Fixed, e768s, Fri Feb 14 11:59:24 2020, OttoTime, RUN, General, N/A, TimeCalibratorRun
|
*** Run Time Calibrator FSC ***
- Run: 506
- MUVI: MUVI3
|
Fixed, e768s, Fri Feb 14 11:47:31 2020, OttoTime, RUN, General, N/A, TimeCalibratorRun
|
*** Run Time Calibrator FSC ***
- Run: 505
- MUVI: MUVI2
|
Fixed, e768s, Fri Feb 14 11:35:44 2020, OttoTime, RUN, General, N/A, TimeCalibratorRun
|
*** Run Time Calibrator FSC ***
- Run: 504
- MUVI: MUVI2
|
Fixed, e768s, Fri Feb 14 11:23:56 2020, OttoTime, RUN, General, N/A, TimeCalibratorRun
|
*** Run Time Calibrator FSC ***
- Run: 503
- MUVI: MUVI1
|
Fixed, e768s, Fri Feb 14 11:11:57 2020, OttoTime, RUN, General, N/A, TimeCalibratorRun
|
*** Run Time Calibrator FSC ***
- Run: 502
- MUVI: MUVI1
|
Fixed, TEST, Wed Feb 12 09:32:17 2020, mugast, RUN, General, 3-alpha, Run 500 - 3alpha - mugast standalone 
|
RUN 500
start: 11/02/2020 @ 19h00
|
Fixed, TEST, Fri Feb 14 10:49:22 2020, Marlène & Franco, RUN, General, 3-alpha, MG11 issues
|
We analyzed the behaviour of the annular. We realized that the 4-peaks behaviour on X was already present last year but only on one sector, now it's on
two sectors. We noticed that there is a correlation between the malfunctioning sectors and the missing strips on Y. Last year there was only a missing
strip on Y, affecting only one sector on X; now 3 strips are missing on Y, affecting 2 sectors on X. We believe that the fact that the boundings are missing |
Fixed, , Fri Feb 14 10:26:46 2020, Franco&Marlène, GRIT, Hardware, , HT BT screenshots before e775s    
|
Detectors are cooled down to 0 deg. |
Fixed, TEST, Wed Feb 12 16:24:52 2020, Marlène, Freddy & Franco, MUST2, Hardware, , Modifications on the electronic scheme
|
We cleaned the electronic scheme.
Now the HF CLEAN VAMOS is taken directly from VAMOS and not re-cleaned by us.
TACs are all ok. Inspection lines have been checked and labeled.
|
Fixed, TEST, Mon Feb 10 18:42:23 2020, Franco&Marlène, GRIT, Hardware, , New MUGAST detectors MG2 MG6
|
MG2 and MG6 have been mounted in the chamber.
MG2 reference number is 3227_10 thickness=468um
|
Fixed, TEST, Thu Jan 23 14:06:51 2020, Franco&Marlène, GRIT, Hardware, 3-alpha, Status of detectors on 23-01-20 6x
|
Short alpha source run (not saved on disk) to test the detectors.
Summary :
|
Fixed, , Wed Jan 22 18:39:54 2020, Marlène & Franco, GRIT, Hardware, , Tests of the new box
|
The new box has been tested together with the new CAEN SY4527 power supply. We connected a laptop with Linux to the power supply by means of an ethernet
cable and communicated with it by opening a browser (http://192.168.0.1, then username: admin and pwd: admin). To control the channels we open a terminal,
go to the folder where we saved the file ChannelsController.jar and run the command: java -jar ChannelsController.jar 192.168.0.1 admin admin
|
Fixed, TEST, Wed Jan 22 15:53:25 2020, Marlène, GRIT, Hardware, , Changes for MUGAST2020
|
Changes were made for MUGAST 2020 :
1- the board for MG10 before (2nd square) is now for MG6 new trap
2- the board for MG9 is put vertical and used for the MUST2 at 90 deg
|
Pending, TEST, Tue Jan 21 17:55:12 2020, Marlène, MUST2, Hardware, , Ovv error on the -5V of CAEN boards   
|
The Ovv error shows on the -5V (first one) of the 3 board (5,3.3,10V for MUST2§squares)
It means that the voltage is over the asked value.
If we ask 5V, we get 5.5V in reality
|
Fixed, TEST, Wed Dec 4 19:19:38 2019, Marlène, GRIT, Hardware, , Modifications of set-up
|
We have installed the new MUFEE board in the reaction chamber for MG2. This board is a new version with ATHED.
We have moved the MG9 board to become the board to read the future MUST2 detector at 90 deg. This board is now perpendicular to the other ones. We have |
Fixed, TEST, Tue Dec 3 18:14:45 2019, Marlène & Franco, GRIT, Hardware, , Pulser test of new board
|
The new board that will be installed for MG2 has been tested and validated with a pulser outside of the chamber. |
Fixed, TEST, Thu Oct 17 12:01:09 2019, Franco & Marlène, GRIT, Hardware, 3-alpha, Test of configuration with 4 MUVI and 3 GAMERS 
|
The electronics set-up for MUGAST2020 has been tested with 4 MUVI on 4 VXI crates, 1GMT for MASTER and 3 GAMERS for slaves.
Configuration file is named mugast.das
|
Fixed, e768s, Sat Aug 3 10:21:04 2019, OttoTime, RUN, General, N/A, TimeCalibratorRun
|
*** Run Time Calibrator FSC ***
- Run: 161
- MUVI: MUVI3
|
Fixed, e768s, Sat Aug 3 09:17:12 2019, OttoTime, RUN, General, N/A, TimeCalibratorRun
|
*** Run Time Calibrator FSC ***
- Run: 161
- MUVI: MUVI3
|
Fixed, e768s, Sat Aug 3 08:13:33 2019, OttoTime, RUN, General, N/A, TimeCalibratorRun
|
*** Run Time Calibrator FSC ***
- Run: 161
- MUVI: MUVI3
|
Fixed, e768s, Sat Aug 3 06:03:51 2019, OttoTime, RUN, General, N/A, TimeCalibratorRun
|
*** Run Time Calibrator FSC ***
- Run: 161
- MUVI: MUVI3
|
|