Fixed, e744, Tue May 22 11:59:06 2018, Valérie, MUST2, Software, N/A, Calimero config GECO 8x
|
|
Fixed, e744, Tue May 22 12:09:27 2018, iulian Fairouz Valerie, MUST2, Software, N/A, Run #354 Calimero T1-T4 strips X    
|
Started run #353
Calimero T1-T4 strips X
Time Calibrator 640 ns ; period 20 ns |
Fixed, e744, Tue May 22 12:10:27 2018, iulian Fairouz Valerie, MUST2, Software, N/A, Run #355 Calimero T1-T4 strips Y
|
Started run #353
Calimero T1-T4 strips Y
Time Calibrator 640 ns ; period 20 ns |
Fixed, e744, Tue May 22 13:40:03 2018, Fairouz Valerie, MUST2, Software, N/A, Run #356 Calimero T1-T4 strips Y
|
Run #356
Time Calibrator Calimero stripYT T1to T4 640ns period 20ns |
Pinned, e744, Tue May 22 18:33:32 2018, Nicolas, Adrien, OTHER, Software, , How to run Online and Convert raw data files
|
[U][B]1. How to run Basic_online or Online code[/B][/U]
[LIST]
[*] From any computer (logged as e748 user), type '[COLOR=pink]online[/COLOR]'. This logs you to the lpcgrit server and the online directory
|
Fixed, e744, Tue May 22 19:36:50 2018, Fairouz, Valérie, MUST2, Software, N/A, Run #359 Calimero T1-T4 strips X 8x
|
Run #359
amplitude of the gene is now set to 10 MeV
Time Calibrator Calimero stripsX T1to T4 640ns period 20ns |
Pending, e744, Tue May 22 19:40:26 2018, Francois, BEAM, General, , Planning of the experiment
|
|
Fixed, e744, Tue May 22 19:40:48 2018, Valérie , MUST2, General, N/A, snapshots for steps in Calimero calibration 11x
|
|
Fixed, e744, Tue May 22 20:06:15 2018, Fairouz, Valerie, edited by Iulian Megumi, MUST2, Software, N/A, Run #360 Calimero T1-T4 strips Y
|
Started run #360
calimero stripsYT T1 to T4 640 ns; period 20 ns
amplitude pulser 10 MeV
|
Fixed, e744, Tue May 22 21:43:31 2018, iulian & megumi, MUST2, Software, N/A, run 361
|
[U][B]run 361[/B][/U]
start 20h38
stop 21h41 |
Fixed, e744, Tue May 22 21:57:51 2018, iulian & megumi, MUST2, Software, N/A, Run #362 Calimero T1-T4 strips X again with solution 
|
Changed the Thresholds on T4. No longer double peaks...
I guess there is some noise on the T4 and he triggers on the noise from time to time => double peaks
Also T1-T3 have have strip 16 near pidestal who is counting a lot (red) |
Fixed, e744, Tue May 22 22:33:09 2018, e744, MUST2, Software, N/A, Run #363 Calimero T1-T4 strips X again with threshold at 50% (same as 40%)
|
[U][B]Run #363[/B][/U]
NB: Thresholds are the ones on the pulser amplitude to inject the signal (Calimero Pulser interface on Geco). see snapshots on the Calimero summary.
Calimero with 50% thresholds on T4. Same spectra as with the thresholds @40%. I think T4 is triggering on cross-talk, so when the thresholds it's high |
Fixed, e744, Tue May 22 22:50:47 2018, e744, MUST2, Software, N/A, Run #365 Calimero strips Y T1-T4 100% Threshold
|
[U][B]Run #365[/B][/U]
Calimero strips Y T1-T4
100% threshold on T4
|
Pinned, e744, Tue May 22 23:42:02 2018, iulian & megumi, EXL-PPT, Software, N/A, Run #366 NARVAL FATAL ERROR
|
RUN #366
EXL PPT triggers
Falat error from NARAVL
|
Fixed, e744, Wed May 23 00:30:34 2018, e744, EXL-PPT, General, N/A, RUN #402, EXL PPT trigger
|
RUN #402
start exl ppt run for checking coincidence
exl ch 17 put in to adc2 ch 32 |
Fixed, e744, Wed May 23 10:52:28 2018, Valérie, EXL-PPT, General, N/A, RUN #405, EXL PPT trigger   
|
RUN #405
start exl ppt run for checking coincidence
|
Fixed, e744, Wed May 23 10:57:13 2018, megumi, DAQ, General, , Photo of the circuit 15x
|
photo of the circuit area |
Fixed, e744, Wed May 23 13:15:47 2018, megumi, DAQ, General, N/A, Time calibration for TDC (EXL)
|
Time calibrator setting
Range 5.12 us
Period 160 ns
|
Pinned, e744, Wed May 23 15:10:15 2018, iulian, OTHER, General, N/A, Procedure to start stop RUNS 
|
|
Fixed, e744, Wed May 23 15:40:58 2018, Charles Houarner Valérie, MUST2, Software, N/A, TESTS MUVI internal pulser T1-T4 strips X    
|
To check the signals and trigger on T4 we used the internal pulser on T4,
from the SYNCHROTEST pulser at 150 Hz.
The signal was set on the channel 5 of each Mate of the MufeeX card of the T4 telescope:
|
|