MUST2 SAMURAI_2018 S3-LEB-LPC SUPERNEMO MUGAST EXPAND SCALP GALATRON HiCARI VELO
  MUGAST  Not logged in MUGAST
Message ID: 5941     Entry time: Mon Mar 9 16:48:25 2020
Author: OttoRun 
ExpNbr: TEST 
Type: RUN 
Category: General 
Status: Fixed 
Target-Source: 3-alpha 
Subject: Run524 
*************************
Run : 524
*************************

Start at 2020-03-09 16:33:27.725289

MUGAST run after recovery of overheating problem.
Mugast is cooled down with the MUST2 cooling system. MUST2 is therefore not biased.

In attachment the raw spectra and the cal spectra using the calibration done before the e775s experiment. Some pedestals seem to have moved.
The energy resolution per strip on X and Y is around 40 keV (I took MG1 in this case).

Stop at 2020-03-09 18:36:44.467422
Stop comment:

Stop at 2020-03-10 16:27:00.079221
Stop comment:

Stop at 2020-03-11 11:00:13.272842
Stop comment:

ELOG Home Page