Subject: Status of the set-up (February 3rd 2021) 
We re-mounted the annular, put the source and cooled down to 0 deg. Voltages and spectra are attached.
MG11 has a very high current (15 uA) but stable. MG9 started from 2 uA and its current is slightly increasing. The MUST2 detectors don't show anymore that oscillation in current that we observed one year ago. MM3 has a higher current (2.9 uA) with respect to the others.
Only 5 sectors are working for the annular, 8 less than last year. We need to check whether it is a problem of the kapton or of the detector itself.