Fixed | e748 |
Sat Jun 17 10:56:09 2017 |
Adrien Matta | BEAM | General | | JC Thomas plans for the beam production |
Here are a few notes by Jean Charles about the expected LISE setting based on 2013 and 2014 16C tuning. |
Fixed | e748 |
Mon Jun 19 16:03:18 2017 |
e748 | EXL | General | | Progress On EXL |
We fiddled around with EXL trying to get a reasonnable 60Co spectrum.
We moved the source to the windows flange, so it is now facing the EXL detector with minimal matters inbetween, although rather far away (around 1m).
We set change the hardware gain on the second module, so the last two channels have the same hardware gain than the rest of the detector. This is an hardware gain of 30.
We set the software gain to 0 overall.
We set the software threshold to 15.
A rought estimate made this a 7 MeV range with a 300 keV threshold.
We are taking data (run 15) to check that. |
Fixed | e748 |
Mon Jun 19 16:04:51 2017 |
e748 | RUN | General | 3-alpha | Run 15 |
Run 15 is started.
Source: We have a 3 alpha source in target position and a 60Co on the window flange.
Trigger: EXL |
Fixed | e748 |
Mon Jun 19 16:36:07 2017 |
e748 | EXL | Hardware | | Gain changed on EXL |
Run 15 online analysis showed waht we believe was 4 MeV range.
So we change the hardware gain on EXL to 2 on the first module and 5 on the second one (channel 17 and 18).
We changed the software gain to 9 on the first module and 6 on the second module.
We left the threshold to there previous values. |
Fixed | e748 |
Mon Jun 19 16:36:31 2017 |
e748 | RUN | General | 60Co | Run 16 |
Run 16 is started.
Source: We have a 3 alpha source in target position and a 60Co on the window flange.
Trigger: EXL
JUNK: playing with threshold a lot. |
Fixed | e748 |
Mon Jun 19 18:02:16 2017 |
e748 | RUN | General | 60Co | Run 17 |
Run 15 is started.
We have a 3 alpha source in target position and a 60Co on the window flange.
Trigger: EXL |
Fixed | e748 |
Mon Jun 19 18:07:55 2017 |
e748 | EXL | Hardware | | Threshold changed on EXL |
Run 15 online analysis showed what we believe was 4 MeV range.
We set the threshold channel by channel:
- 1:5
- 2:5
- 3:6
- 4:4
- 5:6
- 6:7
- 7:6
- 8:6
- 9:5
- 10:4
- 11:5
- 12:4
- 13:6
- 14:4
- 15:5
- 16:5
- 17:6
- 18:6
|
Fixed | e748 |
Tue Jun 20 09:08:46 2017 |
e748 | EXL | General | | EXL Spectrum with 60Co source |
Here is a spectrum of EXL after one night of accumulating stat on 60Co. |
Fixed | e748 |
Wed Jun 21 09:05:26 2017 |
Adrien Matta, Phil Adsley | MUST2 | General | | MUST2 Dead Time |
We cabled the Dead time in the scaler.
We have a 100Hz pulser (entry 2 of U2M) in coincidence with the DT output of the GMT (entry 1), giving a direct reading of the dead time.
Note on MUST2 dead time:
- Without zero suppression the dead time caused by MUVI is 2ms. So this give a 40% DT at 180 counts per second.
- With zero suppression, the DT fall down to 240us (80us cash in to interrupt the CPU and, 80us to process the signal from MUVI to CAS and 80us to send the parameter back to the back plane). So this give a 5-7% DT with 180 counts per second.
- The ADC works in parallel of MUVI, so the DT are not cumulative. |
Fixed | e748 |
Wed Jun 21 09:39:38 2017 |
e748 | DAQ | General | | Backup of config file |
ganacq manip back up. |
Fixed | e748 |
Thu Jun 22 10:24:12 2017 |
e748 | CATS | Software | | Manip config file backup before merge |
Manip config file backup before merge |
Fixed | e748 |
Thu Jun 22 11:10:35 2017 |
Valérian | EXL | Software | | EXL Mapping |
Mapping of the EXL detector in the analysis program versus how it is written on the cables. |
Fixed | e748 |
Thu Jun 22 17:46:45 2017 |
e748 | MUST2 | Hardware | | MUST2 SL cable repair |
Three cables came back from ASCO, two of them were the original spare cables.
Report from ASCO:
- The spare could not work because two pins were soldered together, so the spare never worked
- The other one had a broken pin on the fisher side. Try not to twith the cable when rotating the connector to plug it on the flange.
|
Fixed | e748 |
Thu Jun 22 17:49:20 2017 |
Adrien and Phil | MUST2 | General | | MUST2 telescope test with alpha source |
All Telescope now pass the I2C test.
We ran with an alpha source (attached pictures for the overall sequence),
1. at first the source was at 0deg: T1-4 OK; T6 got half of X strip missing, Y Ok; T7 got half Y missing, X Ok. T5 and 8 see nothing
2. source at 150 deg: T1-4 see nothing; T6-7 see nothing, T8 Ok; T5 got every other strip missing on 3/4 of Y (surely a capton issue).
3. source at 90 deg: T6-7 OK. |
Fixed | e748 |
Thu Jun 22 17:55:52 2017 |
e748 | MUST2 | Hardware | | Surveyor mesurement |
The surveyor are missing some point for the telescope measurement.
We reopen tommorrow to fix T5 and let them retake the missing points. |
Fixed | e748 |
Fri Jun 23 09:18:14 2017 |
e748 | OTHER | General | | Plan for 23rd of June |
We have to open today, I started warming up of the telescope at 9h18 today.
Once open, we have to check on T5 DSSD capton and fix them.
Surveyor will carry out there measurement in the afternoon.
We plan on closing again this evening to perform alpha calibration over the weekend. |
Fixed | e748 |
Fri Jun 23 09:20:33 2017 |
e748 | OTHER | General | | Plan for coming days |
23rd of June:
Fix T5
Surveyor measurement
Mount 60Co in vacuum
24-25th of June:
Pedestal alignement with source facing off
Alpha calibration
26th-27th of June:
Have Calimero Running (Luc started on Wenesday)
MUST2 time calibration
TAC time calibration
EXL calibration with in vacuum source
26th-29th
CHIO test and debug with NUMEXO2 qnd CAEN Digitizer |
Fixed | e748 |
Fri Jun 23 16:37:54 2017 |
Adrien, Laurent, Vincent | OTHER | Hardware | | MUST2 chamber opening |
We opened the chamber and the surveyor took there measurement.
We closed the chamber with a 3 alpha source in the target position and a 60Co source on top.
All is now under vacuum, turbo are on and we are already below 1e-4 mbar. Cooling started, right now alcool at 10deg; copper at 12deg.
T5 capton fix.
I will start the LV tonight or tomorrow morning to run pedestal alignment and calibration over the weekend. |
Fixed | e748 |
Fri Jun 23 20:49:49 2017 |
Adrien | MUST2 | General | | Restart |
Chamber at 2e-5, cooling -9deg outside; -5 on copper LV off.
Starting BT and HV. No trip; normal current (1-2ua on DSSD; ~0.01 on CsI, still high current,9ua at 450V SiLi1 T5, other SiLi at 4-5ua).
DAS Config ok, passing I2C.
Start DAQ Ok storage off: all actor working 500kb/s, 200kb/s
GRU start ok; but receive no buffer. Kill all the daq and das, work ok.
Alpha look nice on all telescope that see the source (1-4 and half of 6/7).
All good for tonigh. let it run with storage off, waiting for the temp to set in before calibration. |
Fixed | e748 |
Sat Jun 24 21:41:30 2017 |
e748 | MUST2 | General | | Pedestal alignement and calibration |
MUSt2 pedestal aligned for all telescope.
Started alpha run with all telescope in trigger as well as EXL.
Sili1 T5 and Sili2 T8 tripped.
EXL back on (LV and HV on) |