MUST2 SAMURAI_2018 S3-LEB-LPC SUPERNEMO MUGAST EXPAND SCALP GALATRON HiCARI VELO
  MUST2, Page 25 of 48  Not logged in MUST2
Fixed, e748, Sun Jul 2 09:54:05 2017, Valérie Lapoux, DAQ, General, CD2, Test of the TAC plast CATS ADC VXI Run78 InspectLinesVXIMust2Run78LTtAiCatsFAG.pngInspectVXIchioADCsignalTplchCATS1Run78.png
 
Fixed, e748, Sun Jul 2 10:31:48 2017, Valérie Lapoux, ONLINE, Hardware, CD2, Change of the electronics for TAC Plast CATS1 
Shift U (Béa +Vlx) entered in the room to check the electronics of Cats1 and of the TAC plast-Cats1
Now we have taken the output of one of the TAC : START=CATS1_FAG - STOP plast (no val)
We send it to a Linear Fifo from which it is duplicated for the ADC channels:
Fixed, e748, Sun Jul 2 10:44:59 2017, Valérie Lapoux, RUN, General, CD2, Run 79 Beam of 18O 8x
run_0079.dat.02-07-17_10h37m20s
Triggers : all MUST2 telescopes 1 to 8.
The pulser is ON again in the DAQ and Scaler Analyzer. See plots.
Fixed, e748, Sun Jul 2 11:38:03 2017, Valérie Lapoux, RUN, General, CD2, Runs 80-81 TESTS change of TAC and triggers 
For runs 80- 81: 
for test purposes, we changed the coincidence mode for CATS: we put FAG instead of the Seq1 coming from the gamer.
Run 80 start 11h33; stop 11h40
Fixed, e748, Sun Jul 2 12:08:56 2017, Valérie Lapoux, RUN, General, CD2, Run 82 
Run82 start at 12h00 stop at 12h09
e748 Exxx    6684672  2 juil. 12:09 run_0082.dat.02-07-17_12h00m43s
Before starting run 82, we enter in the D6 room to restaure the electronics for CATS as it was before
Fixed, e748, Sun Jul 2 12:16:57 2017, Valérie Lapoux, RUN, General, CD2, Runs 83-84 only CATS1 trigger ScalersRun84Cats1trigg.pngConfigTRigCorrelMErgeRun84.png
Run83 start at 12h11 stop at 12h16
 e748 Exxx   21561344  2 juil. 12:16 run_0083.dat.02-07-17_12h11m16s
Fixed, e748, Sun Jul 2 13:22:46 2017, Valérie Lapoux, RUN, General, CD2, Run 85+summary TAC 
Started run #85 at 13h21 stop at 13h31
e748 Exxx    5439488  2 juil. 13:31 run_0085.dat.02-07-17_13h21m57s
Triggers all MUST2
Fixed, e748, Sun Jul 2 13:31:21 2017, Valérie Lapoux, RUN, General, CD2, Run 86 trigger CATS1 CorrelRun86mergeOKruncontrol.pngRun86TacPlastCATStrigCats1.png
Run 86 start at 13h32 stop at 13h43
 e748 Exxx   34340864  2 juil. 13:43 run_0086.dat.02-07-17_13h32m13s
 
Fixed, e748, Sun Jul 2 13:49:20 2017, Valérie Lapoux, RUN, General, CD2, Run 88 trigger CATS1 
Run 88 start at 13h48 
TRIGGER CATS1 div =100 only
18O beam CD2 target
Fixed, e748, Sun Jul 2 14:49:41 2017, Valérie Lapoux, ONLINE, Hardware, CD2, TAC and signals for Plast-CATS1 
Now we return to the situation of 2 TACs for the coding of the same time Plast-Cats1 in the 2 DAQs systems VXI_CHIO and VXI_MUST2.

TAC1 start CATS1val; STOP Plast (delayed by 200ns with DGG) range 1µs OUTPUT signal T_PL_CATS1 VXI_MUST2  ADC add=23
Fixed, e748, Sun Jul 2 14:54:39 2017, Valérie Lapoux, RUN, General, , Runs 90-92 Tuning of the cocktail beam in D4 
Runs 90 to 92 Tuning of LISE 
GMT Trigger IN13=Trig_LISE  (D4)
Fixed, e748, Sun Jul 2 15:13:49 2017, Valérie Lapoux, RUN, General, , Runs 93-94 Tuning of the cocktail beam in D6 ScalerAnalyzerCaviar15h29.png
Runs 93-94 Tuning of LISE 
GMT Trigger IN13=Trig_LISE (=D6)
 e748 Exxx   25493504  2 juil. 15:11 run_0093.dat.02-07-17_15h07m01s
Fixed, e748, Sun Jul 2 17:25:22 2017, Valérie Lapoux, DAQ, Hardware, CD2, ADC parameters-new ones for LISE +Pulser for Correlations DASVXIchioAdcLiseParam.pngConfigDASnew2ndJulyVXIMust2.png
New parameter for LISE in VXI_CHIO ADC_CHIO 
parameter TED4HF address=4 channel 5 (cable label "5" on connected, first branch of ADC_CHIO)
-------------------------------------------------------------------------------------
Fixed, e748, Sun Jul 2 18:19:24 2017, Valérie Lapoux, DAQ, Hardware, CD2, Change of DAQ parameters for LISE DASparamLISEVXIChioNEW.png
 
Fixed, e748, Sun Jul 2 18:27:48 2017, Valérie Lapoux, ONLINE, Software, , Narval configuration, Watcher ports RunControlRun101allwtch.png
New file xml is loaded with various Watchers, the corresponding ports 
Watcher 10205
Watcher  WatRaw 10206
Fixed, e748, Sun Jul 2 19:40:02 2017, Valérie Lapoux, DAQ, General, , Correlation parameters for ChiO in VXI-ChIO VXI_MUST2 Run104ChioCorrelPulser.png
- Parameters
VXI_CHIO
ADC_CHIO_V15 channel 16, add=15
Fixed, e748, Wed Jul 5 11:10:48 2017, Valérie L., MUST2, General, CD2, Geometrical configurations for MUST2 ConfigGEOMImpactMUST2e748.pdf
 
Fixed, e748, Wed Jul 5 11:13:27 2017, Valérie L., MUST2, General, CD2, Power boards HV and LV for MUST2 ConfigHVetLVExpE748MUST2d6v2017.pdf
 
Fixed, e744, Wed May 23 23:37:14 2018, Valérie L., MUST2, General, N/A, MUVI configuration DAS 13x
Configurations files of the MUVI 1 MufeeX and Y.
On MufeeX and Y configuration:
Note that the thresholds on Must2 T1 to T3 is at 400 keV on May 23th Midnight (same as before); and for T4, now it is set at 1 MeV (not 400 keV).
Fixed, e748, Tue Jul 4 02:33:36 2017, Valérie L, shifts U+S Yorick, RUN, Hardware, , GMT channel 8  GMTtriggerRun248.pngGMTtriggerGatconfRun248.png
Apparently Channel 8 of the GMT is not operational. We put telescope T8 in entry 15.
This operation gives good results, see pictures below of Gatconf with 2**(N-1) = 2**14 = 16384 for Run 248.
Triggers GMT are now:

ELOG Home Page