MUST2 SAMURAI_2018 S3-LEB-LPC SUPERNEMO MUGAST EXPAND SCALP GALATRON HiCARI VELO
  MUGAST, Page 125 of 132  Not logged in MUGAST
Status ExpNbr Date Author Type Category Target-Sourcedown Subject Text Attachments
Fixede786s Fri May 31 09:53:14 2019 MarlèneGRITGeneral Alpha source --> status of detectorsMG3T has been fixed by changing the X kapton.

I have removed aluminium foil on square detectors,
  
Fixede786s Mon Jun 3 18:00:10 2019 MarlèneGRITGeneral MUGAST detectors status --> MG4T The problem of MG4T ha sbeen solved by changing
the feedthrough.
The pulser on Y was not working and we were
  
Fixede786s Wed Jun 12 17:59:01 2019 mugastOTHERSoftware Online Energy CalibrationUpdated Online Energy calibration for MUGAST,
based on run_0231 data
 x_E_cal.pngy_E_cal2.png 
Fixede786s Wed Jun 12 18:11:04 2019 mugastOTHERSoftware Online Energy Calibration    
Fixede786s Sun Jun 23 14:37:43 2019 Freddy, Franco and LouisOTHERHardware Temperature error on +2.5v/t3There is a temperature error on the +2.5 V
for MUST2 Telescope. It created error message
due to hv_monitor but it does not shutdown
  
Fixed  Mon Jun 24 14:14:17 2019 Valerian, Adrien - DudouetOFFLINEGeneral p,p' analysis of run49   Screenshot_from_2019-06-24_14-09-39.pngKine46Arpp.pngGamma46Arpp.png 
Fixede786s Tue Jun 25 15:33:40 2019 A LemassonVAMOSGeneral VAMOS Analysis - BrhoVAMOS Analysis : 
VAMOS PID with attached SetAlias.C
 8x 
Fixede786s Thu Jun 27 00:26:37 2019 mugastOFFLINEGeneral statistics until nowMaybe some stat still to recover from better
gate and mass and Z reconstruction.
 cqnvqs_.png 
Fixed  Thu Jul 11 11:40:18 2019 Marlène & FrancoOTHERHardware Some changes in the electronic scheme - We checked the TACs with a time calibrator,
all of them are working. Maybe to check the
ADCs if some timings are not there during
  
Fixede768s Fri Jul 12 17:51:06 2019 mugast (Sylvain, Chloe, François, Marlène, Nicolas)OTHERGeneral Monitoring of beam profile with BAF2In order to provide a efficient way for PCP
to monitor beam profile a new TAC measurement
has set up. 
  
Fixede768s Mon Jul 15 14:29:47 2019 MarlèneMUST2General Masks for MUST2 & Kapton changed for MG5Masks are now correctly mounted

All 4 kaptons from MG5 have been changed
 IMG_20190715_121646_crop_flip.jpgIMG_20190715_122035.jpg 
Fixede768s Tue Jul 16 15:13:01 2019 Jenn, DIegoVAMOSGeneral Brho    E768_BRHO.pdf 
Fixede768s Tue Jul 16 15:32:05 2019 JennOTHERGeneral Shift table google schedule linkPlease, see below the spreadsheet with the
shift schedule.
  
Fixede768s Tue Jul 16 19:21:19 2019 mugastOTHERGeneral Qpole valuesBrho = 0.640 Tm

RMN = 3247,742 Gaus (215.630 A)
  
Fixede768s Tue Jul 16 22:25:51 2019 mugastOTHERGeneral Oxygen 15 decay schemebeta+ decay to the ground state of 15N
gammas @ 511 keV
 15O_decay_scheme.png 
Fixede768s Wed Jul 17 01:51:10 2019 mugastBEAMGeneral Faisceau perduUn Q-pole dans L3 est tombé. L'opérateur a
du mal à le redémarrer. Bertrand et un technicien
ont essayé de réparer mais sans succès. 
  
Fixede768s Wed Jul 17 09:07:19 2019 mugastBEAMGeneral beam settingsvalues as it is now  20190717_061639.jpg 
Fixed  Wed Jul 17 10:25:26 2019 KseniiaMUST2Hardware The burnt crateHere’s what the 48V power supply of the damaged
crate looks like inside
[IMG]elog:5072/1[/IMG]
 58CAA1B9-97A5-4711-AEAA-A4167EC192E3.jpeg 
Fixede768s Wed Jul 17 10:55:23 2019 FrancoisOFFLINEGeneral previous measurement of 15N(7Li,t)    previous.pdf 
Fixede768s Wed Jul 17 11:36:53 2019 JennVAMOSGeneral M/Q valuesValues of M/Q for possible recoils on the
focal plane

         
  

ELOG Home Page