Fixede768s Sun Jul 28 14:57:26 2019 Franco & SylvainBEAMGeneralN/ABeam problems and DC tripThe HV and the vacuum in VAMOS are now recovered
and back to the normal values. The water
tray is 20% filled and under control. Mugast
  
FixedTEST Thu Oct 17 12:01:09 2019 Franco & MarlèneGRITHardware3-alphaTest of configuration with 4 MUVI and 3 GAMERSThe electronics set-up for MUGAST2020 has
been tested with 4 MUVI on 4 VXI crates,
1GMT for MASTER and 3 GAMERS for slaves.
 Mugast_alpha_17102019Mugast_alpha_17102020_2 
Fixede793s Tue Mar 10 18:46:13 2020 Franco & MarlèneGRITHardware Changes in MUGAST- MG6 (FZ) is disconnected.
- MG7 (500um) is dismounted and MG8 (375um)
is replacing it.
 MG_X_raw_11032020.pngMG_Y_raw_11032020.png 
Fixede786s Tue Jun 25 22:20:36 2019 FrancoRUNGeneralCryo-3HeRun58[quote="OttoRun"][quote="OttoRun"]*************************
  [SIZE=3][B]Run  : 58[/B][/SIZE]
*************************
 BRho_run58.png 
Fixede786s Tue Jun 25 23:55:34 2019 FrancoRUNGeneralCryo-3HeRun58[quote="Franco"][quote="OttoRun"][quote="OttoRun"]*************************
  [SIZE=3][B]Run  : 58[/B][/SIZE]
*************************
 scaler_run_58_2.png 
Fixede768s Mon Jul 15 14:38:27 2019 FrancoRUNGeneral3-alphaRun302Attached the calibrated spectra for MUST2
and Mugast for run 302 (no pedestals).
 run0302_must2_cal.pngrun0302_mugast_cal.png 
Fixede768s Mon Jul 15 18:28:19 2019 FrancoRUNGeneral3-alphaRun301Attached the calibrated spectra for MUST2
and Mugast for run 301.
 run0301_must2_cal.pngrun0301_mugast_cal.png 
Fixede768s Tue Jul 16 09:44:53 2019 FrancoRUNGeneral3-alphaRun306Attached the calibrated spectra for Mugast
and MUST2 for run 0306 and the raw spectra
of Mugast.
 run0306_mugast_cal.pngrun0306_must2_cal.pngrun0306_mugast_raw.png 
Fixede768s Tue Jul 16 09:51:56 2019 FrancoRUNGeneral3-alphaRun301Attached the dat file with the position of
the pedestals for each strip of MG3_X
that can be used to correct the pedestal
 pedestals_MG3_STRX_E_RAW.dat 
Fixede768s Thu Jul 18 22:11:57 2019 FrancoOFFLINEGeneral Identification in MugastAttached the spectra obtained in Mugast for
run_0072_0_agata.root.
The colored ones are obtained simply asking
 id_mugast_20Fblackpoints.png 
Fixede768s Fri Jul 19 23:10:18 2019 FrancoRUNGeneral3-alphaRun302Added energy calibration for square detector
(MG 10).
 run302_MG_Ecal.png 
Fixede768s Fri Jul 19 23:12:36 2019 FrancoRUNGeneral3-alphaRun302Added energy calibration for square detector
(MG 10).
 run302_MG_Ecal.png 
Fixede768s Sat Jul 20 05:58:18 2019 FrancoOFFLINEGeneral Time calibration for MG10_XAttached a tentative time calibration for
the square MG10_X from run 89 (to be checked!).
Even applying the calibration the spectrum
 run_0089_MG10_timecal.pngETOF_MG10_run115.png 
Fixede768s Mon Jul 22 21:30:36 2019 FrancoOFFLINEGeneralLiFIdentification in Mugast for Ne in VAMOSDifference in identification in Mugast when
gating on 19Ne and when gating on the two
nearby isotopes (runs from 112 to 121).
 Mugast_19Ne.pngMugast_18Ne_and_20Ne.png 
FixedTEST Tue Feb 18 17:53:36 2020 FrancoGRITGeneral3-alphaAlpha calibration for exp e775s (run 0501)In attachment raw and calibrated spectra for
run 0501 (alpha source) for MUST2 and Mugast.
 E_raw_MM.pngE_raw_MG.pngE_cal_MM.pngE_cal_MG.png 
Fixed  Thu Feb 20 17:58:51 2020 FrancoGRITGeneral Time calibration for e775s expAttached the raw and calibrated time spectra
for the pulser runs. 
 T_raw_MM.pngT_raw_MG.pngT_cal_MM.pngT_cal_MG.png 
FixedTEST Tue Mar 10 11:05:40 2020 FrancoGRITHardware3-alphaTrip in the cooling system after e775s experimentIn the weekend after the experiment e775s
(7-8 March 2020) a problem with the cooling
system of Mugast occurred. In particular,
 run0193_12_heat.pngrun0193_13_heat_MG6vsMG1.pngrun_0524_MG11XYraw.pngrun_0193_9_19_11_MG11XYraw.png 
FixedTEST Tue Feb 2 16:41:17 2021 FrancoVACUUMHardwareN/AMUGAST undervacuum and powered upI put back the supply of MM1 to the channel
1 of the MUST2 box (as in the normal configuration)
and polarized the detectors to make a rapid
  
FixedTEST Wed Feb 3 16:42:41 2021 FrancoGRITGeneral3-alphaStatus of the set-up (February 3rd 2021)We re-mounted the annular, put the source
and cooled down to 0 deg. Voltages and spectra
are attached.
 Voltages_03022021.pngMM_raw_03022021.pngMG_X_raw_03022021.pngMG_Y_raw_03022021.png 
FixedTEST Thu Feb 4 18:16:47 2021 FrancoGRITHardware3-alphaProblems with the annular persistWe opened the chamber to verify the connections
of the kaptons but they were all well connected.
We checked with the tester the conductivity
 8x 