Fixede744 Tue Apr 24 16:22:08 2018 e744 Valérie L.RUNSoftware Narval Merger Parameters   ConfigRunControlNArvalMerger.png 
Fixede744 Wed Apr 25 15:59:51 2018 ValérianEXL-PPTSoftwareN/AMesytech setupTo use the remote settings of the Mesytech
modules : 
[CODE]
  
Fixede744 Wed Apr 25 20:47:59 2018 e744 ValérieMUST2General MUST2 mate Temperature on the DAS interfacetests
The plot:ConfigDASMuvi1Tel4mufeeX.png shows
the situation on the Mufee card when the
 ConfigDASMuvi1Tel4mufeeX.pngConfigDASMuvi1Tel5mufeeXokTemp.png 
Ongoinge744 Thu Apr 26 19:37:24 2018 Valerian and FreddyMUST2HardwareN/ADebug of Must2 telescopesToday we found different problems, no time
on T4 and no signal on T5 and T6 CsI tripping.
After changing cables and a lot of tests
 debug_mm_telescopes.pdfProblem_on_T5_T6_cable_swap.pdf 
Ongoinge744 Sat Apr 28 10:17:22 2018 Valerian and IulianMUST2Hardware3-alphaStatus of MUST2 telescopes 27/04/18Yesterday night we tested T1, T2, T3, T4,
T5, and T6.
 Third_good_run_Energy_and_time_for_5_Must2_detectors_2018-04-27_19-32-23.png 
Ongoinge744 Sat Apr 28 13:27:26 2018 Valerian and IulianMUST2Hardware3-alphamultiples runs with two alpha source on T1 T2 T3 T4 and T5For the following runs we will use two 3 alpha
sources :
  
Ongoinge744 Sat Apr 28 13:36:52 2018 Valerian and IulianOTHERSoftwareN/AProblem with DAS configurationThere is an issue with the GMT slave. When
you click Online and then apply write hardware,
if the message : [U]register error on GMTSlave[/U]
  
Pending  Sat Apr 28 21:12:55 2018 iulianMUST2General3-alphaRun 182 end ; problem with coolingRUN 182 start @13h
Stoped @ 21H
 end_result_alpha_source_28_April.png 
Fixede744 Sun Apr 29 20:37:58 2018 FrancoisCATSGeneral Chronogram   chrono.jpg 
Fixede744 Mon Apr 30 20:38:21 2018 iulianRUNGeneral3-alpha6run 186 3a source 60 co source[U][B]run #186[/B][/U] 

start@11h 30/04
  
Fixede744 Mon Apr 30 20:42:34 2018 iulianRUNGeneral60Corun 188 60 co source[U][B]run #188[/B][/U] [I]60co source EXL+PPT
trigger[/I]
  
Fixede744 Tue May 1 10:47:41 2018 iulian & valerianRUNGeneral60Corun 189 60 Co source[U][B]run #189[/B][/U] [I]60co source EXL+PPT
trigger[/I]
Changed the way of codding PPT. for this
  
Fixede744 Wed May 2 18:11:59 2018 Iulian Marlenne LouisRUNGeneral3-alpha3 alpha testRun191 start 18h stop 18h10 
Run ok, saved config DAS with good polarity
on the thresholds CSI & piedestal allignement
 run_191.png 
Ongoinge744 Thu May 3 08:50:17 2018 FrançoisVACUUMGeneral stop vacuumturbo pumps stopped at 8H45

problem with the cooling system of the M2C
  
Fixede744 Thu May 3 15:04:03 2018 Iulian MarlenneMUST2HardwareN/AX and Y possitions on the detectorTelescope 1 : PCB 5 (ref on pcb: 2274)
Telescope 2 : PCB 4
Telescope 3 : PCB 7
  
Fixede744 Thu May 3 15:20:16 2018 Iulian MarlenneMUST2HardwareN/AX and Y possitions on the detectorTelescope 1 : PCB 5 (ref on pcb: 2274)
Telescope 2 : PCB 4
Telescope 3 : PCB 7
  
Ongoinge744 Thu May 3 19:11:14 2018 Valerian Marlene and FrancoisMUST2General3-alpharun 3 alpha T5[U][B]run #196[/B][/U]
3 alpha source and pedestals supressed
Trigger T5 only
  
Fixede744 Fri May 4 11:20:50 2018 ValerianMUST2General3-alphaT5 : 0° telescope After increasing the voltage on T5 DSSD from
80V to 100V the resolution increased from
~40keV sigma to 20keV sigma.
 StripX_vs_StripY_T5_100V.pdfSi_E_T5_100V.pdf 
Fixede744 Fri May 4 23:58:07 2018 iulianRUNGeneral3-alphaRun #203 alfa source & 60co source[U][B]run #203[/B][/U]

alpha source & 60co source
 run_203_ht.pngrun_203_ppt.pngrun_203_EXL.pngrun_203_must2.png 
Ongoinge744 Sat May 5 11:14:33 2018 Iulian & ValerianRUNGeneral3-alpharun #204 alpha source[U][B]run #204[/B][/U]
alpha source
start @11h15 05/05
 BT_tripped_T5.pngT5_das_no_cooling.pngT5_das_no_cooling2.pngT5_vigru_no_cooling.png20180506_152338.jpg 