Status ExpNbr Date Author Type Category Target-Source Subject
FixedTEST Fri Feb 22 14:36:35 2019 iulianMUST2Hardware3-alphatest new cable for Must2 detectors. Cables OK!
Cables put on Muvi3.
Testing procedure. Use the numbered cables to conect the annular detector (MG11) and check if ok: GECO writting on MUVI3 , ORD Present, DECS present, STOP present.
Couldn't check the spectra as they do not exist.

MUVI3

Entry 1Entry 2Entry 3Entry 4
Cable 1Cable 2Cable 3Cable 4
GECO writting OKOKOKOK
ORD Present OKOKOKOK
DECS Present NONONONO
STOP Present OKOKOKOK
FixedTEST Fri Feb 22 15:54:30 2019 iulian GRITHardware3-alpharun 40 3-alpha night run
Run 40

Alpha source run for trapezoid detectors during the night with pedestals aligned (with HV off) and zero suppression. Cooling at 15°C.
Thresholds at 500 keV.


On MUVI1
ChannelSpectra LabelDetectorVoltage
1 MG1 MG1 140V
2 MG3 MG3 160V
3 MG4 MG4 95V
4 MG5 MG5 150V



On MUVI2:
ChannelSpectra LabelDetectorVoltage
1 MG7 MG7 120V
2 MG9 MG1160V
3 MG10 -- --
4 MG11 -- --


TRIGGERS are:
GMT ch1 --> OR MUVI1
GMT ch2 --> OR MUVI2
GMT ch3 --> OR MUVI3 (nothing on it)

[S]Problem MG5?!
[/S]---> No real problem, just spectra definition on plot was incorrect
PinnedTEST Tue Feb 26 09:30:21 2019 FreddyOTHERHardwareN/ACabling - MUVI / HT-BT / SA-SL
Cabling scheme - to update until final
FixedTEST Tue Feb 26 13:21:32 2019 iulian freddy GRITSoftwareN/Arun 42 Calimero run MUVI 1 on X sectors
run #42
Calimero MUVI1 X


run #41
Calimero MUVI secttor 1X
FixedTEST Tue Feb 26 13:50:36 2019 iulian freddy GRITSoftwareN/Arun 43 Calimero run MUVI 1 on Y sectors
run #43
Calimero MUVI1 Y
FixedTEST Tue Feb 26 14:52:49 2019 iulian freddy GRITSoftwareN/Arun 44 Calimero run MUVI 2 on X sectors (bad run)
run #44 bad run
Calimero MUVI2 X

During the run we noticed that time calibrator is on MUVI1 only .
FixedTEST Fri Mar 1 12:07:44 2019 FreddyGRITGeneral3-alphaEfficiency reconstruction
blue: reconstructed efficiency using alpha run 40 (arbitratily normalized)
red: simulated efficiency

There is a difference in angle for trapezoids -> to be investigated

Remark: Simulation curve seems wrong (maybe a target in detector file not removed)
FixedTEST Mon Mar 4 12:28:26 2019 Valerian and IulianMUST2General3-alphaNo DECS and acq error run 47-48-49. Das Saved after problem fixed
Fixed. Solution:
1) CENTRUM not configured on GENTRUM Master
2) GMT matrix on GMT3 (2nd slave) was all 0

DAS saved after solution found
FixedTEST Mon Mar 4 13:54:24 2019 iulian & ValerianGRITHardware3-alpharun 50 3-alpha short run
RUN 50
First 3a test with all Must2 detectors. HV on all set to 50 V for this test only to avoid any risks
FixedTEST Mon Mar 4 14:34:31 2019 Valerian, Marlène, IulianMUST2GeneralN/AMUST2 positions (based on e755)
Based on experiment e755, the MUST2 config is the following :


T1, PCB10:

X128Y1.......X1
. .
. .
Y128.......


T2, PCB5:

Y128.......X128Y1
. .
. .
.......X1



T3, PCB4:

.......Y128
. .
. .
X1.......X128Y1



T4, PCB7:

X1.......
. .
. .
X128Y1.......Y128
FixedTEST Mon Mar 4 15:30:31 2019 iulian Marlene & ValerianMUST2Hardware3-alpharun 51 3-alpha run with Pedestals Aligned @ 13 deg Measured on cooling system
RUN 51
3a run with Pedestals Aligned but not suppressed
FixedTEST Mon Mar 4 16:24:51 2019 valerian marlèneGRITGeneral RUN52 3 alpha source 4MUST2 no pedestal
RUN52
3 alpha source run with only 4MUST2 (in MUVI3) with pedestal suppression at 13 deg.
Fixed  Mon Mar 4 17:22:29 2019 MarleneGRITGeneral Temperature on MUFEE MUST2 13deg
Temperature given by DAS on MUVI3 (MUST2)
screenshots before run
after run : same values
FixedTEST Tue Mar 5 10:16:03 2019 iulianMUST2SoftwareN/Arun 57 Gamme 300 ns Time Calibration run (Calimero) on MUVI 3 (MUST) X sectors more stats (10000)
RUNS 53,54,55,56 are bad runs

Run 57


Calimero on Muvi3 MM1(X)

Gamme 300 ns

Procedure:

0) HT on the detectors Off
1) Apply Das config
2) Put Das offline
3) Start Geco
4) Activate on each of the sector to calibrate the X side (see bellow image)
5) execute Calimero.exe in calimero directory
6) modify the calimero config on Geco (see bellow near config button)
7) Start the run in RCC
8) Start calimero (start button in Geco)
FixedTEST Tue Mar 5 12:13:41 2019 iulian MarleneMUST2SoftwareN/Arun 60 Gamme 600 ns Time Calibration run (Calimero) on MUVI 3 MM1 X
RUNS 58,59,60 are bad runs

Run 61


Calimero on Muvi3 MM1(X)

Gamme 600 ns

Procedure:

0) HT on the detectors Off
1) Apply Das config
2) Put Das offline
3) Start Geco
4) Activate on each of the sector to calibrate the X side (see bellow image)
5) execute Calimero.exe in calimero directory
6) modify the calimero config on Geco (see bellow near config button)
7) Start the run in RCC
8) Start calimero (start button in Geco)
FixedTEST Tue Mar 5 15:53:04 2019 iulian MarleneMUST2SoftwareN/ACalimero problems
problems with Calimero solved by killing some processes on lpcgrit related with Narval.
to find the processes on lpcgrit:
ps aux | grep Cali*
ps aux | grep Narval

kill -9 all the ids

Other possible solution:

start_acq on lpcgrit
tk followed by kall
exit
FixedTEST Tue Mar 5 16:20:11 2019 iulian MarleneMUST2SoftwareN/Arun 77 Gamme 600 ns Time Calibration run (Calimero) on MUVI 3 MM1 X . good run
Run 76


Calimero on Muvi3 MM1(X)

Gamme 600 ns

Procedure:

0) HT on the detectors Off
1) Apply Das config
2) Put Das offline
3) Start Geco
4) Activate on each of the sector to calibrate the X side (see bellow image)
5) execute Calimero.exe in calimero directory
6) modify the calimero config on Geco (see bellow near config button)
7) Start the run in RCC
8) Start calimero (start button in Geco)
FixedTEST Tue Mar 5 16:42:56 2019 iulian MarleneMUST2SoftwareN/Arun 79 Gamme 600 ns Time Calibration run (Calimero) on MUVI 3 MM2 X . good run
Run 79

Run 78 bad
Calimero on Muvi3 MM2(X)

Gamme 600 ns

Procedure:

0) HT on the detectors Off
1) Apply Das config
2) Put Das offline
3) Start Geco
4) Activate on each of the sector to calibrate the X side (see bellow image)
5) execute Calimero.exe in calimero directory
6) modify the calimero config on Geco (see bellow near config button)
7) Start the run in RCC
8) Start calimero (start button in Geco)
FixedTEST Tue Mar 5 17:11:29 2019 iulian MarleneMUST2SoftwareN/Arun 81 Gamme 600 ns Time Calibration run (Calimero) on MUVI 3 MM3 X . good run
Run 81

Run 80 bad
Calimero on Muvi3 MM3(X)

Gamme 600 ns

Procedure:

0) HT on the detectors Off
1) Apply Das config
2) Put Das offline
3) Start Geco
4) Activate on each of the sector to calibrate the X side (see bellow image)
5) execute Calimero.exe in calimero directory
6) modify the calimero config on Geco (see bellow near config button)
7) Start the run in RCC
8) Start calimero (start button in Geco)
FixedTEST Tue Mar 5 17:48:38 2019 iulian MarleneMUST2SoftwareN/Arun 82 Gamme 600 ns Time Calibration run (Calimero) on MUVI 3 MM4 X . good run
Run 82

Calimero on Muvi3 MM4(X)

Gamme 600 ns