Fixed | e786s |
Wed Jun 26 15:50:36 2019 |
OttoRun | RUN | General | Cryo-3He | Run60 |
OttoRun wrote: |
OttoRun wrote: | *************************
Run : 60
*************************
Start at 2019-06-26 10:21:51.572979
target 700mb
Brho0=0.95 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted
AGATA position +51-25 mm
TRIGGER :
- VAMOS %1000
- CMG
- MM
- CATS2 %1000
Restarted GTS Tree by Emmanuel |
|
No Beam since 15:00. According to PCP there will be no beam for a while. |
Fixed | e786s |
Wed Jun 26 16:59:07 2019 |
Louis | BEAM | General | Cryo-3He | Beam interuption |
Beam stopped from 15h to 17h due to source instabilities |
Fixed | e786s |
Wed Jun 26 18:12:18 2019 |
OttoRun | RUN | General | Cryo-3He | Run62 |
*************************
Run : 62
*************************
Start at 2019-06-26 18:09:42.882502
target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted
AGATA position +51-25 mm
TRIGGER :
- VAMOS %1000
- CMG
- MM
- CATS2 %1000
Crate in System faillure.
Moving FIFO FPMW to other crate. Had to change the cable (3ns to 5ns).
To be accounted for in the TOF
Stop at 2019-06-27 02:19:17.492600
Stop comment: Problems with AGATA, run stopped |
Fixed | e786s |
Wed Jun 26 20:46:20 2019 |
Sylvain, Louis | RUN | General | Cryo-3He | run 62 screenshots |
|
Fixed | e786s |
Thu Jun 27 00:26:37 2019 |
mugast | OFFLINE | General | | statistics until now |
Maybe some stat still to recover from better gate and mass and Z reconstruction.
Andrea |
Fixed | e786s |
Thu Jun 27 03:33:26 2019 |
OttoRun | RUN | General | Cryo-3He | Run63 |
*************************
Run : 63
*************************
Start at 2019-06-27 03:33:09.910218
target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted
AGATA position +51-25 mm
TRIGGER :
- VAMOS %1000
- CMG
- MM
- CATS2 %1000
Crate in System faillure.
Moving FIFO FPMW to other crate. Had to change the cable (3ns to 5ns).
To be accounted for in the TOF
%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
Current ~ 30 kHz.
The cryo target pressure and temperature have slightly increased during the stop for the recover of AGATA, even before the refilling of the LHe tank.
It is probably due to a problem in the refilling of liquid nitrogen.
Stop at 2019-06-27 11:46:17.108352
Stop comment: stop |
Fixed | e786s |
Thu Jun 27 12:28:58 2019 |
OttoRun | RUN | General | Cryo-3He | Run67 |
*************************
Run : 67
*************************
Start at 2019-06-27 12:28:26.879061
target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted
AGATA position +51-25 mm
TRIGGER :
- VAMOS %1000
- CMG
- MM
- CATS2 %1000
Crate in System faillure.
Moving FIFO FPMW to other crate. Had to change the cable (3ns to 5ns).
To be accounted for in the TOF
Stop at 2019-06-27 12:46:41.949098
Stop comment: |
Fixed | e786s |
Thu Jun 27 16:35:45 2019 |
OttoRun | RUN | General | Cryo-3He | Run68 |
*************************
Run : 68
*************************
Start at 2019-06-27 16:35:24.351472
target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted
AGATA position +51-25 mm
TRIGGER :
- VAMOS %1000
- CMG
- MM
- CATS2 %1000 |
Fixed | e786s |
Thu Jun 27 16:41:55 2019 |
OttoRun | RUN | General | Cryo-3He | Run65 |
*************************
Run : 65
*************************
Start at 2019-06-27 11:35:21.078920
target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted
AGATA position +51-25 mm
TRIGGER :
- VAMOS %1000
- CMG
- MM
- CATS2 %1000
Crate in System faillure.
Moving FIFO FPMW to other crate. Had to change the cable (3ns to 5ns).
To be accounted for in the TOF
Stop at 2019-06-27 17:39:22.529443
Stop comment: |
Fixed | e786s |
Thu Jun 27 17:43:15 2019 |
OttoRun | RUN | General | Cryo-3He | Run69 |
*************************
Run : 69
*************************
Start at 2019-06-27 17:42:38.280555
target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted
AGATA position +51-25 mm
TRIGGER :
- CMG
Test with only MUGAST back trigger
Stop at 2019-06-27 17:54:02.829540
Stop comment: |
Fixed | e786s |
Thu Jun 27 17:59:20 2019 |
OttoRun | RUN | General | Cryo-3He | Run70 |
*************************
Run : 70
*************************
Start at 2019-06-27 17:58:29.855367
target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted
AGATA position +51-25 mm
TRIGGER :
- CMG
- M2
- VAMOS%1000
- CATS%1000
back to normal
Stop at 2019-06-27 18:37:51.816766
Stop comment: type your message here |
Fixed | e786s |
Thu Jun 27 18:48:51 2019 |
OttoRun | RUN | General | Cryo-3He | Run71 |
*************************
Run : 71
*************************
Start at 2019-06-27 18:47:56.171341
target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted
AGATA position +51-25 mm
TRIGGER :
- MG
- square (not biased)
- M2
all X on MUVI disabled from trigger
all threshold of MUVI1 at 1 MeV to avoid as much as possible noise
Stop at 2019-06-27 20:23:56.630240
Stop comment: |
Fixed | e786s |
Thu Jun 27 20:26:27 2019 |
OttoRun | RUN | General | Cryo-3He | Run73 |
*************************
Run : 73
*************************
Start at 2019-06-27 20:26:17.916314
target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted
AGATA position +51-25 mm
TRIGGER :
- CMG
- M2
- VAMOS%1000
- CATS%1000
all X on MUVI disabled from trigger
all threshold of MUVI1 at 1 MeV to avoid as much as possible noise
There is no run 72
Stop at 2019-06-27 23:04:45.884452
Stop comment: AGATA DAQ is rejectting all events |
Fixed | e786s |
Thu Jun 27 23:28:30 2019 |
OttoRun | RUN | General | Cryo-3He | Run74 |
*************************
Run : 74
*************************
Start at 2019-06-27 23:25:00.455575
target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted
AGATA position +51-25 mm
TRIGGER :
- CMG
- M2
- VAMOS%1000
- CATS%1000
all X on MUVI disabled from trigger
all threshold of MUVI1 at 1 MeV to avoid as much as possible noise
Stop at 2019-06-28 09:31:40.775271
Stop comment: |
Fixed | e786s |
Fri Jun 28 04:21:48 2019 |
Daniele Mengoni | RUN | General | Cryo-3He | run 74 |
screenshots run 74 attached
frequent error message on the FP DC Wi1 --> defqlt. The elog said so, card tripped, but spectra are still present .. puzzling |
Fixed | e786s |
Fri Jun 28 08:54:50 2019 |
Freddy | RUN | General | Cryo-3He | run 74 |
Daniele Mengoni wrote: | screenshots run 74 attached
frequent error message on the FP DC Wi1 --> defqlt. The elog said so, card tripped, but spectra are still present .. puzzling |
Put the voltage back at 8h00 when i arrived. |
Fixed | e786s |
Fri Jun 28 08:57:21 2019 |
Freddy | CATS | General | Cryo-3He | CATS scaler not compatible with beam rate |
The CATS scaler si now around 100-120 kHz while the beam rate is still 40 kHz.
I checked the CATS signal and there is clearly noise who induce double logic signals. Not all of them are validated by FAG.
We should raise the threshold- call sebastien
9h32 : Stop beam to enter and raise the threshold. |
Fixed | e786s |
Fri Jun 28 09:56:59 2019 |
Bertrand, Valerian and Adrien | BEAM | General | N/A | Dipole stopped in the morning? |
Dipole restart at 10. Shoudl be checked with time stamp. |
Fixed | e786s |
Fri Jun 28 10:00:25 2019 |
OttoRun | RUN | General | Cryo-3He | Run75 |
*************************
Run : 75
*************************
Start at 2019-06-28 09:59:59.505010
target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted
AGATA position +51-25 mm
TRIGGER :
- CMG
- M2
- VAMOS%1000
- CATS%1000
all X on MUVI disabled from trigger
all threshold of MUVI1 at 1 MeV to avoid as much as possible noise
CATS threshold raised
Stop at 2019-06-28 14:33:09.843668
Stop comment: |
Fixed | e786s |
Fri Jun 28 15:25:12 2019 |
OttoRun | RUN | General | Cryo-3He | Run76 |
*************************
Run : 76
*************************
Start at 2019-06-28 15:24:51.867946
target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted
AGATA position +51-25 mm
TRIGGER :
- CMG
- M2
- VAMOS%1000
- CATS%1000
all X on MUVI disabled from trigger
all threshold of MUVI1 at 1 MeV to avoid as much as possible noise
beam back. Q pole issues still
Stop at 2019-06-28 16:55:27.258794
Stop comment: |