Status |
ExpNbr |
Date |
Author |
Type |
Category |
Target-Source |
Subject |
Fixed | e744s |
Thu Apr 18 17:36:56 2019 |
OttoRun | RUN | General | N/A | Run228 |
*************************
Run : 228
*************************
Start at 2019-04-18 17:36:35.317571
MUGAST alpha run no pedestal supression
Stop at 2019-04-18 17:42:58.195742
Stop comment:
This pedestal run was done as a reference, the pedestal were also saved using das and are exactly the same as the one save the 3/04.
This run and all calibration run can be found in /data/mugastX/mugast |
Fixed | e744s |
Thu Apr 18 17:45:45 2019 |
Valerian | OTHER | Hardware | N/A | Start heating and stopping Voltage + Acq |
HV+BT are off and heating is started for Arnaud to open the chamber |
Fixed | e744s |
Wed Apr 24 15:01:01 2019 |
Valerian | OTHER | Hardware | N/A | Electronic scheme on the MUGAST side after experiment e744 |
Attached are schematics of the electronics. |
Fixed | e744s |
Fri Apr 26 11:29:02 2019 |
Valerian | OTHER | Software | N/A | Notes about the electronic |
While doing the electronic scheme, I saw that the Inspection line was inducing a bad signal in the coincidence module between fag and CATS1. See joined image.
Also, I broke a cable between CATS1 Fifo and the delay box, the delay is now 2 ns longer. |
Fixed | TEST |
Mon May 27 16:17:48 2019 |
OttoRun | RUN | General | 3-alpha | Run229 |
*************************
Run : 229
*************************
Start at 2019-05-27 16:17:16.001351
MUGAST alpha run no pedestal supression
test run qfter restarting system. temp is still qround 15deg
Stop at 2019-05-27 16:50:50.508660
Stop comment: source was not in correct position
EDIT: there is no data for this run; the entry was create because the scaler storage was on.
Stop at 2019-05-27 19:09:53.235957
Stop comment: type your message here
Stop at 2019-05-27 19:31:47.881701
Stop comment: type your message here |
Fixed | TEST |
Tue May 28 10:43:58 2019 |
OttoRun | RUN | General | 3-alpha | Run229 |
*************************
Run : 229
*************************
Start at 2019-05-28 10:43:42.122755
MUGAST alpha run no pedestal supression
test run after restarting system. temp is still around 5deg
Stop at 2019-05-28 14:47:46.808600
Stop comment: type your message here
Stop at 2019-05-28 15:22:40.074144
Stop comment: type your message here
Stop at 2019-05-28 15:57:47.785333
Stop comment: type your message here
Stop at 2019-05-28 16:59:06.325170
Stop comment: type your message here
Stop at 2019-05-28 17:14:12.527918
Stop comment: type your message here
Stop at 2019-05-28 17:34:19.526635
Stop comment:
Stop at 2019-05-28 17:52:55.515837
Stop comment: type your message here
Stop at 2019-05-28 18:17:09.389775
Stop comment: type your message here |
Fixed | TEST |
Tue May 28 11:08:10 2019 |
mugast | MUST2 | Hardware | 3-alpha | MG3 is tripping |
MG3T is tripping, tryied to set max current to 20uA but it still trip around 30V. |
Fixed | TEST |
Thu May 30 15:42:13 2019 |
OttoRun | RUN | General | 3-alpha | Run229 |
*************************
Run : 229
*************************
Start at 2019-05-30 15:41:17.803744
MUGAST alpha run pedestal supression
temperature -3 deg (trap) & -4 deg (must2) consigne 5 deg
Stop at 2019-05-30 15:49:50.743195
Stop comment: type your message here |
Fixed | TEST |
Thu May 30 15:55:15 2019 |
OttoRun | RUN | General | 3-alpha | Run229 |
*************************
Run : 229
*************************
Start at 2019-05-30 15:54:16.275808
MUGAST 3alpha run pedestal supression
temperature -3 deg (trap) & -4 deg (must2) consigne 5 deg
no pedestal suppression (but alignement)
Unsual bias for MG4T =130V, I=4 uA because no trigger on Y.
Stop at 2019-05-30 15:57:43.637854
Stop comment: type your message here |
Fixed | TEST |
Thu May 30 15:58:20 2019 |
OttoRun | RUN | General | 3-alpha | Run230 |
*************************
Run : 230
*************************
Start at 2019-05-30 15:57:54.390230
MUGAST 3alpha run pedestal supression
temperature -3 deg (trap) & -4 deg (must2) consigne 5 deg
no pedestal suppression (but alignement)
Unsual bias for MG4T =130V, I=4 uA because no trigger on Y.
same as before except that MG7 had a mate hidden on Y (now it is on)
Stop at 2019-05-30 15:59:04.950841
Stop comment: type your message here
Stop at 2019-05-30 16:01:48.801380
Stop comment: |
Fixed | TEST |
Thu May 30 16:04:09 2019 |
OttoRun | RUN | General | 3-alpha | Run230 |
*************************
Run : 230
*************************
Start at 2019-05-30 16:04:07.241232
MUGAST 3alpha run pedestal supression
temperature -3 deg (trap) & -4 deg (must2) consigne 5 deg
no pedestal suppression (but alignement)
Unsual bias for MG4T =130V, I=4 uA because no trigger on Y.
same as before except that MG7 had a mate hidden on Y (now it is on)
Stop at 2019-05-30 16:06:21.092742
Stop comment: type your message here |
Fixed | TEST |
Thu May 30 16:08:39 2019 |
OttoRun | RUN | General | 3-alpha | Run230 |
*************************
Run : 230
*************************
Start at 2019-05-30 16:08:37.332238
MUGAST 3alpha run pedestal supression
temperature -3 deg (trap) & -4 deg (must2) consigne 5 deg
no pedestal suppression (but alignement)
Unsual bias for MG4T =130V, I=4 uA because no trigger on Y.
same as before except that MG7 had a mate hidden on Y (now it is on)
Stop at 2019-05-30 16:09:38.806154
Stop comment: type your message here |
Fixed | TEST |
Thu May 30 16:10:18 2019 |
OttoRun | RUN | General | 3-alpha | Run230 |
*************************
Run : 230
*************************
Start at 2019-05-30 16:09:44.953775
MUGAST 3alpha run pedestal supression
temperature -3 deg (trap) & -4 deg (must2) consigne 5 deg
no pedestal suppression (but alignement)
MG4T =95V
same as before except that MG7 had a mate hidden on Y (now it is on)
Stop at 2019-05-31 06:39:11.358615
Stop comment: type your message here |
Fixed | e786s |
Fri May 31 09:53:14 2019 |
Marlène | GRIT | General | | Alpha source --> status of detectors |
MG3T has been fixed by changing the X kapton.
I have removed aluminium foil on square detectors, the noise is slightly higher.
MG4T is not working properly : noisy on X but no signal on Y !! I see no triggers on the Y side. I tried to increase bias but it did increase noise. I checked in the last calibration of e744 it was working properly. Also I had to do pedestal realignement for this detector because everything was shifted.
MG7T seem to be slightly different than before. Strips X are hidden from the trigger because it became suddenly noisy. |
Fixed | e786s |
Mon Jun 3 18:00:10 2019 |
Marlène | GRIT | General | | MUGAST detectors status --> MG4T |
The problem of MG4T ha sbeen solved by changing the feedthrough.
The pulser on Y was not working and we were seeing no events (no ORD) because there was a problem with the bias of board. The issue is due to the feedthrough.
So now MG4T (HT &BT) are plugged in a free feedthrough (below MG1T) and I see pulser signals on Y.
The bias seems to apply normally on all detectors (in the dark).
The pulser is seen on each boards.
I have put back aluminium foil on square but I could not test it. |
Fixed | TEST |
Tue Jun 11 18:59:36 2019 |
OttoRun | RUN | General | 3-alpha | Run231 |
*************************
Run : 231
*************************
Start at 2019-06-11 18:58:31.909158
MUGAST 3alpha run pedestal supression
temperature -1 deg (trap) & -4 deg (must2) consigne 5 deg
no pedestal suppression (but alignement)
alpha source in front of cryogenic target, NOT AT TARGET POSITION !!!
Stop at 2019-06-11 21:40:22.107913
Stop comment: |
Fixed | TEST |
Tue Jun 11 21:44:38 2019 |
OttoRun | RUN | General | 3-alpha | Run232 |
*************************
Run : 232
*************************
Start at 2019-06-11 21:44:02.397316
MUGAST 3alpha run pedestal supression
temperature -1 deg (trap) & -4.5 deg (must2) consigne 5 deg
with pedestal suppression
alpha source in front of cryogenic target, NOT AT TARGET POSITION !!!
Stop at 2019-06-12 10:05:20.279209
Stop comment: type your message here |
Fixed | e786s |
Wed Jun 12 17:59:01 2019 |
mugast | OTHER | Software | | Online Energy Calibration |
Updated Online Energy calibration for MUGAST, based on run_0231 data |
Fixed | e786s |
Wed Jun 12 18:11:04 2019 |
mugast | OTHER | Software | | Online Energy Calibration |
|
Fixed | TEST |
Wed Jun 12 18:40:57 2019 |
OttoRun | RUN | General | 3-alpha | Run233 |
*************************
Run : 233
*************************
Start at 2019-06-12 18:40:34.500264
MUGAST 3alpha run pedestal supression
temperature -1 deg (trap) & -4.5 deg (must2) consigne 5 deg
without pedestal suppression
alpha source in front of cryogenic target, NOT AT TARGET POSITION !!!
Stop at 2019-06-13 09:06:36.277090
Stop comment: |