  MUGAST, Page 87 of 132  Not logged in MUGAST
Status ExpNbr Date Author Type Categorydown Target-Source Subject
Fixede786s Wed Jun 26 15:50:36 2019 OttoRunRUNGeneralCryo-3HeRun60

OttoRun wrote:

OttoRun wrote:
*************************
Run : 60
*************************

Start at 2019-06-26 10:21:51.572979

target 700mb
Brho0=0.95 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted

AGATA position +51-25 mm

TRIGGER :
- VAMOS %1000
- CMG
- MM
- CATS2 %1000

Restarted GTS Tree by Emmanuel


No Beam since 15:00. According to PCP there will be no beam for a while.
Attachment 1: run60_6.png
run60_6.png
Fixede786s Wed Jun 26 16:59:07 2019 LouisBEAMGeneralCryo-3HeBeam interuption
Beam stopped from 15h to 17h due to source instabilities
Fixede786s Wed Jun 26 18:12:18 2019 OttoRunRUNGeneralCryo-3HeRun62
*************************
Run : 62
*************************

Start at 2019-06-26 18:09:42.882502

target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted

AGATA position +51-25 mm

TRIGGER :
- VAMOS %1000
- CMG
- MM
- CATS2 %1000

Crate in System faillure.
Moving FIFO FPMW to other crate. Had to change the cable (3ns to 5ns).
To be accounted for in the TOF

Stop at 2019-06-27 02:19:17.492600
Stop comment: Problems with AGATA, run stopped
Fixede786s Wed Jun 26 20:46:20 2019 Sylvain, LouisRUNGeneralCryo-3Herun 62 screenshots
Attachment 1: Vamos10.png
Vamos10.png
Attachment 2: Vamos9.png
Vamos9.png
Attachment 3: Vamos8.png
Vamos8.png
Attachment 4: Vamos7.png
Vamos7.png
Attachment 5: Vamos6.png
Vamos6.png
Attachment 6: Vamos5.png
Vamos5.png
Attachment 7: Vamos4.png
Vamos4.png
Attachment 8: Vamos3.png
Vamos3.png
Attachment 9: Vamos2.png
Vamos2.png
Attachment 10: Vamos1.png
Vamos1.png
Attachment 11: Vamos.png
Vamos.png
Attachment 12: MGToD3.png
MGToD3.png
Attachment 13: MGToD2.png
MGToD2.png
Attachment 14: MGToD1.png
MGToD1.png
Attachment 15: MMToD.png
MMToD.png
Attachment 16: MGImpact.png
MGImpact.png
Attachment 17: MMImpact.png
MMImpact.png
Attachment 18: scaler.png
scaler.png
Fixede786s Thu Jun 27 00:26:37 2019 mugastOFFLINEGeneral statistics until now
Maybe some stat still to recover from better gate and mass and Z reconstruction.

Andrea
Attachment 1: cqnvqs_.png
cqnvqs_.png
Fixede786s Thu Jun 27 03:33:26 2019 OttoRunRUNGeneralCryo-3HeRun63
*************************
Run : 63
*************************

Start at 2019-06-27 03:33:09.910218

target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted

AGATA position +51-25 mm

TRIGGER :
- VAMOS %1000
- CMG
- MM
- CATS2 %1000

Crate in System faillure.
Moving FIFO FPMW to other crate. Had to change the cable (3ns to 5ns).
To be accounted for in the TOF

%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
Current ~ 30 kHz.
The cryo target pressure and temperature have slightly increased during the stop for the recover of AGATA, even before the refilling of the LHe tank.
It is probably due to a problem in the refilling of liquid nitrogen.

Stop at 2019-06-27 11:46:17.108352
Stop comment: stop
Fixede786s Thu Jun 27 12:28:58 2019 OttoRunRUNGeneralCryo-3HeRun67
*************************
Run : 67
*************************

Start at 2019-06-27 12:28:26.879061

target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted

AGATA position +51-25 mm

TRIGGER :
- VAMOS %1000
- CMG
- MM
- CATS2 %1000

Crate in System faillure.
Moving FIFO FPMW to other crate. Had to change the cable (3ns to 5ns).
To be accounted for in the TOF

Stop at 2019-06-27 12:46:41.949098
Stop comment:
Fixede786s Thu Jun 27 16:35:45 2019 OttoRunRUNGeneralCryo-3HeRun68
*************************
Run : 68
*************************

Start at 2019-06-27 16:35:24.351472

target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted

AGATA position +51-25 mm

TRIGGER :
- VAMOS %1000
- CMG
- MM
- CATS2 %1000
Fixede786s Thu Jun 27 16:41:55 2019 OttoRunRUNGeneralCryo-3HeRun65
*************************
Run : 65
*************************

Start at 2019-06-27 11:35:21.078920

target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted

AGATA position +51-25 mm

TRIGGER :
- VAMOS %1000
- CMG
- MM
- CATS2 %1000

Crate in System faillure.
Moving FIFO FPMW to other crate. Had to change the cable (3ns to 5ns).
To be accounted for in the TOF

Stop at 2019-06-27 17:39:22.529443
Stop comment:
Fixede786s Thu Jun 27 17:43:15 2019 OttoRunRUNGeneralCryo-3HeRun69
*************************
Run : 69
*************************

Start at 2019-06-27 17:42:38.280555

target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted

AGATA position +51-25 mm

TRIGGER :
- CMG


Test with only MUGAST back trigger

Stop at 2019-06-27 17:54:02.829540
Stop comment:
Fixede786s Thu Jun 27 17:59:20 2019 OttoRunRUNGeneralCryo-3HeRun70
*************************
Run : 70
*************************

Start at 2019-06-27 17:58:29.855367

target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted

AGATA position +51-25 mm

TRIGGER :
- CMG
- M2
- VAMOS%1000
- CATS%1000

back to normal

Stop at 2019-06-27 18:37:51.816766
Stop comment: type your message here
Fixede786s Thu Jun 27 18:48:51 2019 OttoRunRUNGeneralCryo-3HeRun71
*************************
Run : 71
*************************

Start at 2019-06-27 18:47:56.171341

target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted

AGATA position +51-25 mm

TRIGGER :
- MG
- square (not biased)
- M2


all X on MUVI disabled from trigger
all threshold of MUVI1 at 1 MeV to avoid as much as possible noise

Stop at 2019-06-27 20:23:56.630240
Stop comment:
Fixede786s Thu Jun 27 20:26:27 2019 OttoRunRUNGeneralCryo-3HeRun73
*************************
Run : 73
*************************

Start at 2019-06-27 20:26:17.916314

target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted

AGATA position +51-25 mm

TRIGGER :
- CMG
- M2
- VAMOS%1000
- CATS%1000

all X on MUVI disabled from trigger
all threshold of MUVI1 at 1 MeV to avoid as much as possible noise

There is no run 72

Stop at 2019-06-27 23:04:45.884452
Stop comment: AGATA DAQ is rejectting all events
Fixede786s Thu Jun 27 23:28:30 2019 OttoRunRUNGeneralCryo-3HeRun74
*************************
Run : 74
*************************

Start at 2019-06-27 23:25:00.455575

target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted

AGATA position +51-25 mm

TRIGGER :
- CMG
- M2
- VAMOS%1000
- CATS%1000

all X on MUVI disabled from trigger
all threshold of MUVI1 at 1 MeV to avoid as much as possible noise

Stop at 2019-06-28 09:31:40.775271
Stop comment:
Fixede786s Fri Jun 28 04:21:48 2019 Daniele MengoniRUNGeneralCryo-3Herun 74
screenshots run 74 attached

frequent error message on the FP DC Wi1 --> defqlt. The elog said so, card tripped, but spectra are still present .. puzzling
Attachment 1: Q_vs_Xf_K1.png
Q_vs_Xf_K1.png
Attachment 2: Q_vs_Xf_K2.png
Q_vs_Xf_K2.png
Attachment 3: Q_vs_Xf_K3.png
Q_vs_Xf_K3.png
Attachment 4: Q_vs_Xf_K4.png
Q_vs_Xf_K4.png
Attachment 5: Q_vs_Xf_K5.png
Q_vs_Xf_K5.png
Attachment 6: Q_vs_Xf_K6.png
Q_vs_Xf_K6.png
Attachment 7: Q_vs_Xf_K7.png
Q_vs_Xf_K7.png
Attachment 8: Q_vs_Xf_K8.png
Q_vs_Xf_K8.png
Attachment 9: Q_vs_Xf_K9.png
Q_vs_Xf_K9.png
Attachment 10: Q_vs_Xf_K10.png
Q_vs_Xf_K10.png
Attachment 11: Q_vs_Xf_K11.png
Q_vs_Xf_K11.png
Attachment 12: Q_vs_Xf_K12.png
Q_vs_Xf_K12.png
Attachment 13: Q_vs_Xf_K13.png
Q_vs_Xf_K13.png
Attachment 14: Q_vs_Xf_K14.png
Q_vs_Xf_K14.png
Attachment 15: Q_vs_Xf_K15.png
Q_vs_Xf_K15.png
Attachment 16: Q_vs_Xf_K16.png
Q_vs_Xf_K16.png
Attachment 17: Q_vs_Xf_K17.png
Q_vs_Xf_K17.png
Attachment 18: Q_vs_Xf_K18.png
Q_vs_Xf_K18.png
Fixede786s Fri Jun 28 08:54:50 2019 FreddyRUNGeneralCryo-3Herun 74

Daniele Mengoni wrote:
screenshots run 74 attached

frequent error message on the FP DC Wi1 --> defqlt. The elog said so, card tripped, but spectra are still present .. puzzling




Put the voltage back at 8h00 when i arrived.
Fixede786s Fri Jun 28 08:57:21 2019 FreddyCATSGeneralCryo-3HeCATS scaler not compatible with beam rate
The CATS scaler si now around 100-120 kHz while the beam rate is still 40 kHz.

I checked the CATS signal and there is clearly noise who induce double logic signals. Not all of them are validated by FAG.

We should raise the threshold- call sebastien

9h32 : Stop beam to enter and raise the threshold.
Attachment 1: ScalerCATSproblem.png
ScalerCATSproblem.png
Fixede786s Fri Jun 28 09:56:59 2019 Bertrand, Valerian and AdrienBEAMGeneralN/ADipole stopped in the morning?
Dipole restart at 10. Shoudl be checked with time stamp.
Fixede786s Fri Jun 28 10:00:25 2019 OttoRunRUNGeneralCryo-3HeRun75
*************************
Run : 75
*************************

Start at 2019-06-28 09:59:59.505010

target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted

AGATA position +51-25 mm

TRIGGER :
- CMG
- M2
- VAMOS%1000
- CATS%1000

all X on MUVI disabled from trigger
all threshold of MUVI1 at 1 MeV to avoid as much as possible noise

CATS threshold raised

Stop at 2019-06-28 14:33:09.843668
Stop comment:
Fixede786s Fri Jun 28 15:25:12 2019 OttoRunRUNGeneralCryo-3HeRun76
*************************
Run : 76
*************************

Start at 2019-06-28 15:24:51.867946

target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted

AGATA position +51-25 mm

TRIGGER :
- CMG
- M2
- VAMOS%1000
- CATS%1000

all X on MUVI disabled from trigger
all threshold of MUVI1 at 1 MeV to avoid as much as possible noise

beam back. Q pole issues still

Stop at 2019-06-28 16:55:27.258794
Stop comment: