Fixed, e786s, Wed Jun 26 04:54:07 2019, Freddy, AGATA, General, N/A, AGATA validation problem (solved) 
Since about 4h20, we realized that AGATA validated events and timestamps have suddenly dropped on the AGATA display (yellow bars instead of green).
We checked the counting rate on MUGAST, VAMOS and FAG and there is no obvious problem on this side. Same for the MUGAST/VAMOS data on disk.

We call Emmanuel...

Problem solved by restarting entirely GTS Tree (that was broken).

Restarted at 5h37 with run 59. (it took about 30 min).
Fixed, e786s, Wed Jun 26 05:33:50 2019, OttoRun, RUN, General, Cryo-3He, Run59 
*************************
Run : 59
*************************

Start at 2019-06-26 05:32:53.325868

target 700mb
Brho0=0.95 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted

AGATA position +51-25 mm

TRIGGER :
- VAMOS %1000
- CMG
- MM
- CATS2 %1000

5 to 5h30 - Restarted GTS Tree by Emmanuel
6h00 - Validation broken again

Stop at 2019-06-26 09:44:56.521500
Stop comment: Stopping to Realign GTS Tree
Fixed, e786s, Wed Jun 26 05:58:23 2019, Freddy, OFFLINE, General, N/A, Offline analysis run 56 and 58 run56-58_analysis.pngrun56-59_analysis_gammarays.pngrun56_analysis.pngrun58_analysis.png
Fixed, e786s, Wed Jun 26 07:01:00 2019, Freddy, AGATA, General, N/A, AGATA validation problem - again - during run 59 
30 min after Emmanuel left (around 6h00), the AGATA problem of validation reappeared...

TO BE SOLVED....
Fixed, e786s, Wed Jun 26 07:04:06 2019, Freddy, CATS, General, N/A, Update of CATS calibrations run56_cats_newcal.pngrun_58_cats_newcal.png
New calibrations in :
/home/mugast/analysis/nptool/Projects/calibration/cats/CATS2_X.vamos2019_e786s.npt
/home/mugast/analysis/nptool/Projects/calibration/cats/CATS2_X.vamos2019_e786s.npt

Calibration.txt was modified accordingly.
Attached are the CATS positions for run 56 and 58 (after the central strip problem)
Fixed, e786s, Wed Jun 26 09:56:17 2019, Valerian and Adrien, RUN, General, Cryo-3He, Identification E-TOF MUGAST ETOF_d.pdfETOF_d_vamos.pdf
Light particle ID in MUGAST (no S1 because shitty time).

The red cut is on what we believe is deuterium.

The containt of the cut stop at 8MeV when gating on vamos.
Fixed, e786s, Wed Jun 26 10:29:30 2019, OttoRun, RUN, General, Cryo-3He, Run60 
========================================
*************************
Run : 60
*************************

Start at 2019-06-26 10:21:51.572979

target 700mb
Brho0=0.95 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted

AGATA position +51-25 mm

TRIGGER :
- VAMOS %1000
- CMG
- MM
- CATS2 %1000

Restarted GTS Tree by Emmanuel

Stop at 2019-06-26 17:59:14.820418
Stop comment: change of Brho
Fixed, e786s, Wed Jun 26 13:48:35 2019, mugast, OFFLINE, General, Cryo-3He, analysis Ekine_Ex_run56_58.png
    Fixed, e786s, Wed Jun 26 14:40:58 2019, OttoRun, RUN, General, Cryo-3He, Run60 6x

OttoRun wrote:
*************************
Run : 60
*************************

Start at 2019-06-26 10:21:51.572979

target 700mb
Brho0=0.95 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted

AGATA position +51-25 mm

TRIGGER :
- VAMOS %1000
- CMG
- MM
- CATS2 %1000

Restarted GTS Tree by Emmanuel
    Fixed, e786s, Wed Jun 26 15:48:50 2019, OttoRun, RUN, General, Cryo-3He, Run60 run60_6.png

OttoRun wrote:

OttoRun wrote:
*************************
Run : 60
*************************

Start at 2019-06-26 10:21:51.572979

target 700mb
Brho0=0.95 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted

AGATA position +51-25 mm

TRIGGER :
- VAMOS %1000
- CMG
- MM
- CATS2 %1000

Restarted GTS Tree by Emmanuel


No beeam from 15.00. According to PCP, there will be no beam for a while
    Fixed, e786s, Wed Jun 26 15:50:36 2019, OttoRun, RUN, General, Cryo-3He, Run60 run60_6.png

OttoRun wrote:

OttoRun wrote:
*************************
Run : 60
*************************

Start at 2019-06-26 10:21:51.572979

target 700mb
Brho0=0.95 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted

AGATA position +51-25 mm

TRIGGER :
- VAMOS %1000
- CMG
- MM
- CATS2 %1000

Restarted GTS Tree by Emmanuel


No Beam since 15:00. According to PCP there will be no beam for a while.
Fixed, e786s, Wed Jun 26 16:59:07 2019, Louis, BEAM, General, Cryo-3He, Beam interuption 
Beam stopped from 15h to 17h due to source instabilities
Fixed, e786s, Wed Jun 26 18:12:18 2019, OttoRun, RUN, General, Cryo-3He, Run62 
*************************
Run : 62
*************************

Start at 2019-06-26 18:09:42.882502

target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted

AGATA position +51-25 mm

TRIGGER :
- VAMOS %1000
- CMG
- MM
- CATS2 %1000

Crate in System faillure.
Moving FIFO FPMW to other crate. Had to change the cable (3ns to 5ns).
To be accounted for in the TOF

Stop at 2019-06-27 02:19:17.492600
Stop comment: Problems with AGATA, run stopped
Fixed, e786s, Wed Jun 26 20:46:20 2019, Sylvain, Louis, RUN, General, Cryo-3He, run 62 screenshots 18x
Fixed, e786s, Thu Jun 27 00:26:37 2019, mugast, OFFLINE, General, , statistics until now cqnvqs_.png
Maybe some stat still to recover from better gate and mass and Z reconstruction.

Andrea
Fixed, e786s, Thu Jun 27 03:33:26 2019, OttoRun, RUN, General, Cryo-3He, Run63 
*************************
Run : 63
*************************

Start at 2019-06-27 03:33:09.910218

target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted

AGATA position +51-25 mm

TRIGGER :
- VAMOS %1000
- CMG
- MM
- CATS2 %1000

Crate in System faillure.
Moving FIFO FPMW to other crate. Had to change the cable (3ns to 5ns).
To be accounted for in the TOF

%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
Current ~ 30 kHz.
The cryo target pressure and temperature have slightly increased during the stop for the recover of AGATA, even before the refilling of the LHe tank.
It is probably due to a problem in the refilling of liquid nitrogen.

Stop at 2019-06-27 11:46:17.108352
Stop comment: stop
Fixed, e786s, Thu Jun 27 12:28:58 2019, OttoRun, RUN, General, Cryo-3He, Run67 
*************************
Run : 67
*************************

Start at 2019-06-27 12:28:26.879061

target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted

AGATA position +51-25 mm

TRIGGER :
- VAMOS %1000
- CMG
- MM
- CATS2 %1000

Crate in System faillure.
Moving FIFO FPMW to other crate. Had to change the cable (3ns to 5ns).
To be accounted for in the TOF

Stop at 2019-06-27 12:46:41.949098
Stop comment:
Fixed, e786s, Thu Jun 27 16:35:45 2019, OttoRun, RUN, General, Cryo-3He, Run68 
*************************
Run : 68
*************************

Start at 2019-06-27 16:35:24.351472

target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted

AGATA position +51-25 mm

TRIGGER :
- VAMOS %1000
- CMG
- MM
- CATS2 %1000
Fixed, e786s, Thu Jun 27 16:41:55 2019, OttoRun, RUN, General, Cryo-3He, Run65 
*************************
Run : 65
*************************

Start at 2019-06-27 11:35:21.078920

target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted

AGATA position +51-25 mm

TRIGGER :
- VAMOS %1000
- CMG
- MM
- CATS2 %1000

Crate in System faillure.
Moving FIFO FPMW to other crate. Had to change the cable (3ns to 5ns).
To be accounted for in the TOF

Stop at 2019-06-27 17:39:22.529443
Stop comment:
Fixed, e786s, Thu Jun 27 17:43:15 2019, OttoRun, RUN, General, Cryo-3He, Run69 
*************************
Run : 69
*************************

Start at 2019-06-27 17:42:38.280555

target 700mb
Brho0=0.90 Tm
DC pressure 6 mbar
IC pressure 120 mbar
CATS2 inserted

AGATA position +51-25 mm

TRIGGER :
- CMG


Test with only MUGAST back trigger

Stop at 2019-06-27 17:54:02.829540
Stop comment: