Fixed, e786s, Fri May 31 09:53:14 2019, Marlène, GRIT, General, , Alpha source --> status of detectors 
MG3T has been fixed by changing the X kapton.

I have removed aluminium foil on square detectors, the noise is slightly higher.

MG4T is not working properly : noisy on X but no signal on Y !! I see no triggers on the Y side. I tried to increase bias but it did increase noise. I checked in the last calibration of e744 it was working properly. Also I had to do pedestal realignement for this detector because everything was shifted.

MG7T seem to be slightly different than before. Strips X are hidden from the trigger because it became suddenly noisy.
Fixed, e786s, Mon Jun 3 18:00:10 2019, Marlène, GRIT, General, , MUGAST detectors status --> MG4T  
The problem of MG4T ha sbeen solved by changing the feedthrough.
The pulser on Y was not working and we were seeing no events (no ORD) because there was a problem with the bias of board. The issue is due to the feedthrough.
So now MG4T (HT &BT) are plugged in a free feedthrough (below MG1T) and I see pulser signals on Y.

The bias seems to apply normally on all detectors (in the dark).
The pulser is seen on each boards.
I have put back aluminium foil on square but I could not test it.
Fixed, e786s, Wed Jun 12 17:59:01 2019, mugast, OTHER, Software, , Online Energy Calibration x_E_cal.pngy_E_cal2.png
Updated Online Energy calibration for MUGAST, based on run_0231 data
    Fixed, e786s, Wed Jun 12 18:11:04 2019, mugast, OTHER, Software, , Online Energy Calibration 
Fixed, e786s, Sun Jun 23 14:37:43 2019, Freddy, Franco and Louis, OTHER, Hardware, , Temperature error on +2.5v/t3 
There is a temperature error on the +2.5 V for MUST2 Telescope. It created error message due to hv_monitor but it does not shutdown the voltage.
We modified the code in hv_monitor to not display popup message and elog message if this channel is concerned.
Fixed, , Mon Jun 24 14:14:17 2019, Valerian, Adrien - Dudouet, OFFLINE, General, , p,p' analysis of run49 Screenshot_from_2019-06-24_14-09-39.pngKine46Arpp.pngGamma46Arpp.png
Fixed, e786s, Tue Jun 25 15:33:40 2019, A Lemasson, VAMOS, General, , VAMOS Analysis - Brho 8x
VAMOS Analysis :
VAMOS PID with attached SetAlias.C

Suggestion to change Brho from 1.05 to 0.95 based on the distribution of K ions to optimize acceptance on the K
ions
Fixed, e786s, Thu Jun 27 00:26:37 2019, mugast, OFFLINE, General, , statistics until now cqnvqs_.png
Maybe some stat still to recover from better gate and mass and Z reconstruction.

Andrea
Fixed, , Thu Jul 11 11:40:18 2019, Marlène & Franco, OTHER, Hardware, , Some changes in the electronic scheme  
- We checked the TACs with a time calibrator, all of them are working. Maybe to check the ADCs if some timings are not there during the exp.

- We set the possibility to have MUST2 divided by 100 (on a separate chain) in case of high rate on GMT channel number 7.

- We are now taking as input of the HF the clean HF validated by VAMOS instead of cleaning it on our side through the coincidence with FAG. This is sent to make the STOP of all TACs and MUVIs.

- We sent the OR of MUST2-MUGAST-SQUARES in an additional FIFO with cables of the same length (10 ns) and the output sent to VAMOS and to the TAC called Si-CATS in the acquisition.
Fixed, e768s, Fri Jul 12 17:51:06 2019, mugast (Sylvain, Chloe, François, Marlène, Nicolas), OTHER, General, , Monitoring of beam profile with BAF2 
In order to provide a efficient way for PCP to monitor beam profile a new TAC measurement has set up.

A BAF2 detector has been inserted close to the reaction chamber (in position 7 of AGATA structure frame). Operating voltage of BaF2 is 1800V and the typical signal amplitude is above 100 mV.
The output signal is sent to a TFA (with gain =1). No additional delay was used. The BAF2 trigger signal is made from CFD and used as a start a TAC. TAC range was chosen to 1us.

The trigger signal was also sent to GMT trough input nb 5.




Fixed, e768s, Mon Jul 15 14:29:47 2019, Marlène, MUST2, General, , Masks for MUST2 & Kapton changed for MG5 IMG_20190715_121646_crop_flip.jpgIMG_20190715_122035.jpg
Masks are now correctly mounted

All 4 kaptons from MG5 have been changed
Fixed, e768s, Tue Jul 16 15:13:01 2019, Jenn, DIego, VAMOS, General, , Brho  E768_BRHO.pdf
Fixed, e768s, Tue Jul 16 15:32:05 2019, Jenn, OTHER, General, , Shift table google schedule link 
Please, see below the spreadsheet with the shift schedule.

https://docs.google.com/spreadsheets/d/1NcidygPHZhcBTr5j6TMV7oaz7jJcAoLt4q_sOz813Cg/edit?usp=sharing
Fixed, e768s, Tue Jul 16 19:21:19 2019, mugast, OTHER, General, , Qpole values 
Brho = 0.640 Tm

RMN = 3247,742 Gaus (215.630 A)

Q31 = 18.5 A
Q32 = 20.9 A
Q41 = 170.579 A
Q42 = 150.829 A
Fixed, e768s, Tue Jul 16 22:25:51 2019, mugast, OTHER, General, , Oxygen 15 decay scheme 15O_decay_scheme.png
beta+ decay to the ground state of 15N
gammas @ 511 keV
Fixed, e768s, Wed Jul 17 01:51:10 2019, mugast, BEAM, General, , Faisceau perdu 
Un Q-pole dans L3 est tombé. L'opérateur a du mal à le redémarrer. Bertrand et un technicien ont essayé de réparer mais sans succès.
L'astreinte a été appelée à 2h50.
Fixed, e768s, Wed Jul 17 09:07:19 2019, mugast, BEAM, General, , beam settings 20190717_061639.jpg
values as it is now
Fixed, , Wed Jul 17 10:25:26 2019, Kseniia, MUST2, Hardware, , The burnt crate 58CAA1B9-97A5-4711-AEAA-A4167EC192E3.jpeg
Here’s what the 48V power supply of the damaged crate looks like inside
Fixed, e768s, Wed Jul 17 10:55:23 2019, Francois, OFFLINE, General, , previous measurement of 15N(7Li,t)  previous.pdf
Fixed, e768s, Wed Jul 17 11:36:53 2019, Jenn, VAMOS, General, , M/Q values 

Values of M/Q for possible recoils on the focal plane

          M/Q

15N   15/7   2.14

15N   15/6   2.5 

19F   19/9   2.11

19F   19/8   2.38

20F    20/9  2.22

20F    20/8   2.5