MUST2 SAMURAI_2018 S3-LEB-LPC SUPERNEMO MUGAST EXPAND SCALP GALATRON HiCARI VELO
  MUGAST, Page 77 of 132  Not logged in MUGAST
Fixed, e744s, Sat Apr 13 15:25:56 2019, Marion, Kseniia, Dieter, Didier, Guiseppe, IuliaH, Catalin, ONLINE, General, CH2, Scalers v8 for run122.1 run122_1_scalers_v8.png
"1_" added to screenshot filenames to indicate automatic file opening
Fixed, e744s, Sat Apr 13 17:03:14 2019, Jérémie & François, OFFLINE, General, , first gamma spectrum 35gamma.png
After a few hours run.
Fixed, e744s, Sat Apr 13 17:36:39 2019, mugast, RUN, General, CH2, run 124 
16:30 start run 124

brho is being set
finger slightly changed
Fixed, e744s, Sat Apr 13 18:00:12 2019, Sylvain, RUN, General, CH2, Log & Scalers -- Run 122 at 17:45 20x
Log for run 122 taken at 17h45
Fixed, e744s, Sat Apr 13 19:27:16 2019, Beyhan, Jaromir, OTHER, General, N/A, Spreadsheet with a list of runs up to 122  runs_list.odsruns_list.odsruns_list.odsruns_list.odsruns_list.ods
Resides also in home directory
runs_list.ods

updated with run 121 and vamos %30000 for 122

123 124 125 started

126-134 listed
Fixed, e744s, Sat Apr 13 20:02:36 2019, Sylvain, RUN, General, CH2, Log & Scalers -- Run 122 at 19:45 21x
Log & Scalers -- Run 122 at 19:45
Fixed, e744s, Sat Apr 13 20:16:05 2019, Diego, VAMOS, General, , Vamos Efficiency 
***************************************RUN 118************************************************

Forward (condition by before)
IC 0 95.6667%
IC 1 95.4542%
IC 2 95.513%
IC 3 93.2396%
----------------------
Backward (condition by after)
IC 1 95.9352%
IC 2 94.2267%
IC 3 96.2866%
IC 4 83.0393% //not reliable
----------------------
DC (condition by IC[0])
DC 0 93.0557%
DC 1 92.6187%
DC 2 88.9858%
DC 3 96.6774%
----------------------
DC X (condition by IC[0])
DCX0 93.707%
DCX1 93.0192%
DCX2 94.2166%
DCX3 94.1586%
----------------------
Xf (condition by IC[0])
Xf 95.0046%
----------------------


***************************************RUN 119************************************************
Forward (condition by before)
IC 0 95.2904%
IC 1 95.071%
IC 2 94.9271%
IC 3 93.1055%
----------------------
Backward (condition by after)
IC 1 95.0696%
IC 2 91.6524%
IC 3 90.644%
IC 4 81.9769% //not reliable
----------------------
DC (condition by IC[0])
DC 0 90.7155%
DC 1 90.3863%
DC 2 87.0165%
DC 3 94.7584%
----------------------
DC X (condition by IC[0])
DCX0 88.1504%
DCX1 87.6396%
DCX2 88.7515%
DCX3 88.2139%
----------------------
Xf (condition by IC[0])
Xf 83.22%
----------------------

***************************************RUN 122_0************************************************

Forward (condition by before)
IC 0 95.2443%
IC 1 95.2177%
IC 2 94.8111%
IC 3 92.0637%
----------------------
Backward (condition by after)
IC 1 95.3114%
IC 2 91.8716%
IC 3 90.9907%
IC 4 79.8954% //not reliable
----------------------
DC (condition by IC[0])
DC 0 87.9765%
DC 1 87.8112%
DC 2 83.8229%
DC 3 93.3538%
----------------------
DC X (condition by IC[0])
DCX0 86.4654%
DCX1 86.8069%
DCX2 87.0654%
DCX3 86.5511%
----------------------
Xf (condition by IC[0])
Xf 80.1623%
----------------------

***************************************RUN 122_1************************************************
Forward (condition by before)
IC 0 94.9153%
IC 1 94.9217%
IC 2 94.5962%
IC 3 91.8891%
----------------------
Backward (condition by after)
IC 1 95.0787%
IC 2 91.7874%
IC 3 91.1441%
IC 4 79.9685% //not reliable
----------------------
DC (condition by IC[0])
DC 0 87.4807%
DC 1 87.3732%
DC 2 83.3638%
DC 3 92.8772%
----------------------
DC X (condition by IC[0])
DCX0 85.8206%
DCX1 86.2137%
DCX2 86.5489%
DCX3 85.9571%
----------------------
Xf (condition by IC[0])
Xf 79.0635%
----------------------

There is a continuous decrease of efficiency in the Drift Chamber. The efficiency of IC is also not optimal. I suspect that with this beam intensity (5x10^4 in VAMOS) detectors may start to suffer some damage. If the efficiency keeps decreasing we need to consider to reduce the beam intensity.

In order to check the efficiency, there is a script in /home/mugast/Analysis_e744 called EffCheck.C

Open a root file: root -l RootA/r0XXX_XXXa.root
Load the script: .L EffCheck.C
Launch the script: EffCheck() (the calculation is automatic)
Fixed, e744s, Sat Apr 13 20:37:16 2019, Beyhan, OTHER, General, , Alarm in G2 
Alarm signal from G2.
No beam since 20:30
PCP will have a look in G2, following the procedures.
We do not stop the run for the moment.

20h50 The alarm is given from the captor on the top of the liquid nitrogen dewar of exogam.
It has most probably being set during the detectors filling.

beam back at 20:59
Fixed, e744s, Sat Apr 13 21:23:28 2019, OttoRun, RUN, General, CH2, Run121 
========================================
*************************
Run : 121
*************************

Start at 2019-04-13 06:16:15.350431

Run Conditions:
  • Operator: Mariluz Nikola Nicolas
  • Trigger: MUGAST + Vamos %1000 + CATS %30000
  • Target #3
  • Beam: 14O
  • Intensity: ~2e4
  • Brho = 0.5199 Tm
  • CATS2 in
  • Finger at +12 mm
  • WITH AGATA


Comment:
Start Target 2: 104 um
VAMOS is Open
AGATA Coupling

Stop at 2019-04-14 14:56:42.341083
Stop comment: end of run
Fixed, e744s, Sat Apr 13 21:34:02 2019, Adrien Matta -> update Iulian, RUN, General, N/A, Inconsistent info in the run message-fix explained in reply 
Be careful to fill in correctly all info in the start run message.

There is inconsistency in run 121 comment (target 2) and list (target 3).

Thèse information will be used for the analysis.

Iulian->
This discrepancy has been addressed
Fixed, e744s, Sat Apr 13 21:47:40 2019, Sylvain, RUN, General, CH2, Log & Scalers -- Run 122 at 21:45 20x
    Ongoing, e744s, Sat Apr 13 21:57:48 2019, Adrien Matta, RUN, General, N/A, Inconsistent info in the run message 

Adrien Matta wrote:
Be careful to fill in correctly all info in the start run message.

There is inconsistency in run 121 comment (target 2) and list (target 3).

Thèse information will be used for the analysis.


--> Well.. in entry 4175 it is written target 3...
... ...
    Fixed, e744s, Sat Apr 13 22:04:32 2019, Adrien Matta, RUN, General, N/A, Inconsistent info in the run message 

Adrien Matta wrote:
Be careful to fill in correctly all info in the start run message.

There is inconsistency in run 121 comment (target 2) and list (target 3).

Thèse information will be used for the analysis.


--> The misleading comment is preseved from run 103,
... the last #2 target was at run 110
... Iulian is removing the comments on #2 target that appear untill 121
    Fixed, e744s, Sat Apr 13 22:10:37 2019, OttoRun, RUN, General, CH2, Run121 

OttoRun wrote:
*************************
Run : 121
*************************

Start at 2019-04-13 06:16:15.350431

Run Conditions:
  • Operator: Mariluz Nikola Nicolas
  • Trigger: MUGAST + Vamos %1000 + CATS %30000
  • Target #3
  • Beam: 14O
  • Intensity: ~2e4
  • Brho = 0.5199 Tm
  • CATS2 in
  • Finger at +12 mm
  • WITH AGATA


Comment:
Start Target 2: 104 um
VAMOS is Open
AGATA Coupling
Fixed, e744s, Sat Apr 13 22:40:00 2019, Diego, VAMOS, General, , Saturation in the WIRES of Drift chamber due to Pile-Up DC_Wire_saturation.png
There is Saturation in the Energy of the WIRES due to the Pile-up, it is not urgent to correct it because these parameters are not needed for the reconstruction. Tomorrow morning, when stopping the run, we should change the offsets in NUMEXO in order to better allocate the amplitude of the signals.
Fixed, e744s, Sat Apr 13 23:02:27 2019, Valerian, Freddy and Iulian, OFFLINE, General, CH2, Very preliminary results for two protons TwoProtons_run122_0_1.png
Join are very preliminary spectra for the two protons events gated on 13N in VAMOS with a multiplicity 2 event in MUST2.
    Fixed, e744s, Sat Apr 13 23:25:14 2019, iulian, RUN, General, N/A, Inconsistent info in the run message 
This inconsistency has been addressed

Adrien Matta wrote:

Adrien Matta wrote:
Be careful to fill in correctly all info in the start run message.

There is inconsistency in run 121 comment (target 2) and list (target 3).

Thèse information will be used for the analysis.


--> Well.. in entry 4175 it is written target 3...
... ...
Fixed, e744s, Sun Apr 14 00:06:05 2019, Chloé, Fairouz, Sylvain, Damien, RUN, General, CH2, Log & Scalers -- Run 122 at 00:06 20x
Fixed, e744s, Sun Apr 14 02:04:32 2019, Faïrouz, Chloé, Sylvain & Damien, RUN, General, CH2, Beam lost 19x
Beam stopped because of the HF of C01.
L'astreinte va venir. Arrêt estimé une demi-heure ou plus.
On a remarqué sur le spectre impact qu'on a perdu deux strips de plus sur le MUST2 haut-gauche à partir du run 120 6h00 du matin
Fixed, e744s, Sun Apr 14 04:04:33 2019, Faïrouz, Chloé, Sylvain, Damien & Iulia, RUN, General, CH2, Beam is back @ 3e4 pps in CATS run122_1_scalers_v13.png

ELOG Home Page