 500

start: 11/02/2020 @ 19h00
 MG_all_X_alpha_source_run500MG_all_Y_alpha_source_run500 
FixedTEST Fri Feb 14 10:49:22 2020 Marlène & FrancoRUNGeneral3-alphaMG11 issuesWe analyzed the behaviour of the annular.
We realized that the 4-peaks behaviour on
X was already present last year but only
  
FixedTEST Tue Feb 18 10:00:40 2020 mugastRUNGeneral3-alphaRUN 501 - missing infoRUN 501

alpha run with cooling during the night from
  
FixedTEST Tue Feb 18 17:53:36 2020 FrancoGRITGeneral3-alphaAlpha calibration for exp e775s (run 0501)In attachment raw and calibrated spectra for
run 0501 (alpha source) for MUST2 and Mugast.
 E_raw_MM.pngE_raw_MG.pngE_cal_MM.pngE_cal_MG.png 
Fixede775s Fri Mar 6 17:47:07 2020 OttoRunRUNGeneral3-alphaRun192*************************
  [SIZE=3][B]Run  : 192[/B][/SIZE]
*************************
  
Fixede775s Fri Mar 6 17:52:10 2020 OttoRunRUNGeneral3-alphaRun193*************************
  [SIZE=3][B]Run  : 193[/B][/SIZE]
*************************
  
Fixede775s Mon Mar 9 16:32:46 2020 OttoRunRUNGeneral3-alphaRun200*************************
  [SIZE=3][B]Run  : 200[/B][/SIZE]
*************************
  
FixedTEST Mon Mar 9 16:48:25 2020 OttoRunRUNGeneral3-alphaRun524*************************
  [SIZE=3][B]Run  : 524[/B][/SIZE]
*************************
  
FixedTEST Tue Mar 10 11:05:40 2020 FrancoGRITHardware3-alphaTrip in the cooling system after e775s experimentIn the weekend after the experiment e775s
(7-8 March 2020) a problem with the cooling
system of Mugast occurred. In particular,
 run0193_12_heat.pngrun0193_13_heat_MG6vsMG1.pngrun_0524_MG11XYraw.pngrun_0193_9_19_11_MG11XYraw.png 
Fixede793s Wed Mar 11 19:02:54 2020 mugastGRITGeneral3-alphaAlpha calibration for e793s - run 600========================================
========================================
========================================
  
Fixede793s Wed Mar 11 22:01:18 2020 mugastGRITGeneral3-alphaAlpha calibration - run 600After 2h30' of run we checked the temperatures
of the boards and we noticed that MG7 had
63 degrees on both X and Y, which is very
 MG7X_run600_Eevolution.png 
Fixede793s Thu Mar 12 14:54:13 2020 Marlène & FrancoGRITGeneral3-alphaRun 0601 - alpha source calibrationWe start the alpha source calibration run
with pedestal suppression (except for MG7
and MG9) and cooling at 0 deg.
 HV_run601.png 
Fixede793s Thu Mar 12 17:25:19 2020 mugastGRITGeneral3-alpharun 0601 - alpha source with pedestalsAlpha source calibration run without pedestal
suppression and cooling at 0 deg.
  
FixedTEST Wed Feb 3 16:42:41 2021 FrancoGRITGeneral3-alphaStatus of the set-up (February 3rd 2021)We re-mounted the annular, put the source
and cooled down to 0 deg. Voltages and spectra
are attached.
 Voltages_03022021.pngMM_raw_03022021.pngMG_X_raw_03022021.pngMG_Y_raw_03022021.png 
FixedTEST Thu Feb 4 18:16:47 2021 FrancoGRITHardware3-alphaProblems with the annular persistWe opened the chamber to verify the connections
of the kaptons but they were all well connected.
We checked with the tester the conductivity
 8x 

ELOG Home Page