Message ID: 4489     Entry time: Fri May 31 09:53:14 2019
Author: Marlène 
ExpNbr: e786s 
Type: GRIT 
Category: General 
Status: Fixed 
Target-Source:  
Subject: Alpha source --> status of detectors 
MG3T has been fixed by changing the X kapton.

I have removed aluminium foil on square detectors, the noise is slightly higher.

MG4T is not working properly : noisy on X but no signal on Y !! I see no triggers on the Y side. I tried to increase bias but it did increase noise. I checked in the last calibration of e744 it was working properly. Also I had to do pedestal realignement for this detector because everything was shifted.

MG7T seem to be slightly different than before. Strips X are hidden from the trigger because it became suddenly noisy.