MUST2
SAMURAI_2018
S3-LEB-LPC
SUPERNEMO
MUGAST
EXPAND
SCALP
GALATRON
HiCARI
VELO
|
MUGAST |
Not logged in |
|
|
Message ID: 4949
Entry time: Sat Jul 13 16:28:30 2019
|
Author: |
Diego Ramos |
ExpNbr: |
e768s |
Type: |
OTHER |
Category: |
General |
Status: |
Fixed |
Target-Source: |
N/A |
Subject: |
Proton ToF from VAMOS ToF reconstruction |
|
|
Comparison between the Proton ToF measured directly with the TAC between MUST and CATS (left) and the one recontructed from the time diference between the time MUGAST-VAMOS and the time Target-VAMOS (recontructed from the Brho) (right)
The resolution of the recontructed time looks slightly degraded with respect to the direct measurement but it is certainly much better than using the HF. Still, the resolution of the reconstructed time may be improved with a fine alignment of the different sections of the PPAC and with a better selection of the M/q in VAMOS. |
|
|