MUST2 SAMURAI_2018 S3-LEB-LPC SUPERNEMO MUGAST EXPAND SCALP GALATRON HiCARI VELO
  MUST2, Page 7 of 48  Not logged in MUST2
Fixed, e748, Sun Jul 2 19:40:02 2017, Valérie Lapoux, DAQ, General, , Correlation parameters for ChiO in VXI-ChIO VXI_MUST2 Run104ChioCorrelPulser.png
- Parameters
VXI_CHIO
ADC_CHIO_V15 channel 16, add=15
Fixed, e748, Sun Jul 2 22:03:45 2017, e748, RUN, General, N/A, Run 118 

 -> bidon: Beam in D4 but Trig <> LISE
Fixed, e748, Sun Jul 2 23:20:43 2017, LISE team JCT+ChrS, BEAM, General, N/A, Identification plot LISE Run 119 Capture-2.png
Run 119 2nd July started at 22h20 (around)
Lise trigger
PID ED4_TED4HF without wedge 
Fixed, e748, Mon Jul 3 03:48:57 2017, e748, OFFLINE, General, , Correlation Test on Run 104 run_104_shifting.pdf

We analysed run 104 to check the correlation of events coded in different crates. It is clear now that the events from the MUST2 crate are shifted by +1
with respect to the CHIO crate.
Fixed, e748, Mon Jul 3 06:26:24 2017, e748, OTHER, General, N/A, Damned ELOG  
We just recovered the ELOG after 5-6 hrs out of order
Fixed, e748, Mon Jul 3 06:30:54 2017, e748, BEAM, General, N/A, 12Be beam rate  
Final 12Be beam rate as obtained by JCT with present wedge is about 6.10^3 per euA in D6
expected Imax of 18O is 6-7 euA
Fixed, e748, Mon Jul 3 06:30:58 2017, e748, DAQ, Software, , PROBLEM of conversion for RUN with even number 
[COLOR=red][B]IMPORTANT[/B][/COLOR]

We realized that starting at run 160 (included) all run with an even number (160,162,etc) give an error at conversion:
Fixed, e748, Mon Jul 3 07:04:30 2017, e748, OTHER, General, N/A, tritons 
Count rate in ZD plastic is bout 2;5 times larger than in CATS1 and CATS2, probably due to tritons 
Fixed, e748, Mon Jul 3 07:06:17 2017, e748, RUN, General, CD2, RUN 173 - First Physics with 12Be run173_MUST2_Ecal.pngrun173_MUST2_dEE.pngCapture-1.png
RUN 173

START: 6H11
Fixed, e748, Mon Jul 3 07:52:36 2017, e748, RUN, General, CD2, RUN 175 
[B]RUN 175
[/B]
 
Fixed, e748, Mon Jul 3 07:52:39 2017, e748, RUN, General, CD2, RUN 175 
[B]RUN 175
[/B]
 
Fixed, e748, Mon Jul 3 14:28:09 2017, Franck et Valérian, CHIO, Hardware, N/A, CHIO's electronic scheme Schema_electronique_CHIO.pdf
A quick scheme of the CHIO electronic. The patch outputs are the one from the LISE acquisition room.
Fixed, e748, Mon Jul 3 17:28:05 2017, e748 -Thomas R, Georges V, Sébastien H, CHIO, General, N/A, (Tentative) Solution to recover the CHIO: it works!  P2280767schemeChioElog.JPGP2280756OscilloChio.JPG
An interview of Th, Seb and Georges by Valérie on July 3rd.
Please indicate complements or corrections on this summary of the ionization chamber situation. 
We list here the problems during the last days and the solution adopted to recover it...
Fixed, e748, Mon Jul 3 17:31:08 2017, e748, OFFLINE, General, CD2, BEAM and FRAGMENT PID run_203_PID_SiD6.pdfrun_203_PID_beam.pdfrun_203_PID_reactionfrag.pdf
We used the Si in D6 and the TOF(CATS1-CAVIAR) to identify the beam nuclei.

Using a gate on 12Be and 11Be, we looked at the PID using Qplastic VS TOF(CATS1-CAV) and TOF(CATS1-PL).
Fixed, e748, Mon Jul 3 18:45:41 2017, Valérie L, DAQ, General, N/A, Checking ALL RUNS with Ganil_Tape  GanilTapeRun207.pngMessageRunControlNArval.pngRunControlOnNarvalAllWatcher.png
The problems with the runs which could not be converted during the night was due
to a double header buffers of the runs with an odd number.
Now it affects sometimes also runs with even numbers cf 206 below.
Fixed, e748, Mon Jul 3 19:43:16 2017, Franck, CHIO, Hardware, N/A, CHIO logic signals for TAC and CAEN digitizer 
STOP of CHIO CAEN digitizer = VAL AND Delayed CHIO (Delay = 700 ns). Before run 217, the CHIO was not delayed.
STOP of TAC PLASTval-CHIO = CHIO without delay. Before run 217 it was delayed CHIO.
Fixed, e748, Mon Jul 3 19:59:02 2017, Valérie L, RUN, General, CD2, Series of runs - afternoon/evening 3rd July 

Series of runs during the afternoon with the tuning of the ChIo and the tuning of the line
-------------------------------------------------------------------------------------------------------
Fixed, e748, Mon Jul 3 20:02:46 2017, e748, OFFLINE, General, CD2, run217 Capture.png
PID run 217
trigger = plast
Fixed, e748, Mon Jul 3 20:29:56 2017, Valérie L, RUN, General, CD2, Data runs after LISE tuning ScalersRun222trigPlastCATS1.png
The beam was tuned by the LISE team.
Beam conditions during the run 222 were changed to increase the beam at 1µA.
Beam rate on CATS: 9000/s; Plastic 24000/s FAG ~300/s.
Fixed, e748, Mon Jul 3 20:42:14 2017, Valérie L, RUN, General, CD2, Conditions for data runs, Run 223 TriggRun223AllMust2CatsPlast.png
Triggers for the experiments :MM1...MM8,
CATS1 div, Cats2div Plast div
All divisions on triggers are set at 100.

ELOG Home Page