Fixed, TEST, Fri Feb 22 14:36:35 2019, iulian, MUST2, Hardware, 3-alpha, test new cable for Must2 detectors. Cables OK!
|
Cables put on Muvi3.
Testing procedure. Use the numbered cables to conect the annular detector (MG11) and check if ok: GECO writting on MUVI3 , ORD Present, DECS present, STOP present.
Couldn't check the spectra as they do not exist.
MUVI3
| Entry 1 | Entry 2 | Entry 3 | Entry 4 |
| Cable 1 | Cable 2 | Cable 3 | Cable 4 |
GECO writting | OK | OK | OK | OK |
ORD Present | OK | OK | OK | OK |
DECS Present | NO | NO | NO | NO |
STOP Present | OK | OK | OK | OK |
|
|
Fixed, TEST, Fri Feb 22 15:54:30 2019, iulian , GRIT, Hardware, 3-alpha, run 40 3-alpha night run 
|
Run 40
Alpha source run for trapezoid detectors during the night with pedestals aligned (with HV off) and zero suppression. Cooling at 15°C.
Thresholds at 500 keV.
On MUVI1
Channel | Spectra Label | Detector | Voltage
|
1 | MG1 | MG1 | 140V
|
2 | MG3 | MG3 | 160V
|
3 | MG4 | MG4 | 95V
|
4 | MG5 | MG5 | 150V
|
On MUVI2:
Channel | Spectra Label | Detector | Voltage
|
1 | MG7 | MG7 | 120V
|
2 | MG9 | MG11 | 60V
|
3 | MG10 | -- | --
|
4 | MG11 | -- | --
|
TRIGGERS are:
GMT ch1 --> OR MUVI1
GMT ch2 --> OR MUVI2
GMT ch3 --> OR MUVI3 (nothing on it)
[S]Problem MG5?![/S]---> No real problem, just spectra definition on plot was incorrect |
Pinned, TEST, Tue Feb 26 09:30:21 2019, Freddy, OTHER, Hardware, N/A, Cabling - MUVI / HT-BT / SA-SL
|
Cabling scheme - to update until final |
Fixed, TEST, Tue Feb 26 13:21:32 2019, iulian freddy , GRIT, Software, N/A, run 42 Calimero run MUVI 1 on X sectors
|
run #42
Calimero MUVI1 X
run #41
Calimero MUVI secttor 1X |
Fixed, TEST, Tue Feb 26 13:50:36 2019, iulian freddy , GRIT, Software, N/A, run 43 Calimero run MUVI 1 on Y sectors
|
run #43
Calimero MUVI1 Y |
Fixed, TEST, Tue Feb 26 14:52:49 2019, iulian freddy , GRIT, Software, N/A, run 44 Calimero run MUVI 2 on X sectors (bad run)
|
run #44 bad run
Calimero MUVI2 X
During the run we noticed that time calibrator is on MUVI1 only . |
Fixed, TEST, Fri Mar 1 12:07:44 2019, Freddy, GRIT, General, 3-alpha, Efficiency reconstruction  
|
blue: reconstructed efficiency using alpha run 40 (arbitratily normalized)
red: simulated efficiency
There is a difference in angle for trapezoids -> to be investigated
Remark: Simulation curve seems wrong (maybe a target in detector file not removed) |
Fixed, TEST, Mon Mar 4 12:28:26 2019, Valerian and Iulian, MUST2, General, 3-alpha, No DECS and acq error run 47-48-49. Das Saved after problem fixed
|
Fixed. Solution:
1) CENTRUM not configured on GENTRUM Master
2) GMT matrix on GMT3 (2nd slave) was all 0
DAS saved after solution found |
Fixed, TEST, Mon Mar 4 13:54:24 2019, iulian & Valerian, GRIT, Hardware, 3-alpha, run 50 3-alpha short run 
|
RUN 50
First 3a test with all Must2 detectors. HV on all set to 50 V for this test only to avoid any risks |
Fixed, TEST, Mon Mar 4 14:34:31 2019, Valerian, Marlène, Iulian, MUST2, General, N/A, MUST2 positions (based on e755)
|
Based on experiment e755, the MUST2 config is the following :
T1, PCB10:
X128Y1.......X1
. .
. .
Y128.......
T2, PCB5:
Y128.......X128Y1
. .
. .
.......X1
T3, PCB4:
.......Y128
. .
. .
X1.......X128Y1
T4, PCB7:
X1.......
. .
. .
X128Y1.......Y128 |
Fixed, TEST, Mon Mar 4 15:30:31 2019, iulian Marlene & Valerian, MUST2, Hardware, 3-alpha, run 51 3-alpha run with Pedestals Aligned @ 13 deg Measured on cooling system 
|
RUN 51
3a run with Pedestals Aligned but not suppressed |
Fixed, TEST, Mon Mar 4 16:24:51 2019, valerian marlène, GRIT, General, , RUN52 3 alpha source 4MUST2 no pedestal  
|
RUN52
3 alpha source run with only 4MUST2 (in MUVI3) with pedestal suppression at 13 deg. |
Fixed, , Mon Mar 4 17:22:29 2019, Marlene, GRIT, General, , Temperature on MUFEE MUST2 13deg 8x
|
Temperature given by DAS on MUVI3 (MUST2)
screenshots before run
after run : same values |
Fixed, TEST, Tue Mar 5 10:16:03 2019, iulian, MUST2, Software, N/A, run 57 Gamme 300 ns Time Calibration run (Calimero) on MUVI 3 (MUST) X sectors more stats (10000) 
|
RUNS 53,54,55,56 are bad runs
Run 57
Calimero on Muvi3 MM1(X)
Gamme 300 ns
Procedure:
0) HT on the detectors Off
1) Apply Das config
2) Put Das offline
3) Start Geco
4) Activate on each of the sector to calibrate the X side (see bellow image)
5) execute Calimero.exe in calimero directory
6) modify the calimero config on Geco (see bellow near config button)
7) Start the run in RCC
8) Start calimero (start button in Geco) |
Fixed, TEST, Tue Mar 5 12:13:41 2019, iulian Marlene, MUST2, Software, N/A, run 60 Gamme 600 ns Time Calibration run (Calimero) on MUVI 3 MM1 X 
|
RUNS 58,59,60 are bad runs
Run 61
Calimero on Muvi3 MM1(X)
Gamme 600 ns
Procedure:
0) HT on the detectors Off
1) Apply Das config
2) Put Das offline
3) Start Geco
4) Activate on each of the sector to calibrate the X side (see bellow image)
5) execute Calimero.exe in calimero directory
6) modify the calimero config on Geco (see bellow near config button)
7) Start the run in RCC
8) Start calimero (start button in Geco) |
Fixed, TEST, Tue Mar 5 15:53:04 2019, iulian Marlene, MUST2, Software, N/A, Calimero problems
|
problems with Calimero solved by killing some processes on lpcgrit related with Narval.
to find the processes on lpcgrit:
ps aux | grep Cali*
ps aux | grep Narval
kill -9 all the ids
Other possible solution:
start_acq on lpcgrit
tk followed by kall
exit |
Fixed, TEST, Tue Mar 5 16:20:11 2019, iulian Marlene, MUST2, Software, N/A, run 77 Gamme 600 ns Time Calibration run (Calimero) on MUVI 3 MM1 X . good run
|
Run 76
Calimero on Muvi3 MM1(X)
Gamme 600 ns
Procedure:
0) HT on the detectors Off
1) Apply Das config
2) Put Das offline
3) Start Geco
4) Activate on each of the sector to calibrate the X side (see bellow image)
5) execute Calimero.exe in calimero directory
6) modify the calimero config on Geco (see bellow near config button)
7) Start the run in RCC
8) Start calimero (start button in Geco) |
Fixed, TEST, Tue Mar 5 16:42:56 2019, iulian Marlene, MUST2, Software, N/A, run 79 Gamme 600 ns Time Calibration run (Calimero) on MUVI 3 MM2 X . good run
|
Run 79
Run 78 bad
Calimero on Muvi3 MM2(X)
Gamme 600 ns
Procedure:
0) HT on the detectors Off
1) Apply Das config
2) Put Das offline
3) Start Geco
4) Activate on each of the sector to calibrate the X side (see bellow image)
5) execute Calimero.exe in calimero directory
6) modify the calimero config on Geco (see bellow near config button)
7) Start the run in RCC
8) Start calimero (start button in Geco) |
Fixed, TEST, Tue Mar 5 17:11:29 2019, iulian Marlene, MUST2, Software, N/A, run 81 Gamme 600 ns Time Calibration run (Calimero) on MUVI 3 MM3 X . good run
|
Run 81
Run 80 bad
Calimero on Muvi3 MM3(X)
Gamme 600 ns
Procedure:
0) HT on the detectors Off
1) Apply Das config
2) Put Das offline
3) Start Geco
4) Activate on each of the sector to calibrate the X side (see bellow image)
5) execute Calimero.exe in calimero directory
6) modify the calimero config on Geco (see bellow near config button)
7) Start the run in RCC
8) Start calimero (start button in Geco) |
Fixed, TEST, Tue Mar 5 17:48:38 2019, iulian Marlene, MUST2, Software, N/A, run 82 Gamme 600 ns Time Calibration run (Calimero) on MUVI 3 MM4 X . good run
|
Run 82
Calimero on Muvi3 MM4(X)
Gamme 600 ns |
|