MUST2 SAMURAI_2018 S3-LEB-LPC SUPERNEMO MUGAST EXPAND SCALP GALATRON HiCARI VELO
  MUGAST, Page 69 of 127  Not logged in MUGAST
Fixed, e768s, Fri Jul 12 17:51:06 2019, mugast (Sylvain, Chloe, François, Marlène, Nicolas), OTHER, General, , Monitoring of beam profile with BAF2 
In order to provide a efficient way for PCP to monitor beam profile a new TAC measurement has set up.

A BAF2 detector has been inserted close to the reaction chamber (in position 7 of AGATA structure frame). Operating voltage of BaF2 is 1800V and the typical signal amplitude is above 100 mV.
The output signal is sent to a TFA (with gain =1). No additional delay was used. The BAF2 trigger signal is made from CFD and used as a start a TAC. TAC range was chosen to 1us.

The trigger signal was also sent to GMT trough input nb 5.




Fixed, e768s, Sat Jul 13 16:28:30 2019, Diego Ramos, OTHER, General, N/A, Proton ToF from VAMOS ToF reconstruction TOF_VAMOS.png
Comparison between the Proton ToF measured directly with the TAC between MUST and CATS (left) and the one recontructed from the time diference between the time MUGAST-VAMOS and the time Target-VAMOS (recontructed from the Brho) (right)

The resolution of the recontructed time looks slightly degraded with respect to the direct measurement but it is certainly much better than using the HF. Still, the resolution of the reconstructed time may be improved with a fine alignment of the different sections of the PPAC and with a better selection of the M/q in VAMOS.
Fixed, e768s, Mon Jul 15 12:57:53 2019, Aranaud, Nicolas, Marlène, François, TARGET, General, LiF, Targets position 6x
Targets have been mounted on target ladder and are now in reaction chamber
Fixed, e768s, Mon Jul 15 14:29:47 2019, Marlène, MUST2, General, , Masks for MUST2 & Kapton changed for MG5 IMG_20190715_121646_crop_flip.jpgIMG_20190715_122035.jpg
Masks are now correctly mounted

All 4 kaptons from MG5 have been changed
Fixed, e768s, Mon Jul 15 18:05:14 2019, Iulian, Serge, RUN, General, N/A, Run7 - TAC calibration 
*************************
Run : 7
*************************

Start at 2019-07-15 18:04:51.105052

TAC Time calibrator

range 2.56 um
period 20ns


TACs range settings:
1. VAMOS - AGATA -> 500 ns
2. FAG - HF -> 200 ns
3. CATAS1 - HF -> 500 ns
4. VAMOS - HF -> 2 us
5. MUGAST - VAMOS -> 2 us
6. MUST2 - AGATA -> 2 us
7. MUST2 - HF -> 2 us
8. BAF2 - HF -> 1 us
9. AGATA - HF -> 200 ns

Stop at 2019-07-15 18:16:11.458928
Stop comment: Finish 1st time calibration. Changing ranges now
Fixed, e768s, Mon Jul 15 18:17:49 2019, Iulian, Serge, RUN, General, N/A, Run8 - TAC calibrations 
*************************
Run : 8
*************************

Start at 2019-07-15 18:17:21.942994

TAC Time calibrator

range 2.56 um

10 ns period
Fixed, e768s, Mon Jul 15 18:39:03 2019, Iulian, Serge, RUN, General, N/A, Run9 - TAC calibration 
*************************
Run : 9
*************************

Start at 2019-07-15 18:38:59.133455

TAC Time calibrator

range 2.56 um

10 ns period

Stop at 2019-07-15 19:04:03.255467
Stop comment: type your message here

Stop at 2019-07-15 19:18:09.593144
Stop comment: end of tac time calibration
Fixed, e768s, Mon Jul 15 19:56:52 2019, OttoRun, RUN, General, 3-alpha, Run6 
*************************
Run : 6
*************************

Start at 2019-07-15 17:39:19.237140

3 alpha calibration run
+ MM2 masks
+ suppressed pedestals (new pedestal for all trap detectors)

Square detectors:
+ MG10: not biased
+ MG9 removed
Fixed, e744s, Mon Jul 15 19:57:39 2019, OttoTime, RUN, General, N/A, TimeCalibratorRun 
*** Run Time Calibrator FSC ***
- Run: 144
- MUVI: MUVI1
- Side : MUFEE X (Junction)
- Telescope: 1,
Fixed, e768s, Mon Jul 15 19:59:30 2019, Marlène, Nicolas, Serge, RUN, General, N/A, Run10 - fsc 
*************************
Run : 10
*************************

Start at 2019-07-15 19:57:41.031967

fsc / calibT run

GMT: MUGAST (square detector not biased) + MUST2

Time calibrator:
range -> 640 ns
freq. -> 20 ns
Fixed, e744s, Mon Jul 15 20:03:50 2019, OttoTime, RUN, General, N/A, TimeCalibratorRun 
*** Run Time Calibrator FSC ***
- Run: 144
- MUVI: MUVI1
- Side : MUFEE Y (Ohmic)
- Telescope: 1,
Fixed, e768s, Mon Jul 15 20:07:48 2019, OttoRun, RUN, General, N/A, Run11 
*************************
Run : 11
*************************

Start at 2019-07-15 20:03:52.664898

fsc / calibT run

GMT: MUGAST (square detector not biased) + MUST2

Time calibrator:
range -> 640 ns
freq. -> 20 ns

Stop at 2019-07-15 20:09:19.003266
Stop comment: stop because too many overflow and too little statistics
Fixed, e744s, Mon Jul 15 20:11:55 2019, OttoTime, RUN, General, N/A, TimeCalibratorRun 
*** Run Time Calibrator FSC ***
- Run: 144
- MUVI: MUVI1
- Side : MUFEE X (Junction)
- Telescope: 1,
Fixed, e768s, Mon Jul 15 20:12:30 2019, OttoRun, RUN, General, N/A, Run12 
*************************
Run : 12
*************************

Start at 2019-07-15 20:11:56.531359

fsc / calibT run

internal pulser frequency now set to 300 Hz (was about 1 kHz before)

GMT: MUGAST (square detector not biased) + MUST2

Time calibrator:
range -> 640 ns
freq. -> 20 ns

Stop at 2019-07-15 20:17:56.763955
Stop comment:
Fixed, e744s, Mon Jul 15 20:17:58 2019, OttoTime, RUN, General, N/A, TimeCalibratorRun 
*** Run Time Calibrator FSC ***
- Run: 144
- MUVI: MUVI1
- Side : MUFEE Y (Ohmic)
- Telescope: 1,
Fixed, e768s, Mon Jul 15 20:18:07 2019, Marlène, Nicolas, Serge, RUN, General, N/A, Run13 - fsc 
*************************
Run : 13
*************************

Start at 2019-07-15 20:18:02.650063

fsc / calibT run

internal pulser frequency now set to 300 Hz (was about 1 kHz before)

GMT: MUGAST (square detector not biased) + MUST2

Time calibrator:
range -> 640 ns
freq. -> 20 ns

Stop at 2019-07-15 20:23:54.304478
Stop comment: type your message here
Fixed, e744s, Mon Jul 15 20:23:56 2019, OttoTime, RUN, General, N/A, TimeCalibratorRun 
*** Run Time Calibrator FSC ***
- Run: 144
- MUVI: MUVI2
- Side : MUFEE X (Junction)
- Telescope: 1,
Fixed, e768s, Mon Jul 15 20:24:01 2019, OttoRun, RUN, General, N/A, Run14 
*************************
Run : 14
*************************

Start at 2019-07-15 20:23:58.790182

fsc / calibT run

internal pulser frequency now set to 300 Hz (was about 1 kHz before)

GMT: MUGAST (square detector not biased) + MUST2

Time calibrator:
range -> 640 ns
freq. -> 20 ns

Stop at 2019-07-15 20:30:02.695121
Stop comment:
Fixed, e744s, Mon Jul 15 20:29:56 2019, OttoTime, RUN, General, N/A, TimeCalibratorRun 
*** Run Time Calibrator FSC ***
- Run: 144
- MUVI: MUVI2
- Side : MUFEE Y (Ohmic)
- Telescope: 1,
Fixed, e768s, Mon Jul 15 20:30:04 2019, OttoRun, RUN, General, N/A, Run15 
*************************
Run : 15
*************************

Start at 2019-07-15 20:29:59.325702

fsc / calibT run

internal pulser frequency now set to 300 Hz (was about 1 kHz before)

GMT: MUGAST (square detector not biased) + MUST2

Time calibrator:
range -> 640 ns
freq. -> 20 ns

Stop at 2019-07-15 20:35:53.451006
Stop comment: type your message here

ELOG Home Page